NSK PER-WH-P Istruzioni per l'uso

Tipo
Istruzioni per l'uso

Questo manuale è adatto anche per

1
English
Thank you for purchasing NSK Perio Mate.
This product is a handpiece for cleaning sub- and supragingival tooth surfaces.
Please read this Operation Manual carefully before use to become familiar with operation instructions and
care & maintenance. Keep this Operation Manual for future reference.
-Contents-
1. Features
.......................................................................................................
1
2. User and Indications for Use
.................................................................................
1
3. Precautions for handling and operation
.....................................................................
2
4. Correspondence Coupling
....................................................................................
4
5. Package Contents
.............................................................................................
4
6. Operation
......................................................................................................
5
7. Care After Use
................................................................................................
10
8. Cleaning and Sterilization
...................................................................................
14
9. Periodical Maintenance Checks
.............................................................................
16
10. Troubleshooting
.............................................................................................
17
11. Specifications
...............................................................................................
20
12. Symbols
.....................................................................................................
20
13. Warranty
....................................................................................................
20
14. Option Parts List
............................................................................................
21
15. Spare Parts List
.............................................................................................
21
16. Disposing Product
..........................................................................................
21
1
Features
Ř The 360°twist free swivel coupling enables comfortable and fatigueless operation even during prolonged
procedures, allowing effortless access to difficult areas.
<Powder for cleaning the tooth surface>
Ř NSK Perio Mate Powder used in conjunction with Perio Mate can remove plaque and eliminate sub- and
supragingival biofilm quickly and efficiently.
Ř The main ingredient of Perio Mate Powder is glycine, a type of amino acid.
2
User and Indications for Use
User: Dentist, Dental hygienist
Indications for Use: Perio Mate is for sub- and supragingival plaque removal. When applied subgingivally, it
may be used to support periodontitis and periimplantitis therapy.
English
2
3
Precautions for handling and operation
Please read these precautions carefully and use only as intended or instructed.
Safety instructions are intended to avoid potential hazards that could result in personal injury or damage
to the device. Safety instructions are classified as follows in accordance with the seriousness of the risk.
Degree of Risk
Hazard that could result in serious injury or damage to the device if the
safety instructions are not correctly followed.
Hazard that could result in light or moderate injury or damage to the device
if the safety instructions are not correctly followed.
General product specification information highlighted to avoid product
malfunction and performance reduction.
Class
WARNING
CAUTION
NOTICE
WARNING
Ř If any hypersensitivity such as eruption and dermatitis develops due to treatment using this product,
stop using the product and seek medical attention.
Ř The Perio Mate system can be used in periodontal pockets. Therefore, attention is required for patients
with emphysema. If emphysema develops due to treatment using this product, immediately stop using
the product and seek medical attention.
Ř This instrument is for use by dentists or hygienists only.
Ř Do not use on the following patients
1) Those with respiratory diseases.
2) Those with pathologically deep periodontal pockets (6 mm or more) or mucosal lesion. Prolonged
application of the spray to soft tissue or salivary glands may cause powder air to enter, which may
cause emphysema.
3) Those with severe ulcers in digestive organs.
4) Those with hepatic dysfunction.
5) Those with lung or cardiac disturbance or ventilatory impairment.
6) Those with damage or abnormality in the oral cavity.
7) Those with hyperemia, hemorrhage, or inflammation in the oral cavity.
8) Those that tend to develop an inflammation or sores in the oral mucous membrane.
9) Those with allergies.
10) Depending on the patient’s constitution, allergy symptoms or stomatitis may be caused in very rare
cases. If such symptoms are detected, stop using the product immediately.
Ř Do not use Perio Mate with the patients listed below. When used in periodontal pockets for these
patients, this product may develop bacillemia.
-Patients with endocardial inflammation
-Patients pregnant or lactating
-Patients with a contagious illness or disease
-Patients with an immune deficiency disease (neutropenia, granulocytopenia, agranulocytosis,
diabetes, hemophilia)
-Patients under irradiation treatment, chemotherapy, or administration of an antibiotic
Ř Use this product only for patients with healthy periodontal pockets for maintenance after treatment of
a gum disease. When used for any patients other than those, this product may cause emphysema.
