AND 162C-BA-BC BA Series Analytical Balances Guida utente

Tipo
Guida utente
(This document is included in the packing contents for the balance.)
Quick Start Guide
BA Series
Analytical Balances
English (EN)
Guide de démarrage rapide
Série BA
Balances analytiques
Français (FR)
Schnellstartanleitung
Serie BA
Analysenwaage
Deutsch (DE)
Guía de inicio rápido
Serie BA
Básculas analíticas
Español (ES)
Guida introduttiva
Serie BA
Bilance analitiche
Italiano (IT)
快速入門指南
BA 系列
分析天平
中文
(CT)
1WMPD40042
94
B
© 2021 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated into
any language in any form by any means without the written permission of A&D Company,
Limited.
The contents of this publication and the specifications of the instrument covered by this
publication are subject to change for improvement without notice.
3
Quick start guide BA Series
Analytical balances .......................................................................... 4
Guide de démarrage rapideSérie BA
Balances analytiques ..................................................................... 19
SchnellstartanleitungSerie BA
Analysenwaage ............................................................................. 35
Guía de inicio rápidoSerie BA
Básculas analíticas ........................................................................ 53
Guida introduttiva Serie BA
Bilance analitiche ........................................................................... 70
快速入門指南
BA
系列
分析天平 ........................................................................................ 86
EN
FR
DE
ES
IT
CT
EN
4
1. Quick Start Guide
Thank you for purchasing an A&D BA series analytical balance. This quick start guide describes how to
install the BA series balance, and it covers the basic functions and main operations of the balance.
For further information about the BA series balance, please refer to the separate instruction manuals
listed in "1.1. Detailed manuals".
1.1. Detailed manuals
Detailed manuals are available for download from the A&D website ( https://www.aandd.jp ):
BA Series Instruction Manual
This is the manual to help you understand the functions and operations of the BA series in detail and
make full use of them.
Ionizer (static eliminator) Instruction Manual
This is the manual describing the functions of the AX-ION-25 ionizer and instructions on use.
2. Assembly and Installation
Caution
A BA series analytical balance is a precision instrument, and it should be carefully unpacked.
It is advisable to store the packing materials so that they can be used when transporting the balance.
The contents of the package vary depending on the product. Refer to the illustration of the packing
contents and make sure that everything is included.
Do not connect the AC adapter to the balance until the balance is assembled and installed.
Use the dedicated AC adapter specified for the balance.
Do not connect the included AC adapter to other devices.
If you use the wrong AC adapter, the balance and other equipment may not work properly.
Make sure that the AC adapter is unplugged before you connect the breeze break unit and the ionizer.
The BA series analytical balance is composed of the weighing unit and display unit. It is not possible
to replace just one of the units.
EN
5
2.1. Unpacking
Packing contents common to BA series
Packing contents for BA-6E / BA-6DE
Packing contents for BA-225 / BA-225D / BA-125D
Pan B, Φ85
Breeze break ring B
Dust plate B
Breeze break front glass
Breeze break doors
Weighing unit
Display unit protection cover
AC adapter ID labels
This document
USB cable
(approx. 1.8 m)
Breeze break
top door
Breeze break
bottom plate
Tweezers
Display unit
Cable B
(approx. 1.0 m)
Cable A (approx. 0.2 m)
AC adapter
(approx. 1.5 m)
Attach the AC adapter labels to the adapter.
Pan A, Φ25
Pan support A
Breeze break unit
Breeze break
front glass
Dust plate A
Breeze break ring A
Small glass breeze break set
Side panes (3 pieces)
Top pane
Base stand
Aluminum round pans for analysis (10 pieces each)
Large pan,
Φ15, 0.8 mL
Breeze break doors
Medium pan,
Φ12, 0.3 mL
Small pan,
Φ8, 0.05 mL
Cable C
(approx. 0.6 m)
Ionizer
Weighing pan for filters
EN
6
2.2. Assembly
The BA-6E is used here for the example of assembly. Other models can be assembled in the same way.
Step 1 Pull out the breeze break
locking handles.
Step 2 Place the breeze break unit on
the weighing unit.
Step 3 Push in the breeze break
locking handles to secure the
units to each other.
