Microlife BP A6 PC Manuale utente

Categoria
Unità di pressione sanguigna
Tipo
Manuale utente
Microlife Corporation
9F, 431, RuiGuang Road, NeiHu
Taipei 11492, Taiwan, China
www.microlife.com
Microlife UAB
P. Lukšio g. 32
08222 Vilnius
Lithuania
IB BP A6 PC S-V10-1 4823
Revison Date: 2023-11-10
Microlife BP A6 PC
EN  1
ES  11
FR  22
IT  33
DE  44
TR  55
PT  65
NL  76
GR  87
AR  99
Microlife BP A6 PC BP A6 PC
BP A6 PC
Guarantee Card
Name of Purchaser / Nom de l'acheteur / Nome
del rivenditore / Name des Käufers / Alıcının adı /
Nombre del comprador / Nome do comprador /
Naam koper / Ονοματεπώνυμο αγοραστή /
Serial Number / Numéro de série / Numero di
serie / Serien-Nr. / Seri numarası / Número de
serie / Número de série / Serienummer / Αριθμός
σειράς /
Date of Purchase / Date d'achat / Data d’acquisto /
Kaufdatum / Satın alma tarihi / Fecha de compra /
Data da compra / Datum van aankoop /
Ημερομηνία αγοράς /
Specialist Dealer / Revendeur / Categoria riven-
ditore / Fachhändler / Uzman satıcı / Distribuidor
especializado / Revendedor autorizado / Specialist
Dealer / Εξειδικευμένος αντιπρόσωποςa /
BP W100 ١٢ AR
ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
(ﻲﻔﻠﺨﻟا ءﺎﻄﻐﻟا ﺮﻈﻧا) ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
:يﺮﺘﺸﻤﻟا ﻢﺳا
:ﻞﻳدﻮﻤﻟا
:ﻞﺴﻠﺴﺘﻟا ﻢﻗر
:ءاﺮﺸﻟا ﺦﻳرﺎﺗ
:ﺺﺘﺨﻤﻟا ﺮﺟﺎﺘﻟا
BP W100 ١٢ AR
ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
(ﻲﻔﻠﺨﻟا ءﺎﻄﻐﻟا ﺮﻈﻧا) ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
:يﺮﺘﺸﻤﻟا ﻢﺳا
:ﻞﻳدﻮﻤﻟا
:ﻞﺴﻠﺴﺘﻟا ﻢﻗر
:ءاﺮﺸﻟا ﺦﻳرﺎﺗ
:ﺺﺘﺨﻤﻟا ﺮﺟﺎﺘﻟا
BP W100 ١٢ AR
ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
(ﻲﻔﻠﺨﻟا ءﺎﻄﻐﻟا ﺮﻈﻧا) ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
:يﺮﺘﺸﻤﻟا ﻢﺳا
:ﻞﻳدﻮﻤﻟا
:ﻞﺴﻠﺴﺘﻟا ﻢﻗر
:ءاﺮﺸﻟا ﺦﻳرﺎﺗ
:ﺺﺘﺨﻤﻟا ﺮﺟﺎﺘﻟا
BP W100 ١٢ AR
ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
(ﻲﻔﻠﺨﻟا ءﺎﻄﻐﻟا ﺮﻈﻧا) ﺔﻟﺎﻔﻜﻟا ﺔﻗﺎﻄﺑ
:يﺮﺘﺸﻤﻟا ﻢﺳا
:ﻞﻳدﻮﻤﻟا
:ﻞﺴﻠﺴﺘﻟا ﻢﻗر
:ءاﺮﺸﻟا ﺦﻳرﺎﺗ
:ﺺﺘﺨﻤﻟا ﺮﺟﺎﺘﻟا
1Microlife BP A6 PC EN
Display
Microlife BP A6 PC EN
1START/STOP Button
2Display
3Cuff socket
4Mains Adapter Socket
5Battery compartment
6Cuff
7Cuff connector
6AFIB/MAM Switch
9User switch
AT Time button
AK M-button (memory)
AM + «Forward» Button
AL - «Backward» Button
AN Lock switch
AO USB Port
AP Date/Time
AQ Systolic value
AR Diastolic value
AS Pulse rate
BT Battery display
BK Traffic light indicator
BL Stored value
BM Pulse indicator
BN Cuff check indicator
BO Atrial Fibrillation Indicator (AFIB)
BP AFIB/MAM Mode
BQ Arm movement indicator
BR User indicator
BS AFIB/MAM Interval time
Read the important information in these
instructions for use before using this
device. Follow the instructions for use for
your safety and keep it for future reference.
Type BF applied part
Keep dry
Manufacturer
Batteries and electronic devices must be
disposed of in accordance with the locally
applicable regulations, not with domestic
waste.
Authorized representative
in the European Community
Catalogue number
Serial number
(YYYY-MM-DD-SSSSS;
year-month-day-serial number)
Caution
Humidity limitation for operating and
storage
Temperature limitation for operating or
storage
Medical device
Keep away from children of age 0 - 3
CE Marking of Conformity
S
N
N
2
Intended use:
This oscillometric blood pressure monitor is intended for
measuring non-invasive blood pressure in people aged 12 years or
older.
It is clinically validated in patients with hypertension, hypotension,
diabetes, pregnancy, pre-eclampsia, atherosclerosis, end-stage
renal disease, obesity and the elderly.
The device can detect an irregular pulse suggestive of Atrial Fibril-
lation (AF). Please note that the device is not intended to diagnose
AF. A diagnosis of AF can only be confirmed by ECG. The patient
is advised to see a physician.
Dear Customer,
This device was developed in collaboration with physicians and
clinical tests carried out prove its measurement accuracy to be of
a very high standard.*
Microlife AFIB detection is the world’s leading digital blood pres-
sure measurement technology for the detection of atrial fibrillation
(AF) and arterial hypertension. These are the two top risk factors
of getting a stroke or heart disease. It is important to detect AF and
hypertension at an early stage, even though you may not experi-
ence any symptoms. AF screening in general and thus also with
the Microlife AFIB algorithm, is recommended for people of 65
years and older. The AFIB algorithm indicates that atrial fibrillation
may be present. For this reason, it is recommended that you visit
your doctor when the device gives an AFIB signal during your
blood pressure measurement. The AFIB algorithm of Microlife has
been clinically investigated by several prominent clinical investiga-
tors and showed that the device detects patients with AFIB at a
certainty of 97-100%. 1,2
If you have any questions, problems or want to order spare parts
please contact your local Microlife-Customer Service. Your dealer
or pharmacy will be able to give you the address of the Microlife
dealer in your country. Alternatively, visit the internet at
www.microlife.com where you will find a wealth of invaluable infor-
mation on our products.
Stay healthy – Microlife Corporation!
* This device uses the same measuring technology as the award
winning «BP 3BTO-A» model tested according to the British and
Irish Hypertension Society (BIHS) protocol.
1 Kearley K, Selwood M, Van den Bruel A, Thompson M, Mant D,
Hobbs FR et al.: Triage tests for identifying atrial fibrillation in
primary care: a diagnostic accuracy study comparing single-lead
ECG and modified BP monitors. BMJ Open 2014; 4:e004565.