Ř Use this product only in periodontal pockets that meet the following conditions:
-Periodontal pockets caused by a gum disease
-Periodontal pockets of 6 mm or less
Ř Check the depth of periodontal pockets with a measurement probe or roentgenography before use.
3
English
Ř Use Perio Mate Powder as cleaning powder. If other powder is used, the Perio Mate Nozzle Tip will
be damaged and it will increase the risk of damaging substantia ossea and dentine or causing
emphysema. We shall assume no liability for any failures caused by the use of other powder.
Ř During use, both the operator and the assistant should always wear protective glasses and protective
face masks. Also, continuously vacuum and collect ejected powder during use. In case any powder
gets in the eyes wash out immediately with copious amount of water and consult an ophthalmologist.
Ř Set supply air pressure at hand to 0.3 - 0.4 MPa (3.0 - 4.0 kgf/cm
2
). If this product is used with
incorrect pressure, the operator, assistant or patient may be injured due to a damaged Powder Case
Cover.
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize it since it may be damaged or
contaminated.
Ř Make sure that the Nozzle Tip package is not open or damaged before use. If the package is open or
damaged, do not use that Nozzle Tip as it may be damaged or contaminated.
Ř Follow the use-by date on the Nozzle Tip package. If the use-by date has lapsed, do not use that
Nozzle Tip as it may be damaged or contaminated.
CAUTION
Ř Read this Operation Manual before use to fully understand the product functions, and file for future
reference.
Ř When operating the product always consider the safety of the patient.
Ř The end user shall be responsible for any judgment that relates to the application of this product to a
patient.
Ř Users are responsible for the operational control, maintenance and continual inspection of this product.
Ř Do not attempt to disassemble the product nor tamper with the mechanism except as recommend by
NSK in this Operation Manual.
Ř This product is a precision instrument. Do not allow any impact on to the product. Do not drop the
product. Even the slightest impact may damage and deform the product.
Ř Should the product function abnormally, cease operation immediately and contact your Authorized NSK
Dealer.
Ř Be very careful to avoid cleaning powder entering a patient’s mucosal areas (eyes, nose etc) other
than the oral cavity. Also, protect the face with a towel, etc., or protective glasses, etc., to prevent entry
of cleaning powder into eyes.
*If the patient wears contact lenses, have the patient wear safety glasses or remove his/her contact
lenses before treatment.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the air may cause
internal solidification of the cleaning powder.
Ř Securely attach the Powder Case Cap. Do not open the Powder Case Cap during use. The Powder Cap
and internal powder may be ejected since pressure is high in the Powder Case.
Ř Before connecting the Powder Case to the Handpiece Hose, be sure to blow off water remaining
on the Hose connection with dry air. Residual moisture may enter the supply air, leading to internal
solidification of the cleaning powder.
Ř If too much cleaning powder is put into the Powder Case, the proper amount of powder may not be
sprayed immediately after the start of use.
Ř The Perio Mate Powder has the property of easily scattering because of its fine particles. Fill the
chamber slowly and be careful to prevent the powder from scattering.
Ř Powder stored for a long while in opened packages can attract moisture and the powder may clog in
the Nozzle, etc. Make sure to use new dry powder. Securely close the lid of the Perio Mate Powder
container after filling the powder.
Ř Apply Vaseline, etc., to the patient’s lips to prevent drying or chapping at the corners of the mouth.
Ř Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
4
CAUTION
4
Correspondence Coupling
Joint Type Model
NSK PHATELUS Coupling, MACH Coupling PER-PTL-P
KaVo
®
MULTIflex
®
, MULTIflex
®
LUX Coupling Type PER-KV-P
W&H
®
Roto Quick
®
Coupling Type PER-WH-P
Bien-Air
®
Unifix
®
, Unifix
®
L Coupling Type PER-BA-P
Sirona
®
Quick Coupling Type PER-SR-P
ISO 9168 Type2 (Midwest 4-hole) Joint PER-M4-P
ISO 9168 Type1 (Borden 2-hole) Joint PER-B2-P
5
Package Contents
Handpiece 1 Set
Powder Case (With Powder Case Cap) 1 Set
Perio Mate Nozzle Tip 40 pcs.
