Equinox Systems 4000 Manuale utente

Tipo
Manuale utente
eQuinox
TM
Model 4000 Oxygen Concentrator
eQuinox
TM
Modèle 4000 Concentrateur d’oxygène
eQuinox
TM
Modell 4000 Sauerstoffkonzentrator
eQuinox
TM
Modelo 4000 Concentrador de oxígeno
eQuinox
TM
Modello 4000 Concentratore di ossigeno
eQuinox
TM
Modelo 4000 Concentrador de oxigénio
eQuinox
TM
Model 4000 Zuurstofconcentrator
User Manual
Manuel d’utilisation
Bedienungsanleitung
Manual del usuario
Manuale dell’utente
Manual do Utilizador
Gebruikershandleiding
2 — 4805-C1 C (English)
Symbol Definition Symbol Definition
Read user manual before operation. See user manual
for instructions.
FAA Approved Symbol: The U.S. Federal Aviation
Administration (FAA) has approved this device for
use on-board commercial aircraft.
No Smoking Icon: Do not smoke near unit. Flow Setting Indicator
Warnings / ALERT (Yellow) Indicator Pulse Mode Operation
ALARM (Red) Indicator Device operating normally; power button
External Power Indicator Increase Flow Setting
Battery Pack (battery) Status Gauge Decrease Flow Setting
Battery Pack Symbol Use no oil or grease.
A
Amperes No open or naked ames.
Alternating Current No serviceable parts inside. Do not open cover.
IEC Icon for DC Power (Input) Warm Surface.
0
2
Oxygen Output Do not get wet.
Type BF Applied Part (degree of protection against
electric shock)
IPX1
Drip Proof Equipment-IPX1: The eQuinox Battery
Pack (Battery) and AC Power Supply provide
protection against the harmful effects of the ingress
of liquids. (IPX1, per IEC 60529)
Recycle Symbol Name and address of manufacturer
This symbol is to remind the equipment owners
to return it to a recycling facility at the end of its
life, per Waste Electrical and Electronic Equipment
(WEEE) Directive.
Authorized representative in the European
Community
IP22
Drip and solid ingress proof: The eQuinox provides
protection against the harmful effects of the ingress
of liquids and solid foreign objects in accordance
with IEC 60529 at the rating of IP22.
Class II equipment
Certied for both the U.S. and Canadian markets, to
the applicable U.S. and Canadian standards.
This device complies with the requirements of
Directive 93/42/EEC concerning medical devices. It
therefore bears the CE marking as shown.
User Controls & System Status Indicators
eQuinox (Model 4000) User Manual — 3
Contents
User Controls & System Status Indicators ..............................................................................................2
Quick Start Guide ......................................................................................................................................4
Indications for Use ....................................................................................................................................7
Contraindications for Use .........................................................................................................................7
Safety Guidelines ......................................................................................................................................7
Recommended Operating Environments ................................................................................................9
Operating Instructions ............................................................................................................................10
Voice Module Operation .........................................................................................................................13
First Time eQuinox Operation ................................................................................................................14
User Troubleshooting Table ...................................................................................................................16
Cleaning, Care & Routine Maintenance .................................................................................................19
EMC Tables ...............................................................................................................................................21
General Airline Travel Information ........................................................................................................25
Battery Pack & Airline Travel ..................................................................................................................25
4 — 4805-C1 C (English)
Quick Start Guide
AC Power Supply
12VDC Cable
eQuinox Cart
Battery Pack
(battery)
1
Unpack Your eQuinox
2
Getting to Know Your eQuinox
Review all Warnings, Cautions and additional device
information in the rest of this manual.
Become familiar with the key features of the eQuinox and the
User Control Panel.
BACK
Battery
Pack Latch
Exhaust
Vent
Battery
Pack
Service
Cabinet
FRONT
Control
Panel
Handle
External Power
Receptacle
Oxygen
Outlet Port
FAA
Approval
Icon
Rating Label &
Serial Number
Location
Air inlet
Filter
eQuinox (Model 4000) User Manual — 5
ON/OFF Button (Green) Indicator
Press and hold the “ON/OFF” button for two (2) seconds to
power ON your eQuinox. A brief audible and visual power on
self-test will occur. Check that all the indicators are lit up and
the buzzer sounds for three (3) seconds. Press and hold the “ON/
OFF” button for two (2) seconds to power OFF your eQuinox.
To connect to AC power source: Insert the AC Power Supply
cord securely into the External Power Receptacle on the right
side of the device. Plug the AC power cord into the power
supply and into an AC outlet.
When the eQuinox is plugged in properly, a green indicator
on the AC Power Supply will light up. The External Power
Indicator on the user control panel will also light up.
Install the Battery: Slide the battery into the empty compartment
until it locks into position. It will be ush with the front of the
device.
When the battery is properly installed and the eQuinox is either
ON or plugged into AC power, the Battery Pack Status Gauge
will appear.
3
Powering on the eQuinox
The eQuinox User Control Panel
displays important operating
information.
ON/OFF Button (Green) Indicator: This button powers
the device ON or OFF. The Green Indicator is lit up when
the device is ON and functioning properly.
Increase or Decrease Flow Setting Buttons: Use these
buttons to set the flow to your prescribed setting.
Delivery Mode Button and Indicator: The button switch-
es between Continuous Flow and Pulse Dose Mode. The
Pulse Dose Mode activates autoSAT Technology. This
allows a significant increase in the operating time while
powered by the battery pack (battery). When the Pulse
Dose Mode is activated, the green Pulse Dose Mode
Indicator lights up. A pulse of oxygen is delivered and
the green indicator blinks with each inspiratory effort.
ALERT (Yellow) Indicator—When lit up, this indicates a
low priority awareness condition or Caution. Continue
to use your system. Refer to the Troubleshooting Table
for the correct action, or contact your Home Care
Provider. A flashing yellow indicates a medium priority
alert. A prompt response is necessary. (Typically low
oxygen or low battery.)
ALARM (Red) Indicator —Indicates a high priority alarm
condition. An immediate response is necessary. Refer
to the Troubleshooting Table and contact your home
care provider. (Typically loss of power, low oxygen, or
malfunction.)
Information Display.
• C or P – “C” indicates Continuous Flow mode and
“P” indicates Pulse Dose mode.
Time on battery – The display will cycle every 20
seconds when running on battery power to indicate the
estimated time left on the battery
• Setting (continuous ow or pulse dose) – each time
you power the device ON, the last mode and/or setting
has been saved and will be used at start-up.
Battery Pack (battery) Status Gauge: This indicator
displays the charge remaining in the battery. Each of
the five horizontal gray bars represents approximately
20% of the total battery charge. When the battery is
charging, the charge indicator bars will blink in a wa-
terfall-type fashion. If the battery is not installed, or if it
is incorrectly installed, the Battery Pack (battery) Status
Gauge will not light up or will flash on and off.