2 Wiesel J, Arbesfeld B, Schechter D: Comparison of the Microlife
blood pressure monitor with the Omron blood pressure monitor for
detecting atrial fibrillation. Am J Cardiol 2014; 114:1046-1048.
Table of contents
1. Important facts about blood pressure and self-measure-
ment
2. How do I evaluate my blood pressure
3. Important Facts about Atrial Fibrillation (AF)
What is Atrial Fibrillation (AF)?
Who should be screened for Atrial Fibrillation?
Microlife AFIB detection provides a convenient way to screen
for AF (only in AFIB/MAM mode)
Risk factors you can control
4. Using the device for the first time
Inserting the batteries
Setting the date and time
Selecting the correct cuff
Selecting standard or AFIB/MAM mode
AFIB/MAM mode (highly recommended)
5. Taking a blood pressure measurement
Checklist for taking a reliable measurement
How not to store a reading
6. Appearance of the Atrial Fibrillation Indicator for early
Detection (Active only in AFIB/MAM mode)
7. Traffic light indicator in the display
8. PC-Link functions
9. Data memory
Viewing the stored values
Memory full
Clearing all values
10. Battery indicator and battery change
Low battery
Flat battery - replacement
Which batteries and which procedure?
Using rechargeable batteries
11. Using a mains adapter
12. Error Messages
3Microlife BP A6 PC EN
13. Safety, care, accuracy test and disposal
Device care
Cleaning the cuff
Accuracy test
Disposal
14. Guarantee
15. Technical Specifications
1. Important facts about blood pressure and self-
measurement
Blood pressure is the pressure of the blood flowing in the
arteries generated by the pumping of the heart. Two values, the
systolic (upper) value and the diastolic (lower) value, are
always measured.
Permanently high blood pressure values can damage your
health and must be treated by your doctor!
Always discuss your values with your doctor and tell them if you
have noticed anything unusual or feel unsure. Never rely on
single blood pressure readings.
There are several causes of excessively high blood pressure
values. Your doctor will explain them in more detail and offer
treatment where appropriate.
Under no circumstances should you alter the dosages of
drugs or initiate a treatment without consulting your
doctor.
Depending on physical exertion and condition, blood pressure
is subject to wide fluctuations as the day progresses. You
should therefore take your measurements in the same
quiet conditions and when you feel relaxed! Take at least
two readings every time (in the morning: before taking medica-
tions and eating / in the evening: before going to bed, bathing
or taking medication) and average the measurements.
It is quite normal for two measurements taken in quick succes-
sion to produce significantly different results.
Deviations between measurements taken by your doctor or in
the pharmacy and those taken at home are quite normal, as
these situations are completely different.
Several measurements provide much more reliable informa-
tion about your blood pressure than just one single measure-
ment.
Leave a small break of 5 minutes between two measurements.
If you are pregnant, you should monitor your blood pressure
regularly as it can change drastically during this time.
2. How do I evaluate my blood pressure
Table for classifying home blood pressure values in adults in
accordance with the international Guidelines (ESH, ESC, JSH).
Data in mmHg.
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
3. Important Facts about Atrial Fibrillation (AF)
What is Atrial Fibrillation (AF)?
Normally, your heart contracts and relaxes to a regular beat.
Certain cells in your heart produce electrical signals that cause the
heart to contract and pump blood. Atrial fibrillation occurs when
rapid, disorganized electrical signals are present in the heart’s two
upper chambers, called the atria; causing them to contract irregu-
larly (this is called fibrillation). Atrial fibrillation is the most common
form of heart arrhythmia. It often causes no symptoms, yet it
significantly increases your risk of stroke. You’ll need a doctor to
help you control the problem.
This monitor is specially tested for use in pregnancy and
pre-eclampsia. When you detect unusual high readings in
pregnancy, you should measure after a short while again
(eg. 1 hour). If the reading is still too high, consult your
doctor or gynecologist.
Range Systolic Diastolic Recommendation
1. blood pressure
normal
< 120 < 74 Self-check
2. blood pressure
optimum
120 - 129 74 - 79 Self-check
3. blood pressure
elevated
130 - 134 80 - 84 Self-check
4. blood pressure
too high
135 - 159 85 - 99 Seek medical
advice
5. blood pressure
dangerously
high
≥ 160 ≥ 100 Urgently seek
medical advice!
4
Who should be screened for Atrial Fibrillation?
AF screening is recommended for people over 65 years of age,
since the chance of having a stroke increases with age. AF
screening is also recommended for people from the age of 50
years who have high blood pressure (e.g. SYS higher than 159 or
DIA higher than 99) as well as those with diabetes, coronary heart
failure or for those who have previously had a stroke.
In young people or in pregnancy AF screening is not recom-
mended as it could generate false results and unnecessary
anxiety. In addition, young individuals with AF have a low risk of
getting stroke as compared to elder people.
For more information visit our website: www.microlife.com/afib.
Microlife AFIB detection provides a convenient way to screen
for AF (only in AFIB/MAM mode)
Knowing your blood pressure and knowing whether you or your
family members have AF can help reduce the risk of stroke. Micro-
life AFIB detection provides a convenient way to screen for AF
whilst taking your blood pressure.
Risk factors you can control
Early diagnosis of AF followed by adequate treatment can signifi-
cantly reduce the risk of getting stroke. Knowing your blood pres-
sure and knowing whether you have AF is the first step in proactive
stroke prevention.
4. Using the device for the first time
Inserting the batteries
Switch the lock switch AN to «unlock» position. The battery
compartment 5 is on the bottom of the device. Insert the batteries
(4 x 1.5 V, size AAA), thereby observing the indicated polarity.
Setting the date and time
1. After the new batteries are fitted, the year number flashes in the
display. You can set the year by pressing either the «+» AM or
the «-» AL button. To confirm and then set the month, press the
time button AT.
2. Press the «+» AM or the «-» AL button to set the month. Press
the time button AT to confirm and then set the day.
3. Follow the instructions above to set the day, hour and minutes.
4. Once you have set the minutes and pressed the time button, the
date and time are set and the time is displayed.
5. If you want to change the date and time, press and hold the time
button for approx. 3 seconds until the year number starts to
flash. Now you can enter the new values as described above.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Contact your local Microlife Service if the enclosed cuff 6 does
not fit.
Connect the cuff to the device by inserting the cuff connector 7
into the cuff socket 3 as far as it will go.
Selecting standard or AFIB/MAM mode
This device enables you to select either standard (standard single
measurement) or AFIB/MAM mode (automatic triple measure-
ment). To select standard mode, slide the AFIB/MAM switch 6 on
the side of the device downwards to position «1» and to select
AFIB/MAM mode, slide this switch upwards to position «3».
AFIB/MAM mode (highly recommended)
In AFIB/MAM mode, 3 measurements are automatically taken in
succession and the result is then automatically analysed and
displayed. Because blood pressure constantly fluctuates, a result
determined in this way is more reliable than one produced by a
single measurement. AF detection is only activated in AFIB/MAM
mode.