Perio Mate Powder 1
Maintenance Kit
Cleaning File (For Nozzle tip, For Internal Nozzle) 1
Cleaning Wire (S) (For the root of the Nozzle) 1
Cleaning Wire (L) (for other uses) 1
Cleaning Brush 1
Perio Mate Tip Remover 1
Ring Wrench 1
Turbine Oil 1
Blower Nozzle 1
(The shape varies depending on the Joint Type.)
Spare parts
Powder Case Cover 1
O-Ring (Handpiece End) 1 Set
O-Ring (Coupling End) (NSK PHATELUS coupling ONLY) 1 Set
Ř The Nozzle, Handpiece, Powder Case, and Blower Nozzle cannot be used for NSK
Prophy-Mate neo.
Ř Perform regular function and maintenance checks.
Ř If the device has not been used for a long time, check that it functions correctly and safely before
using it on a patient.
Ř U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
KaVo
®
and MULTIflex
®
are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany.
Sirona
®
is a registered trademark of Sirona Dental System GmbH, Germany.
W&H
®
and Roto Quick
®
are registered trademarks of W&H dentalwerk Burmoos GmbH, Austria.
Bien-Air
®
and Unifix
®
are registered trademarks of Bien-Air Dental S.A., Switzerland.
5
English
WARNING
CAUTION
Component Names
6
Operation
Ř Avoid prolonged continuous use, and allow the patient to rinse at appropriate times.
Ř Never impact the Powder Case and Powder Case Cap as this may cause the Powder
Case or Powder Case Cap to scratch or crack.
Ř Never take the Perio Mate apart during operation. Never pull the Handpiece Release
Collar or the coupling release during operation. Attempting to do this may cause
components of the Perio Mate to unexpectedly detach and the Perio Mate Powder to
scatter.
Ř Do not open the Powder Case Cap while using the Perio Mate. Powder may scatter.
Ř Use this product by setting the pressure to 0.3 - 0.4 MPa (3.0 - 4.0 kgf/cm
2
). Using the
product at an excessive pressure increases the risk of causing emphysema.
Ř Thoroughly dry the Perio Mate and all its components inside and out before use.
Insufficient drying may cause clogging of powder.
Ř Be very careful to avoid Perio Mate Powder entering patient’s mucosal areas (eyes,
nose etc) other than the oral cavity. Also, protect the face with a towel, etc., or protective
glasses, etc., to prevent entry of powder into eyes.
Ř If chemicals, oils, etc, adhere to the Powder Case Cover, wipe them off immediately as
they could cause damage.
Cover Holder
Powder Case Cover
Rubber Packing
Powder Case Cap
Hose Connection
Handpiece Joint
O-Ring
Flow Volume
Adjustment Ring
Handpiece Release Collar
Powder Case
Handpiece
Nozzle
Nozzle Tip
Internal Nozzle
6
WARNING
CAUTION
CAUTION
6-1 Air Pressure setting
Measure the supply pressure at the handpiece
point and set the pressure to the value specified
on the specification table. (Fig. 1)
*For Multi Gauge information refer to Option Parts
List.
Ř Do not exceed the optimum pressure specified on the specification table.
Ř Excess air pressure could cause the Powder Case and Powder Case Cap to crack or
fracture, which could lead to injury.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the
air may cause internal solidification of the cleaning powder.
Ř Before connecting each part, thoroughly dry the Coupling, Handpiece, Hose Joint section
of the Powder Case, inside of the Powder Case, and Nozzle. Also, perform idle running
with air alone to expel any internal water.
6-2 Connecting the Powder Case to the Handpiece Hose
Drive Air
Pressure
Fig. 1
Handpiece
Exhaust Air
Pressure
Chip Air Pressure
Water Pressure
Internal Nozzle
Powder Case
Powder Case Cap
Coupling
Handpiece connection
AIR
AIR
AIR
AIR
AIR
AIR
7
English
Connect the hose to the Powder Case.
