TensCare MYND MIGRAINE RELIEF Manuale utente

Tipo
Manuale utente
0
1
Dear Customer,
Thank you for choosing Mynd. TensCare
stands for high-quality, thoroughly tested
products for the applications in the areas
of gentle electrotherapy, muscle toning,
continence management and pain relief
during labour.
Please read these instructions for use
carefully and keep them for later use, be
sure to make them accessible to other
users and observe the information they
contain.
Best regards,
Your TensCare Team
2
CONTENTS
1. INTENDED USE ................................. 5
2. CONTRAINDICATIONS, WARNINGS
& CAUTIONS ............................................... 5
3. PROGRAMMES ............................. 12
3.1. PROGRAMME SETTINGS ... 12
3.2. GEL PAD ADVICE .............. 13
4. CONTENT ....................................... 14
5. UNIT INFORMATION ...................... 14
5.1. CONTROLS ......................... 14
5.2. OPERATING INSTRUCTIONS
15
6. SETTING UP AND USING MYND ... 17
6.1. CHARGING THE BATTERY . 17
6.2. PREPARING FOR SESSION 21
6.3. TREATMENT SESSION ......... 23
6.4. AFTER YOUR TREATMENT
SESSION 24
7. CLEANING & STORAGE ............... 24
8. EMC ............................................... 25
9. DISPOSAL OF WASTE ELECTRICAL
AND ELECTRONIC PRODUCTS (WEEE) ... 25
10. ACCESSORIES ......................... 26
11. WARRANTY .............................. 27
12. TROUBLESHOOTING ................ 30
13. GENERAL SPECIFICATION ...... 31
3
SYMBOLS USED
TYPE BF APPLIED PART: Equipment
providing a degree of protection
against electric shock, with isolated
applied part. Indicates that this
device has
conductive contact
with the end user.
This symb
ol on the unit means
“Refer to instructions for use”.
Temperature Limitation: indicates
the temperature limits to which the
medical device can be safely
exposed.
Lot Number: indicates the
manufacturer’s batch code so that
the batch or lot can be identified.
Humidity Limitation: indicates the
humidity limits to which the medical
device can be safely exposed.
Serial Number: indicates the
manufacturer’s serial number so
that a specific medical device can
be identified.
Do not dispose in household waste.
Catalogue Number: indicates the
manufacturer’s catalogue number
so that the device can be
identified.
Atmospheric Pressure: indicates the
atmospheric limits to which the
medica
l device can be safely
exposed.
Medical Device
4
Manufacturer Symbol
Date of Manufacture: indicates the
date which the medical device
was manufactured. This is included
within the serial number found on
the travel case, either as
“E/Year/Number” (YY/123456) or
“E/Month/Year/Number”
(MM/YY/123456).
CE Mark
This medical device is indicated for
home use.
This medical device is not water
resistant and should be protected
from liquids.
The first number 2: Protected
against access to hazardous parts
with a finger, and the jointed test
finger of 12 mm ø, 80 mm length,
shall have adequate clearance
from hazardous parts, and
protected against solid foreign
objects of 12.5 mm ø and greater.
The second number 2: Protected
against vertically falling water
drops when enclosure is tilted up to
15˚. Vertical
ly falling drops shall
have no harmful effects when the
enclosure is tilted at any angle up
to 15˚ on either side of the vertical.
5
1. INTENDED USE
MYND is a medical device
designed to be used in the
home healthcare environment
for the treatment and prevention of
frequent episodic and chronic migraine
headaches.
It is suitable for use by all who can control
the device and understand the
instructions. It may be used by a child
from the age of 8 if supervised by an
adult, but there is currently no evidence
of effectiveness for users under the age of
18.
2. CONTRAINDICATIONS,
WARNINGS & CAUTIONS
In this manual:
A Contraindication is used when
a device should not be used
because the risk of use clearly
outweighs any foreseeable benefits and
may result in serious injury or death.
A Warning is used when failure to
follow the instructions may result
in serious injury or death.