When you select the 3 measurements, the AFIB/MAM-symbol
BN appears in the display.
The bottom, right hand section of the display shows a 1, 2 or 3
to indicate which of the 3 measurements is currently being
taken.
There is a break of 15 seconds between the measurements.
A count down indicates the remaining time.
The individual results are not displayed. Your blood pressure
will only be displayed after all 3 measurements are taken.
Do not remove the cuff between measurements.
If one of the individual measurements was questionable, a
fourth one is automatically taken.
Cuff size for circumference of upper arm
S 17 - 22 cm
M 22 - 32 cm
M - L 22 - 42 cm
L 32 - 42 cm
L - XL 32 - 52 cm
Only use Microlife cuffs.
5Microlife BP A6 PC EN
5. Taking a blood pressure measurement
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down on a back-supported chair and relax for 5 minutes.
Keep your feet flat on the floor and do not cross your legs.
3. Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in
the future. The arm with the higher blood pressure should be
measured.
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the
cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the
elbow.
The artery mark on the cuff (ca.3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
6. Slide the lock switch AN down to the «unlock» position. Press
the START/STOP button 1 to start measuring.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and
the pressure falls gradually. If the required pressure was not
reached, the device will automatically pump some more air into
the cuff.
9. During the measurement, the pulse indicator BM flashes in the
display.
10.The result, comprising the systolic AQ and the diastolic AR
blood pressure and the pulse rate AS are displayed. Note also
the explanations on further display symbols in this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
How not to store a reading
As soon as the reading is displayed press and hold the START/
STOP button 1 until «M» BL is flashing. Confirm to delete the
reading by pressing the M-button AK.
6. Appearance of the Atrial Fibrillation Indicator for
early Detection (Active only in AFIB/MAM mode)
This device is able to detect atrial fibrillation (AF). This symbol BO
indicates that atrial fibrillation was detected during the measure-
ment. Please refer to the next paragraph for information regarding
the consultation with your doctor.
You can stop the measurement at any time by pressing the
ON/OFF button or open the cuff (e.g. if you feel uneasy or
an unpleasant pressure sensation).
If the systolic blood pressure is known to be very high,
it can be an advantage to set the pressure individually.
Press the START/STOP 1button after the monitor has
been pumped up to a level of approx. 30 mmHg (shown on
the display). Keep the button pressed until the pressure is
about 40 mmHg above the expected systolic value – then
release the button.
Information for the doctor on frequent appearance of the
atrial fibrillation indicator
This device is an oscillometric blood pressure monitor that also
analyses pulse irregularity during measurement. The device is
clinically tested.
The AFIB symbol is displayed after the measurement, if atrial
fibrillation occurred during measuring. If the AFIB symbol
appears after having performed a full blood pressure measure-
ment episode (triplicate measurements), the patient is advised
to perform another measurement episode (triplicate measure-
ments). If the AFIB symbol appears again, we recommend the
patient to seek medical advice.
If the AFIB-symbol appears on the screen of the blood pressure
monitor, it indicates the possible presence of atrial fibrillation.
The atrial fibrillation diagnosis however, must be made by a
cardiologist based on ECG interpretation.
In the presence of atrial fibrillation the diastolic blood pres-
sure value may not be accurate.
6
7. Traffic light indicator in the display
The bars on the left-hand edge of the display BK show you the
range within which the indicated blood pressure value lies.
Depending on the height of the bar, the readout value is either
within the optimum (green), elevated (yellow), too high (orange) or
dangerously high (red) range. The classification corresponds to
the 4 ranges in the table as defined by the international guidelines
(ESH, ESC, JSH), as described in «Section 1.».
8. PC-Link functions
This device can be used in conjunction with a personal computer
(PC) running the Microlife Blood Pressure Analyzer+ (BPA+) soft-
ware. The memory data can be transferred to the PC by
connecting the monitor via a cable.
If no download-voucher and cable is included download the BPA+
software from www.microlife.com/software and use a USB cable
with a Mini-B 5 pin connector.
9. Data memory
This device automatically stores up to 99 measurement values for
each of the 2 users.
Viewing the stored values
Select either user 1 or 2 with the user switch 9.
Switch the lock switch AN to «unlock» position. Press the M-button
AK briefly. The display first shows «M» BL and «28A», which
stands for the average of all stored values.
Pressing the «+» AM or the «-» AL button repeatedly enables you
to move from one stored value to another. Press the M-button
again to exit the memory mode.
Memory full
Clearing all values
Make sure the correct user is activated.
1. Firstly unlock the device AN, then select either 1 or 2 with the
user switch 9.
2. Hold down the M-button AK until «CL» appears and then
release the button.
3. Press the M-button while «CL» is flashing to permanently clear
all values of the selected user.
10.Battery indicator and battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
BT will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol BT will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 5 on the bottom of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in Section
«4. Using the device for the first time».
Which batteries and which procedure?
Keep the arm still during measuring to avoid false readings.
This device may not or wrongly detect atrial fibrillation in
people with pacemakers or defibrillators.
During the connection, the device is completely controlled
by the computer.
Pay attention that the maximum memory capacity of 99
memories per user is not exceeded. When the 99 memory
is full, the oldest value is automatically overwritten with
the 100th value. Values should be evaluated by a doctor
before the memory capacity is reached otherwise data will
be lost.
Cancel deletion: press START/STOP button 1 while
«CL» is flashing.
Individual values cannot be cleared.
The memory retains all values although date and time must
be reset – the year number therefore flashes automatically
after the batteries are replaced.
The memory retains all values although date and time must
be reset – the year number therefore flashes automatically
after the batteries are replaced.
Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
7Microlife BP A6 PC EN
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
11.Using a mains adapter
You can operate this device using the Microlife mains adapter
(DC 6V, 600 mA).
1. Plug the adapter cable into the mains adapter socket 4 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
12.Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
13.Safety, care, accuracy test and disposal
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
Error Description Potential cause and remedy
«Err 1» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
BQ
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
BN
No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«Err 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for taking a reliable measurement and
then repeat the measurement.*
«Err 6» AFIB/MAM
Mode
There were too many errors during the
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
taking a reliable measurement and then
repeat the measurement.*
«HI» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO» Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Safety and protection
Error Description Potential cause and remedy
8
This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient’s symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient’s feedback.
Calling a doctor or an ambulance is advised if needed.
Do not use this device if the patient’s condition meets the following
contra-indications, to avoid inaccurate measurements or injuries.
The device is not intended for measuring blood pressure in
pediatric patients of age younger than 12 years old (children,
infant, or neonates).
Presence of significant cardiac arrhythmia during measurement
may interfere with blood pressure measurement and affect the
reliability of blood pressure readings. Consult with your doctor
about whether the device is suitable for use in this case.
The device measures blood pressure using a pressured cuff. If
the measuring limb suffers from injuries (for example open
wounds) or under conditions or treatments (for example intrave-
nous drip) making it unsuitable for surface contact or pressur-
ization, do not use the device, to avoid worsening of the injuries
or conditions.
