Abbott Tendyne TENDV-SP-33A Instructions For Use Manual

Tipo
Instructions For Use Manual
IFU, TENDYNE
TM
Mitral Valve System
Page 1 of 284
Tendyne
Mitral Valve System
Instructions for Use
Brugsanvisning
Gebruiksaanwijzing
Mode d’emploi
Gebrauchsanweisung
Οδηγίες χρήσης
Istruzioni per l’uso
Bruksanvisning
Instruções de utilização
Instrucciones de uso
Bruksanvisningen
IFU, TENDYNE
TM
Mitral Valve System
Page 2 of 284
EN: English ................................................................................................................................................................................ 5
DA: Dansk ................................................................................................................................................................................ 28
NL: Nederlands ........................................................................................................................................................................ 51
FR : Français ........................................................................................................................................................................... 77
DE: Deutsch ........................................................................................................................................................................... 103
EL: Ελληνικά ........................................................................................................................................................................... 129
IT: Italiano .............................................................................................................................................................................. 156
NO: Norsk .............................................................................................................................................................................. 182
PT: Português ........................................................................................................................................................................ 206
ES: Español ............................................................................................................................................................................ 233
SV: Svenska ........................................................................................................................................................................... 259
IFU, TENDYNE
TM
Mitral Valve System
Page 3 of 284
EN: English
WARNING: Read all instructions carefully. Failure to follow these instructions, warnings, and precautions may lead to
device damage or patient injury. Use of the Tendyne
TM
Mitral Valve System is restricted to those physicians who have
undergone training on the use of this product. The implanting physician should be experienced in transapical
procedures.
DA: Dansk
ADVARSEL: Læs alle instruktioner grundigt. Hvis disse instruktioner, advarsler og forholdsregler ikke følges, kan det
medføre skade på enheden eller patienten. Tendyne
TM
mitralklapsystemet må kun anvendes af læger, der har
gennemgået uddannelse i brugen af produktet. Den læge, der udfører implantationen, bør have erfaring med
transapikale indgreb.
NL: Nederlands
WAARSCHUWING: Lees alle instructies zorgvuldig door. Het hulpmiddel kan mogelijk beschadigd raken of de patiënt
kan mogelijk letsel oplopen als deze instructies, waarschuwingen en voorzorgsmaatregelen niet worden opgevolgd. Het
Tendyne
TM
-mitralisklepsysteem mag uitsluitend worden gebruikt door artsen die zijn opgeleid in het gebruik van dit
product. De uitvoerend arts moet ervaring hebben met transapicale ingrepen.
FR : Français
AVERTISSEMENT : lisez attentivement toutes les instructions. Le non-respect de ces instructions, avertissements et
précautions peut endommager le dispositif ou blesser le patient. Le système de valve mitrale Tendyne
TM
doit être
utilisé uniquement par des médecins ayant suivi une formation sur l’utilisation de ce produit. Le médecin qui réalise
l’implantation doit maîtriser les procédures transapicales.
DE: Deutsch
WARNUNG: Alle Anweisungen sorgfältig lesen. Werden diese Anweisungen, Warnhinweise und Vorsichtsmaßnahmen
missachtet, kann es zu Schäden an den Geräten oder Verletzungen des Patienten kommen. Das Tendyne
TM
Mitralklappensystem darf ausschließlich von Ärzten verwendet werden, die im Umgang mit diesem Produkt geschult
sind. Der implantierende Arzt muss in der Durchführung transapikaler Verfahren erfahren sein.
EL: Ελληνικά
ΠΡΟΕΙΔΟΠΟΙΗΣΗ: Διαβάστε όλες τις οδηγίες προσεκτικά. Αν δεν ακολουθήσετε αυτές τις οδηγίες, προειδοποιήσεις
και προφυλάξεις, μπορεί να προκληθεί βλάβη στη συσκευή ή τραυματισμός των ασθενών. Η χρήση του συστήματος
μιτροειδούς βαλβίδας Tendyne
TM
περιορίζεται μόνο στους ιατρούς που έχουν εκπαιδευτεί στη χρήση του εν λόγω
προϊόντος. Ο ιατρός που θα πραγματοποιήσει την εμφύτευση πρέπει να έχει πείρα στις διακορυφαίες διαδικασίες.
IT: Italiano
AVVERTENZA: leggere con attenzione tutte le istruzioni. L’inosservanza di queste istruzioni, avvertenze e precauzioni
può provocare danni al dispositivo o lesioni al paziente. L’uso del sistema valvolare mitralico Tendyne
TM
è riservato ai
medici che hanno ricevuto formazione per l’uso di questo prodotto. Il medico che esegue l’impianto deve essere
esperto nelle procedure transapicali.
IFU, TENDYNE
TM
Mitral Valve System
Page 4 of 284
NO: Norsk
ADVARSEL: Les alle instruksjoner nøye. Unnlatelse av å følge disse anvisningene, advarslene og forholdsreglene kan
føre til skader på anordningen eller pasienten. Bruk av Tendyne
TM
-mitralklaffsystemet skal begrenses til leger som har
fått opplæring i bruken av dette produktet. Implantasjonslegen må ha erfaring innen transapikale prosedyrer.
PT: Português
AVISO: Leia todas as instruções atentamente. O incumprimento destas instruções, avisos e precauções pode danificar o
dispositivo ou lesionar o paciente. O sistema de válvula mitral Tendyne
TM
destina-se a ser utilizado exclusivamente por
médicos com formação sobre a utilização deste produto. O médico responsável pela implantação deve ser experiente
em procedimentos transapicais.
ES: Español
ADVERTENCIA: Lea todas las instrucciones con atención. Si no sigue estas instrucciones, advertencias y precauciones
puede ocasionar daños al dispositivo o lesiones al paciente. El uso del sistema de válvula mitral Tendyne
TM
está
restringido a médicos formados en el uso de este producto. El médico que lo implante debe tener experiencia en
procedimientos transapicales.
SV: Svenska
VARNING! Läs alla instruktioner noggrant. Underlåtenhet att följa dessa instruktioner, varningar och
försiktighetsåtgärder kan leda till skada på enheten eller patientskada. Tendyne
TM
-mitralisklaffsystem får endast
användas av läkare som genomgått utbildning för användning av denna produkt. Den implanterande läkaren ska ha
erfarenhet av transapikala ingrepp.