External Power Indicator: When the eQuinox is properly
plugged in and is using the AC or 12VDC Cable, this
indicator will light up on the User Control Panel.
Buzzer: An audible alarm alerts you to the operating condition of the
device, either a warning or failure, and confirms a valid key press.
Voice Module: If activated, a voice indicator will alert you of the op-
erating condition of the device, changes in power source, and warning
conditions or failures.
Step 1 Step 2 Step 3
6 — 4805-C1 C (English)
Line up the eQuinox with the shaped oor of the cart so that it
sits at and upright. Line up the large threaded screw to the insert
at the back of the eQuinox. Then hand-tighten the knob on the
cart until the cart is tightly secured to the device. Depress the
push-button on the telescoping cart handle and pull upward to
adjust the height of the handle.
Your unit is equipped with an Air Inlet Filter (extra lter
provided). Check daily and clean once a week. Instructions for
cleaning are on page 19.
Connect the oxygen supply tubing to the oxygen outlet and
connect the oxygen cannula per the instructions provided with
the cannula, or according to the instructions from your home care
provider.
Breathe normally through the nasal cannula.
When operating on battery power, using the Pulse Dose Mode
will result in longer duration-of-use.
Air Inlet Filter
6
Begin Using the eQuinox
7
eQuinox Cart Operation
Flow Mode Button and Indicator
Press the Flow Mode button to select your desired ow mode.
Pressing this button repeatedly will switch you back and forth
between continuous ow and pulse dose modes. In Continuous
Flow Mode, the oxygen is provided at a constant ow rate
between 0.5 and 3.0 LPM, indicated by the letter C. In Pulse
Dose Mode, oxygen is supplied in a bolus at the beginning of
each inhalation. Pulse mode settings 1–9 provide bolus sizes
ranging from 16mL to 192mL and are designated by the letter
P on the display. Pulse dose can be adjusted for sensitivity and
breath bolus delivery. Flow settings in both modes should be
prescribed for you by your physician.
Increase or Decrease Flow Setting Button
Using the + or - buttons, adjust the setting to the prescription
supplied by your physician.
Continuous Flow Mode Display
Pulse Dose Mode Display
4
Select Flow Delivery Mode
5
Adjust Flow Setting to Prescribed Level
WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE
PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE
FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO
DO SO BY A LICENSED CLINICIAN. THE PORTABLE OXYGEN CONCEN-
TRATOR MAY BE USED DURING SLEEP UNDER THE RECOMMENDA-
TION OF A QUALIFIED CLINICIAN.
Air inlet
Filter
eQuinox (Model 4000) User Manual — 7
Important!
Safety Instructions are defined as follows:
Indications for Use
The eQuinox is intended for the administration of supplemental
oxygen. The device is not intended for life support, nor does it
provide any patient monitoring capabilities.
Contraindications for Use
Safety Guidelines
WARNING: IMPORTANT SAFETY INFORMATION FOR HAZ-
ARDS THAT MIGHT CAUSE SERIOUS INJURY.
CAUTION: Important information for preventing damage
to the eQuinox.
Note: Information needing special attention.
WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE
PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE
FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO
DO SO BY A LICENSED CLINICIAN. THE PORTABLE OXYGEN CONCEN-
TRATOR MAY BE USED DURING SLEEP UNDER THE RECOMMENDA-
TION OF A QUALIFIED CLINICIAN.
WARNING: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE OR
RENTAL BY ORDER OF A PHYSICIAN OR OTHER LICENSED HEALTH
CARE PROVIDER.
WARNING: THIS UNIT IS NOT TO BE USED FOR LIFE SUPPORT.
GERIATRIC, PEDIATRIC, OR ANY OTHER PATIENT UNABLE TO COM-
MUNICATE DISCOMFORT WHILE USING THIS DEVICE MAY REQUIRE
ADDITIONAL MONITORING. PATIENTS WITH HEARING AND/OR
SIGHT IMPAIRMENT(S) MAY NEED ASSISTANCE WITH MONITORING
ALARMS.
WARNING: NO MODIFICATION OF THIS EQUIPMENT IS
PERMITTED
WARNING: IN CERTAIN CIRCUMSTANCES, THE USE OF
NON-PRESCRIBED OXYGEN CAN BE HAZARDOUS. THIS
DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A
PHYSICIAN.
WARNING: NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANES-
THETICS.
WARNING: AS WITH ANY ELECTRICALLY POWERED DEVICE, THE
USER MAY EXPERIENCE PERIODS OF NON-OPERATION AS A RESULT
OF ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE
EQUINOX SERVICED BY A QUALIFIED TECHNICIAN. THE EQUINOX
IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE
ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPO-
RARY INTERRUPTION.
WARNING: THE MANUFACTURER RECOMMENDS AN ALTER-
NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT
OF A POWER OUTAGE, ALARM CONDITION, OR MECHANI-
CAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT
PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED.
WARNING: THIS DEVICE SUPPLIES HIGH-CONCENTRATION OXYGEN
THAT PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR
OPEN FLAMES WITHIN THE SAME ROOM OF (1) THIS DEVICE, OR (2)
ANY OXYGEN-CARRYING ACCESSORY. FAILURE TO OBSERVE THIS
WARNING CAN RESULT IN SEVERE FIRE, PROPERTY DAMAGE, AND/
OR CAUSE PHYSICAL INJURY OR DEATH.
WARNING: DO NOT OPERATE UNIT IN A RESTRICTED OR CONFINED
SPACE (I.E., A SMALL CASE OR HANDBAG) WHERE VENTILATION
CAN BE LIMITED. THIS CAN CAUSE THE OXYGEN CONCENTRATOR TO
OVERHEAT AND IMPAIR PERFORMANCE.
WARNING: THE CONCENTRATOR SHOULD BE LOCATED AS TO AVOID
SMOKE, POLLUTANTS OR FUMES.
WARNING: THE USE OF SOME OXYGEN ADMINISTRATING ACCES-
SORIES NOT SPECIFIED FOR THIS OXYGEN CONCENTRATOR MAY
IMPAIR ITS PERFORMANCE. RECOMMENDED ACCESSORIES ARE
REFERENCED WITHIN THIS MANUAL.
WARNING: IF THE OXYGEN CONCENTRATOR HAS BEEN DROPPED,
DAMAGED OR EXPOSED TO WATER, PLEASE CONTACT YOUR HOME
CARE PROVIDER FOR INSPECTION OR POSSIBLE REPAIR OF THE
DEVICE. DO NOT USE THE OXYGEN CONCENTRATOR IF IT HAS A
DAMAGED POWER CORD OR PLUG.
WARNING: PROPERLY SECURE, BELT OR OTHERWISE RESTRAIN THE
OXYGEN CONCENTRATOR WHEN IN A VEHICLE DURING TRANSPORT
TO PREVENT DAMAGE OR INJURY.
WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR
OUTLET VENTS TO BECOME BLOCKED. DO NOT DROP OR INSERT ANY
OBJECTS INTO ANY OPENINGS ON THE DEVICE. THIS CAN CAUSE
THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR PERFOR-
MANCE.
WARNING: DO NOT OVERFILL THE OPTIONAL HUMIDIFIER. FILL THE
OPTIONAL HUMIDIFIER WITH WATER ONLY TO THE LEVEL SHOWN BY
THE MANUFACTURER OF THE HUMIDIFIER.
WARNING: DO NOT OPERATE THE EQUINOX AND AMBULATE WHILE
A HUMIDIFIER IS ATTACHED. REMOVE THE HUMIDIFIER BOTTLE BE-
FORE WALKING. DO NOT LAY THE EQUINOX DOWN WHILE ATTACHED
TO A HUMIDIFIER BOTTLE.
WARNING: THE US DEPARTMENT OF TRANSPORTATION (DOT) AND
UNITED NATIONS (UN) REGULATIONS REQUIRE THE REMOVAL OF
THE BATTERY FROM THE DEVICE FOR ALL INTERNATIONAL AIRLINE
TRAVEL WHEN THE OXYGEN CONCENTRATOR IS CHECKED AS LUG-
GAGE. WHEN SHIPPING THE OXYGEN CONCENTRATOR, THE BATTERY
MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED
PROPERLY.
WARNING: DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST
READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE
TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT
YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS
EQUIPMENT; OTHERWISE, INJURY OR DAMAGE MAY RESULT.
8 — 4805-C1 C (English)
WARNING: IF YOU FEEL DISCOMFORT OR ARE EXPERIENC-
ING A MEDICAL EMERGENCY, SEEK MEDICAL ASSISTANCE
IMMEDIATELY.
WARNING: OPERATING THE OXYGEN CONCENTRATOR OUTSIDE OF
THE OPERATIONAL TEMPERATURE SPECIFICATIONS CAN LIMIT THE
CONCENTRATOR’S ABILITY TO MEET OXYGEN CONCENTRATION
SPECIFICATION. REFER TO THE SPECIFICATION SECTION OF THIS
MANUAL FOR TEMPERATURE LIMITS.
WARNING: USE NO OIL, GREASE, OR PETROLEUM-BASED OR OTHER
FLAMMABLE PRODUCTS WITH THE OXYGEN-CARRYING ACCESSO-
RIES OR THE OXYGEN CONCENTRATOR. ONLY WATER BASED, OXY-
GEN COMPATIBLE LOTIONS OR SALVES SHOULD BE USED. OXYGEN
ACCELERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES.
WARNING: THE INCORRECT USE OF THE BATTERY CAN CAUSE THE
BATTERY TO GET HOT, IGNITE, AND MAY CAUSE SERIOUS INJURY. BE
SURE NOT TO PIERCE, STRIKE, STEP ON, DROP THE BATTERY, OR OTH-
ERWISE SUBJECT THE BATTERY TO STRONG IMPACTS OR SHOCKS.
THE USE OF A DAMAGED BATTERY MAY CAUSE PERSONAL INJURY.
WARNING: THE OXYGEN CONCENTRATOR SHOULD NOT BE USED
ADJACENT TO OR STACKED WITH OTHER EQUIPMENT. IF ADJACENT
OR STACKED USE IS UNAVOIDABLE, THE DEVICE SHOULD BE OB-
SERVED TO VERIFY NORMAL OPERATION.
CAUTION: In the event of an alarm or you observe the
Oxygen Concentrator is not working properly; consult
the Troubleshooting section in this manual. If you cannot
resolve the problem, consult your Equipment Provider.
CAUTION: To prevent a void warranty, follow all manufacturers’
instructions.
WARNING: SMOKING WHILE USING OXYGEN IS THE
NUMBER ONE CAUSE OF FIRE INJURIES AND RELATED
DEATHS. YOU MUST FOLLOW THESE SAFETY WARNINGS:
WARNING: DO NOT ALLOW SMOKING, CANDLES, OR OPEN FLAMES
IN THE SAME ROOM WITH THE DEVICE OR THE OXYGEN-CARRYING
ACCESSORIES.
WARNING: SMOKING WHILE WEARING AN OXYGEN CANNULA MAY
CAUSE FACIAL BURNS AND POSSIBLY DEATH.
WARNING: IF YOU SMOKE, THESE 3 STEPS MAY SAVE YOUR LIFE:
TURN OFF THE OXYGEN CONCENTRATOR, TAKE OFF THE CANNULA,
AND LEAVE THE ROOM WHERE THIS DEVICE IS LOCATED.
WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE
PROMINENTLY DISPLAYED IN THE HOME, OR WHERE OXYGEN IS IN
USE. PATIENTS AND THEIR CAREGIVERS MUST BE INFORMED ABOUT
THE DANGERS OF SMOKING IN THE PRESENCE OF, OR WHILE USING,
MEDICAL OXYGEN.
WARNING: DO NOT USE YOUR OXYGEN CONCENTRATOR IN THE
PRESENCE OF FLAMMABLE GASES. THIS CAN RESULT IN RAPID
BURNING CAUSING PROPERTY DAMAGE, BODILY INJURIES OR
DEATH.
WARNING: REMOVING THE CANNULA AND PUTTING IT ON CLOTH-
ING, BEDDING, SOFAS, OR OTHER CUSHION MATERIAL WILL CAUSE
A FLASH FIRE WHEN EXPOSED TO A CIGARETTE, HEAT SOURCE, OR
FLAME.
WARNING: DO NOT LEAVE A NASAL CANNULA ON OR UNDER
CLOTHING, BED COVERINGS OR CHAIR CUSHIONS. IF THE UNIT IS
TURNED ON BUT NOT IN USE, THE OXYGEN WILL MAKE THE MA-
TERIAL FLAMMABLE. SET THE OXYGEN CONCENTRATOR TO THE 0
(OFF) POSITION WHEN THE OXYGEN CONCENTRATOR IS NOT IN USE.
WARNING: DO NOT ATTEMPT ANY MAINTENANCE OTHER
THAN THE POSSIBLE SOLUTIONS LISTED WITHIN THIS
MANUAL. DO NOT REMOVE COVERS, ONLY YOUR EQUIP-
MENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD
REMOVE THE COVERS OR SERVICE THIS DEVICE.
WARNING: ONLY ACCESSORIES RECOMMENDED BY THE MANUFAC-
TURER. USE OF ANY OTHER MAY BE HAZARDOUS, CAUSE SERIOUS
DAMAGE TO YOUR OXYGEN CONCENTRATOR AND WILL VOID THE
WARRANTY .
CAUTION: Always disconnect AC power supply from the
wall before disconnection the AC power supply from the
oxygen concentrator.
