SETTING THE CHAMBER TEMPERATURE AND CO
2
DENSITY
Table below shows the basic procedure for setting the chamber temperature and CO
2
density. The
upper limit alarm temperature setting is also shown in the table. Perform key operations in the sequence
indicated in the table. The example in the table is based on the assumption that the desired temperature
is 37
o
C and CO
2
density is 5%. Adjustment of the upper limit regulator should be executed after the
chamber temperature reaches the stable condition.
Note:
• The incubator is set at the factory so that the chamber temperature is 37
o
C and CO
2
density is 0%.
• When starting the incubator for the first time or after not using it for an extended period of time, allow at
least 8 hours for the chamber temperature, humidity, and CO
2
sensor to become stable after setting the
desired chamber temperature at 0% CO
2
density. Then change the setting to the desired CO
2
density.
Basic operation sequence (Example: Chamber temperature: 33
o
C, CO
2
density: 5%)
Description of operation Key operated Indication after operation
1 Turn the power switch ON. ----
The current chamber temperature is displayed
in the digital temperature indicator.
2 Press set key. SET The left digit blinks.
When pressed,
the changeable digit is shifted.
3
By pressing numerical value shift
key and digit shift key,
set the figure to 37.0.
When pressed,
the figure of settable digit changes.
4 Press enter key. ENT
Set temperature is memorized.
Left digit in the digital
CO
2
density indicator blinks.
When pressed, the changeable digit is shifted.
5
By pressing digit shift key and
numerical value shift key, set the
figure to 05.0.
When pressed,
the figure of settable digit changes.
6 Press enter key. ENT Set CO
2
density is memorized.
7
(Executed after the chamber
temperature reaches the stable
condition)
Adjust upper limit regulator so that
the alarm temp. is 1
o
C higher than
set temperature.
----
In the digital CO
2
density indicator, HI is displayed.
In the digital temperature indicator, upper limit
temp. is displayed.
The upper limit temp. can be changed
by turning upper limit regulator.
8 Press enter key. ENT
This is the end of set mode and the indicators
display current temperature and CO
2
density.
Note:
● In each set mode, if the change of the setting is not necessary, pressing set key (SET) skips to next set
mode.
● When the CO
2
density is set to 00.0, the control is OFF regardless of chamber density.
● The upper limit temperature set value will change when the regulator is turned even if the incubator is
not in set mode, because the alarm circuit is an independent circuit.
● In each set mode, the indicator returns to the current temperature and CO
2
density display mode
automatically when 90 seconds has passed without any key operation.
18