IFU, TENDYNE
TM
Mitral Valve System
Page 5 of 284
EN: English
Graphical Symbols for Medical Device Labeling
Consult instructions for use
Follow instructions for use on this
website
MR conditional
Temperature limit
Do not use if package is damaged
Contents
Contents (component included with
device)
Do not re-use
Date of manufacture
Use-by date
Do not resterilize
Sterilized using ethylene oxide
Sterilized using liquid chemical
sterilants
Batch code
Catalog number
Serial number
Manufacturer
Authorized representative in the
European Community
Keep away from sunlight
Keep dry
Storage solution: Glutaraldehyde
Non-pyrogenic
Non-sterile
Clean and sterilize prior to each use
Unique device identification number
Caution
Conformité Européenne (European
Conformity). Affixed in accordance
with European Council Directive
93/42/EEC and 2011/65/EU. Hereby,
Abbott Medical declares that this
device is in compliance with the
essential requirements and other
relevant provisions of this directive.
IFU, TENDYNE
TM
Mitral Valve System
Page 6 of 284
Device Description
The Tendyne
TM
Mitral Valve System consists of the
Tendyne Mitral Valve with Apical Pad and a Tendyne
Mitral Valve Delivery System that facilitates placement
of the valve.
1.1. Tendyne
TM
Mitral Valve with Apical Pad
The Tendyne Mitral Valve is a bioprosthesis intended for
transapical implantation within the native mitral valve.
The valve includes two (2) parts; a porcine pericardial
bioprosthetic valve and an Apical Pad (connected by a
braided tether, Figure 1). The valve has three (3) porcine
pericardial tissue leaflets sewn onto a circular, self-
expanding frame made of nitinol. The inner frame is
sewn inside a self-expanding nitinol outer frame. The
outer frame helps orient the prosthesis inside the native
mitral valve. There is a radiopaque marker at A1 that
can be visualized under fluoroscopy to confirm
orientation of the valve. The outer-frame cuff is raised
along the anterior aspect and is referred to as the A2
region. The valve is designed for radial orientation such
that the anterior aspect of the cuff rests upon the
aortomitral continuity. This orientation aligns the cuff
with the anterior portion of the native mitral valve.
The frame is covered with a PET (polyethylene
terephthalate) fabric cuff that provides the sealing
surface within the native annulus. The self-expanding
prosthesis is connected to a braided fiber tether made
of polyethylene, intended to stabilize the valve by
passing through the left ventricular myocardium near
the apex, where it is fastened to an Apical Pad on the
epicardium.
Figure 1. Tendyne
TM
Mitral Valve with Apical Pad
The valve is fully repositionable and retrievable
intraoperatively. Repositioning allows optimization of
the valve position following deployment, and retrieval
allows use of an alternative valve size if the initial valve
does not perform adequately.
The Tendyne valve is available in several sizes, as listed
in Table 1, to accommodate differing cardiac anatomies.
Two (2) sealing heights are available: Standard Profile
designated by “SP” or Low Profile, designated by “LP”.
The SP valve has an effective orifice area (EOA) of
approximately 3.0 cm
2
. The LP valve has an EOA of
approximately 2.2 cm
2
. The LP valve is designed to
accommodate patients with smaller native annulus
dimensions and to reduce left ventricular outflow tract
(LVOT) obstruction. The valve size should be selected to
provide proper fit for paravalvular sealing, device
stability and LVOT area. Standard imaging techniques
should be used to assess the native mitral valve
dimensions and left ventricular dimensions for valve
sizing. Refer to Section 11 for Patient Selection and
Treatment.
IFU, TENDYNE
TM
Mitral Valve System
Page 7 of 284
The Apical Pad is available in a large and a small size. It is
comprised of a PEEK (polyether ether ketone) button
covered by a double velour polyester (PET) fabric. The
nitinol leader and tether are threaded through the
Apical Pad to affix the pad to the epicardial surface of
the heart near the apex. The cam portion of the PEEK
button is moveable to lock and unlock the Apical Pad to
the tether.
Table 1: TENDYNE
TM
Mitral Valve Sizes
Model
Description
Valve
Size
TENDV-LP-29S
Tendyne
TM
Mitral Valve LP
29S
TENDV-LP-29L
Tendyne
TM
Mitral Valve LP
29L
TENDV-SP-33A
Tendyne
TM
Mitral Valve
33A
TENDV-SP-33B
Tendyne
TM
Mitral Valve
33B
TENDV-SP-33C
Tendyne
TM
Mitral Valve
33C
TENDV-SP-33S
Tendyne
TM
Mitral Valve
33S
TENDV-LP-33S
Tendyne
TM
Mitral Valve LP
33S
TENDV-SP-33M
Tendyne
TM
Mitral Valve
33M
TENDV-LP-33M
Tendyne
TM
Mitral Valve LP
33M
TENDV-SP-33L
Tendyne
TM
Mitral Valve
33L
TENDV-SP-35S
Tendyne
TM
Mitral Valve
35S
TENDV-LP-35S
Tendyne
TM
Mitral Valve LP
35S
TENDV-SP-35M
Tendyne
TM
Mitral Valve
35M
TENDV-LP-35M
Tendyne
TM
Mitral Valve LP
35M
Model
Description
Valve
Size
TENDV-SP-35L
Tendyne
TM
Mitral Valve
35L
TENDV-SP-37S
Tendyne
TM
Mitral Valve
37S
TENDV-LP-37S
Tendyne
TM
Mitral Valve LP
37S
TENDV-SP-37M
Tendyne
TM
Mitral Valve
37M
TENDV-LP-37M
Tendyne
TM
Mitral Valve LP
37M
TENDV-SP-37L
Tendyne
TM
Mitral Valve
37L
TENDV-SP-39S
Tendyne
TM
Mitral Valve
39S
TENDV-SP-39M
Tendyne
TM
Mitral Valve
39M
TENDV-SP-41S
Tendyne
TM
Mitral Valve
41S
1.2. Tendyne
TM
Mitral Valve Delivery System
The Tendyne Mitral Valve Delivery System consists of a
Tendyne
TM
Loading System, Tendyne
TM
Delivery System,
Tendyne
TM
Pad Positioning System, Tendyne
TM
Stand
Components and non-sterile, reusable Tendyne
TM
Loading Stand and Tendyne
TM
Weight. The Tendyne
TM
Retrieval System is also available to retrieve the valve
intraoperatively, in the event the valve implant is
suboptimal.
1.2.1. Tendyne
TM
Loading System
The Tendyne Loading System (Figure 2) collapses the
Tendyne Mitral Valve into a loading tube. The A2
portion of the Tendyne Mitral Valve aligns with black
marks on the Loading System. The System includes a
Closed Cap, Centering Cap Assembly (Centering Cap,
Centering Rod, Centering Cone, and Thumb Screw),
loading funnel, loading tube, loading handle, and tether
clip.