Step 4 Assemble the weighing pan.
For BA-6E / BA-6DE, assemble
the five parts of the weighing
pan in the correct position by
referring to the cross-sectional
view.
In addition to the standard
weighing pan (Pan A), you can
also use the weighing pan for
filters. For details, refer to the
detailed instruction manual
listed in “1.1. Detailed
manuals”.
For BA-225 / BA-225D / BA-
125D, assemble the four parts
of the weighing pan in the
correct position by referring to
the cross-sectional view.
Breeze break unit
Step 1
Step 3
Step 2
Weighing unit
For BA-225 / BA-225D / BA-125D
For BA-6E / BA-6DE
Pan A
Pan support A
Breeze break
ring A
Dust plate A
Pan B
Breeze break
ring B
Dust plate B
Breeze break
bottom plate
Cross-sectional view
BA-6E / BA-6DE
Weighing pan
for filters
Cross-sectional view
BA
-225 / BA-225D
BA-125D
EN
7
Step 5 Assemble the glass breeze
break.
Assemble the BA-6E / BA-6DE
glass breeze break in the
weighing chamber. Align the
side panes with the notch of the
base stand.
Step 6 Insert the breeze break top door
from the handle side.
Step 7 Assemble the breeze break
doors on the left and right sides.
7a Insert the breeze break doors
on the left and right sides.
7b Secure the breeze break doors
with the latches.
Step 8 Assemble the breeze break
front glass.
8a Insert the breeze break front
glass.
8b Secure the breeze break front
glass with the latches.
Notch
Example of the small breeze
break with the right side open.
Base stand
Step 5
Top pane
Side panes
(3 pieces)
Base stand
Side pane
Step 7a
Step 6
Step 8b
Step 7b
Step 8a
Step 7a
Step 8b
Step 7b
EN
8
Step 9 Using the included cable A and cable B, connect the "weighing unit" and "breeze break unit"
and the "display unit" and "weighing unit" respectively, paying attention to the direction of the
arrow on each cable.
Caution Make sure to unplug the AC adapter before connecting.
Step 10 Connect the breeze break unit and ionizer with the included cable C.
Caution The ionizer operates when the AC adapter is connected to the balance.
Step 11 Connect the included AC adapter to the balance.
BA-225 / BA-225D / BA-125D
BA-6E / BA-6DE
Display unit
Ionizer
Step 9
Step 11
Cable A
Cable B
AC adapter
Weighing unit
Grounding terminal
Serial number
Step 10
Cable C
EN
9
3. Installation Considerations, Preparation and Precautions
Prepare the following installation conditions in order to bring out the full performance of the balance.
Install the balance in an environment where the temperature and humidity are not excessive. The
best operating temperature is about 20°C ±2°C at about 45% to 60% RH relative humidity.
Install the balance where it is free of dust.
The weighing table should be solid. (An anti-vibration table or stone table is ideal)
Place the balance on a horizontal table, and make sure that it is not tilted.
Install the balance in a stable location, avoiding vibration and shock. Corners of rooms on the first
floor are best, as they are less prone to vibration.
Install the balance where it is not affected by heaters or air conditioners.
Install the balance where it is not exposed to direct sunlight.
Install the balance away from equipment which produces
magnetic fields.
Level the balance with the leveling feet and bubble spirit level.
Refer to “3.1. How to adjust the level of the balance”.
Be sure to warm up the balance before use for at least an hour, or at least four hours for BA-6E / BA-
6DE, with the AC adapter connected to the power supply.
Adjust the sensitivity of the balance before using it for the first time or after having moved it to another
location so that accurate weighing can be performed. For details, refer to the detailed instruction
manual listed in “1.1. Detailed manuals”.
Caution Do not install the balance where flammable or corrosive gas is present.
3.1. How to adjust the level of the balance
Level the balance by adjusting the leveling feet so that the bubble of the bubble spirit level is centered
in the red circle.
When the bubble is off to the left.
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the right.
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the backward
position.
Turn both leveling feet on the front in the
clockwise direction at the same time.
When the bubble is off to the forward
position.
Turn both leveling feet on the front in the counter
clockwise direction at the same time.