Patient motions during measurement may interfere with the
measurement process and influence results.
Avoid taking measurements of patients with conditions,
diseases, and susceptible to environment conditions that lead
to incontrollable motions (e.g. trembling or shivering) and
inability to communicate clearly (for example children and
unconscious patients).
The device uses oscillometric method to determine blood pres-
sure. The arm being measure should have normal perfusion.
The device is not intended to be used on a limb with restricted
or impaired blood circulation. If you suffer with perfusion or
blood disorders, consult your doctor before using the device.
Avoid taking measurement on the arm on the side of a mastec-
tomy or lymph node clearance.
Do not use this device in a moving vehicle (for example in a car
or on an aircraft).
Indicates a potentially hazardous situation, which if not avoided,
could result in death or serious injury.
This device may only be used for the intended uses described
in this Instructions for Use. The manufacturer cannot be held
liable for damage caused by incorrect application.
Do not change the patient medication and treatment based the
result of one or multiple measurements. Treatment and medica-
tion changes should be prescribed only by a medical profes-
sional.
Inspect the device, cuff, and other parts for damage. DO NOT
USE the device, cuff or parts if they appear damaged or oper-
ating abnormally.
Blood flow of the arm is temporarily interrupted during measure-
ment. Extended interruption of blood flow reduces peripheral
circulation and may cause tissue injury. Beware of signs (for
example tissue discoloration) of impeded peripheral circulation
if taking measurements continuously or for an extended period
of time.
Prolonged exposure to cuff pressure will reduce peripheral
perfusion and may lead to injury. Avoid situations of extended
cuff pressurization beyond normal measurements. In the case
of abnormally long pressurization, abort the measurement or
loose the cuff to depressurize the cuff.
Do not use this device in oxygen rich environment or near flam-
mable gas.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Contra-indications
WARNING
9Microlife BP A6 PC EN
The device is not water resistant or water proof. Do not spill or
immerse the device in water or other liquids.
Do not dissemble or attempt to service the device, accessory
and parts, during use or in storage. Access to the device
internal hardware and software is prohibited. Unauthorized
access and servicing of the device, during use or in storage,
may compromise the safety and performance of the device.
Keep the device away from children and people incapable of
operating the device. Beware of the risks of accidental ingestion
of small parts and of strangulation with the cables and tubes of
this device and accessories.
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient, or
cause damage to the device or other property.
The device is intended only for measuring blood pressure at
upper arm. Do not measure other sites because the reading
does not reflect your blood pressure accurately.
After a measurement is completed, loosen the cuff and rest for
> 5 minutes to restore limb perfusion, before taking another
measurement.
Do not use this device with other medical electrical (ME) equip-
ment simultaneously. This may cause device malfunction or
measurement inaccuracies.
Do not use this device in proximity of high frequency (HF)
surgical equipment, magnetic resonance imaging (MRI) equip-
ment, and computerized tomography (CT) scanners. This may
cause device malfunction and measurement inaccuracies.
Use and store the device, cuff and parts in temperature and
humidity conditions specified in the «Technical Specifications».
Usage and storage of the device, cuff and parts in conditions
outside ranges given in the «Technical Specifications» may
results in device malfunction and the safety of usage.
Protect the device and accessories from the following to avoid
damaging the device:
water, other liquids, and moisture
extreme temperatures
impacts and vibrations
direct sunlight
contamination and dust
This device has 2 user settings. Please ensure it is cleaned and
disinfected between users to prevent any cross contamination.
Stop using this device and cuff and consult with your doctor if
you experience skin irritation or discomfort.
Electromagnetic Compatibility Information
This device is compliant with EN60601-1-2: 2015 Electromagnetic
Disturbances standard.
This device is not certified to be used in vicinity of High Frequency
(HF) medical equipment.
Do not use this device close to strong electromagnetic fields and
portable radio frequency communication devices (for example
microwave oven and mobile devices). Keep a minimum distance
of 0.3 m from such devices when using this device.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soap-
suds.
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
14.Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
CAUTION
WARNING: Under no circumstances may you wash the
inner bladder!
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
10
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Batteries, power adapter
(optional).
The cuff is covered by a functional guarantee (bladder tightness)
for 2 years.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website: www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
15.Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating
conditions:
10 - 40 °C / 50 - 104 °F
15 - 90 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 90 % relative maximum humidity
Weight: 354 g (including batteries)
Dimensions: 160 x 80 x 32 mm
Cuff size: from 17 - 52 cm according to the cuff
sizes (see «Using the device for the first
time»)
Measuring proce-
dure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V
diastolic
Measurement range: SYS: 60 - 255 mmHg
DIA: 40 - 200mmHg
Pulse: 40 - 199 beats per minute
Cuff pressure display
range:
0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source: 4 x 1.5 V alkaline batteries; size AAA
Mains adapter DC 6V, 600 mA
(optional)
Battery lifetime: approx. 400 measurements
(using new batteries)
IP Class: IP 20
Reference to
standards:
IEC 80601-2-30; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000 measure-
ments, whichever comes first
11Microlife BP A6 PC ES
Pantalla
Microlife BP A6 PC ES
1Botón START/STOP (Encendido/Apagado)
2Pantalla
3Enchufe para manguito
4Enchufe para adaptador de corriente
5Compartimento de baterías
6Brazalete
7Conector del manguito
6Interruptor AFIB/MAM
9Interruptor de cambio de usuario
AT Botón Hora
AK Botón M (Memoria)
AM Botón «+» (avance)
AL Botón «-» (retroceso)
AN Interruptor de bloqueo
AO Puerto USB
AP Fecha/Hora
AQ Valor sistólico
AR Valor diastólico
AS Frecuencia del pulso
BT Indicador de batería
BK Indicador de semáforo
BL Valor guardado
BM Indicador del pulso
BN Indicador de comprobación del manguito
BO Indicador de la fibrilación auricular (AFIB)
BP Modo AFIB/MAM
BQ Indicador de movimiento del brazo
BR Indicador de usuario
BS Tiempo de intervalo AFIB/MAM
Antes de utilizar el dispositivo lea estas
instrucciones de uso, ya que contienen
información importante. Siga las instruc-
ciones por su seguridad y consérvelas para
poder consultarlas en el futuro.
Pieza aplicada tipo BF
Mantener en lugar seco
Fabricante
Las baterías y los dispositivos electrónicos
se deben eliminar según indique la norma-
tiva local pertinente y no se deben
desechar junto con la basura doméstica.
Representante autorizado en la Comu-
nidad Europea
Número de referencia
Número de serie (AAAA-MM-DD-SSSSS;
año-mes-día-número de serie)
Precaución
Limitación de humedad para funciona-
miento y almacenamiento
Limitación de temperatura para funciona-
miento o almacenamiento
Dispositivo médico
Mantenga el dispositivo fuera del alcance
de los niños de 0 a 3 años de edad.
Marca de conformidad CE
S
N
N
12
Uso previsto:
Este monitor de presión arterial oscilométrico está diseñado para
medir la presión arterial no invasiva en personas de 12 años o
más.
