Ohaus ISHD23HDG Manuale utente

Tipo
Manuale utente

Questo manuale è adatto anche per

E
N
EN
ES
FR
DE
IT
CZ
DA
NL
HU
PL
PT
SV
KR
CN
Incubating Shaker
ISHD16HDG
ISHD23HDG
ISHD23CDG
Instruction Manual
Incubating Shaker EN-1
Table of Contents
1. INTRODUCTION ...................................................................................................................................... 2
1.1. Safety Information .......................................................................................................................... 2
1.2. Intended Use .................................................................................................................................. 2
1.3. Package Contents .......................................................................................................................... 3
1.4. Installation ...................................................................................................................................... 3
1.5. Overview ........................................................................................................................................ 4
1.6. Control Panel .................................................................................................................................. 7
2 OPERATION ............................................................................................................................................. 8
2.1 Operation Instruction ...................................................................................................................... 8
2.2 Load Sensing Function ................................................................................................................... 9
2.2.1 Beeper Preference (Muting Audible Alarm) ....................................................................... 10
2.2.2 Temperature Calibration Procedure (Single Point) ............................................................ 10
2.2.3 To restore unit to factory setting ......................................................................................... 10
2.2.4 Speed Calibration Procedure ............................................................................................. 10
2.2.5 RS-232 Serial Port ............................................................................................................. 10
2.3 ISHD23CDG Upper Ambient Control Limit ................................................................................... 10
2.3.1 Heat/Cool Setting Procedure to Adjust Upper Ambient Control Limit ................................ 11
2.3.2 Additional ISHD23CDG Notes: .......................................................................................... 11
3 MAINTENANCE ...................................................................................................................................... 11
3.1 Troubleshooting ............................................................................................................................ 12
3.2 Service Information ...................................................................................................................... 12
4 TECHNICAL DATA ................................................................................................................................. 13
5 COMPLIANCE ........................................................................................................................................ 13
EN-2 Incubating Shaker
1. INTRODUCTION
This manual contains installation, operation and maintenance instructions for the Ohaus shakers. Please
read the manual completely before using.
1.1. Safety Information
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings.
Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false
results.
WARNING
For a hazardous situation with medium risk, possibly resulting in severe injuries or
death if not avoided.
CAUTION
For a hazardous situation with low risk, resulting in damage to the device or the
property or in loss of data, or minor or medium injuries if not avoided.
ATTENTION
For important information about the product. May lead to equipment damage if not
avoided.
NOTE
For useful information about the product.
Warning Symbols
General
hazard
Caution, hot
surface
Electrical
shock hazard
Safety Precautions
Please read the entire instruction manual before operating the Incubating Orbital Shaker.
WARNING: DO NOT use the Incubating Orbital Shaker in a hazardous atmosphere or with
hazardous materials for which the unit was not designed. In addition, the user should be aware
that the protection provided by the equipment might be impaired if used with accessories not
provided or recommended by the manufacturer, or used in a manner not specified by the
manufacturer.
Always operate unit on a level surface for best performance and maximum safety.
DO NOT lift unit by the tray, front bezel or lid.
CAUTION! To avoid electrical shock, completely cut off power to the unit by disconnecting the
power cord from the unit or unplug from the wall outlet. Disconnect unit from the power supply
prior to maintenance and servicing.
Spills should be removed promptly.
DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.
CAUTION! The caution hot indicator light warns that the temperature of the air in the chamber is
above 40°C. The light will illuminate and remain lit when the temperature of the air in the chamber
reaches approximately 40°C. When the heat is turned off, the caution hot indicator light will stay lit
until the temperature of the air in the chamber is less than 40°C.
Earth Ground – Protective Conductor Terminal.
Alternating Current
1.2. Intended Use
The Ohaus shaker is intended for general laboratory use. Safety cannot be guaranteed if used outside of
the intended use.