CAUTION: Always place oxygen supply tubing and power cords in a
manner that prevents a trip hazard.
CAUTION: When using the Oxygen Concentrator in an automobile,
boat, or on other DC sources with the 12VDC Cable, make sure that
the vehicle is started and running before connecting the Oxygen
Concentrator. If the external power light on the eQuinox does not
illuminate and requires resetting, disconnect the 12VDC Cable from
the DC outlet, restart your vehicle, and then reconnect your 12VDC
Cable into the DC outlet. Failure to follow these instructions can
result in not supplying power to the Oxygen Concentrator.
CAUTION: When the automobile in which you are using the Oxygen
Concentrator unit is turned off, disconnect and remove the device
from the automobile. Do not store the Oxygen Concentrator in a
very hot or cold automobile or in other similar, high-or low-tem-
perature environments.
Note: Portable and mobile RF communications equipment can effect
medical electrical equipment.
WARNING: THE MANUFACTURER RECOMMENDS AN ALTER-
NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT
OF A POWER OUTAGE, ALARM CONDITION, OR MECHANI-
CAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT
PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED.
WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR
OUTLET VENTS TO BECOME BLOCKED. DO NOT DROP OR INSERT
ANY OBJECTS INTO ANY OPENINGS ON THE DEVICE. THIS CAN
CAUSE THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR
PERFORMANCE.
eQuinox (Model 4000) User Manual — 9
Recommended Operating Environments
For correct use of the device refer to the following chart. The
chart provides important information about the recommended
operating environments, or operating conditions.
Specifications
Operating
Temperature
32º F to 104º F (0º C to 40º C)
Operating
Humidity
10% - 95% at an 82.4° F (28° C) dew point
Transport/
Storage
Temperature
-13°F to 158°F (-25°C to 70°C)
Humidity: Up to 95% Non-Condensing
-4°F to 122°F (-20°C to 50°C) for battery packs
Electrical Use no extension cords. Use no electrical outlets con-
trolled by a switch.
Altitude 0 – 13,123 feet (0 - 4,000 meters)
Placement DO NOT block the air inlet (top back cover) or the exhaust
vent (back bottom right). Place device a minimum of 3
inches (7.5 cm) away from walls, draperies, furniture, etc.
Environment Must be smoke, pollutant, and fume free.
Operating
Time
24-hours a day when connected to an external AC or DC
power source.
Proper Placement of the eQuinox
Pick a location for the device that avoids the intake of smoke,
fumes and pollutants. Correct positioning of the device should
allow intake of air through the air inlet lter at the top rear of
the cabinet. It should also allow exhaust air to freely leave the
exhaust vent at the bottom right of the device.
Place the device so that the alarms can be heard.
Position the oxygen supply tubing in such a way that it does not
kink or occlude.
Note: If the Oxygen Concentrator has been stored for an extended period
of time outside its normal operating temperature range, the unit should
be allowed to return to normal operating temperature before being
turned on. (Refer to the Specifications section in this manual.)
Note: When the automobile in which you are using the Oxygen Con-
centrator unit is turned off, disconnect and remove the device from the
automobile. Do not store the Oxygen Concentrator in a very hot or cold
automobile or in other similar, high-or low-temperature environments.
DO NOT leave the Oxygen Concentrator or the Power Supply plugged
into the vehicle if the ignition is in the off position. Doing so may drain
the vehicle’s battery.
WARNING: DO NOT USE YOUR OXYGEN CONCENTRATOR
IN THE PRESENCE OF FLAMMABLE GASES. THIS CAN
RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE,
BODILY INJURIES OR DEATH. USE NO OIL, GREASE, OR PETRO-
LEUM-BASED OR OTHER FLAMMABLE PRODUCTS WITH THE OXY-
GEN-CARRYING ACCESSORIES OR THE OXYGEN CONCENTRATOR.
ONLY WATER BASED, OXYGEN COMPATIBLE LOTIONS OR SALVES
SHOULD BE USED. OXYGEN ACCELERATES THE COMBUSTION OF
FLAMMABLE SUBSTANCES.
WARNING: THIS DEVICE SUPPLIES HIGH-CONCENTRATION OXYGEN
THAT PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR
OPEN FLAMES WITHIN THE SAME ROOM OF (1) THIS DEVICE, OR (2)
ANY OXYGEN-CARRYING ACCESSORY. FAILURE TO OBSERVE THIS
WARNING CAN RESULT IN SEVERE FIRE, PROPERTY DAMAGE, AND/
OR CAUSE PHYSICAL INJURY OR DEATH.
CAUTION: Do not operate unit in a restricted or confined
space (i.e., a small case or handbag) where ventilation
can be limited. This can cause the Oxygen Concentrator
to overheat and impair performance. Do not allow either the air
intake or the air outlet vents to become blocked. This can cause the
Oxygen Concentrator to overheat and impair performance.
10 — 4805-C1 C (English)
Operating Instructions
Before Operating
This user manual is your reference to help you use and maintain
the device. If you have any questions or concerns please call
your home care provider.
Important! DO NOT try to use the eQuinox without rst reading
the Safety Guidelines section of this manual. Please follow all
of the operating instructions. Please observe all Warnings on the
device and in the User Manual. In order to decrease the risk of
re, personal injury and serious damage to the eQuinox, please
follow all of the safety precautions.
Powering on your eQuinox
The device can be operated from three (3) different power
sources:
AC Power Supply
• 12VDC Cable
• Rechargeable Battery Pack (Battery)
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND
POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER
APPLIED WITH A DAMP (NOT WET) CLOTH OR SPONGE,
AND THEN WIPE ALL SURFACES DRY. DO NOT ALLOW ANY LIQUID TO
GET INSIDE THE DEVICE. PAY SPECIAL ATTENTION TO THE OXYGEN
OUTLET FOR THE CANNULA CONNECTION TO MAKE SURE IT RE-
MAINS FREE OF DUST, WATER, AND PARTICLES.
CAUTION: Do not allow either the air intake or the air out-
let vents to become blocked. This can cause the Oxygen
Concentrator to overheat and impair performance. Do not
operate the Oxygen Concentrator without the air intake filter in
place. If a second filter is provided, insert the “replacement” filter
before you clean the dirty filter. Clean the dirty filter in a warm
soap and water solution then dry thoroughly prior to use.
AC POWER SUPPLY
The eQuinox includes a universal AC Power Supply for use at
home or wherever standard AC power is available. To connect to
an AC power source, plug the AC Power Supply cord securely
into the recessed external connector. Also make sure that the cord
from the AC Power Supply is plugged into an AC outlet. When
the eQuinox is plugged in correctly, a green indicator on the AC
Power Supply and the External Power Indicator will light up. The
indicator
on the user control panel will also light up. The
CAIRE AC Power Supply PN 4836 is the main power source for
the eQuinox portable oxygen concentrator. It is a very important
part of the overall system.