Está clínicamente validado en pacientes con hipertensión, hipo-
tensión, diabetes, embarazo, preeclampsia, aterosclerosis, enfer-
medad renal en etapa terminal, obesidad y adultos mayores.
El dispositivo puede detectar un pulso irregular sugestivo de fibri-
lación atrial (FA). Tenga en cuenta que el dispositivo no está dise-
ñado para diagnosticar FA. El diagnóstico de FA solo puede
confirmarse mediante ECG. Se le aconseja al paciente que vea a
un médico.
Estimado cliente,
Este dispositivo ha sido desarrollado en colaboración con médicos
y su muy alta precisión ha sido probada en ensayos clínicos.*
La detección AFIB es la tecnología de medición digital de presión
arterial líder en el mundo para la detección de la fibrilación atrial
(FA) y la hipertensión arterial. Estos son los dos principales
factores de riesgo de sufrir un derrame cerebral o una enfermedad
cardíaca. Es importante detectar la FA y la hipertensión en una
etapa temprana, aunque es posible que no experimente ningún
síntoma. La detección de FA en general y por lo tanto, también con
el algoritmo Microlife AFIB, se recomienda para personas de 65
años en adelante. El algoritmo AFIB indica que la fibrilación atrial
puede estar presente. Por esta razón, se recomienda que visite a
su médico cuando el dispositivo emita una señal AFIB durante la
medición de la presión arterial. El algoritmo AFIB de Microlife ha
sido investigado clínicamente por varios investigadores clínicos
destacados y demostró que el dispositivo detecta pacientes con
AFIB con una certeza de 97-100%. 1,2
Si tiene cualquier pregunta o problema o si desea pedir piezas de
recambio, no dude en contactar con su servicio de atención al
cliente de Microlife local. Su distribuidor o farmacia podrá indicarle
la dirección del distribuidor de Microlife en su país. También puede
visitarnos en Internet en www.microlife.com donde encontrará
información útil sobre nuestros productos.
¡Cuide su salud con Microlife Corporation!
* Este dispositivo usa la misma tecnología de medición que el
premiado modelo «BP 3BTO-A» probado de acuerdo con el proto-
colo de la Sociedad Británica e Irlandesa de Hipertensión (BIHS).
1 Kearley K, Selwood M, Van den Bruel A, Thompson M, Mant D,
Hobbs FR et al.: Triage tests for identifying atrial fibrillation in
primary care: a diagnostic accuracy study comparing single-lead
ECG and modified BP monitors. BMJ Open 2014; 4:e004565.
2 Wiesel J, Arbesfeld B, Schechter D: Comparison of the Microlife
blood pressure monitor with the Omron blood pressure monitor for
detecting atrial fibrillation. Am J Cardiol 2014; 114:1046-1048.
Índice
1. Datos importantes acerca de la presión arterial y la auto-
medición
2. ¿Cómo he de evaluar mi presión arterial?
3. Datos importantes sobre la fibrilación auricular (FA)
¿Qué es fibrilación auricular (FA)?
¿Quién debe ser examinado para fibrilación atrial?
Microlife AFIB permite un control cómodo de la FA mientras
está midiendo su presión arterial (sólo en modo AFIB/MAM)
Factores de riesgos que le permite controlar
4. Uso del dispositivo por primera vez
Colocar las baterías
Ajuste de fecha y hora
Elegir el manguito correcto
Seleccionando el modo estándar o AFIB/MAM
Modo AFIB/MAM (altamente recomendado)
5. Medición de la presión arterial
Lista de chequeo para efectuar una medición fiable
Procedimiento para no guardar la lectura
6. Aspecto del indicador de fibrilación atrial para la detec-
ción temprana (activo solo en modo AFIB/MAM)
7. Indicador de semáforo en la pantalla
8. Funciones de conexión con el PC
9. Memoria de datos
Ver los valores guardados
Memoria llena
Borrar todos los valores
10. Indicador de baterías y cambio de baterías
Baterías con poca carga
Baterías descargadas – cambio
¿Qué baterías y qué procedimiento?
Uso de baterías recargables
11. Uso de un adaptador de corriente
12. Mensajes de error
13Microlife BP A6 PC ES
13. Seguridad, cuidado, control de precisión y eliminación de
residuos
Cuidado del dispositivo
Limpieza del brazalete
Control de precisión
Eliminación de residuos
14. Garantía
15. Especificaciones técnicas
1. Datos importantes acerca de la presión arterial y la
automedición
La presión arterial es la presión de la sangre que circula por
las arterias, generada por el bombeo del corazón. Se miden
siempre dos valores, el valor sistólico (superior) y el valor
diastólico (inferior).
¡Una presión arterial permanentemente alta puede perju-
dicar su salud y debe ser tratada por su médico!
Consulte siempre con su médico los valores medidos y comén-
tele si ha notado algo inusual o si tiene alguna duda. No confíe
nunca en una sola lectura de presión arterial.
La presión arterial excesivamente alta puede deberse a
distintas causas. Su médico se lo explicará con más detalle y le
recetará un tratamiento cuando sea necesario.
Bajo ninguna circunstancia debe modificar las dosis de
medicamentos o iniciar un tratamiento sin consultar a su
médico
La presión arterial está sujeta a grandes fluctuaciones a lo largo
del día dependiendo del esfuerzo y el estado físico. Así pues,
debe tomársela siempre en reposo y después de haberse
relajado. Realice al menos dos lecturas cada vez (por la
mañana: antes de tomar la medicación y de comer / por la
noche: antes de acostarse, bañarse o tomar la medicación) y
haga la media de ambos resultados.
Es completamente normal que dos mediciones tomadas en
rápida sucesión puedan producir resultados que difieran
significativamente.
Las diferencias entre las mediciones tomadas por su médico
o en la farmacia y las tomadas en casa son muy normales,
debido a que estas situaciones son completamente diferentes.
Varias mediciones proporcionan información más fiable sobre
la presión arterial que una sola medición.
Deje pasar un breve intervalo de tiempo de 5 minutos entre
dos mediciones.
Si está embarazada, debe controlar su presión arterial con
regularidad, ya que puede cambiar drásticamente durante este
tiempo.
2. ¿Cómo he de evaluar mi presión arterial?
Tabla de clasificación de los valores de la presión arterial en
adultos según las Normas Internacionales (ESH, ESC, JSH).
Datos en mmHg.
El valor superior es el que determina la evaluación. Ejemplo: un
valor de la presión arterial de 140/80 mm Hg o 130/90 mm Hg
indica «presión arterial demasiado alta».
3. Datos importantes sobre la fibrilación auricular
(FA)
¿Qué es fibrilación auricular (FA)?
Normalmente, el corazón se contrae y relaja con un ritmo regular.