PER-PTL-P
Designed for when NSK FM-CL (B2/B3), PTL-CL-LED (6 pins), or PTL-CL-FV (5 Hole) coupling is used.
Insert each coupling into the Hose Connection of the Powder Case.
-PER-M4-P, PER-B2-P
Used when the hose of Midwest 4 Hole (M4) or Borden 2 Hole (B2) is used. Align each pipe at the Hose
Connection of the Powder Case with the ports in the hose. Carefully insert the pipe straight into the hose
until it stops and securely tighten the Hose Nut.
-PER-KV-P, PER-WH-P, PER-BA-P, PER-SR-P
Used when KaVo
®
, W&H
®
, Bien-Air
®
or Sirona
®
coupling etc., is used. Check with the instruction
manual of the coupling you have and connect Perio Mate accordingly.
6-4 Check before treatment
Before use, check by operating the device away from a patient. If any abnormality is detected such
as vibration, sound or other during checking, or when using, stop use immediately and contact the
Authorized NSK Dealer.
6-5 Powder filling
Remove the Powder Case Cap and fill the Perio Mate Powder to the position marked ‘Max’ in the
Powder Case. (Approx. 10 g of powder can be filled.) (Fig. 3)
Securely close the Perio Mate Powder container lid.
Securely close the Powder Case Cap. (Fig. 4)
6-3 Connecting the Handpiece to the Powder Case
Pull back the Handpiece Release Collar on the
Powder Case and push the handpiece straight on
to the connection. Release the Handpiece Release
Collar once the handpiece is in place. (Fig. 2)
Once connected gently pull handpiece to ensure it
is securely connected.
For Perio Mate Powder information refer to Spare Parts List.
Fig. 3 Fig. 4
Fig. 2
Handpiece Release Collar
Replacement dose of Perio Mate Powder
CLOSE
8
CAUTION
Ř If excess cleaning powder is put into the Powder Chamber, the correct amount of
powder may not be sprayed immediately after the start of use.
Ř The Perio Mate Powder has the property of easily scattering because of its fine particles.
Perform filling slowly and be careful to prevent the powder from scattering.
Ř Do not use any powder stored for a long while in opened packages. Do not use leftover
powder. Using such powder may cause clogging of the nozzle.
Ř Use only Perio Mate Powder. Other powders have different particle sizes and will clog
and not perform as expected. Using other powders may also cause significant damage
to the Nozzle Tip.
Ř Always securely close the container lid of the Perio Mate Powder. Failure to do so will
attract moisture which may cause clogging.
Ř Use Perio Mate Powder as cleaning powder. If other powder is used, the Perio Mate
Nozzle Tip will be damaged and it will increase the risk of damaging substantia ossea
and dentine or causing emphysema. NSK shall assume no liability for any failures
caused by the use of other powders.
Ř If powder adheres to the Powder Case and/or the threaded portion of Powder Case Cap,
clean it immediately or it will become difficult to close the Cap.
6-6 Operation
1) Attach the Nozzle Tip to the Nozzle of the Perio
Mate Handpiece.
2) After filling the powder, spray it into a spittoon,
etc., for one second and confirm that powder
is mixed with water and air, and spraying is
appropriately performed.
3) Insert the Nozzle Tip into a periodontal pocket
by 3 mm or more. (Fig. 6)
If the Nozzle Tip is inserted less than 3 mm,
powder may not be sprayed into the periodontal
pocket efficiently.
4) Start operation. Spray powder in the same
periodontal pocket for up to 5 seconds or on
the same tooth for up to 20 seconds. Do not
spray longer than this specified time.
5) After treatment, remove the Nozzle Tip from
the Perio Mate Handpiece using the Nozzle Tip
Remover and dispose of it. (Fig. 7)
Fig. 5
Fig. 6
Fig. 7
Nozzle Tip
Nozzle Tip Remover
9
English
NOTICE
WARNING
Ř The powder spray volume can be adjusted with the Flow Volume Adjustment Ring (Fig. 8).