Leveling foot
Bubble spirit level
Bubble
spirit level
Up
Down
Leveling
foot
The balance is level
(Correct weighing)
Leveling
foot
The balance is tilted
(Inaccurate weighing).
Up
Down
Bubble
Red circle
EN
10
4. Display and Key Panel (Basic Operation)
Lit display
Blinking display
Keys
Data memory number
IR sensor indicator
Response indicator (lights for 30 seconds after start
of weighing)
Preset tare mark
Gross mark
Net mark
Processing indicator
Gross zero mark
Stabilization indicator
Interval output mode
standby indicator
USB connecting mark
Displays weighing value, stored data, and items
Standby indicator
Data number being displayed
Processing indicator
Automatic sensitivity adjustment notice
Interval output mode active indicator
Unit display
SAMPLE
key
ON/OFF
key
Left IR sensor
Right IR sensor
MODE
key
CAL
key
PRINT
key
TARE
key
RE-ZERO
key
Shock indicator
EN
11
Key operation
Key operations affect how the balance functions. Normal
key operation during measurement is “Press and release
the key immediately” or “Press the key”. Please do not
“Press and hold the key (for 2 seconds)” unless required.
Key
When pressed and released
When pressed and held (for 2 seconds)
Turns the display on and off.
When the display is turned off, only the
standby indicator is displayed.
When the display is turned on, weighing is
possible.
If the password function is enabled, you will
be prompted to enter the password when the
display is turned on. For details, refer to the
separate instruction manual listed in “1.1.
Detailed manuals".
The ON:OFF key is active at any time,
and pressing this key during operation
always turns off the display.
Switches the IR sensor on and off.
In weighing mode, press the key to turn the
digit for readability on and off.
In counting or percent mode, press the key
to enter the sample storing mode.
Enters the function table mode. Please
refer to the separate instruction manual
listed in “1.1. Detailed manuals".
Runs the repeatability check function when
pressed and held for another 2 seconds
after the function table menu is displayed.
For details, refer to the separate instruction
manual listed in “1.1. Detailed manuals".
Switches the weighing units stored in the
function table.
Enters "Changing Weighing Speed" mode.
Performs sensitivity adjustment of the
balance using the internal weight.
Displays the menu related to sensitivity
adjustment.
Stores or outputs the weighing value when
stable according to the function table
settings.
(Data is output at factory setting.)
Enters mode to change the unit weight
registration number in counting mode.
The following settings are possible with the
function table:
Outputs "Title block" and "End block" for
GLP/GMP report.
Displays the data memory menu.
Performs tare operation. -
Sets the displayed value to zero. -
IR sensor (touchless sensor).
It reacts when you bring your hand close to
it.
Opening and closing of the breeze break
door are assigned.
For details, refer to "5. IR Sensors and
Auto Doors".
-
Press the key
(Press and release
the key immediately.)
Press and hold
the key for 2
seconds.
EN
12
5. IR Sensors and Auto Doors
5.1. IR sensors
BA series analytical balances are equipped with IR sensors that allow operation without directly touching
the balance display. At factory setting, the IR sensors on the left and right of the display are assigned to
open and close the breeze break doors. By pressing and holding the ON:OFF key (for about 2
seconds), you can switch the IR sensors on and off with a single touch.
5.2. Auto doors
BA series analytical balances are equipped with auto doors that allow you to open/close the breeze break
without touching the doors.
The IR sensors on the left and right of the display are assigned settings to open/close the breeze break
doors on the right and left respectively.
At factory setting, each breeze break door opens to the position it was previously opened to.
It is also possible to change which sensors the breeze break doors are assigned to and for the doors to
be fully open or partially open in the function table of the balance.
For details, refer to the separate instruction manual listed in “1.1. Detailed manuals”.
Opening the breeze break door
Step 1. If you want to open the breeze break door on the left side (or on the right side) when it is closed,
hold your hand over the right IR sensor (or the left IR sensor).
Step 2. The detection buzzer sounds and the breeze break door on the left (or right) side opens.
Closing the breeze break door
Step 1. If you want to close the breeze break door on the left side (or on the right side) when it is open,
hold your hand over the right IR sensor (or the left IR sensor).
Step 2. The detection buzzer sounds and the breeze break door on the left (or right) side closes.