Incubating Shaker EN-3
1.3. Package Contents
Incubating Orbital Shaker
Non-skid rubber mat
Power cord
Instruction manual
1.4. Installation
Upon receiving the Ohaus Incubating Orbital Shaker, check to ensure that no damage has occurred during
shipment. It is important that any damage that occurred in transport is detected at the time of unpacking. If
you do find such damage the carrier must be notified immediately.
After unpacking, it requires at least two (2) people to lift the Incubating Orbital Shaker, from the bottom, to
place on a level bench or table, away from explosive vapors. It is preferable to use a hydraulic lift or other
appropriate equipment when handling the unit.
Do not lift unit by the front bezel. Ensure that the surface on which the unit is placed will withstand typical
heat produced by the unit. Always place the unit on a sturdy work surface.
The Incubating Orbital Shaker is supplied with a power cord that is inserted into the IEC connector on the
back or side of the unit first, then it can be plugged into a properly grounded outlet. The 120V unit plugs into
a 120 volt, 50/60 Hz source. The 230V unit plugs into a 230 volt, 50/60 Hz source.
This device complies with Part 15 rules.
EN-4 Incubating Shaker
1.5. Overview
Model:
ISHD16HDG
Controls:
see section 1.6
Overall dimensions (L x W x H):
25.5 x 14 x 16” (64.8 x 35.6 x 40.6cm)
Interior dimensions (L x W x H):
13.4 x 12 x 9.5” (34 x 30.5 x 24.1cm)
Tray dimensions (L x W):
13 x 11” (33 x 27.9cm)
Power:
120V, 50/60 Hz / 230V 50/60 Hz
Power consumption:
450 watts
Fuses:
5mm x 20mm, 5 amp quick acting
Temperature range:
ambient +5°C to 65°C
Temperature uniformity:
±0.5°C at 37°C
Speed range:
15 to 500rpm
Speed accuracy:
above 100rpm: ±1% of set speed / below 100rpm: ±1rpm
Timer:
1 second to 9999 minutes (increased in 1 second
increments)
Orbit:
0.75” (19mm)
Net Weight:
79 lbs (35.9 kg)
Ship weight:
83lbs (37.7kg)
Capacity:
~35lbs (16kg) @ 75rpm / ~5lbs (2.3kg) @ 500rpm
Figure 1-1 ISHD16HDG
Incubating Shaker EN-5
Model:
ISHD23HDG
Controls:
see section 1.6
Overall dimensions (L x W x H):
32.1 x 26.6 x 23.5” (81.5 x 67.6 x 59.7cm)
Interior dimensions (L x W x H):
20.6 x 24.8 x 17” (52.3 x 63 x 43.2cm)
Tray dimensions (L x W):
18 x 18” (45.7 x 45.7cm)
Power:
120V 50/60Hz / 230V 50/60Hz
Power consumption:
800 watts
Fuses:
5mm x 20mm, 8 amp quick acting
Temperature range:
ambient +5°C to 65°C
Temperature uniformity:
±0.5°C at 37°C
Speed range:
15 to 500rpm
Speed accuracy:
above 100rpm: ±1% of set speed / below 100rpm: ±1rpm
Timer:
1 second to 9999 minutes (increased in 1 second
increments)
Orbit:
1” (25mm)
Net Weight:
224 lbs (101.6 kg)
Ship weight:
228lbs (103.4kg)
Capacity:
~50lbs (22.7kg) @ 125rpm / ~10lbs (4.53kg) @ 500rpm
Communication:
RS232 Interface
Figure 1-2 ISHD23HDG
EN-6 Incubating Shaker
Model:
ISHD23CDG
Controls:
see section 1.6
Overall dimensions (L x W x H):
41.1 x 26.6 x 23.5” (104.4 x 67.6 x 59.7cm)
Interior dimensions (L x W x H):
20.6 x 24.8 x 17” (52.3 x 63 x 43.2cm)
Tray dimensions (L x W):
18 x 18” (45.7 x 45.7cm)
Power:
120V 50/60Hz / 230V 50/60Hz
Power consumption:
800 watts
Fuses:
5mm x 20mm, 10 amp quick acting
Refrigerant:
R404A, 7.1oz. by weight
Temperature range:
ambient -15°C to 65°C
Temperature uniformity:
±0.5°C at 37°C
Speed range:
15 to 500rpm
Speed accuracy:
above 100rpm: ±1% of set speed / below 100rpm: ±1rpm
Timer:
1 second to 9999 minutes (increased in 1 second
increments)
Orbit:
1” (25mm)
Net Weight:
290 lbs (131.7kg)
Ship weight:
295lbs (133.8kg)
Capacity:
~50lbs (22.7kg) @ 125rpm / ~10lbs (4.53kg) @ 500rpm
Communication:
RS232 Interface
Figure 1-2 ISHD23CDG
Incubating Shaker EN-7
1.6. Control Panel
The front panel of the Incubating Orbital Shaker contains all the controls and displays needed to operate the
unit