Start by setting it to Min. then adjust according to the application and patient condition.
Ř After adjusting the powder spray amount with the Flow Volume Adjustment Ring,
clogging may occur due to the powder condition. (for example, when moisture is
contained) In such a case, the problem may be resolved by again setting the Flow
Volume Adjustment Ring to Max, then adjusting down.
Ř Securely insert the Tip all the way by referring to the Fig. 9. If the Tip is inserted
incorrectly, the Nozzle Tip will come off and accidental ingestion may result.
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize the tip since it may be
damaged or contaminated. If the Nozzle Tip is reused, sufficient performance may not
be exerted due to nozzle deformation during removal, or accidental ingestion may result
due to detachment of the Nozzle Tip during use.
Ř Do not use this product in the same area immediately after scaling treatment of a tooth
surface in a periodontal pocket.
Ř Do not insert the Nozzle Tip into the same periodontal pocket during the same treatment
more than once.
Ř Powder should be sprayed for 5 seconds at maximum in the same periodontal pocket (20
seconds at maximum for one tooth). Never spray powder longer than this specified time.
Ř Do not strongly force the Nozzle Tip into a periodontal pocket.
Ř Make sure to use Perio Mate Powder.
Ř Do not spray powder while the Nozzle Tip end is located at the bottom of a periodontal
pocket. Emphysema may be caused.
Ř Never spray powder toward a human body. Spray only toward a tooth surface.
Ř Do not use the Nozzle Tip in any places other than periodontal pockets.
Ř If the hose swells or air or powder leaks from connections between the Powder Case
and Handpiece, during use, immediately stop use.
Ř Do not use a deformed or damaged Nozzle Tip. It may detach during use.
Ř Excessive air pressure supply may cause emphysema. Make sure the air pressure is
correctly set, as specified. Set the spray air pressure to the extent that the spray slightly
flows out from the periodontal pocket, even at maximum. Adjust the spray air pressure
within an adequate supply air pressure while checking the patient’s condition.
Ř Use the product while spraying water.
Ř Make sure to use sufficient evacuation vacuum to prevent a patient from ingesting a
large amount of powder. Avoid continuous use for a long period of time and allow the
patient to rinse out the mouth as necessary.
Fig. 9
Fig. 8
Water Flow
10
WARNING
WARNING
CAUTION
Operation without Nozzle Tip
Perio Mate can also perform cleaning of
supragingival tooth surfaces or around marginal
gingiva. Remove the Nozzle Tip from the Perio
Mate Handpiece and perform cleaning. During
operation, use the Nozzle around 10mm away
from the tooth surface at angles of 30 to 60
degrees. (Fig. 10) Do not spray powder directly to
the oral mucosa and do not insert the Nozzle into
periodontal pockets.
Ř When using Perio Mate without the Nozzle Tip, avoid use on the oral mucosa.
Ř If organic solvents, etc., adhere to the Powder Case Cap, this may result in damage.
Ř The Powder Case Cover will deteriorate over time under the influence of ultra violet
rays and other environmental factors and is therefore considered a consumable item.
Replace the powder case cover every six months.
Ř Make sure to perform 7-1 Discard of Perio Mate Powder - 7-4 Lubrication after each
use. If powder remains in the Powder Case or Handpiece, it will solidify which may
cause clogging.
Ř Remove the Powder Case Cover and Rubber Packing from the Cover Holder once a
week to remove powder caught in the clearance. Check that there are no cracks or
scratches on the Powder Case Cover at that time.
Ř Never use the Powder Case Cover with cracks or scratches. If used in such a condition
damage and injury may result from any broken pieces.
7
Care After Use
After each patient maintain the product as follows.
Ř Even after the foot control pedal is released, powder may still come out until the air
pressure in the Powder Case is expelled. After each operation, vacuum excess powder
until the powder ejection stops in order to prevent it from entering the oral cavity of the
patient.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the
air may cause internal solidification of the cleaning powder.
Ř Use only drinkable water for supply water.