Operation example
Left IR sensor: Opens/closes the right breeze
break door.
Example for BA-6E / BA-6DE
Right IR sensor: Opens/closes the left
breeze break door.
EN
13
6. Sensitivity Adjustment
6.1. Automatic sensitivity adjustment
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal weight. For details, refer to the separate instruction
manual listed in “1.1. Detailed manuals”.
6.2. Sensitivity adjustment using the internal weight
Sensitivity adjustment using the internal weight can be performed with one key press.
Step 1. Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6E/BA-6DE, with the AC adapter connected to the power supply.
Step 2. Press the CAL key. The balance displays .
Step 3. The balance automatically performs sensitivity adjustment using the internal weight. Do not
apply vibration and the like to the balance.
Step 4. After sensitivity adjustment, the balance will output the “sensitivity adjustment report” and store
the data as set in the function table.
For details, refer to the separate instruction manual listed in “1.1. Detailed manuals”.
Step 5. The balance returns automatically to weighing mode.
6.3. Sensitivity adjustment using an external weight
This function performs sensitivity adjustment of the balance using your external weight.
For details, refer to the separate instruction manual listed in “1.1. Detailed manuals”.
EN
14
7. Error Displays (Error Codes)
Display Code
Description
EC,E11
Stability error
Weighing value is unstable and "zero display", "sensitivity adjustment", etc.
cannot be executed. Check around the pan. R
efer to the separate
instruction manual listed in “1.1. Detailed manuals”.
Improve the
environment of the installation location (vibration, drafts, static electricity,
etc.). To return to weighing mode, press the CAL key.
Out of the setting range
The value to be set exceeds the setting range.
Set again within the setting range.
Malfunction of the internal memory element of the balance
If this error continues to be displayed, repair is necessary. Please contact
your local dealer for repair.
EC,E16
Internal weight error
Applying the internal weight does not yield a change in the mass value as
specified.
Confirm that there is nothing on the pan and perform the operation from the
beginning.
EC,E17
Internal weight error
The internal weight
application mechanism does not function properly.
Perform the operation from the beginning.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, repair is necessary.
Please contact your local dealer for repair.
Weighing unit error
Data from the weighing unit has not been received. Check the connection
between the weighing unit and the display unit, and turn the power off and
then on again.
Breeze break unit error
Communication with the breeze break unit is not possible.
Unplug the power supply and reconnect the cable between the breeze
break unit and the weighing unit.
Serial number error
The serial numbers of the weighing unit and the display unit do not match.
Reconnect with the correct combination.
EC,E20
Sensitivity adjustment weight error (Positive value)
The sensitivity adjustment
weight is too heavy. Check around the pan.
Check the mass value of the weight.
To return to weighing mode, press the CAL key.
EN
15
Display Code
Description
EC,E21
Sensitivity adjustment weight error (Negative value)
The sensitivity adjustment weight is too light. Check around the pan. Check
the mass value of the weight.
To return to weighing mode, press the CAL key.
Overload error
A sample beyond the balance weighing capacity has been placed on the
pan. Remove the object from the pan.
Weighing pan error
The weighing
value is too light. Check that the weighing pan is installed
correctly. Set the weighing pan correctly. Perform sensitivity adjustment.
Sample mass error
The sample is too light to be stored as a sample mass for the counting mode
or percent mode. The sample cannot be used.
Unit weight error
The sample mass for the counting mode is too light. Storing and using it for
counting may cause a counting error. Add samples to reach the specified
number and press the PRINT key. Although pressing the PRINT key
without adding samples will put the balance in counting mode, make sure
to add samples for accurate counting.
Clock battery error
The clock backup battery has been depleted. Press any key and set the
time and date. Even if the clock backup battery is depleted, the clock and
calendar function
works normally as long as the balance is powered with
the AC adapter. If this error appears frequently, contact your local dealer for
repair.
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Check that the AC adapter is the one supplied with the balance.
Repeatability error
The standard deviation (SD) of repeatability exceeded 50 d. Review the
installation environment of the balance.
displayed in repeatability display.
displayed in minimum weighing value (reference value) display.
For details, refer to the separate instruction manual listed in”1.1.
Detailed
manuals”.