Note: Use only electrical voltage specified on the specification label
affixed to the device.
AC Power Supply
General Charging Information
The eQuinox runs on external AC power, DC power, or a
rechargeable battery. The system includes a “smart battery
charger.” This recharges the battery whenever the eQuinox is
connected to an outside power source. The system monitors and
controls the power source and the battery charger.
The eQuinox charges the battery when using AC or DC power
when the battery temperature is within safe charging temperature.
When external power is removed, the device automatically
switches over to the battery, as long as the battery is installed and
charged. When connected to an AC or DC outlet again, the device
runs on the external power source and recharges the battery.
Line up the power plug with the recessed external connector to
correctly plug in the AC Supply or 12VDC cable. The recessed
power connector is on the right side of the device when facing the
Control Panel. Push the power plug into the recessed connector.
Step 1 Step 2 Step 3
eQuinox (Model 4000) User Manual — 11
CAUTION: Use only accessories recommended by the
manufacturer. Use of any other may be hazardous, cause
serious damage to your oxygen concentrator and will
void the warranty.
CAUTION: Do not use extension cords with this unit or connect too
many plugs into the same electrical outlet. The use of extension
cords could adversely affect the performance of the device. Too
many plugs into one outlet can result in an overload to the electri-
cal panel causing the breaker/fuse to activate or fire if the breaker
or fuse fails to operate.
CAUTION: Use of cables and adapters other than those specified,
with the exception of cables and adapters sold by the manufacturer
of the medical electrical equipment as replacement parts for in-
ternal components, may result in increased emissions of decreased
immunity of the Oxygen Concentrator.
12VDC Cable
12VDC Cable
A 12VDC Cable allows the system to operate from DC outlets,
such as those found in motor vehicles.
1. Start your vehicle.
2. Insert the 12 Volt DC cord into the recessed power
receptacle on the side of the device.
3. Insert the 12VDC Cable plug into the DC power outlet in
your motor vehicle.
When the device is properly connected and receiving power
from the DC power source, the green External Power Indicator
on the control panel will illuminate.
Note: While operating the device from a 12VDC Cable, the Battery may
not charge.
Note: When operating the eQuinox from a 12VDC Cable, it is recommend-
ed that the Pulse Dose Mode be used if prescribed to use a pulse dose
device.
Note: To ensure that the device is utilizing the DC power, check that the
External Power Indicator is illuminated. If a beep is heard, this could indi-
cate a loss of external power. Check if the green External Power Indicator
LED is on or off. If the battery is removed on DC power, it will ensure no
loss of charge while the unit is running. If the indicator is not illuminated,
the battery is powering the device.
Note: Actual performance of the eQuinox may vary depending upon usage
conditions of the motor vehicle. Some vehicles may not be able to support
full functionality on DC power. In the event the vehicle cannot provide
enough power to operate at the current setting, the eQuinox will first
stop charging the battery and continue to run at the prescribed flow rate.
If the vehicle still cannot provide enough power, the eQuinox will then
lower the flow rate if the flow rate is 2.0LPM or greater. If the flow rate is
less than 2.0LPM, the eQuinox will shut down.
If low flow rates are experienced, remove all other electronic devices
connected to the power circuit (GPS, cell phone charger, etc.). If sufficient
power is restored, flow rates are not automatically increased back to the
original setting; user will need to manually adjust. If the eQuinox is still
not operating at the prescribed flow rate, attempt to remove the battery
pack.
WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE
AIR OUTLET VENTS TO BECOME BLOCKED. THIS CAN CAUSE
THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR
PERFORMANCE.
WARNING: IF THE VEHICLE’S DC POWER SOURCE DROPS BELOW 11.5
VOLTS, THE EQUINOX WILL REVERT TO THE BATTERY OPERATION (IF
THE BATTERY IS PRESENT AND CHARGED).
CAUTION: When the automobile in which you are using
the Oxygen Concentrator unit is turned off, disconnect
and remove the device from the automobile. Do not store
the Oxygen Concentrator in a very hot or cold automobile or in
other similar, high-or low-temperature environments. DO NOT leave
the Oxygen Concentrator or the Power Supply plugged into the
vehicle if the ignition is in the OFF position. Doing so may drain the
vehicle’s battery.
eQuinox DC Supply Connection
When the eQuinox is connected to the DC supply in a vehicle, the
following applies:
The eQuinox will run at all continuous flow settings 0.5–
3.0LPM and all pulse settings 1–9 while being operated with
the DC Supply in a car.
The battery will only charge if there is sufficient power
remaining after the device is running at the prescribed flow
rate. The eQuinox will dedicate the power supplied to run
the unit first. Any power left over will be used as a trickle
charge to re-charge the battery. If the eQuinox battery is being
charged, the battery symbol will waterfall. If the eQuinox
battery is not charging, the battery symbol will be displayed
solid.
WARNING: DO NOT TOUCH THE BATTERY CONTACTS AND
THE PATIENT AT THE SAME TIME WHEN ENERGIZING THE
UNIT WITH THE EXTERNAL POWER SUPPLY (AC POWER
SUPPLY) AND WITH NO BATTERY PACK INSTALLED.
12 — 4805-C1 C (English)
Initial Battery Charging
The new Battery that comes with your eQuinox is not fully
charged when it is shipped from the factory. Before using your
eQuinox for the rst time, you must rst fully charge the Battery.
Plug in the AC Power Supply and correctly install the Battery.
Allow the Battery to completely charge. The Battery is fully
charged when the Battery Pack Status Gauge on the Control
Panel no longer cascades and remains solid.
The eQuinox can be used while the Battery recharges. It may
take up to 4 hours to charge, depending on the ow setting. The
Battery will also charge when the eQuinox is plugged into AC
power and turned off.
CAUTION: Only use the Manufacturer’s provided batter-
ies. For proper battery disposal, contact your Equipment
Provider or your local government agency for disposal
requirements.
* Batteries are recyclable. *
In the event of an AC or DC power interruption, the eQuinox will
beep and automatically switch to Battery operation, as long as a
charged Battery is installed. When external power is returned, the
Battery will automatically start charging again. If the Battery is
missing or is fully discharged when the AC power is interrupted,
the eQuinox will shut down and alarm. The Battery is fully
discharged when the Battery Pack Status Gauge on the control
panel appears empty. The device will display a yellow light and
sound a low battery alarm. This will continue until the device
completely shuts down with a red light and constant alarm. The
audible alert may continue for up to 5 minutes until the device
shuts down or external power is restored.
Note: Press and hold the eQuinox ON/OFF button for a minimum of 5
seconds, to silence the Loss of Power alarm. Connecting the eQuinox to
an external AC power supply for 2 minutes will also eliminate this alarm.