Fig. 10
30 - 60 degrees
10mm
11
English
CAUTION
7-1 Discard of Perio Mate Powder
1) Open the powder case cap, and remove the
remaining cleaning powder. (Fig. 11)
2) Perform idle running with air alone to discharge
water and powder in the Nozzle. (Fig. 12)
3) Blow off the powder remaining in the powder
case with air. (Fig. 13)
4) Wipe the Nozzle and handpiece with a soft cloth
saturated in alcohol to remove saliva and other
physical contamination.
Ř In case of failure, make sure to contact your Authorized NSK Dealer for repairs.
Ř After use always thoroughly remove powder from the product then perform cleaning,
washing and sterilization. Ensure the instrument is totally dry for storage.
Ř Make sure to perform Ultrasonic Cleaning once a week to prevent clogging of powder.
(See 8-1 Ultrasonic Cleaning.)
Ř Never lubricate Perio Mate. (However, lubricate the O-rings and Flow Volume Adjustment
Ring with the supplied turbine oil. (See 7-4 Lubrication.)
Fig. 12
Fig. 11
Fig. 13
AIR
12
CAUTION
7-2 Removal of powder from the Powder Case, Handpiece and Nozzle
1) Remove the Powder Case from the hose.
2) Pull back the Handpiece Release Collar and
remove the handpiece.
3) Insert the Nozzle into the Ring Wrench and
loosen the Nozzle. Remove the Nozzle from the
Handpiece. (Fig. 14)
4) Fit the supplied Blower Nozzle to the coupling
and attach to the handpiece. Run idle with no
water more than 5 sec. to clear lines. (Fig. 15)
7-3 Cleaning and washing
1) Remove powder from the Handpiece and
Powder Case connections and the Adjustment
Ring section.
Clean the inside of the Handpiece connection
with the supplied cleaning brush. (Fig. 17: Left)
Clean the Powder Case connections with the
supplied cleaning brush in the same way.
(Fig. 17: Right)
5) Connect the Blower Nozzle to the Nozzle end
(Fig. 16) and idle with air alone. Residual
powder in the Nozzle will be discharged.
Ř When idling, hold the Nozzle and Handpiece tight and be careful not to allow the Nozzle
to be ejected by air pressure.
Ř When idling, stop water supply and idle with air alone.
*The Blower Nozzle in the product package depends on the joint type you purchase. Do not use a blower nozzle
other than the supplied blower nozzle.
Fig. 17
Fig. 14
Fig. 16
Fig. 15
Ring Wrench
Powder Case
Nozzle
LOOSEN
Handpiece
Blower Nozzle
Coupling Handpiece
Blower Nozzle
(KaVo
®
MULTIflex
®
,
MULTIflex
®
LUX Coupling)
Blower Nozzle
(PHATELUS, MACH)
13
English
CAUTION
*If powder adheres to the O-ring section, remove
the O-rings, and carefully clean the O-rings and
grooves. (Fig. 18)
2) Remove the Rubber Packing and disassemble
the Powder Case Cap. (Fig. 19)
3) Wash away the Perio Mate Powder adhering
to the inside of the Powder Case/Powder Case
Cap under running water. (Fig. 20)
4) Blow water off the Powder Case Cover, Rubber
Packing, and the Cover Holder by air and dry.
Then, assemble.
Ř If cleaning powder remains in the Powder Case, powder will solidify, and clogging may
occur.
Ř Do not immerse the Powder Case or the Handpiece, with residual powder, into water.
The powder will solidify, and clogging may occur.
Fig. 20
Fig. 19
Fig. 18
Rubber Packing
14
CAUTION
CAUTION
CAUTION
7-4 Lubrication
1) Dry the O-ring section of the Powder Case and the Adjustment Ring section, apply a drop of the
supplied turbine oil and rub it in with your finger, etc. (Fig. 21, 22)
2) Perform autoclave sterilization.
Ř If you fail to lubricate the O-ring section on the Handpiece side, the Joint section will be
damaged and this may cause the Handpiece connection to be difficult to rotate. It may
also cause water leakage.
