Lovibond 00386452 - MD1x0/200 - Urea T Manuale utente

Tipo
Manuale utente
Lovibond® Water Testing
Tintometer® Group
Manual of Methods
MD 100 • MD 110 • MD 200
Urea
EN
Manual of Methods
Page 4
DE
Methodenhandbuch
Seite 18
ES
Manual de Métodos
Página 32
FR
Méthodes Manuel
Page 46
IT
Manuale dei Metodi
Pagina 60
PT
Métodos Manual
Página 74
NL
Handboek Methoden
Zijde 88
ZH
方法手册
Page 102
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Urea T / M390
ENMethod Reference Book1.06
EN
Urea T M390
0.1 - 2.5 mg/L Urea Ur1
Indophenol / Urease
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
UREA Reagent 1 15 mL 459300
UREA Reagent 2 10 mL 459400
Ammonia No. 1 Tablet / 100 512580BT
Ammonia No. 1 Tablet / 250 512581BT
Ammonia No. 2 Tablet / 100 512590BT
Ammonia No. 2 Tablet / 250 512591BT
Set Ammonia No. 1/No. 2 100 Pc.#100 each 517611BT
Set Ammonia No. 1/No. 2 250 Pc.#250 each 517612BT
Ammonia Conditioning Powder Powder / 26 g 460170
Urea Pretreat (compensates for the interference
of free Chlorine up to 2 mg/l)
Tablet / 100 516110BT
UREA Reagent Set 1 Set 517800BT
Preparation
1. The temperature of the sample should be between 20 °C and 30 °C.
2. The analysis must take place within one hour after taking the sample at the latest.
3. With the analysis of sea water samples, before the addition of Ammonia No. 1
Tablet, twoscoops of ammonium conditioning powder must be added to the sample
and dissolved by swirling.
Remarks
1. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2
Tablet has been added.
2. Ammonium and chloramines are accounted for in the urea determination.
Urea T / M390
ENMethod Reference Book1.0 7
EN
Determination of Urea with Tablet and Liquid Reagent
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
If free chlorine (HOCl)
is present, add a UREA
PRETREAT tablet.
Crush tablet(s) by rotating
slightly.
Close vial(s).
Urea T / M390
ENMethod Reference Book1.08
EN
2
Dissolve tablet(s) by
inverting.
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 2 drops Urea Reagenz
1.
1
Close vial(s). Invert several times to mix
the contents.
Add 1 drops Urea Reagenz
2.
Close vial(s). Invert several times to mix
the contents.
Press the ENTER button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
24 mm
Wait for 5 minute(s)
reaction time.
Add AMMONIA
No.1 tablet .
Crush tablet(s) by rotating
slightly.
Urea T / M390
ENMethod Reference Book1.0 9
EN
24 mm
Add AMMONIA
No.2 tablet .
Crush tablet(s) by rotating
slightly.
Close vial(s).
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
10 min
Wait for 10 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Urea appears on the display.
Urea T / M390
ENMethod Reference Book1.010
EN
Chemical Method
Indophenol / Urease
Appendix
Interferences
Persistant Interferences
Concentrations above 2 mg/L urea can lead to results within the measuring range. In
this case, the water sample must be diluted with water that is free from urea and the
measurement must be repeated (plausibility test).
Removeable Interferences
A UREA PRETREAT Tablet eliminates the interference of free chlorine up to 2 mg/L
(two tablets up to 4 mg/L, 3 tablets up to 6 mg/L).
Interference from / [mg/L]
Cl22
Bibliography
R.J. Creno, R.E. Wenk, P. Bohling, Automated Micromeasurement of Urea Using Urease
and the Berthelot Reaction, American Journal of Clinical Pathology (1970), 54 (6), p.
828-832
# including stirring rod, 10 cm
Urea T / M391
ENMethod Reference Book1.012
EN
Urea T M391
0.2 - 5 mg/L Ureai) Ur2
Indophenol / Urease
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
UREA Reagent 1 15 mL 459300
UREA Reagent 2 10 mL 459400
Ammonia No. 1 Tablet / 100 512580BT
Ammonia No. 1 Tablet / 250 512581BT
Ammonia No. 2 Tablet / 100 512590BT
Ammonia No. 2 Tablet / 250 512591BT
Set Ammonia No. 1/No. 2 100 Pc.#100 each 517611BT
Set Ammonia No. 1/No. 2 250 Pc.#250 each 517612BT
Ammonia Conditioning Powder Powder / 26 g 460170
Urea Pretreat (compensates for the interference
of free Chlorine up to 2 mg/l)
Tablet / 100 516110BT
UREA Reagent Set 1 Set 517800BT
Preparation
1. With the analysis of sea water samples, before the addition of Ammonia No. 1
Tablet, two scoops of ammonium conditioning powder must be added to the sample
and dissolved by swirling.
Urea T / M391
ENMethod Reference Book1.0 13
EN
Determination of Urea with Tablet and Liquid Reagent
Select the method on the device.
24 mm
5 mL
Sample
Put 5 mL sample and
5 mL of deionised water
in the sample vessel.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
If free chlorine (HOCl)