IFU, TENDYNE
TM
Mitral Valve System
Page 8 of 284
Figure 2. Tendyne
TM
Loading System
1. Closed Cap
2. Centering Cap Assembly (Centering Cap, Centering Rod,
Centering Cone, and Thumb Screw)
3. Loading Funnel
4. Loading Tube
5. Loading Handle
6. Tether Clip (not shown)
1.2.2. Tendyne
TM
Delivery System
The Tendyne Delivery System (Figure 3) introduces the
sheath tip through the ventricular myocardium to the
mitral annulus to facilitate deployment of the valve. If
necessary, the Delivery System can partially recapture
and reposition the valve after full deployment. The
Delivery System handle includes a Delivery Sheath,
Loading Tube (part of the Loading System), Delivery
Handle, Collapsible Dilator, and Introducer Needle.
Figure 3. Tendyne
TM
Delivery System
1. Delivery Sheath
2. Loading Tube (Part of Loading System)
3. Delivery Handle
4. Collapsible Dilator
5. Introducer Needle
1.2.3. Tendyne
TM
Pad Positioning System
The Tendyne Pad Positioning System (Figure 4) fastens
the Apical Pad to the braided tether at the epicardial
surface. It employs a Tether Load System (TLS) that
holds the tether securely with a pin while tension load is
applied.
Figure 4. Tendyne
TM
Pad Positioning System
1. Positioning Handle
2. Pad Arms
3. Pad Release Knob
4. Lock Wheel
5. Travel Indicator and Indicator Window
6. Lead Screw Release Knob and Button
7. Tether Load System (TLS)
8. Extension Tube
9. Pin
1.2.4. Non-Sterile Reusable Components
The Tendyne Stand and Tendyne Weight are provided
non-sterile and are reusable. They must be cleaned and
sterilized prior to each use. Follow the cleaning and
sterilization instructions provided in Section 10.14.
IFU, TENDYNE
TM
Mitral Valve System
Page 9 of 284
Figure 5. Tendyne
TM
Loading Stand
1. Loading Stand Base
2. Left Stand Post
3. Right Stand Post
1.2.5. Tendyne
TM
Sterile Components
The following additional system components are
provided sterile for use in conjunction with the Tendyne
Mitral Valve System: Tendyne Stand Components
(angled clip, rotating clip, rotating clip head, bottom
loading clip, thumb screws), 1-way and 3-way stopcocks,
hemostasis adapter, Tuohy-Borst valve, valve pocket air
evacuation tube, tether flushing assembly, verification
tether, extension tube, tether clip, and pressure
transducer.
1.3. Tendyne
TM
Retrieval System
The Tendyne Retrieval System (Figure 6) recaptures and
collapses the valve for removal from the body during
the procedure if the size or final position relative to the
native annulus is suboptimal.
Figure 6. Tendyne
TM
Retrieval System
1. Retrieval Tool Assembly
2. Cable Tie
3. Tuohy Valve
4. Tether Piercing Pin
5. Retrieval Dilator Assembly
6. Retrieval Tool Assembly Rod
7. Scale Markings
8. Retrieval Sheath Assembly
9. Cable Tie Retention Feature
10. Retrieval Front Plate
11. Actuator
12. Slide Bar Locking Mechanism
13. Bottom Thumbscrew
14. Retaining Ring
15. Retaining Ring Thumbscrew
16. Allen Wrench Set (not shown)
17. Insulated Wire Clippers (not shown)
18. 1-Way Stopcock (not shown)
19. Extension tube (not shown)
Indications
The Tendyne Mitral Valve System is indicated for the
treatment of the native mitral valve without prior mitral
valve intervention in patients with moderate-to-severe
or severe mitral valve regurgitation (MR Grade ≥3+), a
life expectancy less than 5 years, left ventricular ejection
fraction (LVEF) ≥ 30%, left ventricular end-diastolic
dimension (LVEDD) ≤ 7.0 cm, who do not have severe
mitral annular calcification and are deemed not suitable
for surgical repair or replacement by a multi-disciplinary
heart team who have:
Primary MR and are at prohibitive surgical risk,
deemed not suitable for transcatheter repair by
a multidisciplinary heart team and have left
ventricular end-systolic dimension (LVESD) >
3.0cm, or
Secondary MR and are symptomatic despite
maximally-tolerated guideline directed medical
therapy (including cardiac resynchronization
therapy, if indicated).
Tendyne Apical Pad: The Tendyne Apical Pad is
intended to secure the valve after Tendyne Mitral Valve
delivery and system withdrawal.
Tendyne Delivery System: The Tendyne Delivery
System is intended for delivery of the Tendyne Mitral
Valve.
Tendyne Loading System: The Tendyne Loading System
is intended for loading of the Tendyne Mitral Valve.
Tendyne Pad Positioning System: The Tendyne Pad
Positioning System is intended for securement of the
Apical Pad.
Tendyne Retrieval System: The Tendyne Retrieval
System is intended for retrieving the Tendyne Mitral
Valve during the procedure, if required.
IFU, TENDYNE
TM
Mitral Valve System
Page 10 of 284
Tendyne Weight: The Tendyne Weight is a reusable
accessory intended for preparation of the Pad
Positioning System.
Tendyne Stand: The Tendyne Stand is a reusable
accessory intended to hold the Loading and Delivery
Systems during preparation of the Tendyne Mitral Valve
System.
Tendyne Stand Components: The Tendyne Stand
Components are sterile accessories used with the
Tendyne Stand during preparation of the Tendyne
Mitral Valve System.
Contraindications
The Tendyne
TM
Mitral Valve System is contraindicated
for patients:
Who cannot tolerate procedural anticoagulation or
post-procedural antiplatelet/anticoagulation regimen
With sepsis including active endocarditis
With evidence of left ventricular or atrial thrombus
With vegetation or mass on the mitral valve
With a thin or fragile apex rendering the subject
unsuitable for a transapical procedure
With hypersensitivity to nickel or titanium
4. Warnings and Precautions
4.1. Warnings
The valve implant procedure should only be performed
where emergency mitral valve surgery can be
performed promptly.
All sterile supplied products are for single use only. Do
not reuse, reprocess, or re-sterilize components that are
provided sterile. Reuse, reprocessing, or re-sterilization
of these components may compromise the structural
integrity of the device and/or create a risk of
contamination of the device, which could result in
patient injury, illness, or death.