A. Standby button/standby indicator light: The standby indicator light will illuminate when the unit is
plugged in. The unit will be in standby mode. Press the standby button to activate the temperature,
speed and time functions. The standby indicator light will shut off and the temperature, speed and time
displays will illuminate. Press the standby button again and the unit will once again be in standby mode.
B. Temperature display: Displays the actual/set-point temperatures in conjunction with the actual/set-
point indicator lights.
C. Up/down arrows for set-point control. On/off button starts/stops the heating function.
D. The heat indicator light will be illuminated when the unit is heating.
E. Actual indicator light: Illuminates when the temperature displayed is the actual temperature of the air
in the chamber.
F. Set-point indicator light: Illuminates when the set-point temperature is displayed.
G. Caution hot indicator light: Illuminates when the air temperature of the chamber is above 40°C
(104°F).
H. Speed display: Displays the speed of the shaker.
I. Up/down arrows for set-point control. On/off button starts/stops shaking function.
J. The speed indicator light will be illuminated when the unit is shaking.
K. Time display: Displays accumulated time (continuous mode) or how much time is remaining (timed
mode). The display range is from 0 to 9,999 minutes in one (1) second increments. The display will
indicate minutes and seconds until the timer reaches 99 minutes and 59 seconds (99:59), then the
display will automatically display minutes up to 9,999.
L. Up/down arrows for set-point control. On/off button starts/stops the time function.
A
J
C
D
E
I
L
F
G
B
H
K
EN-8 Incubating Shaker
2 OPERATION
2.1 Operation Instruction
The Incubating Orbital Shaker has been designed for the temperature, speed and time functions to work
independently of one another. The temperature and speed can be reset without resetting the timer and the
timer can be stopped and started without interrupting the heating and shaking functions.
1. Getting ready:
a) Plug the power cord into a properly grounded outlet. The standby indicator light will illuminate,
verifying power to the unit.
b) Press the standby button to move the unit from standby mode. The standby indicator light will turn
off and the temperature, speed and time displays will illuminate, displaying the previously used
settings.
NOTE: ISHD23CDGunits have a toggle switch on the left side. This switch must be in the “on”
position (depress “I”) for the unit to function properly.
2. Setting temperature:
a) Press the up/down arrows below the temperature display until you reach the desired temperature.
When you release the button, the display will blink off and then on indicating the new set temperature
has been accepted.
b) Press the on/off button to start the heating function. The indicator light below the temperature
display will illuminate to indicate the heating function is in use and remain lit until heating has
ceased.
c) Temperature adjustments can be made without interrupting heating by using the up/down arrows
below the temperature display. After the change has been made and you release the button, the
display will blink off and then on indicating the new set temperature has been accepted.
d) To stop the heating function, press the on/off button below the temperature display. The heat
indicator light will turn off.
CAUTION HOT indicator:
The caution hot indicator light warns that the temperature of the air in the chamber is above
40°C (104°F). The light will illuminate and remain lit when the temperature of the air in the
chamber reaches approximately 40°C (104°F). When the heat is turned off, the caution hot
indicator light will stay lit until the temperature of the air in the chamber is less than 40°C
(104°F).