Blinking (alternately)
Full memory
The number of stored weighing values has reached the upper limit. In order
to store a
new weighing value, it is necessary to delete data. For details,
refer to the separate instruction manual "1.1. Detailed manuals”.
Blinking (alternately)
Full memory
The stored sensitivity adjustment /
calibration test history has reached 50
results. In order to store a new result, the oldest history will be deleted.
For
details, refer to the separate instruction manual "1.1. Detailed manuals”.
EN
16
Display Code
Description
EC,E00
Communications error
A protocol error occurred in communications.
Check the format, baud rate, etc.
EC,E01
Undefined command error
An undefined command was found.
Check the transmitted command.
EC,E02
Not ready
The received command cannot be executed:
(e.g.) Q command was received when not in weighing mode.
(e.g.) Q command was received while re-zeroing.
Adjust the delay time to transmit a command.
EC,E03
Timeout error
With the timeout parameter set to , there was a wait time of
approximately 1 second or more while receiving command characters.
Check the communication.
EC,E04
Character length error
The number of characters in the received command has exceeded the
limit. Check the command to transmit.
EC,E06
Format error
The description of the received command is incorrect:
(e.g.) The number of digits of numerical values is incorrect.
(e.g.) There are alphabet characters among the numerical values.
Check the transmitted command.
EC,E07
Parameter setting error
The value of the received command has exceeded the allowed value.
Check the setting range of the numerical value of the command.
7.1. Other display
The automatic sensitivity adjustment notice (the indicator blinking) indicates
that the automatic sensitivity adjustment will start.
If the balance is not used for a certain period of time with this indicator blinking, the
balance automatically performs sensitivity adjustment using the internal weight. (The
blinking period depends on the operating environment.)
Tips Although it is possible to continue using the balance even while this indicator is
blinking, use after sensitivity adjustment is completed is advisable in order to maintain
the weighing accuracy.
EN
17
8. Ionizer (AX-ION-25)
The AX-ION-25 ionizer comes standard with the BA-6E / BA-6DE. The ionizer removes static electricity
by irradiating the target object with positive or negative ions, which are generated from four discharge
electrodes by DC corona discharge. Ordinarily, insulators such as powders, filters, and weighing paper
tend to be charged when the humidity is 45% RH or less, and an error of a few milligrams may occur
during weighing. By removing static from the weighing object with the ionizer, it is possible to eliminate
the error in the weighing value due to charging, and correct weighing can be performed. For details, refer
to the separate instruction manual in "1.1. Detailed manuals".
8.1. Using the ionizer
Step 1. Make sure that the AC adapter of the balance is not connected, and then connect the balance
and the ionizer (as noted in "2. Assembly and Installation").
Step 2. Connect the AC adapter, plug it in, and then turn on the power. The power lamp of the ionizer
lights up.
Step 3. Place the target object within the effective range of static elimination.
Step 4. Hold your hand over the IR sensor (touchless infrared proximity sensor) on the front of the
ionizer to start static elimination. The ION lamp indicating that static elimination is in progress
lights up.
Step 5. Refer to the figure below for the effective range of static elimination, and perform static
elimination. At factory setting, when 3 seconds have passed from the start of static elimination,
the ION lamp indicating that static elimination is in progress turns off and static elimination stops.