Care should be exercised in patients with
hypersensitivities or allergies to aldehyde, silicone,
polymeric materials and/or contrast.
Inspect all product prior to use. Do not use the valve or
any component of the System if the seal or sterile
barrier is compromised or if the product has been
dropped, damaged, or mishandled in any way.
Do not use if the ‘USE BY’ date specified on the package
has elapsed.
Do not manipulate or handle the valve with sharp or
pointed objects.
If a valve is inserted into a patient and then retrieved, it
may not be reused. Do not attempt to reload it on the
same or any other Delivery System.
Failure to properly clean and/or re-sterilize the reusable
stand and reusable weight prior to use may result in
infections, user injury, or patient injury.
The Collapsible Dilator is coated with a hydrophilic
coating at the distal end of the device for a length of 8.0
centimeters. Please refer to Section 10.5 for further
information on how to prepare and use this device to
ensure it performs as intended. Failure to abide by the
warnings in this labeling might result in damage to the
device coating, which may necessitate intervention or
result in serious adverse events.
4.2. Pre-implantation Precautions
Pre-procedural planning, including transesophageal
echocardiography and contrast-enhanced cardiac CT,
should be performed to evaluate anatomic compatibility
with the Tendyne prosthesis.
The safety and performance of the Tendyne system has
not been established for patients with coronary artery
disease requiring intervention, severe tricuspid
regurgitation, tricuspid valve disease requiring surgery,
severe right ventricular dysfunction, or hypertrophic or
restrictive cardiomyopathy.
Do not use the valve if there is any damage to the
container (e.g., cracked jar or lid, leakage, broken or
missing seals). Do not use the valve if the container is
leaking or if the glutaraldehyde storage solution does
not completely cover the valve.
Do not freeze or expose to extreme heat. Check the
temperature indicator prior to use. Do not use if the
temperature indicator identifies exposure to a high or
low temperature.
The valve and the glutaraldehyde storage solution are
sterile. The outside of the valve container is non-sterile
and must not be placed in the sterile field.
Exposure to glutaraldehyde may cause irritation of the
skin, eyes, nose, and throat. Avoid prolonged or
repeated exposure to the vapors. Use only with
IFU, TENDYNE
TM
Mitral Valve System
Page 11 of 284
adequate ventilation. If skin contact occurs, immediately
flush the affected area with water. In the event of eye
contact, flush with water and seek medical attention
immediately. For more information about
glutaraldehyde exposure, request a Safety Data Sheet
(SDS) from Abbott.
Do not use the valve without thoroughly rinsing it, as
directed. Adequate rinsing of the valve with sterile
heparinized saline is mandatory before implantation
(see instructions in Section 10.3).
During rinsing, do not touch the leaflets or squeeze the
valve.
Do not expose the valve to solutions other than the
storage and rinse solutions.
Do not add antibiotics or any other substance (other
than heparin) to either the storage or rinse solutions.
Do not allow the valve to dry. Maintain tissue hydration
with irrigation or immersion.
If a valve is damaged during loading, do not use or
attempt to repair.
Failure to properly clean and dry the reusable stand and
reusable weight may lead to a reduction in device life.
Do not use steel wool, wire brushes, pipe cleaners, or
abrasive detergents. Do not use saline to clean or soak
the device. Do not use chlorine bleach.
4.3. Implantation Precautions
The valve is to be used in conjunction with the Tendyne
Delivery and Retrieval Systems only.
Do not introduce air into the System. Ensure that the
components remain free of air by flushing with sterile
saline.
Do not deform the valve more than what is necessary
for loading and implantation.
Do not use Delivery System or Retrieval System
components if they have been dropped, damaged, or
mishandled in any way.
5. Magnetic Resonance (MR) Safety
5.1. MR Conditional
Non-clinical testing has demonstrated that the valve can
be scanned safely under the following conditions:
Static magnetic field of 1.5-Tesla (1.5 T) or 3-Tesla (3 T)
Maximum spatial gradient field of 4,000 G/cm
(40.00 T/m)
Maximum MR System reported, whole body averaged
specific absorption rate (SAR) of 2.0 W/kg (Normal
Operating Mode).
5.2. Radio Frequency (RF) Heating
Under the scan conditions defined in Section 5, the
valve is expected to produce a maximum temperature
rise of <7.5°C after 15 minutes of continuous scanning.
This level of heating is not expected to be associated
with any adverse physiological effect considering the
heat dissipation caused by the blood flow at the
mitral valve.
CAUTION: The RF heating behavior does not scale with
static field strength. Devices that do not exhibit
detectable heating at one field strength may exhibit
high values of localized heating at another field
strength.
5.3. MR Artifact
In non-clinical testing, the image artifact caused by the
device extends approximately 0.9 cm from the valve
when imaged with a gradient-echo pulse sequence in a
3 T MRI System.
Limitations
A chest condition preventing transapical access
7. Potential Adverse Events
Adverse events potentially associated with use of
the Tendyne Mitral Valve System may include, but
are not limited to, the following:
Adverse foreign body response
Adverse reaction to anesthesia
Allergic reaction
Anemia
Annular dissection
Aortic insufficiency
Atrial or ventricular injury
Atrial Septal Defect (resulting from mitral
valvuloplasty, if performed)
Bioprosthetic valve dysfunction
IFU, TENDYNE
TM
Mitral Valve System
Page 12 of 284
Bleeding complications
Blood loss which may require transfusion
Cardiac arrest
Cardiac perforation
Conduction defect with or without need for
pacemaker
Cardiac arrhythmia, atrial or ventricular
Damage to cardiac tissue and/or structures
Death
Decreased LV function and/or cardiac output
Device embolism
Device erosion, migration or malposition
Device thrombosis
Embolism (air, blood clot, calcium, tissue, etc.)
Endocarditis
Esophageal irritation, stricture, or perforation
Fever
Foreign body response
Heart Failure, new or worsening
Hematoma
Hemolysis
Hypotension
Infection / Sepsis
Leaflet, chordal, papillary or ventricular rupture
(resulting from mitral valvuloplasty, if
performed)
Liver failure
Mitral valve injury
Mitral valve prolapse / stenosis
Myocardial infarction
Obstruction
Pain
Pleural effusion
Pulmonary embolism
Pulmonary hypertension
Paravalvular leak
Pericardial effusion / tamponade
Renal insufficiency or failure
Respiratory difficulty, insufficiency or failure
Stroke or transient ischemic attack
Tear or damage to device
Vascular and access-related complications
Worsening of mitral regurgitation (MR)
8. How Supplied
8.1. Sterile
The valve is supplied sterile and non-pyrogenic for single
use only. The valve is sterilized with and stored in
glutaraldehyde solution for preservation of porcine
tissue components. The valve is supplied in a storage
container with a screw cap closure and a tamper-
evident seal.