3. Setting speed:
a) Press the up/down arrows below the speed display until you reach the desired speed. When you
release the button, the display will blink off and then on indicating the new set speed has been
accepted.
b) Press the on/off button to start the shaking function. The indicator light below the speed display will
illuminate and blink until the set-point is reached. Once the set-point is reached the light will stop
blinking and remain lit until shaking has ceased. The microprocessor controlled ramping feature
slowly increases speed until the set-point is reached which helps to avoid splashing, and provides
excellent low end control.
c) Speed adjustments can be made without interrupting shaking by using the up/down arrows below
the speed display. After the change has been made and you release the button, the display will
blink off and then on indicating the new set speed has been accepted.
d) To stop the shaking function, press the on/off button below the speed display. The speed indicator
light will turn off.
4. Setting time to zero (0:00) and continuous mode: Accumulated time.
a) Press and hold the on/off button below the time display. After three (3) seconds the display will
indicate the previous set time.
Incubating Shaker EN-9
b) Simultaneously press both the up and the down arrows, the display will indicate zero (0:00). The
unit time is now set to zero (0:00) minutes. Alternately, you can use the up/down arrows to get to
zero (0:00).
c) Press the on/off button below the time display. The display will indicate accumulated time. The
up/down arrows will become inactive. To stop timer, press the on/off button again.
IMPORTANT: This will NOT interrupt the shaking or heating functions. Press the on/off
button
below the speed or temp displays to interrupt that function.
d) To reset, press and hold the on/off button below the time display. After three (3) seconds the
display will indicate the previous set time, which was zero (0:00).
5. Setting timed mode: Programmed time.
a) Press the up/down arrows below the time display until you reach the desired time.
b) Start this function by pressing the on/off button below the time display. The unit will run for the
selected time, the up/down arrows will become inactive while the timer is running. The unit will stop
shaking/heating when time display reaches zero (0:00). Four (4) audible beeps will indicate the
count down function is complete. The time display will default back to the set time. To repeat for the
same time, simply press the on/off button again.
c) To interrupt an automatic timing cycle before it is completed, press the on/off button below the time
display. The display will flash off and on to indicate the time function is on “hold”.
IMPORTANT: This will NOT interrupt the shaking/heating functions. Press the on/off button below
the speed or temp displays to interrupt these functions. Restart the timer by pressing the on/off
button below the time display. Unit will continue counting down to zero (0:00). When the display
reaches zero (0:00), you will hear the four (4) audible beeps that indicate the count down function is
complete and the shaking/heating function will cease.
6. Turning unit off:
To turn the unit off, press the standby button. The temperature, speed and time displays will be blank,
the standby indicator light will illuminate. The Incubating Orbital Shaker should be kept in standby mode
when not in use. To completely cut off power to the unit, disconnect the power cord from the unit or
unplug from the wall outlet.
Operating tips
Opening the lid on the Incubating Orbital Shaker will cause the unit to pause shaking and/or
heating. Close the lid and the unit will resume shaking and/or heating at current settings.
Centering your sample and even weight distribution on the tray helps with balance and stability.
When possible, samples should be covered to prevent excess condensation inside the incubation
chamber. Should condensation occur, the use of a desiccant is recommended.
As a safety feature, a built-in program will shut power off to the motor if the tray is prevented from
rotating, or the unit is overloaded beyond its recommended weight capacity.
The shaker will automatically restart after a power interruption. Built-in memory maintains the last
used temperature, speed and time settings during a power interruption.
2.2 Load Sensing Function
The Incubating Orbital Shaker is equipped with a load sensing function that can be activated by the user.
This function provides protection against improper positioning of load and maximum load being exceeded.
When activated, the unit will automatically sense improper load conditions and slow to a safe running speed,
then display that speed followed by an E04 error message on the speed display. The unit will also beep
three (3) times every 60 seconds until the error is reset by pressing the speed on/off button.