WARNING: FOR OXYGEN CONCENTRATORS EQUIPPED WITH
BATTERIES: STORE IN A COOL AND DRY LOCATION TO HELP
ENSURE THE LONGEVITY OF YOUR BATTERY. STORING YOUR
OXYGEN CONCENTRATOR FOR EXTENDED PERIODS OF TIME AT HIGH
TEMPERATURES OR WITH A FULLY CHARGED OR COMPLETELY DIS-
CHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE. DO
NOT ATTEMPT TO OPEN THE BATTERY; THERE ARE NO SERVICEABLE
PARTS INSIDE THE BATTERY. KEEP BATTERIES AWAY FROM CHILDREN.
Rechargeable Battery Pack (Battery)
The eQuinox can be run by the rechargeable
Battery Pack (Battery), which is supplied with
the device.
Installing the Battery: Line up the Battery so it
will slide into the empty Power Compartment.
Push the Battery into the empty Power
Compartment until it locks (clicks) into
position. It will be ush with the front of the
device. When the Battery is properly installed
and the eQuinox is ON, the Battery Status
Gauge will show on the Control Panel. (If you
DO NOT have a Battery installed, the Battery
Pack Status Gauge will not light up.) The unit
will still run on external power.
To remove the Battery, push the release
button on the top of the battery pack. While holding DOWN the
release button, pull the Battery away from the eQuinox.
Battery Operating Times: The battery operating time is shown
on the Control Panel, in both symbol and alpha-numeric formats.
When running on battery power, the battery status gauge will be
shown at all times. The display screen will cycle to display the
hours/minutes of battery life left. A number of factors, such as
ow setting, Pulse or Continuous Flow Mode, and breath rate
will affect the operating time. The table Typical New Battery
Pack Operating Times provides operating time estimates for
the eQuinox using a new, fully charged Battery based on ow
settings and indicated operating conditions.
Note: The eQuinox comes with a 12-cell battery when shipped from the
factory. The 24-cell battery is an optional accessory for longer battery
durations.
Rechargeable Battery
Pack (Battery)
Typical New Battery Pack Operating Times
Operating Type Battery Type
12-cell 24-cell
Continuous
Mode
1 LPM 2.16 hrs (2 hr, 9 min) 4.84 hrs (4 hrs, 50 min)
2 LPM 1.2 hours (1 hour, 12 2.75 hrs (2 hrs, 45 min)
3 LPM 0.8 hrs (48 min) 1.83 hrs (1 hr, 49 min)
Pulse Mode
(12 BPM)
16 mL (1.0) 2.62 hrs (2 hrs, 37 min) 5.94 hrs (5 hrs, 56 min)
48 mL (3.0) 2.38 hrs (2 hrs, 22 min) 5.39 hrs (5 hrs, 23 min)
96 mL (6.0) 1.70 hrs (1 hr, 41 min) 3.85 hrs (3 hrs, 51 min)
Note: Operating the Oxygen Concentrator outside of its normal operating
temperature range can affect performance and decrease battery run time
and/or increase battery charge time. (Refer to the Specifications section in
this manual.) You will achieve longer operating time on the Battery if you
operate your eQuinox in the Pulse Mode (Physician approved mode).
eQuinox (Model 4000) User Manual — 13
Typical Battery Recharge Time
It will normally take between 2 and 5 hours to recharge a fully
drained Battery back to 80% capacity, depending on the ow
setting being used. When the battery is fully charged, the battery
gauge will stop cascading. The battery can be charged in the unit
when the eQuinox is turned off and plugged in.
If the Battery becomes too warm during discharging, recharging
will not begin until the Battery sufciently cools. Removing the
Battery may speed up the cooling process.
The Battery can recharge from both AC and DC power sources.
The Battery will also charge from the optional Desktop Charger.
Enable/Disable the Voice Module
1. To enable/disable the voice module function, power on the
unit by pressing the ON/Standby Key.
2. After the unit has powered on, press the No Smoking Icon
and Hidden Key once.
3. Within 1–2 seconds, the SPKR = ON or SPKR = OFF menu
item will be present.
4. Press the Increase or Decrease Flow Rate Key to toggle the
selection to ON (enable voice module) or OFF (disable voice
module).
5. Within 3–5 seconds, the option selected will become active.
Note: This will only disable the voice module function once. When the
device is turned off and device is reset, the voice module function will
automatically be enabled.
Increase/Decrease the Volume of Vocal Notifications
1. To increase or decrease the volume of vocal notifications,
press the No Smoking Icon & Hidden Key twice.
2. The Volume = # - (Volume Setting Menu) will be present.
3. Press the Increase or Decrease Flow Rate Key to increase or
decrease the volume level from 1 to 10.
4. After selecting the desired loudness, the option will become
active/set within 3–5 seconds.
Changing Voice Module Language
1. To change the voice module language, press the No
Smoking Icon & Hidden Key five times to access the language
options.
2. The current activated language will be indicated on the
display. For example, if the voice module is speaking English
when accessing the language options menu, “English” will be
the text that appears on the display.
3. Press the Increase or Decrease Flow Rate Key to scroll
through the languages available.
4. Once the desired language is displayed, it will automatically
be set within 3–5 seconds.
Voice Module Operation
Voice Module Operational Instructions
The eQuinox can provide verbal notications or alarms and
operating status. These verbal notications supplement the
information on the light emitting diode (LED) display. The
following information provides instructions on how to turn the
voice module function on and off, adjust loudness, and change
language.
ON/Standby
Key
ON/Standby
Indicator
(Green LED)
Alert Indicator
(Yellow LED)
Alarm Indicator
(Red LED)
External
Power
Present
Indicator
(Green LED)
Decrease Flow
Rate Key
LCD
Display
Flow Mode
Indicator
(Green LED)
Flow Mode
Key
No
Smoking
Icon &
Hidden Key
Increase Flow
Rate Key
14 — 4805-C1 C (English)
Step 1: Positioning Your eQuinox for Use
Place the eQuinox in a well-ventilated, well lit area. Be sure the
air inlet and exhaust vents in the back are not blocked.
Position the eQuinox so that all
audible and visual indicators or
alarms can be easily seen and
heard.
Be sure the air inlet lter is in
place before operating your
eQuinox. If the air inlet lter is
missing, contact your home care
provider. If the air inlet lter is
dirty, wash with warm soapy
water. Then rinse it with clear
water and allow it to dry before
placing the air inlet lter back in
the device. If necessary, replace it
with a new, clean air inlet lter.
Plug the device into an AC Power
outlet, a DC Power source, or be sure there is a fully charged
Battery installed.
Step 2: Power ON the Device and Allow it to
Warm-Up
Press and hold the “ON/OFF” Button for two (2) seconds to
power ON your eQuinox. A brief visual power-on self-test will
occur.
Note: If there are three beeps when the device is first connected to power
or powered on, this indicates a low battery condition for the 9-volt battery
inside the cabinet of the eQuinox. Please contact your home care provider
and request that the 9-volt battery be replaced. If three beeps continue
intermittently after the unit has been power on and running for several
minutes, this is not a low 9V battery condition. Contact your healthcare
provider for service.