The Apical Pad is supplied sterile and non-pyrogenic for
single use only. The Apical Pad is sterilized with ethylene
oxide gas.
The Tendyne
TM
Loading System, Tendyne Delivery
System, Tendyne Pad Positioning System, and Tendyne
Retrieval System are supplied sterile for single use only.
They are sterilized with ethylene oxide gas.
Additional system components provided sterile are
single use only and include the Tendyne Stand
Components (angled clip, rotating clip, rotating clip
head, bottom loading clip, thumb screws), 1-way and 3-
way stopcocks, hemostasis adapter, Tuohy- Borst valve,
valve pocket de-air tube, tether flushing assembly,
verification tether, extension tube, tether clip, and
pressure transducer. They are sterilized with ethylene
oxide gas.
8.2. Non-Sterile
The Tendyne
TM
Stand and Tendyne Weight are provided
non-sterile and are reusable. These components must
be cleaned and sterilized prior to each use. Follow the
cleaning and sterilization instructions provided in
Section 10.14.
IFU, TENDYNE
TM
Mitral Valve System
Page 13 of 284
8.3. Storage
Store the Tendyne valve in a cool, dry environment. Do
not expose the bioprosthesis to extreme low or extreme
high temperature (<5°C [<41°F] or >47°C [>116°F]).
CAUTION: Check the temperature indicator
prior to use. Do not use if the temperature
indicator identifies exposure to a high or low
temperature.
Appropriate inventory control should be maintained so
that Tendyne valves, Apical Pads, Delivery Systems and
system components with earlier ‘USE BY’ dates are used
preferentially.
9. Required Equipment (Not Provided)
Standard cardiac catheterization lab equipment
Transesophageal and transthoracic
echocardiographic equipment
Three (3) 1-Liter sterile rinsing basins
Two (2) Liters heparinized saline
Two (2) 1000 mL sterile saline bags with
pressure infusion bag and IV pole
20 35cc Luer-lock syringes
0.035” (0.89 mm) guidewire
8F introducer sheath, short length
7F or 8F single lumen balloon tip catheter
compatible with 0.035” (0.89 mm) guidewire
Hemodynamic system interface cable
Long sterile probe cover
30 cc Balloon inflation device
10. Operator’s Instructions
10.1. Setup Loading Stand and Stand Components
Before use, clean and sterilize stand, posts and weight
per Section 10.14. Ensure that the Stand Base and Stand
Posts are free of damage or corrosion. If damage or
corrosion is detected, do not use.
1. Assemble the stand by threading the Left Stand
Post into the left post hole in the Stand Base
and the Right Stand post into the right post
hole.
2. Ensure that the Stand Posts are fully threaded
into the post holes.
3. Remove sterile tray from Stand Components
Package. Inspect components and remove
from tray.
Figure 7. Tendyne
TM
Stand Components sterile tray
component layout.
1. Bottom Loading Clip - Clip 2
2. Angled Clip - Clip 3
3. De-air tube
4. Tether Flushing Assembly
5. Thumbscrews
6. Rotating Clip - Clip 1
7. Rotating Clip Head
8. Verification Tether
4. Attach the Rotating Clip - Clip 1 to the Rotating
Clip Head using a thumb screw.
5. Attach remaining thumb screws to the Rotating
Clip Clip 1, Bottom Loading Clip Clip 2 and
the Angled Clip Clip 3.
6. Attach Bottom Loading Clip - Clip 2 to the Right
Stand Post and align with the lower groove
marked 2. Attach the Rotating Clip Clip 3
assembly to the Right Stand Post at the upper
groove marked 1. Attach the Angled Clip
Clip 3 to the Left Stand Post at the groove
marked 3.
NOTE: Install the clips with the numbers and
tether grooves facing upward.
7. Line the clips up with the respective numbering
and circumferential grooves on the stand
posts.
IFU, TENDYNE
TM
Mitral Valve System
Page 14 of 284
10.2. Removing the Valve from the Storage Container
The valve package or container should not be opened
until the valve sizing is confirmed.
WARNING: Do not use the valve if the tamper-evident
seal is damaged, broken, or missing, or if fluid is
leaking from the jar.
1. Assure that the labeling on the valve container
indicates the proper valve model selected for
the case.
2. Using blunt-nosed sterile forceps, carefully
remove the valve holder from the jar. Use the
designated tab on the valve holder.
CAUTION: Do not place the non-sterile exterior
of the jar containing the valve in the sterile
field.
3. Carefully cut the stay suture and grasp the
valve at the cuff. Remove the valve from the
holder.
CAUTION: Manage tether and leader to assure
they remain sterile.
4. Drain the valve completely.
5. Carefully inspect the valve.
CAUTION: Do not use the valve if there is any
sign of damage or deterioration.
10.3. Rinsing the Valve
Follow sterile technique during valve preparation and
implantation.
1. On a table within the sterile field, prepare
three (3) sterile basins for rinsing the valve,
each with 600 800 mL of sterile heparinized
saline (1000 units/L).
CAUTION: Do not add any other solutions,
drugs, chemicals, or antibiotics to the rinse
solution, as irreparable damage to the leaflet
tissue, which may not be apparent under visual
inspection, may result.
2. Fully immerse the valve in the heparinized
saline solution in the first basin and rinse the
valve for two (2) minutes.
3. Use a 20-35 mL syringe to perform a 20 mL
flush of each of the three (3) leaflet pockets to
remove residual glutaraldehyde.
CAUTION: Do not touch the leaflets or squeeze
the valve during rinsing.
4. Transfer the valve to the second basin and
rinse the valve for a second time for two (2)
minutes.
5. Perform another 20 mL flush of each of the
three (3) leaflet pockets.
6. Transfer the valve to the third basin and rinse
the valve for a third time for two (2) minutes.
7. Perform another 20 mL flush of each of the
three (3) leaflet pockets.
8. Flush the braided tether thoroughly to remove
residual glutaraldehyde.
9. Insert the tether leader into the Tuohy-Borst
valve on the tether flushing assembly until the
end abuts the valve stem. Close the Tuohy-
Borst valve.
10. Attach a 20-35 mL syringe to the Y-connector
and aspirate the air from the tether.