To activate the load sensing function use the following steps:
1 Place the unit in standby mode.
2 Press and hold the speed on/off button and then press and release the standby button. The unit will
beep two (2) times, confirming the function is enabled.
3 To restore normal operation, remove AC power to the unit for ten (10) seconds and then restore.
EN-10 Incubating Shaker
If the E04 error occurs, be sure the load is within the maximum specification and properly balanced (centered
on tray) and/or reduce sample size/speed before restarting the unit. If the E04 error occurs due to
acceptable sample vibration or another vibration source, the vibration sensing function can be disabled as
described above.
2.2.1 Beeper Preference (Muting Audible Alarm)
To silence beeper operation (except for error codes), with the unit in standby mode, press and hold the time
on/off button and press the standby button. Release the standby button first, and then release the on/off
button. To restore normal beeper operation, remove AC power to unit for ten (10) seconds and then restore.
2.2.2 Temperature Calibration Procedure (Single Point)
This procedure is used to fine tune and calibrate the Incubating Orbital Shaker at a specific temperature
setting. This process may be repeated for up to three (3) separate set-points. If a fourth calibration set-point
is entered, the first set-point entered will be overwritten.
1 Turn unit on.
2 Set desired temperature.
3 Stabilize one (1) hour or more, measuring the temperature with a temperature probe/thermometer
centrally located in the chamber.
4 Press and hold standby button, then press the temperature up arrow once. The unit will beep two (2)
times, confirming calibration mode. The display will now be flashing.
5 Press the temperature up/down arrows until the display on the unit matches the external temperature
probe/thermometer. (Example: Desired temperature is 37ºC. Set unit at 37ºC per step 2. Follow steps 3
& 4. Display reads 37ºC and the external temperature probe/thermometer reads higher at 39ºC. Push
the temperature up arrow so the display will match the external temperature device and also read 39ºC.
By doing this a biased offset for 37ºC will be used when unit is set to 37ºC.)
6 Press standby button to exit calibration mode and return to normal heating.
This process may be repeated at the same set-point, multiple times for fine tuning if desired.
The unit will now use the biased offset for that specific temperature setting and increase or decrease
temperature accordingly to bring the chamber temperature to set-point. The decimal point of the display
will flash to indicate a biased offset is being used. All other temperature settings will use the standard
internal calibration. This offset will be stored in memory and retained until reset.
2.2.3 To restore unit to factory setting
Press and hold the standby button while pressing the temperature down arrow once. The reset will be
confirmed with two (2) audible beeps. Press the standby button to exit calibration mode and return to normal
heating.
2.2.4 Speed Calibration Procedure
This procedure is used to self calibrate the symphony™ Orbital Shakers.The tray should be free of any
samples, vessels, and accessories prior to calibrating.
1 Turn unit on. Speed and time displays will be illuminated.
2 Press and hold the standby button and momentarily press the speed on/off button. The speed display
should read. “CAL”.
3 The unit will run for approximately one (1) minute and automatically calibrate.
2.2.5 RS-232 Serial Port
The RS-232 serial port provides two-way communications for data logging and unit control. If you need
additional details, please contact your Ohaus representative for support.
NOTE: To comply with EMI/EMC requirements, the length of the external cable shall not exceed 3 meters.
2.3 ISHD23CDG Upper Ambient Control Limit
Please follow the procedure below if setting the ISHD23CDG unit’s temperature at/near the room’s ambient
temperature. For optimum temperature control in the chamber of the ISHD23CDG unit, the upper ambient
control limit is to be set a minimum of 3ºC above the measured room temperature.
Example: Measured (ambient) temperature is 22ºC; the upper ambient control limit should be set at a
minimum of 25ºC.