A Caution is used when failure to
follow the instructions may result
6
in a minor or moderate injury, or damage
to the device or other property.
Notes are used to provide
clarification or recommendation.
CONTRAINDICATIONS:
Do not use if you have unresolved brain
or facial injuries.
Do not use MYND on any other areas
apart from the forehead. Use on the
chest or neck may affect heart or
implanted devices and may affect
epileptic seizure threshold. May also
affect body worn monitoring devices.
Do not use MYND if you have a cardiac
pacemaker, implanted defibrillator, or
other implanted metallic or electronic
device. Such use could cause electric
shock, burns, electrical interference, or
death.
Do NOT use if you have, in the area being
treated: active or suspected cancer or
undiagnosed pain with a history of
cancer. Stimulation directly through a
confirmed or suspected malignancy
should be avoided as it may stimulate
growth and promote spread of cancer
cells.
WARNINGS:
After a MYND session, wait an hour before
undertaking an activity that requires
7
concentration. You may experience
drowsiness or lack of concentration.
Do not place electrode pads on broken
skin. The electrode pads could
encourage infection.
Do not use MYND in close proximity (less
than 1 meter) to short wave or
microwave therapy equipment as it may
produce instability in MYND output.
It can be dangerous to connect MYND
with devices not described in this user
manual.
WARNING: Do not modify this equipment
without authorization of the
manufacturer.
Keep away from children under 5 years of
age. Long cord - risk of strangulation in
infants.
Do not use simultaneously with high
frequency hospital
diagnostic/therapeutic equipment.
Doing so may result in burns at the site of
the electrodes and possible damage to
device.
Do not apply MYND near the thorax
because the introduction of electrical
current may increase the risk of cardiac
fibrillation.
Stimulation should not be applied directly
on the eyes, covering the mouth, on the
front of the neck (especially the carotid
sinus), or from electrodes placed on the
upper back or crossing over the heart.
8
Do not apply stimulation across your
chest because the introduction of
electrical current into the chest may
cause rhythm disturbances to your heart,
which could be lethal.
This device is only used for forehead
surface skin. Do not use the device for
any purpose other than this intended use.
CAUTIONS:
Do not ignore any allergic reaction to the
electrode pads: If a skin irritation
develops, stop using MYND, as a small
number of users may have an adverse
reaction to the gel
Trigeminal neuralgia and Ophthalmic
shingles make facial nerves
hypersensitive, so use of MYND could be
painful.
Caution should be used if you have a
bleeding disorder as stimulation may
increase blood flow to the stimulated
region.
Not intended for use in an oxygen rich
environment.
Not intended for use in conjunction with
flammable anaesthetics or flammable
agents.
The patient is an intended operator.
Do not service and maintain the device
while in use with a patient.
9
Changing the gel is the only
maintenance the patient should perform.
Maintenance and all repairs should only
be carried out by an authorized agency.
The manufacturer will not be held
responsible for the results of maintenance
or repairs by unauthorized persons.
If necessary, we will provide circuit
diagrams, component part lists or other
information that will assist authorized
service personnel to repair the device.
Degraded gel pads can negatively
impact performance or cause other
problems.
Keep this device out of the reach of
children pets and pests.
Do not put the lead wire on or wrapped
around the neck.
SIDE EFFECTS
Intolerance to the paresthesia
caused by electrical stimulation
Intolerance to the strong pressure
on dental, forehead, cervical pains
and tension-type headache
Arousal and sleep changes
Vomiting and nausea at the end of
stimulation
Increased tinnitus during the session
Mild pain over one eyebrow,
numbness at the back of the head,
feeling of abrupt electrical
variation.
10
Discomfort during treatment: A
small portion of users may find
the stimulation to be very
uncomfortable. This feeling usually goes
away as you become used to it.
However, for some patients this
discomfort remains. If so the patient
should reduce the intesity.
Headache after Programme 2:
Very rarely you may experience
a headache after using
Programme 2. This is short-lived but may
make use of Programme 2 difficult for
some users.