11. Perform a forward and backward flush of the
tether three (3) times or until no air bubbles
are identified exiting the tether flushing
assembly. Loosen Tuohy-Borst valve and
remove the tether and leader from the tether
flushing assembly.
12. After rinsing, leave the valve fully immersed in
the third basin until it is ready to be loaded.
CAUTION: The valve should be kept hydrated
throughout the rest of the preparation
procedure to prevent the tissue from drying.
WARNING: Throughout preparation of the
components, remove air using syringes, stopcocks, and
flushing with positive pressure saline as needed. It is
imperative that all components be purged of air to
prevent introducing air emboli into the heart.
10.4. Preparing the Pad Positioning System
1. Place sterile weight on table. Inspect for
evidence of damage or corrosion. If damage or
corrosion is detected, do not use.
2. Assure that the labeling on the pouch indicates
the Apical Pad model that is selected for the
case. Remove jar containing Apical Pad from
sterile pouch. Open jar and inspect Apical Pad.
Do not use if damage is detected.
3. Remove sterile tray from Tendyne
TM
Pad
Positioning System package. Assemble
components on sterile table.
IFU, TENDYNE
TM
Mitral Valve System
Page 15 of 284
Figure 8. Tendyne
TM
Pad Positioning System sterile tray
component layout.
1. Positioning Handle
2. Pressure Transducer
3. 3-way Stopcock
4. 1-way Stopcock
5. Tether Load System (TLS)
4. With the Release Button on the Lead Screw
Knob fully depressed, set the indicator to “15
relative to the printed scale on the Indicator
Window.
Figure 9. Positioning Handle knobs and actuators.
1. Pad Release Knob
2. Lock Wheel
3. Lock Wheel Release Button
4. Indicator Window
5. Lead Screw Knob
6. Lead Screw Release Button
5. Rotate the Pad Release Knob to open the arms.
6. Attach the Apical Pad with the hole in the outer
ring of the Apical Pad facing up.
7. Rotate the Pad Release Knob in the opposite
direction to close the arms and secure the
Apical Pad onto the Positioning Handle.
8. Advance the pin cam to the opening on the
Apical Pad by activating the Lock Wheel
Release Button and ensuring that the “D”
shaped pin cam engages with the opening.
9. Check pin cam movement by rotating the Lock
Wheel to the full lock position and back to the
full unlocked position.
10. Attach a 20-35mL syringe filled with 10 mL of
sterile saline to the Tether Load System (TLS).
11. With the syringe positioned vertically above
the TLS, pull vacuum to evacuate air, then push
fluid into the System.
12. Repeat this process until minimal bubbles are
observed. Remove syringe from TLS.
13. Align the Pressure Transducer to the holder
without twisting the pressure tube.
Do not clip into the holder.
14. Attach the 1-way stopcock onto the Pressure
Transducer. Re-attach the syringe to the 1-way
stopcock.
15. Hold the System vertically and repeat the air
evacuation procedure.
16. Once air is evacuated, slightly pressurize the
System, lock the 1-way stopcock, and then
remove the syringe.
CAUTION: Do not over pressurize the TLS. Very
little pressure is needed.
17. Clip the Pressure Transducer into the holder
and place the System on a level surface.
Connect Pressure Transducer cable to
interconnect cable.
18. Open the 1-way stopcock to atmosphere and
zero.
19. After the System is zeroed, close the 1-way
stopcock.
20. Verify prep by attaching the weight to the TLS
using the Verification Tether. Insert leader
through the knotted loop on the tether
braided. Place loop of the Verification Tether
around groove in the 2 lb. Weight and tighten.
Insert the leader of the Verification Tether
through the threaded hole on the bottom of
the TLS until the tether braid is clear of the TLS.
Slide tab on the TLS Pin to pin the tether braid.
IFU, TENDYNE
TM
Mitral Valve System
Page 16 of 284
Holding the TLS from the bottom surface, lift
until the weight is clear of the table.
If weight does not verify an acceptable signal,
repeat steps 7 to 12 once. If weight does not
verify an acceptable signal after a second
evacuation, discard unit start with a new
System.
2 lb. weight pressure monitor 120 mmHg
(+/- 15 mmHg)
21. Remove the Verification Tether by retracting
the slide tab to release the TLS Pin and pull the
Verification Tether through the TLS, and attach
the TLS to the Positioning Handle.
CAUTION: Ensure the TLS Pin is not engaged
after the TLS has been attached to the
Positioning Handle. If the TLS Pin is engaged it
may interfere with the nitinol leader joint and
could result in damage to the Pin.
10.5. Preparing the Delivery System Handle and
Collapsible Dilator
1. Remove sterile tray from Delivery System
packaging. Inspect and remove contents to
sterile table.
Figure 10. Tendyne Delivery sterile tray component
layout.
1. Delivery Sheath
2. Introducer Needle
3. Collapsible Dilator
4. Delivery Handle
5. Hemostasis Valve Adapter
6. Tuohy-Borst with extension
7. 1-way Stopcock
8. 3-way Stopcock
2. Attach the Hemostasis Valve to the Delivery
Sheath side port.
3. Attach a 3-way stopcock to the Collapsible
Dilator hub.
4. Carefully remove the protector from the
Collapsible Dilator.
5. Prepare diluted contrast solution
(30/70 contrast/saline).
6. Fill a balloon inflation device with at least
5-10 ml of the contrast solution.
7. Attach the balloon inflation device to the 3-way
stopcock.
8. Using standard balloon preparation
techniques, prepare the Collapsible Dilator to
remove air and close the 3-way stopcock.
9. Refill the inflation device with 20 ml of contrast
solution.
10. Insert the Collapsible Dilator into the
Hemostasis Valve on the Delivery Sheath until
the strain relief on the Collapsible Dilator hub
abuts the Hemostasis Valve housing.
10.6. Preparing the Loading System
1. Remove sterile tray from Loading System
Package. Inspect components and remove
from tray.
Figure 11. Tendyne Loading System sterile tray component
layout.
1. Centering Cap Assembly
2. Tether Clip
3. Loading Tube
4. Loading Funnel
IFU, TENDYNE
TM
Mitral Valve System
Page 17 of 284
5. Small Centering Cone (for loading Tendyne Mitral Valve
LP only)
6. Closed Cap
7. Loading Handle Extension Tube
8. Loading Handle
9. 1-way Stopcock (not shown)
10. 3-way Stopcock (not shown)
11. Tuohy-Borst with extension (not shown)
2. Attach the Loading Tube to the Loading Funnel.
Ensure the Loading Tube black indicator and
Funnel cutout indicator are aligned. Ensure the
interfacing keys are aligned and the interfacing
O-ring is fully seated.