Ř After autoclave sterilization, the Flow Volume Adjustment Ring can be difficult to rotate
in some rare cases. However, it may return to the normal state once you rotate it. If the
Adjustment Ring cannot rotate properly, contact your Authorized NSK Dealer.
Ř Do not use the Thermo Disinfector for the Powder Case, Powder Case Cap and Nozzle.
Ř Make sure to perform Ultrasonic Cleaning once a week or after the remedy described in
10. Troubleshooting to prevent clogging of powder.
8
Cleaning and Sterilization
8-1 Ultrasonic Cleaning
Only the handpiece can be washed via Thermo Disinfector.
1) After disassembling, perform cleaning. (See 7-1 Discard of Perio Mate Powder - 7-3 Cleaning and
Washing.)
2) Pass through the supplied Ring Wrench from the Nozzle end, loosen the Nozzle, and remove the
Nozzle from the Handpiece.
3) Perform Ultrasonic Cleaning with hot water for the Handpiece, Powder Case and Nozzle.
4) After cleaning, blow waterdrops from all parts by air to dry them.
5) Reassemble the parts following the disassembly procedures in reverse.
6) Connect the Powder Case and Hose with the cover not attached and discharge air. This allows water
in the Internal Nozzle to be discharged. Then thoroughly dry the inside.
OIL
Fig. 21 Fig. 22
OILOIL
15
English
CAUTION
CAUTION
NOTICE
Ř When attaching the Nozzle, lightly thread the Nozzle screw into the Handpiece by hand
at first and then securely tighten it with the supplied Ring Wrench.
Ř After Ultrasonic Cleaning, blow remaining water from the inside with a syringe and
sufficiently dry the parts before use.
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize it since it may be
damaged or contaminated.
Ř Before use, thoroughly dry all parts.
Ř Remove the Powder Case Cap from the Powder Case, and then sterilize by autoclave.
Sterilization with the Powder Case Cap attached may cause breakage of the Powder
Case Cap.
Ř Do not autoclave the product with other instruments even when it is in a pouch. This is
to prevent possible discoloration and damage to the product from chemical residue from
other instruments.
Ř If the sterilizer chamber temperature may exceed 135°C during the dry cycle then
delete the dry cycle. Even in this case, sufficient drying of the product is required.
Ř Do not wipe with, or clean, or immerse in, high acid water or sterilizing solutions.
Ř Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation,
and sunlight. The air should be free from dust, salt and sulphur.
Ř Do not heat or cool the product too quickly. Rapid change in temperature could cause
damage to the product.
Ř The product can be sterilized only by autoclave sterilization.
Ř Do not touch the product immediately after autoclaving as it will be very hot and must
remain in a sterile condition.
8-2 Sterilization
Sterilize the product by autoclave sterilization. Remove the Nozzle Tip and dispose of powder after each
patient and sterilize as noted below.
All parts of this product can be autoclaved.
Autoclave Procedure
1) Disassemble the product and clean it. (See 7-1 Discard of Perio Mate Powder - 7-4 Lubrication.)
2) Insert into an autoclave pouch. Seal the pouch.
3) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
4) Keep the product in the autoclave pouch to maintain sterility until it is used. To use, perform
reassembly in reverse order to disassembly.
Ř NSK recommends Class B sterilizers as stated in EN13060.
16
9
Periodical Maintenance Checks
Perform periodical maintenance checks every three months, referring to the check sheet below. If any
abnormalities are found, contact your Authorized NSK Dealer.
Points to check Details
Powder Case Cap Check that there are no large/deep scratches.
Powder Spray After filling the powder chamber, spray powder into a spittoon, etc., and
confirm that powder has not leaked from connections other than the Nozzle
Tip connection.
Water Spray After filling the powder chamber, spray powder into a spittoon, etc., and
confirm that water has not leaked from connections other than those of the
Nozzle Tip connection.
Adjustment Ring After filling the powder chamber, spray powder into a spittoon, etc., and
confirm that powder spray amount changes when the Flow Volume Adjustment
Ring is switched from Min. to Max.