is present, add a UREA
PRETREAT tablet.
Crush tablet(s) by rotating
slightly.
Close vial(s).
Urea T / M391
ENMethod Reference Book1.014
EN
2
Dissolve tablet(s) by
inverting.
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 2 drops UREA
Reagenz 1.
1
Close vial(s). Invert several times to mix
the contents.
Add 1 drops UREA
Reagenz 2.
Close vial(s). Invert several times to mix
the contents.
Press the ENTER button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
24 mm
Wait for 5 minute(s)
reaction time.
Add AMMONIA No.
1 tablet .
Crush tablet(s) by rotating
slightly.
Urea T / M391
ENMethod Reference Book1.0 15
EN
24 mm
Add AMMONIA No.
2 tablet .
Crush tablet(s) by rotating
slightly.
Close vial(s).
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
10 min
Wait for 10 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Urea appears on the display.
Urea T / M391
ENMethod Reference Book1.016
EN
Chemical Method
Indophenol / Urease
i) high range by dilution | # including stirring rod, 10 cm
DE Methodenhandbuch 01/20
KS4.3 T
0,1 - 4 mmol/l KS4.3
Säure / Indikator
20
S:4.3
Instrumentspezifische Informationen
Der Test kann auf den folgenden Geräten durchgeführt werden. Zusätzlich sind die
benötigte Küvette und der Absorptionsbereich der Photometer angegeben.
Geräte Küvette λMessbereich
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0,1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0,1 - 4 mmol/l KS4.3
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
Alka-M-Photometer Tablette / 100 513210BT
Alka-M-Photometer Tablette / 250 513211BT
Anwendungsbereich
Abwasserbehandlung
Trinkwasseraufbereitung
Rohwasserbehandlung
Anmerkungen
1. Die Begrie Alkalität-m, m-Wert, Gesamtalkalität und Säurekapazität KS4.3 sind
identisch.
2. Die exakte Einhaltung des Probevolumens von 10 ml ist für die Genauigkeit des
Analysenergebnisses entscheidend.
Displayanzeige
im MD 100 MD 110
/ MD 200
Barcode zur
Methodenerkennung
Methoden Name
Messbereich
Chemische
Methode
Methodennummer
Sprachkürzel nach
ISO 639-1 Revisionsstand
KS4.3 T / 20
Leitfaden für den Anwender
ENMethod Reference Book18
DE
Durchführung der Bestimmung Säurekapazität KS4.3 mit Tablette
Die Methode im Gerät auswählen.
Für diese Methode muss bei folgenden Geräten keine ZERO-Messung durchgeführt
werden: XD 7000, XD 7500
10 ml
24-mm-Küvette mit 10 ml
Probe füllen.
Küvette(n) verschließen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Tablette(n) durch Um-
schwenken lösen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Test
Taste TEST (XD: START)
drücken.
In der Anzeige erscheint das Ergebnis als Säurekapazität KS4.3.
Durchführung der
Messung
KS4.3 T / 20
DE Methodenhandbuch 01/20
Leitfaden für den Anwender
ENMethod Reference Book 19
DE
Harnstoff T / M390
DEMethod Reference Book1.020
DE
Harnstoff T M390
0,1 - 2,5 mg/L Urea Ur1
Indophenol / Urease
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
UREA Reagenz 1 15 mL 459300
UREA Reagenz 2 10 mL 459400
Ammonia No. 1 Tablette / 100 512580BT
Ammonia No. 1 Tablette / 250 512581BT
Ammonia No. 2 Tablette / 100 512590BT
Ammonia No. 2 Tablette / 250 512591BT
Set Ammonia No. 1/No. 2#je 100 517611BT
Set Ammonia No. 1/No. 2#je 250 517612BT
Ammonium Konditionierpulver Pulver / 26 g 460170
Urea Pretreat (eliminiert die Störung von freiem
Chlor bis zu 2 mg/l)
Tablette / 100 516110BT
UREA Reagenzien Set 1 Satz 517800BT
Vorbereitung
1. Die Probentemperatur sollte zwischen 20 °C und 30 °C liegen.
2. Die Analyse spätestens eine Stunde nach der Probennahme durchführen.
3. Bei der Analyse von Meerwasserproben muss vor Zugabe der Ammonia No. 1
Tablette zwei Messlöffel Ammonium Konditionierungs-Pulver zur Probe gegeben
und durch Schwenken aufgelöst werden.
Anmerkungen
1. Die AMMONIA No. 1 Tablette löst sich erst vollständig nach Zugabe der AMMONIA
No. 2 Tablette auf.
2. Ammonium und Chloramine werden bei der Harnstoffbestimmung miterfasst.
Harnstoff T / M390
DEMethod Reference Book1.0 21
DE
Durchführung der Bestimmung Harnstoff mit Tablette und
Flüssigreagenz
Die Methode im Gerät auswählen.
10 mL
Sample
24-mm-Küvette mit 10 mL
Probe füllen.
Küvette(n) verschließen. Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Zero
Taste ZERO drücken. Küvette aus dem
Messschacht nehmen.
24 mm
Bei Anwesenheit von
freiem Chlor (HOCl)
eine UREA PRETREAT
Tablette zugeben.
Tablette(n) unter leichter
Drehung zerdrücken.
Küvette(n) verschließen.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117

Lovibond 00386452 - MD1x0/200 - Urea T Manuale utente

Tipo
Manuale utente