3. Feed the tether through the Loading Funnel
and Loading Tube until the valve is seated
inside the Loading Funnel. Approximately
align the anterior aspect of the valve cuff
(raised portion) to the Loading Funnel cutout
(Figure 13).
4. Attach the Closed Cap onto Loading Funnel.
5. Attach the Closed Cap Extension Tube to the
Closed Cap and attach a 1-way stopcock to the
free end of the Closed Cap Extension Tube.
6. Mount the Loading Funnel, Loading Tube, and
Closed Cap into the stand clips oriented with
the Closed Cap down (in Bottom Loading Clip -
Clip 2) and the Loading Tube up (in Rotating
Clip - Clip 1). Make sure that the black
alignment indicators are visible.
7. Hang a 1000 mL saline bag pressurized to 150
250 mmHg and attach the fluid line to the
Closed Cap/Closed Cap Extension Tube
assembly. Make sure that all luer connections
are tight and secure.
8. Fill the assembly with saline up to a level even
with the open end of the Loading Tube and
shut the 1-way stopcock.
9. Tap the Loading Funnel and Loading Tube to
remove any air bubbles behind the valve.
10. Attach a Tuohy-Borst with extension to the
Loading Handle Extension Tube, then attach a
3-way stopcock to the side port tubing luer on
the Tuohy-Borst valve.
Figure 12. Tendyne
TM
Loading Handle.
1. Loading Handle Knob
2. Knob Retention Thumb Screw (not shown, behind
Loading Handle Knob)
11. Verify that the line on the Loading Handle Knob
and the line on the Loading Handle are aligned.
The Loading Handle Knob can be rotated until
the two (2) are aligned.
12. Assure that the Tuohy-Borst valve is open. If
closed, turn the rotator in the counter-
clockwise direction to open the valve.
13. Thread the tether through the Loading Handle,
ensuring the side of the Loading Handle with
the knob is aligned with the black indicator line
on the Loading Tube.
14. Attach the Loading Handle to Loading Tube.
15. Use the saline bag to continue filling the
remainder of the Loading System until saline
exits the Loading Handle Extension Tube and
fittings. If leaks are observed, reconfirm keying
of Loading Handle.
16. Close the Tuohy-Borst valve and all stopcocks.
17. Turn off the flush and remove the saline line
from the Closed Cap Extension Tube.
18. Invert Loading System and secure to the stand
by securing the Loading Tube into the Rotating
Clip - Clip 1 and Loading Handle into the
Bottom Loading Clip - Clip 2.
19. Remove Closed Cap from assembly.
20. Place the Tether Clip onto the tether leader at
the end of the Tuohy-Borst valve.
21. Apply tension to the tether leader and slide the
Tether Clip into contact with the Tuohy-Borst
valve.
22. Use syringe with de-air (air evacuation) tube to
remove air from the valve pockets.
IFU, TENDYNE
TM
Mitral Valve System
Page 18 of 284
CAUTION: Do not damage the pocket fabric or
the leaflets.
23. Align the anterior aspect of the valve cuff
(raised portion) to the Loading Funnel cutout
(Figure 13).
Figure 13. Top down view of valve in Loading Funnel
1. Raised portion of valve
2. Funnel Cutout
24. If a Tendyne Mitral Valve LP is being loaded,
replace the Centering Cone and attach the
Small Inner Centering Cone to the Centering
Rod.
If a standard Tendyne Mitral Valve is being
loaded, do not replace the Centering Cone.
Screw the Centering Cap assembly onto the
Loading Funnel with the Centering Cone fully
retracted, then, while holding the Centering
Rod loosen the thumb screw and gently lower
the Centering Cone in contact with the valve.
10.7. Loading the Valve
1. Rotate the Loading Handle Knob
counterclockwise to compress the valve into
the Loading Tube.
2. Ensure the Centering Cone supports the valve
as it is compressed by gently applying force to
the Centering Rod until the rod reaches its
stop, and then tighten the thumb screw.
3. Continue rotating the Loading Handle Knob
until the stop is reached and the valve is fully
loaded.
CAUTION: Note time upon completion of this
step. Do not keep the valve loaded in the
Loading Tube longer than 15 minutes. If the
valve is in the Loading Tube longer than
15 minutes, discard the valve and prep a new
valve.
4. Confirm that the tether leader connection is
still visible in the Loading Handle extension
tube.
CAUTION: If the tether leader is not visible in
the Loading Handle extension tube when the
valve is fully loaded, discard the valve and prep
a new valve.
5. Remove Centering Cap assembly. Place Closed
Cap on Loading Funnel.
10.8. Moving the Loading Tube to the Delivery System
Handle
1. Remove the Loading Funnel from the Loading
Tube by unthreading the Loading Funnel collet
connecting the Loading Funnel to the Loading
Tube.
2. Turn the Loading Funnel to the side and keep
the funnel over the Loading Stand Base to
control saline spillage.
3. Ensure the Delivery Sheath side port aligns
with the Loading Tube black mark and attach
the sheath to the Loading Tube by tightening
the collet.
4. Attach saline line to the Hemostasis Adapter on
the Sheath Side Port.
5. Fill the sheath with saline.
NOTE: Do not turn off the saline flush to the
sheath side port.
6. Inflate the collapsible dilator to 4-5 atm to seal
off the end of the sheath, while continuing to
flush through the sheath side port.
7. Close all stopcocks to turn off the flush.
8. Place the sheath and Loading Tube assembly
into the Angled Clip Clip 3.
9. Remove the tether clip and loosen the Tuohy-
Borst valve.
10. Pull the Loading Handle Knob to disengage
while turning the knob clockwise one complete
rotation. Remove the Knob Retention Thumb
Screw and remove the Loading Handle Knob
from the Loading Handle.
IFU, TENDYNE
TM
Mitral Valve System
Page 19 of 284
11. Pull the tether until taut, loosen the Loading
Tube collet, and then remove the Loading
Handle.
12. Alternate pulling the leader and handle to fully
remove tether and prevent entanglement.
NOTE: Verify the tether is not tangled by looking
through the opening where the knob was
removed on the Loading Handle.
13. Tap the sheath and Loading Tube to remove
any air bubbles that may reside behind the
valve after loading. The flush may also be
turned on and off during this step to replace
any saline lost during Loading Handle removal.
Figure 14. Tendyne
TM
Delivery System. Upper image: top view.
Lower image: bottom view.