17
English
10
Troubleshooting
Trouble Check Item Cause Remedy
Air and powder do not
come out.
Or air and powder come
out but the pressure is
not strong enough and
cleaning power is also
weak.
When the Nozzle Tip
is removed and air is
supplied, air and powder
come out from the
Nozzle section.
The inside of the Nozzle
Tip is clogged with
powder.
Replace the Nozzle Tip.
When the Handpiece
is removed and air is
supplied, air and powder
come out from the
Handpiece connection.
The inside of the
Handpiece is clogged
with powder.
Loosen the Nozzle screw with the supplied Ring Wrench and remove the Nozzle from
the Handpiece. Clean the Nozzle base with the supplied Cleaning wire (small) and
Nozzle end with the cleaning file. In addition, clean the inside of the Handpiece with
the Cleaning wire (large). If powder and air do not flow after performing the above
cleaning, perform Ultrasonic Cleaning with hot water. (See 8-1 Ultrasonic Cleaning.)
Cleaning wire (small)
Cleaning file
Cleaning wire (large)
18
When the Handpiece
is removed and air is
supplied, air and powder
do not come out of the
Powder Case.
The inside of the Powder
Case is clogged with
powder.
Insert the Cleaning wire (large) into the center hole on the end of the Handpiece
section connection of the Powder Case while turning the Wire and clean the inside. In
addition, clean the Internal Nozzle in the Powder Case with the cleaning file. If powder
and air do not flow after performing the above cleaning, perform Ultrasonic Cleaning
with hot water. (See 8-1 Ultrasonic Cleaning.)
Powder has partially
solidified in the
Handpiece or in some
part in the Powder Case,
reducing the passage
width.
Powder has partially
solidified with water, etc.,
in the middle of the air
passage.
Clean with the supplied Cleaning wire or cleaning file (see the above figure) and blow
air to remove the remaining powder inside. If the problem is not resolved, perform
Ultrasonic Cleaning with hot water. Then, thoroughly dry the inside before use. (See
8-1 Ultrasonic Cleaning.)
Powder has partially
solidified in the Powder
Case.
Powder has long been
opened and exposed to
air.
Replace powder with new powder.
Cleaning file
Cleaning wire (large)
19
English
Water is mixed in supply
air.
If you use the model for PHATELUS coupling or MACH coupling, confirm that the
O-ring is not damaged in the coupling joint section. If the O-ring is damaged, replace
it. If your model is other than those for PHATELUS and MACH couplings, confirm the
coupling for use.
After O-ring replacement, securely tighten the Taper Ring. If tightening of the Taper
Ring is not sufficient, the Powder Case may be ejected during use.
Open the filter in the air supply circuit, and the drain in the compressor, and check that
water has not accumulated.
Water leaks from
between the Handpiece
and Powder Case.
Check the O-ring in the
Handpiece joint section
of the Powder Case.
The O-ring is scratched. Replace the O-ring in the Handpiece joint section of the Powder Case, apply a drop
of the supplied turbine oil and rub it in with your finger, etc. (See 7-3 Cleaning and
Washing - 7-4 Lubrication.)
It is difficult to rotate
the connection between
the Powder Case and
Handpiece connections.
Check O-rings in the
Powder Case and
Handpiece joint section.
Powder is caught in
the O-ring grooves
and connection of the
coupling section has
become tight.
Remove O-rings from the Handpiece joint section of the Powder Case and carefully
remove the powder adhered to the grooves and O-rings with the supplied cleaning
brush. Then, thoroughly wipe off the Handpiece joint section with a cloth soaked with
alcohol. Install the O-rings, apply a drop of the supplied turbine oil and rub it in with
your finger, etc. (See 7-3 Cleaning and Washing - 7-4 Lubrication.)
*If the trouble is not resolved by any of the above remedies, contact your Authorized NSK Dealer for service repair.
Loosen
Taper Ring
O-ring
O-ring grooves
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133

NSK PER-WH-P Istruzioni per l'uso

Tipo
Istruzioni per l'uso
Questo manuale è adatto anche per