Neck pain: At the start of
treatment you may tense your
neck muscles when you are
using MYND, and this can lead to an
aching neck. If this happens, try resting
your head on a pillow during treatment.
Note: Electrode pads supplied
have no measurable latex
content.
IF ANY OF THE SIDE EFFECTS BECOME
UNACCEPTABLE, PLEASE STOP USING THE
DEVICE IMMEDIATELY.
TO KEEP YOUR DEVICE IN GOOD
WORKING ORDER, OBSERVE THE
FOLLOWING ADDITIONAL CAUTIONS:
Caution: Do not immerse your device in
water or place it close to excessive heat
such as a fireplace or radiant heater or
sources of high humidity such as a
nebulizer or kettle as this may cause it to
cease to operate correctly.
11
Caution: Keep the device away from
sunlight, as long-term exposure to sunlight
may affect the rubber causing it to
become less elastic and crack.
Caution: Keep the device away from lint
and dust, as long-term exposure to lint or
dust may affect the socket
Caution: Temperature & Relative
Humidity of storage: -10°C to +60°C, 15%
to 75% R.H. Temperature & Relative
Humidity of transportation: -10°C to
+60°C, 15% to 75% R.H.
Caution: There are no user serviceable
parts. Do not attempt to open or modify
the device. This may affect the safe
operation of the unit and will invalidate
the warranty.
Caution: Silicone and backside gels on
the surface of the device may come into
contact with the patient or operator. If
allergic to these materials, avoid contact
and discontinue use.
12
3. PROGRAMMES
3.1. PROGRAMME SETTINGS
Programme 1 is designed to treat
migraine attacks and has a duration of 60
minutes. It should be applied when the
migraine is already present or, preferably,
when the first signs of the next migraine
attack are felt.
Note: Some migraine patients
develop Allodynia with their
migraine attacks. This makes any
skin stimulation, including MYND, feel
painful.
Programme 1 - Treatment
Maximum intensity
(across a load
resistance of 500 Ω)
20 mA (Peak)
Pulse frequency
110 Hz
Pulse width
200 μs
Pulse duration
200 μs
Session duration
60 minutes
Programme 2 - Prevention
Maximum intensity
(across a load
resistance of 500 Ω)
20 mA (Peak)
Pulse frequency
55 Hz
Pulse width
200 μs
Pulse duration
200 μs
Session duration
20 minutes
13
Programme 2 is designed to prevent
migraines and will last for 20 minutes. This
programme should be used once a day.
Regular use helps to reduce the
frequency of migraine attacks.
3.2. GEL PAD ADVICE
Gel pads are reusable but for single
patient use.
In order to get the best conductivity
through the gel pads, always make
sure they are in good condition and
sticky.
Make sure you clean the forehead
before use.
After use, always replace the clear
plastic liner on top of the gel pads on
the device. Keep the device and gel
pads in the travel case.
Replace the gel pads when they lose
their adhesion.
A bad connection can cause
discomfort and irritation of the skin.
The shelf life of an unopened packet
of gel pads is 2 years. This can be
affected by very high temperatures or
very low humidity.
14
4. CONTENT
The pack contains:
1 x MYND unit
3 x Reusable gel pads
1 x Charging cable
1 x Travel case
1 x Mirror
1 x Instruction manual
1 x Quickstart guide
5. UNIT INFORMATION
5.1. CONTROLS
ON/OFF
PROGRAMME
SELECTOR
INTENSITY
DOWN
INTENSITY
UP
INTENSITY
LEDS
15
5.2. OPERATING INSTRUCTIONS
ON/OFF
To turn the unit ON, press the
ON/OFF button of the unit and
hold for 1 second. You will hear a
long beep and the white LED lights will
switch on.
The unit will turn OFF automatically when
the session is complete (60 minutes in
programme 1 and 20 minutes in
programme 2). A long beep will indicate
the end of the session, the intensity will
gradually decrease and the device
switches off.
If you wish to manually turn the unit OFF,
press and hold the ON/OFF button for 3
seconds. You will hear a long beep (1
second) and the white LEDs light will
switch off.