Incubating Shaker EN-11
With the unit at the heat/cool setting as described below, and in cool mode (“cool” on time display), a
temperature setting is displayed in the temperature window. The default temperature setting of the
ISHD23CDG is 28ºC. This is the factory set upper ambient control limit. This setting can be adjusted
between 20ºC and 32ºC by the user.
Any unit set point above the upper limit setting will heat only.
Any unit set point below 20ºC will cool only.
Any set point between 20ºC and the selected upper limit setting, up to 32ºC, the chiller and heater
will both control chamber temperature.
2.3.1 Heat/Cool Setting Procedure to Adjust Upper Ambient Control Limit
1 Put unit in standby mode.
2 Press and hold the temperature down, speed down and time down arrows simultaneously, now press
the standby button. Release all standby button and then release all 3 of the down arrows
simultaneously.
3 Time display will now read “cool” and the temperature display will read 28°C. If the time display reads
“heat” press the time up/down arrow until the word “cool” appears in the display.
4 Now you can set the upper ambient control limit. Using the temperature up/down arrows adjust this
temperature to be 3°C above the measured ambient temperature.
5 Once the desired temperature is selected, press the standby button to return to normal operation.
6 This procedure does not require recalibrating the unit.
2.3.2 Additional ISHD23CDG Notes:
ISHD23CDG units have a toggle switch on the left side. This switch must be in the “on” position
(depress “I”) for the unit to function properly.
For ISHD23CDG units, make sure the chiller’s hose is positioned to drain freely (into a sink when
possible).
Do not allow the end of the chiller hose to become submerged.
If chiller hose is positioned in a drain container, the
container
should have a volume of at least 2 gallons
if the unit will be unattended for 48 hours.
There are two strips of adhesive on the bottom of the rubber mat. To prevent the mat from sliding on the
tray during operation, remove the plastic film to expose the adhesive and firmly press the mat down
onto the tray to secure.
3 MAINTENANCE
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause undesired
operation.
The Incubating Orbital Shaker is built for long, trouble-free, dependable service. No lubrication or other
technical user maintenance is required. However at least every three (3) months you should:
Unplug the unit.
Remove any accumulated dirt from the base and tray.
Check all accessible items to make sure they are properly tightened.
The unit should be given the care normally required for any electrical appliance. Avoid wetting or
unnecessary exposure to fumes. Spills should be removed promptly.
Do not use a cleaning agent or solvent on the front panel or lid, which is abrasive or harmful to plastics, nor
one which is flammable. Always ensure the power is disconnected from the unit prior to any cleaning. If the
unit ever requires service, contact your Ohaus representative.
EN-12 Incubating Shaker
3.1 Troubleshooting
The following table lists common problems and possible causes and remedies. If the problem persists,
contact OHAUS or your authorized dealer.
.
Problem
Cause
Solution
Unit fails to power
on
Missing or blown fuse
Check and replace fuse if necessary. If problem
persists, please contact your Ohaus
representative for repair.
Unit is excessively
noisy
Rattling or ticking sounds may
indicate a loose screw on the
tray
Ensure the tray is secured tightly. If problem
persists, please contact your Ohaus
representative for repair.
Unit not shaking at
proper speed
-
Perform speed calibration. If problem persists,
please contact your Ohaus representative for
repair.
Unit not heating to
proper temperature
-
Perform single point calibration. If problem
persists, please contact your Ohaus
representative for repair.
E1
RTD open or ±C
temperature deviation from
set-point (after unit has
stabilized)
This error cannot be fixed by the end users. If
problem persists, please contact your Ohaus
representative for repair.
E2
RTD shorted or temperature
below 0°C
This error cannot be fixed by the end users. If
problem persists, please contact your Ohaus
representative for repair.
E3
Mechanical obstruction
Drive system failure
Ceased bearing
Drive belt broken
Remove mechanical obstruction. If problem
persists, the reason may be the drive system.
Please contact your Ohaus representative for
repair.
E4
Improper positioning of load or
Maximum load exceeded
Ensure the load is evenly distributed and does
not exceed the maximum load capacity for the
unit. See “Load Sensing Function” section.
