12
Troubleshooting
Problem Possible causes Solution
Pump does
not work
• Empty battery • Recharge battery
Low suction
force
• Air leakage at connection
positions (Valve, diaphragm,
silicone breast shield, tubing)
• The output air path of breast
pump is blocked, e.g. the
suction pipe is pressed or
bent, or there are sundries in
the pipe
• No cause found
• Check all connecting
parts to make sure they
are installed properly
• Straighten or clean air
tubing
• Please contact supplier
No suction
• Valve is not installed or
damaged due to
deformation
• No case found
• Reinstall or replace
valve
• Please contact supplier
Warnings
• This product is a precision electronic appliance. Please do not use it during
bathing. Please keep it in a cool, dry and ventilated place, avoid direct
sunlight, and keep it out of the reach of children.
• Unplug the power adapter after use.
• This product is for personal use only. Do not use it between more than one
person to avoid cross infection.
• Do not use a breast pump during pregnancy, because early breastmilk
expression will stimulate uterine contraction which may lead to abortion.
• Do not forcibly remove breast shield from breast. Please power off and
gently release the seal between the breast and the breast shield, and
then remove breast shield from the breast.
• When disassembling or assembling the pump, make sure to assemble
each part according to the Installation Guide. When disassembling and
cleaning, please be careful not to misplace/lose the small parts.
• Set the suction level according to individual physique to ensure
comfortable milk expression.
• Please clean with water and disinfect before use.