Ciertas células del corazón producen señales eléctricas que
hacen que el órgano se contraiga y bombee la sangre. La fibrila-
ción auricular se produce cuando en las dos aurículas (las
cámaras superiores del corazón) se dan señales rápidas e irregu-
lares. Esto hace que las aurículas se contraigan de manera irre-
gular (fibrilación). La fibrilación auricular es la forma más común
de arritmia cardiaca. Si bien a menudo no se manifiesta mediante
Este monitor está especialmente probado para su uso en el
embarazo y la preeclampsia. Cuando detecta altas lecturas
inusuales en el embarazo, debe medir después de un
tiempo breve (por ejemplo, 1 hora). Si la lectura es
demasiado alta, consulte a su médico o ginecólogo.
Nivel Sistólico Diastólico Recomendación
1. presión arterial
normal
< 120 < 74 Autocontrol
2. presión arterial
óptima
120 - 129 74 - 79 Autocontrol
3. presión arterial
elevada
130 - 134 80 - 84 Autocontrol
4. presión arterial
demasiado alta
135 - 159 85 - 99 Acudir al médico
5. presión arterial
peligrosamente
alta
≥ 160 ≥ 100 ¡Acudir urgente-
mente al médico!
14
síntomas, hace que el riesgo de infarto cerebral aumente signifi-
cativamente. Necesitará la ayuda de un médico para controlar
este problema.
¿Quién debe ser examinado para fibrilación atrial?
La detección de FA se recomienda para personas mayores de 65
años, ya que la posibilidad de tener un accidente cerebrovascular
aumenta con la edad. La detección de FA también se recomienda
para personas mayores de 50 años que tienen presión arterial alta
(por ejemplo, SYS superior a 159 o DIA superior a 99), así como
aquellos con diabetes, insuficiencia cardíaca coronaria o para
aquellos que han tenido un accidente cerebrovascular.
En personas jóvenes o en el embarazo no se recomienda el
examen de detección de FA, ya que podría generar resultados
falsos y ansiedad innecesaria. Además, las personas jóvenes con
fibrilación atrial tienen un riesgo bajo de sufrir un derrame cerebral
en comparación con las personas mayores.
Para más información, visite nuestra web: www.microlife.com/afib.
Microlife AFIB permite un control cómodo de la FA mientras
está midiendo su presión arterial (sólo en modo AFIB/MAM)
Conocer que su presión arterial y si usted o algún miembro de su
familia tienen FA puede ayudar a reducir el riesgo de sufrir un
infarto. Microlife AFIB le permite llevar un control cómodo de la FA
mientras está midiendo su presión arterial.
Factores de riesgos que le permite controlar
El diagnóstico precoz de FA seguido de un tratamiento adecuado
puede reducir significativamente el riesgo de sufrir un accidente
cerebrovascular. Conocer su presión arterial y saber si tiene FA es
el primer paso en la prevención proactiva del accidente cerebro-
vascular.
4. Uso del dispositivo por primera vez
Colocar las baterías
Desbloquee el interruptor de bloqueo AN. El compartimento de las
baterías 5 está en la parte inferior del dispositivo. Inserte las
baterías (4 x 1,5 V, tamaño AAA) de acuerdo con las marcas de
polaridad indicadas.
Ajuste de fecha y hora
1. Una vez activadas las baterías, el número del año parpadea en
la pantalla. Puede seleccionar el año pulsando el botón «+» AM
o el botón «-» AL. Para confirmarlo y ajustar a continuación el
mes, pulse el botón Hora AT.
2. Pulse el botón «+» AM o el botón «-» AL para seleccionar el
mes. Pulse el botón Hora AT para confirmar y seleccione el día.
3. Por favor, siga las instrucciones anteriores para ajustar el día,
la hora y los minutos.
4. Una vez que haya ajustado los minutos y pulsado el botón Hora,
la fecha y la hora quedan establecidas y se visualiza la hora.
5. Si desea cambiar la fecha y la hora, pulse y mantenga pulsado
el botón Hora durante aprox. 3 segundos hasta que comience
a parpadear el número del año. Ahora, podrá introducir los
nuevos valores tal como se ha descrito anteriormente.
Elegir el manguito correcto
Microlife ofrece diferentes tamaños de manguitos. Elija el tamaño
de manguito adecuado según la circunferencia de su brazo
(medida de forma ajustada en el centro del brazo).
Contacte su servicio al cliente de Microlife local, si el manguito
6 incluido no es el adecuado para usted.
Conecte el manguito al dispositivo insertando el conector 7
del manguito en el enchufe 3 del manguito hasta que no entre
más.
Seleccionando el modo estándar o AFIB/MAM
Este dispositivo le permite seleccionar el modo estándar (medi-
ción sencilla estándar) o el modo AFIB/MAM (medición triple auto-
mática). Para seleccionar el modo estándar, deslice el interruptor
AFIB/MAM 6 situado en la parte lateral del dispositivo hacia
abajo a la posición «1», y para seleccionar el modo AFIB/MAM,
deslice el interruptor hacia arriba a la posición «3».
Modo AFIB/MAM (altamente recomendado)
En el modo AFIB/MAM, se efectúan automáticamente 3 medi-
ciones sucesivas y, después, se analiza y se visualiza automática-
mente el resultado. Debido a que la presión arterial fluctúa cons-
tantemente, un resultado determinado de esta manera es más
fiable que uno obtenido por una medición sencilla. El detector de
FA se activa en el modo de AFIB/MAM.
Tamaño de manguito para una circunferencia del brazo
S 17 - 22 cm
M 22 - 32 cm
M - L 22 - 42 cm
L 32 - 42 cm
L - XL 32 - 52 cm
Use únicamente manguitos Microlife.
15Microlife BP A6 PC ES
Cando seleccione las 3 mediciones, el símbolo AFIB/MAM BN
aparecerá en la pantalla.
En la parte inferior izquierda de la pantalla aparecen 1, 2 o 3
para indicar cuál de las 3 mediciones se está realizando en ese
momento.
Hay un descanso de 15 segundos entre las mediciones. Una
cuenta regresiva indica el tiempo restante.
Los resultados individuales no se visualizan. Su presión arterial
sólo se visualizará una vez que se hayan efectuado las 3 medi-
ciones.
No retire el manguito entre las mediciones.
Si una de las mediciones individuales ha sido dudosa, se toma
automáticamente una cuarta.
5. Medición de la presión arterial
Lista de chequeo para efectuar una medición fiable
1. Evite comer o fumar, así como cualquier otra actividad inmedia-
tamente antes de la medición.
2. Siéntese en una silla con respaldo y relájese durante 5 minutos.
Mantenga los pies planos sobre el piso y no cruce las piernas.
3. Efectúe la medición siempre en el mismo brazo (normal-
mente en el izquierdo). Es recomendable que su médico realice
dos mediciones en cada brazo durante la primera visita para
determinar de qué brazo se tomarán las mediciones en el
futuro. Las mediciones se deben efectuar en el brazo con
presión arterial más alta.
4. Quítese las prendas de vestir ajustadas a su brazo. Para evitar
la constricción, no se deben arremangar las mangas de
camisa, ya que no interfieren con el manguito si se encuentran
en contacto con el brazo.
5. Asegúrese de utilizar el tamaño correcto del manguito
(marcado en el manguito).