Verify load has not exceeded the maximum
allowable weight. Reduce load if necessary. If
error code resumes, contact your Ohaus
representative for repair.
E6
Over temperature error
This error cannot be fixed by the end user.
Please contact your Ohaus representative for
repair.
3.2 Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS
service agent. For service assistance or technical support in the United States call toll-free 1-800-672-7722
ext. 7852 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be available to
provide assistance. Outside the USA, please visit our website, www.ohaus.com to locate the Ohaus office
nearest you.
Serial Number: __________________________
Date of Purchase: __________________________
Supplier: __________________________
Incubating Shaker EN-13
4 TECHNICAL DATA
Operating Conditions: Indoor use only.
Temperature: 15 to 32°C (59 to 90°F)
Humidity: maximum 80% relative humidity, non-condensing
Altitude: 0 to 6,562 ft (2000 M) above sea level
Non-Operating Storage:
Temperature: -20 to 65°C (-4 to 149°F)
Humidity: maximum 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664
5 COMPLIANCE
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark
Standard
This product complies with the applicable harmonized standards of EU Directives
2011/65/EU (RoHS), 2014/30/EU (EMC) and 2014/35/EU (LVD). The EU Declaration of
Conformity is available online at www.ohaus.com/ce.
This product complies with the EU Directive 2012/19/EU (WEEE). Please dispose of
this product in accordance with local regulations at the collecting point specified for
electrical and electronic equipment.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
c
EN 61326-1
CAN/CSA-C22.2 No. 61010-1, CAN/CSA-C22.2 No. 61010-2-010, CAN/CSA-C22.2 No.
61010-2-051
UL 61010-1, UL 61010-2-010, UL 61010-2-051
ISED Canada Compliance Statement
This Class A digital apparatus complies with Canadian ICES-003.
ISO 9001 Registration
The management system governing the production of this product is ISO 9001 certified.
EN-14 Incubating Shaker
FCC Supplier Declaration of Conformity
Unintentional Radiator per 47CFR Part B
Trade Name: OHAUS CORPORATION
Model or Family identification: ISHD16HDG / ISHD23HDG / ISHD23CDG
Party issuing Supplier’s Declaration of Conformity:
Ohaus Instruments (Changzhou) Co., Ltd.
2F, 22 Block, 538 West Hehai Road, Xinbei District, Changzhou
Jiangsu 213022,
China
Phone: +86 519 85287270
Responsible Party U.S. Contact Information:
Ohaus Corporation
7 Campus Drive, Suite 310
Parsippany, NJ 07054
United States
Phone: +1 973 377 9000
Web: www.ohaus.com
FCC Compliance Statement:
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may
cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the interference at his own expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
Incubating Shaker EN-15
LIMITED WARRANTY
OHAUS products are warranted against defects in materials and workmanship from the date of delivery
through the duration of the warranty period. During the warranty period OHAUS will repair, or, at its option,
replace any component(s) that proves to be defective at no charge, provided that the product is returned,
freight prepaid, to OHAUS.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to
radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result
of service or modification by other than OHAUS. In lieu of a properly returned warranty registration card, the
warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied
warranty is given by OHAUS Corporation. OHAUS Corporation shall not be liable for any consequential
damages.
As warranty legislation differs from state to state and country to country, please contact OHAUS or your
local OHAUS dealer for further details.