The eQuinox has an oxygen concentration status indicator
(OCSI) built in to the device. The OCSI continually monitors
the oxygen output of the device. It uses a green light (normal),
yellow light (alert), red light (alarm) and an audible alarm.
These indicators will all illuminate during the power-on
self-test at start-up. After rst powering ON the device,
please allow up to ve (5) minutes for the device to reach its
performance specications. When only the green indicator
light is illuminated, the device has reached its performance
specications.
Step 3: Connect the Oxygen Supply Tubing or Nasal
Cannula to the Oxygen Outlet
Replace the cannula and oxygen supply tubing regularly, as
recommended by your home care provider. Ask your home care
provider for cannula and supply tubing cleaning and replacement
information.
Note: Ensure the cannula is fully inserted and secure. This ensures that the
Oxygen Concentrator can properly detect inspiration for oxygen delivery.
During inhalation, you should hear or feel oxygen flow to the prongs of
the nasal cannula.
Step 4: Select the Flow Mode
Press the Flow Mode button to choose the mode. Pressing this
button more than once will switch you back and forth between
Pulse Dose Mode and Continuous Flow Mode. The correct mode
and setting should be prescribed for you by your physician.
Continuous Flow Mode Operation:
When running in the Continuous Flow Mode, a constant supply
of oxygen will ow through your tubing and nasal cannula. The
oxygen is measured in measured in liters per minute (LPM). You
may use a humidier when in Continuous Flow mode only. DO
NOT tilt, transport, or lay the eQuinox down while a humidier
bottle is attached.
Pulse Dose Mode Operation:
When running in Pulse Mode, a bolus of oxygen, measured in
milliliters (mL), is delivered upon inhalation. DO NOT use a
humidier in the Pulse Dose Mode.
When operating in pulse dose mode, the eQuinox supplies a
continuous ow of oxygen every 15 seconds if a breath is not
detected. The device will stop continuous ow and repeatedly
check for an inspiratory effort every 15 seconds. Once an
inspiration is detected, the device will deliver a measured bolus
again. To leave Pulse Dose Mode and start Continuous Flow
Mode, push the ow mode button.
Have your physician prescribe you the correct pulse dose to make
sure you receive the proper oxygen saturation. Take into account
your needs at rest, when you exercise, and when you travel at
high elevation. Contact your care provider for prescription and
boarding forms.
First Time eQuinox Operation
Air inlet Filter
Exhaust Vent
eQuinox (Model 4000) User Manual — 15
Pulse Dose Mode and autoSat Features
Your eQuinox has a feature called autoSAT
®
. It delivers a
consistent-sized bolus of oxygen up to 40 breaths per minute
(see Max Breath Rate table below). As your breath rate
increases, such as when you walk, the autoSAT feature servo-
controls the device, delivering the set bolus size. autoSAT
Technology is proprietary to CAIRE and is used during Pulse
Mode only.
Max Breath Rate
Bolus Volume (+/- 15%)
Pulse Dose Setting
Bolus Size
(± 15%) mL
Max Breath Rate
1 16 40
2 32 40
3 48 40
4 64 40
5 80 37
6 96 31
7 128 23
8 160 18
9 192 15
Step 5: Adjust the Flow Setting to the Prescribed
Level
Using the Flow Setting Buttons, adjust the ow setting to the
prescribed setting supplied by your physician.
WARNING: IT IS VERY IMPORTANT TO SELECT ONLY THE
PRESCRIBED LEVEL OF OXYGEN. DO NOT CHANGE THE
FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO
DO SO BY A LICENSED CLINICIAN.
Note: Bolus volume decreases as breath rate exceeds published maxi-
mum breath rate
Note: The pulse dose setting 1–9 is not equal to LPM.
WARNING: PULSE DOSE MODE SETTINGS MUST BE DE-
TERMINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR
NEEDS AT REST, DURING EXERCISE AND WHEN TRAVELING.
WARNING: THE EQUINOX MAY NOT BE ABLE TO DETECT ALL IN-
SPIRATORY EFFORTS IN PULSE MODE. IF THE EQUINOX DOES NOT
RESPOND TO YOUR INSPIRATION EFFORTS, ASK YOUR HOME CARE
PROVIDER TO CHECK THE SENSITIVITY SETTING.
Note: To Equipment Provider: The following oxygen administration acces-
sories are recommended for use with the Oxygen Concentrator:
• Nasal Cannula with 7 feet (2.1 m) of tubing: PN 5408-SEQ
• For any additional recommended accessories, please see the Accesso-
ries Catalog (PN ML-LOX0010) available on www.CAIREmedical.com.
• CAIRE offers the OxySafe™ as an optional accessory. This is intended
to be used in conjunction with the eQuinox concentrator. For customers
in regions requiring compliance to EN ISO 8359:1996-Ammendment
1:2012, this accessory will meet this need.
The OxySafe™ is a thermal fuse to stop the ow of gas in the event
that the downstream cannula or oxygen tubing is ignited and burns
to the OxySafe. It is placed in-line with the nasal cannula or oxygen
tubing between the patient and the oxygen outlet of the Equniox.
For proper use of the OxySafe, always refer to the manufacturer’s
instructions (included with each OxySafe kit).
CAIRE offers an OxySafe™ kits that includes a cannula headpiece,
OxySafe™, and 7’ tubing: PN 20628667
Step 6: Begin Using Your eQuinox
Breathe normally through the nasal cannula.
Note: You will achieve longer operating time on the Battery if you oper-
ate your eQuinox in the Pulse Mode (Physician approved mode).
Step 7: eQuinox Cart Operation
The eQuinox Cart is designed for use on walking-type surfaces.
Line up the eQuinox with the oor of the cart so it sits upright
and level.
Line up the large threaded screw to the insert at the back of
the eQuinox. Then hand-tighten the knob on the cart until the
eQuinox Cart is tightly secured to the device.
Push down the push-button on the eQuinox Cart handle to adjust
the height of the handle.
Note: DO NOT operate the eQuinox and ambulate while a humidifier is
attached. Remove the humidifier bottle before walking. DO NOT lay the
eQuinox down while attached to a humidifier bottle.
Step 8: Power Off the Device
Press and hold the “ON/OFF” Button for two (2) seconds to
power OFF your eQuinox.
Store the eQuinox in a cool and dry location for a period no
longer than 1 month. The eQuinox must run for at least 2 hours
per month to ensure proper function of the unit.
16 — 4805-C1 C (English)
User Troubleshooting Table
Do Not Ignore Alarms
WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUT-
AGE, ALARM CONDITION, OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE
SYSTEM REQUIRED.
Yellow Light
Symptom Possible Cause Your Action
The yellow light is on solid. The
alarm is giving 1 beep every 30
seconds.