Effective range of static elimination
Top view
0 cm 5 cm 7 cm 10 cm
Distance
Side view
1 sec 1.5 sec 3 sec
Static elimination time
Effective range
Effective range
Ionizer
(AX-ION-25)
IR SENSOR (touchless infrared proximity sensor)
Power lamp (lights up when the power is turned on)
ION lamp (lights up during static elimination)
EN
18
9. Specifications
BA-6E
BA-6DE
Weighing capacity 6.2 g
6.2 g
2.1 g
Maximum display 6.200084 g
6.20008 g
2.100009 g
Readability 0.001 mg
0.01 mg
0.001 mg
Repeatability
(Standard deviation)
0.0030 mg (6 g) 0.01 mg (6 g)
0.0010 mg (1 g) 0.0025 mg (1 g)
Linearity ±0.010 mg
±0.02 mg
±0.010 mg
Display
Liquid crystal display (monochrome)
Operation
Key switches
Communication
RS-232C (printer, PLC, etc.), USB Type Mini-B (PC)
Stereo jack (external switch)
BA-225
BA-225D
BA-125D
Weighing capacity 220 g
220 g
120 g
51 g
51 g
Maximum display 220.00084 g
220.0008 g
120.0008 g
51.00009 g 51.00009 g
Readability 0.01 mg
0.1 mg
0.1 mg
0.01 mg 0.01 mg
Repeatability
(Standard deviation)
0.03 mg
(200 g) 0.1 mg
(200 g) 0.1 mg
(100 g)
0.015 mg
(50 g)
0.025 mg
(50 g)
0.030 mg
(50 g)
Linearity
±0.15 mg
±0.2 mg
±0.2 mg
Display
Liquid crystal display (monochrome)
Operation
Key switches
Communication
RS-232C (printer, PLC, etc.), USB Type Mini-B (PC)
Stereo jack (external switch)
FR
19
1. Guide de démarrage rapide
Nous vous remercions d'avoir acheté une balance analytique A&D de la série BA. Ce guide de démarrage
rapide décrit comment installer la balance de la série BA, et il couvre les fonctions basiques et les
opérations principales de la balance.
Pour d'informations sur la balance série BA, se référer aux manuels d'instructions séparés répertoriés
dans « 1.1. Manuels détaillés ».
1.1. Manuels détaillés
Les manuels détaillés peuvent être téléchargés depuis le site web d'A&D ( https://www.aandd.jp ) :
BA Series Instruction Manual (Manuel d'instructions de la série BA)
Ce manuel vous aidera à comprendre en détail les fonctions et les opérations de la série BA et à les
utiliser intégralement.
Ionizer (static eliminator) Instruction Manual (Manuel d'instructions de l'ionisateur
(éliminateur d'électricité statique))
Il s'agit du manuel décrivant les fonctions de l'ionisateur AX-ION-25 et les instructions d'utilisation.
2. Montage et installation
Attention
Une balance analytique de la série BA est un instrument de précision et elle doit être déballée avec
soin.
Il est conseillé de stocker le matériel d'emballage afin de pouvoir l'utiliser lors du transport de la
balance.
Le contenu de l'emballage varie en fonction du produit. Se référer à l'illustration du contenu de
l'emballage et s'assurer que tout est inclus.
Ne pas connecter l'adaptateur CA à la balance avant que la balance ne soit montée et installée.
Utiliser l'adaptateur CAdié spécifié pour la balance.
Ne pas connecter l'adaptateur CA fourni sur d'autres dispositifs.
Si un adaptateur CA incorrect est utilisé, la balance et les autres équipements peuvent ne pas
fonctionner correctement.
S'assurer que l'adaptateur CA est débranché avant de connecter l'unité de coupe-brise et l'ionisateur.
La balance analytique de la série BA est composée de l'unité de pesage et de l'unité d'affichage. Il
n'est pas possible de remplacer seulement une de ces unités.
FR
20
2.1. Déballage
Contenu de l'emballage ordinaire de la série BA
Contenu de l'emballage pour BA-6E / BA-6DE
Contenu de l'emballage pour BA-225 / BA-225D / BA-125D
Plateau B, Φ25
Bague du coupe-brise B
Plaque antipoussière B
Vitre avant du coupe-brise
Portes du coupe-brise
Unité de coupe-brise
Unité de pesage
Couvercle de protection de
l'unité d'affichage
Ce document
Câble USB
(environ 1,8 m)
Plaque
inférieure du
coupe-brise
Porte
supérieure du
coupe-brise
Pince
Unité
d'affichage
Câble B
(environ 1,0 m)
Câble A (environ 0,2 m)
Adaptateur CA
(environ 1,5 m)
Jeu de petites vitres de coupe-brise
Volet supérieur
Socle
Plateaux circulaires en aluminium (10 pièces de chaque)
Grand
plateau, Φ15,
0,8 mL
Portes du coupe-brise
Plateau
moyen, Φ12,
0,3 mL
Petit plateau,
Φ8, 0,05 mL
Câble C
(environ 0,6 m)
Ionisateur
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104

AND 162C-BA-BC BA Series Analytical Balances Guida utente

Tipo
Guida utente