Agitador de incubación ES-1
Tabla de contenido
1. INTRODUCCIÓN ...................................................................................................................................... 2
1.1. Información de seguridad ............................................................................................................... 2
1.2. Uso previsto ................................................................................................................................... 3
1.3. Contenido del paquete ................................................................................................................... 3
1.4. Instalación ...................................................................................................................................... 3
1.5. Visión general ................................................................................................................................. 4
1.6. Panel de control ............................................................................................................................. 7
2 OPERACIÓN ............................................................................................................................................ 8
2.1 Instrucciones de operación ............................................................................................................ 8
2.2 FUNCIÓN DE DETECCIÓN DE CARGA ..................................................................................... 10
2.2.1 Preferencia de beeper (Silenciamiento de la alarma audible) ........................................... 10
2.2.2 Procedimiento de calibración de temperatura (Punto único) ............................................. 10
2.2.3 Para restaurar la unidad a la configuración de fábrica: ...................................................... 11
2.2.4 Procedimiento de calibración de la velocidad: ................................................................... 11
2.2.5 Puerto serial RS-232 .......................................................................................................... 11
2.3 ISHD23CDG Límite de control ambiental superior ....................................................................... 11
2.3.1 Procedimiento de ajuste de calor/frío para ajustar el límite superior de control ambiental 11
2.3.2 Notas adicionales del ISHD23CDG: .................................................................................. 11
3 MANTENIMIENTO .................................................................................................................................. 12
3.1 Solución de problemas ................................................................................................................. 13
3.2 Servicio de información ................................................................................................................ 13
4 DATOS TÉCNICOS ................................................................................................................................ 14
5 CUMPLIMIENTO .................................................................................................................................... 14
ES-2 Agitador de incubación
1. INTRODUCCIÓN
Este manual contiene instrucciones de instalación, operación y mantenimiento para el agitador Ohaus. Lea
el manual completamente antes de usar.
1.1. Información de seguridad
Las notas de seguridad están marcadas con palabras de señales y símbolos de advertencia. Estos
muestran los problemas y advertencias de seguridad. Ignorar las notas de seguridad puede causar lesiones
personales, daños al instrumento, funcionamiento incorrecto y resultados incorrectos.
ADVERTENCIA
Para una situación peligrosa con riesgo medio, posiblemente cause lesiones graves
o la muerte si no s evita.
PRECAUCIÓN
Para una situación peligrosa con riesgo bajo, que cause daños al dispositivo, a la
propiedad, una pérdida de datos, o lesiones menores o moderadas si no se evita.
ATENCIÓN
Para información importante sobre el producto. Puede causar daño al equipo si no se
evita.
NOTA
Para información útil sobre el producto.
mbolos de advertencia
Peligro general
Precaución,
superficie caliente
Peligro de
descarga eléctrica
Instrucciones de seguridad
Lea todo el manual de instrucciones antes de utilizar el Agitador orbital de incubación.
¡ADVERTENCIA! NO use el agitador orbital de incubación en una atmósfera peligrosa o con
materiales peligrosos para los que la unidad no fue diseñada. Además, el usuario debe saber que
la protección provista por el equipo puede verse afectada si se usa con accesorios no provistos o
recomendados por el fabricante , o usado de una manera no especificada por el fabricante.
Siempre opere la unidad en una superficie nivelada para obtener el mejor rendimiento y la
máxima seguridad.
NO levante la unidad por la bandeja, el marco frontal o la tapa.
¡PRECAUCIÓN! Para evitar descargas eléctricas, desconecte completamente la alimentación de
la unidad desconectando el cable de alimentación de la unidad o desenchúfela de la toma de
corriente. Desconecte la unidad de la fuente de alimentación antes del mantenimiento y servicio.
Los derrames deben eliminarse de inmediato. NO sumerja la unidad para limpiarla.
NO opere la unidad si muestra signos de daños eléctricos o mecánicos.
¡PRECAUCIÓN! La luz indicadora de precaución caliente advierte que la temperatura del aire en
la cámara es superior a 40 ° C. La luz se iluminará y permanecerá encendida cuando la
temperatura del aire en la cámara alcance aproximadamente 40 ° C. Cuando se apaga el calor, la
luz indicadora de precaución permanecerá encendida hasta que la temperatura del aire en la
cámara sea inferior a 40 ° C.
Tierra - Terminal de conductor de protección
Corriente alterna
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124

Ohaus ISHD23HDG Manuale utente

Tipo
Manuale utente
Questo manuale è adatto anche per