Coloque el manguito de tal forma que esté ajustado, pero
que no quede demasiado apretado.
Asegúrese de que el manguito esté colocado 1-2 cm por
encima del codo.
La marca de la arteria en el manguito (barra larga de 3 cm)
debe descansar sobre la arteria que va en el lado interior del
brazo.
Coloque el brazo de tal forma que esté relajado.
Asegúrese de que el manguito se encuentre a la misma
altura que su corazón.
6. Desbloquee el interruptor de bloqueo AN. Pulse el botón
START/STOP 1 para comenzar la medición.
7. Ahora, el manguito se inflará automáticamente. Relájese, no se
mueva y no tense los músculos del brazo hasta que se visua-
lice el resultado de la medición. Respire normalmente y no
hable.
8. Una vez alcanzada la presión correcta, el inflado se detiene y
la presión disminuye gradualmente. Si no se alcanzó la presión
necesaria, el dispositivo bombeará automáticamente más aire
al manguito.
9. Durante la medición, el indicador de pulso BM parpadea en la
pantalla.
10.Se visualiza el resultado compuesto por la presión arterial
sistólica AQ, diastólica AR, y la frecuencia del pulso AS. Tenga
en cuenta también las explicaciones de otras indicaciones en
este manual.
11.Retire el manguito cuando la medición haya terminado.
12.Desconecte el dispositivo. (El tensiómetro se apaga automáti-
camente al cabo de aprox. 1 min).
Procedimiento para no guardar la lectura
Cuando la lectura aparezca en la pantalla mantenga pulsado el
botón START/STOP 1 hasta que «M» BL parpadee. Confirme
que desea borrar la lectura pulsando el botón M AK.
6. Aspecto del indicador de fibrilación atrial para la
detección temprana (activo solo en modo AFIB/
MAM)
Este dispositivo puede detectar la fibrilación atrial (FA). Este
símbolo BO indica que se detectó fibrilación atrial durante la medi-
ción. Consulte el párrafo siguiente para obtener información sobre
la consulta con su médico.
La medición se puede detener en cualquier momento
presionando el botón START/STOP (p.ej. si no se
encuentra bien o en caso de tener una sensación de
presión desagradable).
Cuando se sabe que la presión arterial sistólica es muy
alta, puede resultar ventajoso ajustar la presión individual-
mente. Pulse el botón START/STOP
1
una vez que el
tensiómetro se haya inflado hasta un nivel de aprox. 30
mmHg (representado en la pantalla). Mantenga pulsado el
botón hasta que la presión esté aprox. 40 mmHg por encima
del valor sistólico esperado - entonces suelte el botón.
16
7. Indicador de semáforo en la pantalla
Las barras en el borde izquierdo de la pantalla BK muestran el
rango de valores de la presión arterial. Dependiendo de la altura
de la barra, el valor indicado es óptimo (verde), elevado (amarillo),
demasiado alto (naranja) o peligrosamente alto (rojo). Esta clasifi-
cación se corresponde con los 4 rangos definidos por las normas
internacionales (ESH, ESC, JSH) e indicados en la tabla del
«Apartado 1.».
8. Funciones de conexión con el PC
Este dispositivo se puede utilizar junto con un ordenador personal
(PC) con el software Microlife Blood Pressure Analyzer+ (BPA+).
Los datos almacenados se pueden transferir al PC conectando el
tensiómetro al PC mediante un cable.
Si el cable y el cupon de descarga no están incluidos puede
descargar el software BPA+ a través de www.microlife.com/
software y utilice un cable USB estándar con conector Mini-B de 5
pines.
9. Memoria de datos
Este dispositivo almacena automáticamente hasta 99 mediciones
por cada usuario.
Ver los valores guardados
Seleccioné usuario 1 o 2 con el interruptor de cambio de usuario
9.
Cambie el interruptor de bloqueo AN a la posición «desbloqueo».
Presione brevemente el botón M AK. La pantalla muestra primero
«M» BL y «28A», que representa el promedio de todos los valores
almacenados.
Pulsar los botones «+» AM o «-» AL repetidamente le permite ir de
un valor almacenado a otro. Presione el botón M de nuevo y salga
del modo memoria.
Memoria llena
Borrar todos los valores
Asegúrese que el usuario correcto está activado.
1. Primero desbloquee el dispositivo AN. Seleccione 1 ó 2 con el
interruptor de cambio de usuario 9.
2. Mantenga pulsado el botón M AK hasta que en la pantalla
aparezca «CL». Deje de pulsar el botón.
3. Mantenga pulsado el botón M mientras «CL» esté parpa-
deando para borrar permanente todos los valores del usuario
seleccionado.
Información para su médico en caso de indicación de
fibrilación auricular frecuente
Este dispositivo es un tensiómetro oscilométrico que también
analiza la frecuencia de las pulsaciones durante la medición. El
dispositivo ha sido probado clínicamente.
El símbolo AFIB se muestra después de la medición, si se
produjo fibrilación atrial durante la medición. Si aparece el
símbolo AFIB después de haber realizado un episodio completo
de medición de la presión sanguínea (mediciones por tripli-
cado), se recomienda al paciente realizar otro episodio de
medición (mediciones por triplicado). Si el símbolo AFIB
aparece nuevamente, recomendamos que el paciente busque
asistencia médica.
Si el símbolo AFIB aparece en la pantalla del monitor de presión
arterial, indica la posible presencia de fibrilación atrial. Sin
embargo, el diagnóstico de fibrilación atrial debe ser realizado
por un cardiólogo según la interpretación del ECG.
En presencia de fibrilación auricular, el valor de la presión
arterial diastólica puede no ser exacto.
Mantenga el brazo quieto durante la medición para evitar
lecturas falsas.
Este dispositivo puede detectar o no detectar la fibrilación
atrial en personas con marcapasos o desfibriladores.
Durante la conexión, el dispositivo es controlado completa-
mente por el ordenador.
Asegúrese de que no exceder la capacidad máxima de 99
memorias de almacenamiento por usuario. Cuando la
memoria está llena, el valor más antiguo es reempla-
zado por el valor número 100 Los valores deberán ser
evaluados por un médico antes de alcanzar la capacidad
máxima de la memoria - en caso contrario, se perderán los
datos.
Cancelar el borrado: pulse el botón START/STOP 1
mientras «CL» esté parpadeando.
No es posible borrar valores individuales.
17Microlife BP A6 PC ES
10.Indicador de baterías y cambio de baterías
Baterías con poca carga
Cuando las baterías están gastadas aproximadamente ¾, el
símbolo de baterías BT parpadea al momento de encender el
dispositivo (se visualiza una batería parcialmente cargada).
Aunque el dispositivo seguirá midiendo de manera fiable, tenga a
mano baterías de recambio.
Baterías descargadas – cambio
Cuando las baterías están gastadas, el símbolo de baterías BT
parpadea al momento de encender el dispositivo (se visualiza una
batería descargada). Ya no se puede realizar ninguna medicn y
es preciso cambiar las baterías.