Note: Always check unit is OFF
before applying or removing the
MYND device from the
forehead.
Note: Always clean the
forehead before using the
device.
PROGRAMME CONTROL
The ON/OFF button is also the
programme select button. MYND has two
pre-set programmes.
16
Every time the unit is switched on, the
programme 1 is automatically selected. If
you wish to change to programme 2
press and release the ON/OFF button.
Programme 2 selection will be indicated
by two short beeps.
To change back to programme 1, press
and release the ON/OFF button. A single
beep will indicate programme 1 has
been selected.
Every time a new programme is selected,
the strength level is reverted to zero. This
is a safety feature to alleviate any sudden
feeling of a surge, as each programme
gives a different sensation.
STRENGTH CONTROLS
The buttons marked ˄, ˅ are the strength
controls.
To increase the intensity in steps of 0.5 mA
press and release the ˄ button. To
increase automatically, press and hold
the ˄ button (fast increase in steps of 0.5
mA). To decrease the intensity in steps of
0.5 mA press and release the ˅ button. To
decrease automatically, press and hold
the ˅ button (fast decrease in steps of 0.5
mA).
A short beep will be heard with each
press of the strength controls.
When using the strength controls,
reaching maximum or minimum intensity
will be indicated by three quick beeps.
17
You may feel nothing over the first few
stages of increase. Continue pressing
until the sensation is strong but
comfortable. Further increases during use
may be necessary if your body becomes
used to the sensation. Note that the
maximum intensity of 20mA is sufficient to
have a therapeutic effect, even if you
feel like this intensity has become
tolerated.
INTENSITY LEDS
The unit has 40 levels of strength. The
strength level will be indicated by the
three LEDs on the bottom of the device.
Each LED gets brighter as the intensity
increases. The minimum and maximum
intensities are as follows:
Minimum intensityone dim LED
Maximum intensitythree bright LEDS
6. SETTING UP AND USING
MYND
6.1. CHARGING THE BATTERY
MYND is powered by a built-in
rechargeable Li-ion battery. A separate
charging cable with universal USB plug is
included in the kit.
18
When the battery is low, the central LED
will flash white. This would start about 4
hours before the battery is empty (4 hours
on stand-by).
Warning: Use only the charging
cable supplied. Use of other
chargers could be hazardous
and will negate the warranty.
To charge the battery:
Connect the USB cable to a mobile
phone or tablet charging socket or PC
USB port.
Whilst the battery is charging, the central
LED light will flash white slowly and
repeatedly.
The device will take approximately 2-2.5
hours to charge.
When the battery is fully charged, the
white LED light will stop flashing and
remain turned on.
Note: To minimise the reduction
in battery capacity the device
should be recharged yearly,
regardless of usage.
Note: The device cannot be
used while being charged.
Note: Charging voltage: DC 5V,
charging current: 60mA.
Note: Use adapters of the
following specifications.
Output: DC 5 V; 1A
19
Rated input voltage shall not
exceed 250 V
Classification of protection against
electric shock: Class II
IP classification: at least IP21
The adapter need to comply with
IEC 60601-1 or IEC 60950-1
Battery Life
The battery life will last approximately 22
hours at 50% maximum output intensity.
Battery Warnings
This product is equipped with a Lithium-
ion battery. Battery is fixed inside the
device and so device warnings also
apply to the battery. Failure to follow
these instructions could cause the lithium-
ion battery to leak acid, become hot,
explode or ignite and cause injury and
/or damage.
Do not pierce, open, disassemble or use
in a humid and / or corrosive
environment.
Do not expose to temperatures above
60 ° C (140F).
Do not put, store or leave near heat
sources, in direct sunlight, in a high
temperature location, in a pressurized
container or in a microwave oven.
Do not immerse in water or seawater and
do not apply to wet skin.
Do not short circuit.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140

TensCare MYND MIGRAINE RELIEF Manuale utente

Tipo
Manuale utente