1. Sheath Retraction Knob Knob 3
2. Tether Retention Knob Knob 2
3. Delivery Push Knob Knob 1
4. Top Alignment Window
5. Bottom Alignment Window
14. Confirm the sheath retraction mechanism of
the Delivery Handle is properly positioned by
checking the Bottom Alignment Window to
assure that the tab is centered. Adjust travel
using the Sheath Retraction Knob Knob 3 if
necessary.
15. Confirm the Delivery Push Knob Knob 1 is
fully rotated counterclockwise.
NOTE: Do not over rotate and damage the
mechanism.
16. Attach a Tuohy-Borst with side extension to the
luer connector on the end of the Delivery
Handle and then attach a 1-way stopcock to
the Extension Tube on the Tuohy-Borst valve.
17. Ensure that the Tuohy-Borst valve is open.
Carefully inspect the valve assembly at the
braided tether to nitinol leader joint and verify
that there is no evidence of damage.
CAUTION: Do not use the valve if there is any
sign of damage or deterioration.
18. Feed the tether through the Delivery Handle
until the tether leader exits the hub of the
Delivery Handle and Tuohy-Borst valve.
19. Maintain tension on the tether and attach the
Delivery Handle to the Loading Tube, ensuring
the Delivery Handle is in proper alignment with
the Loading Tube. Clip one end of the Delivery
Handle into Rotating Clip - Clip 1.
The black indicator on the Loading Tube should
align with the Top Alignment Window.
20. Place the tether leader into the retention
openings on the hub of the Delivery Handle.
NOTE: Four (4) to six (6) inches (10 to 15
centimeters) of leader should be in the hub for
retention.
21. Make sure that all luer connections are tight
and secure. Flush the Delivery Handle with
saline until saline exits the Tuohy-Borst valve
and 1-way stopcock. Close the Tuohy-Borst
valve and the 1-way stopcock.
22. Remove the shipping mandrel from the
collapsible dilator and flush the dilator lumen
with sterile saline
23. Verify the Tether Retention Knob - Knob 2 is
not engaged with the tether leader.
10.9. Preparing for the Transapical Procedure
The valve should be implanted with sterile
technique using primarily transesophageal
echocardiographic guidance. In addition,
transthoracic echocardiographic and fluoroscopic
imaging modalities may be used.
WARNING: Only physicians who have
undergone training on use of the TENDYNE
TM
Mitral Valve System should operate this
product.
1. Prepare the thoracic access site using standard
practice.
2. Determine the proper trajectory from the
ventricular myocardium to the mitral valve
using echocardiography guidance.
3. Determine proper apical access site using
transesophageal echocardiography. Puncture
the myocardium with the introducer needle
and insert a standard 0.035” (0.89 mm) guide
IFU, TENDYNE
TM
Mitral Valve System
Page 20 of 284
wire. With wire in the left ventricle, remove
introducer needle.
WARNING: Do not attempt to re-insert a
partially or completely withdrawn needle.
4. Insert an 8F sheath into the left ventricle and
remove dilator. Insert a balloon tip catheter
over the wire and through the sheath into the
left ventricle. Using the inflated balloon,
establish an entanglement free pathway from
the ventricular access site to the left atrium
leading with the inflated balloon tip catheter.
5. Advance the guide wire across the mitral valve
and into the left atrium. Deflate the balloon
and remove the balloon tipped catheter from
the sheath off the wire. Remove the sheath
from the heart and off the wire. Maintain wire
position in the left atrium.
10.10. Deploying the Valve
1. Run a continuous saline flush through the
Delivery Handle. Confirm that the pressure is
between 150-250 mmHg.
2. Insert the collapsible dilator and sheath
through the myocardium over the 0.035”
(0.89 mm) guide wire.
3. Advance the Delivery Handle using
echocardiography and fluoroscopy guidance to
maintain proper trajectory and to confirm final
sheath depth is above the mitral annulus.
CAUTION: Maintain end of sheath
approximately 1 centimeter (cm) above mitral
annulus to ensure device does not interact with
atrial wall.
4. Deflate the collapsible dilator for at least
5 seconds and remove. If resistance is
encountered, confirm the dilator has fully
deflated. Remove the 0.035” guidewire from
the heart through the hemostasis valve.
5. Stop the flush and evacuate air from the
sheath through the hemostasis adapter side
port.
6. Restart flush.
7. Rotate the Delivery Push Knob Knob 1 on the
Delivery Handle in the direction indicated by
the arrows to push the valve out of the Loading
Tube into the sheath.
8. Continue rotating Delivery Push Knob Knob 1
until the indicator line is seen in the Top
Alignment Window, indicating that the valve is
at the end of the sheath.
Confirm on echo that the valve is at the end of
the sheath.
9. Turn off the flush.
10. Pull the tether taut and tighten the Tether
Retention Knob - Knob 2 at the end of the
Delivery Handle to secure the tether leader.
Use fingertips only. Do not overtighten.
11. Ensure the tip of the sheath is at the level of
the mitral annular plane.
12. Rotate Sheath Retraction Knob - Knob 3 in the
direction indicated by the arrows to retract the
sheath and initiate valve deployment.
CAUTION: Ensure that Delivery Push Knob
Knob 1 does not rotate.
10.11. Orienting the Valve
1. Ensure that the valve is at the level of the
mitral annulus and that the cuff has been
deployed a sufficient amount to correctly
identify the orientation of the valve A2 cuff.
2. Confirm with echocardiography that the valve
is oriented correctly, ensuring that the raised
anterior aspect (straight segment of the valve)
is radially oriented toward the aortomitral
continuity.
Fluoroscopy may also be used to confirm
orientation, using pre- determined projection
angles from pre-operative CT scan and properly
aligning the A1 radiopaque marker on the
valve.
3. Continue to rotate Sheath Retraction Knob -
Knob 3 in the direction of the arrow until a
hard stop.
4. Rotate Delivery Push Knob Knob 1 in the
direction opposite the indicator arrow one-half
turn to deploy the valve fully.
5. If the valve needs to be repositioned, use the
Delivery Handle to recapture the valve stem
and nitinol struts by rotating the Sheath
Retraction Knob - Knob 3 in the direction of the
indicator arrows.
CAUTION: Confirm native leaflets are not
captured between the sheath and valve.
6. Once recaptured, push the valve into the left
atrium, reorient radially as needed, re-orient
axially within the annulus and redeploy by
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284

Abbott Tendyne TENDV-SP-33A Instructions For Use Manual

Tipo
Instructions For Use Manual