1. Abra el compartimento de baterías 5 situado en el fondo del
dispositivo.
2. Sustituya las baterías; asegúrese de que la polaridad sea la
correcta, tal como lo indican los signos en el compartimiento.
3. Para ajustar la fecha y la hora, siga el procedimiento descrito
en el Apartado «4. Uso del dispositivo por primera vez».
¿Qué baterías y qué procedimiento?
Uso de baterías recargables
Este dispositivo también se puede usar con baterías recargables.
11.Uso de un adaptador de corriente
Este dispositivo se puede hacer funcionar utilizando el adaptador
de corriente de Microlife (DC 6V, 600 mA).
1. Inserte el cable del adaptador en el enchufe para el adaptador
de corriente 4 situado en el dispositivo.
2. Inserte la clavija del adaptador en una caja de enchufe.
Mientras esté conectado el adaptador de corriente, no se
consume corriente de las baterías.
12.Mensajes de error
Si se produce un error durante la medición, esta se interrumpe y
se visualiza un mensaje de error, p. ej., «Err 3».
La memoria retiene todos los valores, aunque haya que
resetear la fecha y la hora; para este fin, el número del año
parpadea automáticamente después de cambiar las
baterías.
La memoria retiene todos los valores, aunque haya que
resetear la fecha y la hora; para este fin, el número del año
parpadea automáticamente después de cambiar las
baterías.
Utilice 4 baterías alcalinas nuevas de tamaño AAA, de
1,5 V, de larga duración.
No utilice baterías caducadas.
Si no va a usar el dispositivo durante un período prolon-
gado, extraiga las baterías.
Utilice únicamente baterías reutilizables del tipo «NiMH».
Las baterías se deben extraer y recargar, cuando aparezca
el símbolo de batería (batería descargada). No deben
permanecer en el interior del dispositivo, ya que se pueden
dañar (descarga total como resultado de un uso poco
frecuente del dispositivo, incluso estando apagado).
Retire siempre las baterías recargables si no va a usar el
dispositivo en una semana o más.
¡Las baterías NO se pueden recargar dentro del
tensiómetro! Recargue las baterías en un cargador externo
y observe la información relativa a la recarga, el cuidado y
a la duración.
Utilice únicamente el adaptador de corriente de Microlife
disponible como accesorio original, apropiado para su
voltaje de alimentación.
Asegúrese de que no estén dañados ni el adaptador ni el
cable.
Error Descripción Posible causa y solución
«Err 1» Señal
demasiado
débil
Las señales del pulso en el manguito son
demasiado débiles. Vuelva a colocar el
manguito y repita la medición.*
«Err 2»
BQ
Señal de
error
Durante la medición se han detectado
señales de error por el manguito,
causadas, por ejemplo, por el movi-
miento o la contracción de un músculo.
Repita la medición manteniendo el
brazo quieto.
«Err 3»
BN
No hay
presión en el
manguito
No se puede generar una presión adec-
uada en el manguito. Se puede haber
producido una fuga. Compruebe que el
manguito esté conectado correcta-
mente y que no esté demasiado suelto.
Cambie las baterías si fuese necesario.
Repita la medición.
18
* Por favor, consulte inmediatamente a su médico, si este o cual-
quier otro problema ocurre repetidamente.
13.Seguridad, cuidado, control de precisión y elimi-
nación de residuos
Siga las instrucciones de uso. Este documento proporciona
información importante de operación y seguridad del producto
con respecto a este dispositivo. Lea atentamente este docu-
mento antes de usar el dispositivo y consérvelo para futuras
consultas.
Este dispositivo sólo se debe utilizar para los propósitos
descritos en estas instrucciones. El fabricante no se responsa-
biliza de ningún daño causado por un uso inadecuado.
El dispositivo está integrado por componentes sensibles y se
debe tratar con cuidado. Respete las indicaciones de almace-
namiento y funcionamiento que se describen en el apartado
«Especificaciones técnicas».
Los brazaletes son sensibles y deben tratarse cuidadosamente.
Infle el manguito únicamente cuando está colocado correcta-
mente en el brazo.
No utilice el dispositivo si cree que puede estar roto o detecta
alguna anomalía.
Nunca abra el dispositivo.
Lea las instrucciones de seguridad adicionales en los disposi-
tivos individuales de este manual.
El resultado de medición dado por este dispositivo no es un
diagnóstico. No está reemplazando la necesidad para la
consulta de un médico, especialmente si no coincide con los
síntomas del paciente. No confíe sólo en el resultado de la
medición, considere siempre otros síntomas potencialmente
presentes y la retroalimentación del paciente. Se aconseja
llamar a un médico o una ambulancia si es necesario.
A fin de evitar mediciones imprecisas o lesiones, no utilice este
dispositivo si el estado del paciente coincide con las contraindica-
ciones siguientes.
El dispositivo no está indicado para tomar la presión de
pacientes pediátricos de menos de 12 años de edad (niños,
niños pequeños o neonatos).
La presencia de una arritmia cardiaca elevada durante el
proceso puede interferir en la medición de la presión arterial y
afectar a la fiabilidad de los resultados. Consulte a su médico
si puede utilizar el dispositivo en este caso.
El dispositivo mide la presión arterial mediante un manguito de
presión. Si el miembro donde se va a tomar la presión presenta
una lesión (por ejemplo heridas abiertas), alteraciones o está
sometido a tratamiento (por ejemplo con una vía intravenosa)
que desaconsejen el contacto de superficie o la presurización,
no utilice el dispositivo para evitar que empeoren las lesiones o
la situación.
Los movimientos del paciente durante la medición pueden
interferir en el proceso e influir en los resultados.
«Err 5» Resultado
anormal
Las señales de medición son impre-
cisas y, por ello, no se puede visualizar
ningún resultado. Lea la lista de
chequeo para efectuar mediciones
fiables y repita la medición.*
«Err 6» Modo AFIB/
MAM
Ha habido demasiados errores durante
la medición en el modo AFIB/MAM, por
lo que es imposible obtener un resul-
tado final. Lea la lista de chequeo para
efectuar mediciones fiables y repita la
medición.*
«HI» Pulso o
presión de
manguito
demasiado
alto
La presión en el manguito es
demasiado alta (superior a 299 mm Hg)
o el pulso es demasiado alto (más de
200 latidos por minuto). Relájese
durante 5 minutos y repita la medición.*
«LO» Pulso
demasiado
bajo
El pulso es demasiado bajo (menos de
40 latidos por minuto). Repita la
medición.*
Si cree que los resultados son inusuales, por favor, lea
detenidamente la información en el «Apartado 1.».
Seguridad y protección
Error Descripción Posible causa y solución
Asegúrese de que los niños no utilicen el dispositivo sin
supervisión, puesto que podrían tragarse algunas de las
piezas más pequeñas. Tenga en cuenta el riesgo de
estrangulamiento en caso de que este dispositivo se sumi-
nistre con cables o tubos.
Contraindicaciones
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109

Microlife BP A6 PC Manuale utente

Categoria
Unità di pressione sanguigna
Tipo
Manuale utente