The eQuinox has detected a
problem with the connec-
tion to the battery pack.
1) Remove battery pack and re-install to ensure that it is secure in the concentrator.
2) If possible, install a separate battery pack in the eQuinox.
3) If this problem persists, contact healthcare provider.
The yellow light is flashing. The
battery icon is flashing. The
alarm is giving 1 beeps every 2
minutes.
The eQuinox battery pack
is warm.
1) Re-attach the eQuinox to external power or install a fully-charged battery pack in the
eQuinox. Allow warm battery pack to cool outside the concentrator for 30 minutes.
2) Re-charge the battery pack using AC power
3) Replace the battery pack if another is available.
4) If the condition persists, contact your healthcare provider.
The yellow light is flashing. The
battery icon is flashing. The
alarm is giving 2 beeps every
30 seconds.
The eQuinox battery pack
voltage is low.
1) Plug eQuinox into AC power supply and recharge or replace battery pack with fully
charged battery pack.
The yellow light is flashing. The
alarm is giving 1 beep every 2
minutes.
The eQuinox has detected a
flow restriction causing low
or blocked flow.
1) Clean and replace cabinet inlet filter
2) Ensure that the cannula is not kinked or blocked. If used with a humidifier bottle,
ensure that it is filled properly and not creating a blockage.
3) Ensure that the eQuinox has proper ventilation. It needs to be at least 3 inches from
any surface to ensure the vents aren’t blocked.
4) If the problem persists, switch to an alternate source of oxygen and contact health-
care provider for assistance.
The yellow light is flashing.
There is no audible alarm.
The eQuinox has detected
low oxygen levels.
1) Ensure the air intake filter is not clogged or restricted. Clean and replace the filter if
necessary.
2) Ensure the eQuinox is in a well ventilated area. Make sure there are at least 3 inches
between the back of the eQuinox and any obstructions (furniture, curtain, etc.)
3) If operating in the car, be sure the back of the eQuinox is facing outward in the seat.
4) If the condition persists, switch to an alternate source of oxygen and contact your
healthcare provider immediately.
eQuinox (Model 4000) User Manual — 17
Red Light
Symptom Possible Cause Your Action
The red light is flashing.
The alarm is giving 3 beeps
every 2 minutes.
The eQuinox has detected
low oxygen levels
1) Ensure the air intake filter is not clogged or restricted. Clean and replace the filter if
necessary.
2) Ensure the eQuinox is in a well ventilated area. Make sure there are at least 3 inches
between the back of the eQuinox and any obstructions (furniture, curtain, etc.)
3) If operating in the car, be sure the back of the eQuinox is facing outward in the seat.
4) If the condition persists, switch to an alternate source of oxygen and contact your
healthcare provider immediately.
The red light is flashing. The
alarm is giving a constant
beep. The LCD displays
********. The eQuinox is
not delivery oxygen and will
not power on.
The eQuinox has lost power
temporarily while running on
external power.
1) Inspect the external (AC or DC) power supply and ensure that its connections are
secure at the eQuinox, the transformer, and the external power outlet.
2) If possible, attempt to use another electrical (AC or DC) outlet to power the eQuinox.
3) If the condition persists, switch to an alternate source of oxygen and contact your
healthcare provider immediately.
The eQuinox has lost power
due to a depleted battery
pack (battery pack) charge or
an overheated battery pack.
1) Re-attach the eQuinox to external power or install a fully-charged battery pack in the
eQuinox. Allow warm battery pack to cool outside the concentrator for 30 minutes.
2) Re-charge the battery pack using AC power
3) Replace the battery pack if another is available.
4) If the condition persists, contact your healthcare provider.
The red light is solid. The
display screen says “FAIL”.
The eQuinox is not produc-
ing delivering oxygen and
will not power on.
The eQuinox has experienced
a system malfunction.
1) Remove the battery pack and disconnect the AC or DC power supply from the eQuinox.
The red light should disappear and the FAIL message will disappear from the screen when
the power sources are removed. Wait approximately 20 seconds, reconnect power sourc-
es, and attempt to power on the eQuinox again.
2) If the eQuinox does power on, monitor it to determine if the FAIL message occurs
again. If the FAIL message occurs again, contact your healthcare provider immediately.
3) If the eQuinox does not power back on, contact your healthcare provided immediately.
18 — 4805-C1 C (English)
Other Alarm Conditions
Symptom Possible Cause Your Action
While in pulse mode,
the pulse mode LED
flashes quickly for 15
seconds.
The eQuinox is
unable to detect
the user’s breathing
effort. See page
15 for pulse mode
information.
1) Pulse mode may not work correctly if the user has any condition that causes blocked/restricted nasal
passages. (Illness, deviated septum, etc.)
2) Ensure cannula is no longer than 7 feet in length.
3) Ensure that there are no kinks or blockage in cannula tubing.
4) If a humidifier bottle is connected, remove the humidifier bottle and connect cannula directly to outlet
port. Pulse mode will not work with a humidifier bottle installed.
5) If the condition persists, contact your healthcare provider.
Battery pack is drain-
ing while plugged in to
AC/DC power supply
The eQuinox is not
receiving power
from the attached
power supply.
1) Verify that the outlet is providing power. Do not connect the power supply to a dimmer circuit or a
power strip.
2) Check that cable connections on power supplies are secure to the wall/vehicle and concentrator. If
using the AC power supply, check the cable connection on the power supply transformer as well.
3) Ensure that the power verification lights are on.
- There will be a green light on the power supply box.
- There will be a green light on the eQuinox concentrator control panel in the shape of a power cord.
4) If power verification LEDs are not lit, remove all connections of the power supply for 20 seconds and
reconnect.
5) If the condition persists, contact your healthcare provider.
The Battery pack
(battery) charge is not
lasting as long as it
should.
The battery pack
was not fully
charged
1) Connect to AC power to recharge the battery. Verify that the battery charges for 2-5 hours and the
battery icon is full and not flashing before use.
2) Refer to the actions for “Battery Pack is Not Charging” Symptom below.
The eQuinox was
not operating at the
anticipated flow
rate.
1) Ensure that you are using the flow rate prescribed by your physician and that pulse or continuous flow
is being used as required.
The battery pack is
not preforming to
specifications.
1) Attempt to drain the battery pack completely. Do this by running the eQuinox on battery power until
the machine shuts off completely. Then remove the battery and allow it to cool for approximately 30
minutes. After this time, re-insert the battery in the eQuinox and recharge it using AC power.
2) If the battery still does not last as long as it should after following step 1, contact your healthcare
provider.
The Battery Pack (bat-
tery) is not charging.
The eQuinox is not
receiving external
AC power to charge
the battery.
1) Verify that the outlet is providing power. Do not connect the power supply to a dimmer circuit or a
power strip.
2) Check that cable connections on power supplies are secure to the wall/vehicle and concentrator. If
using the AC power supply, check the cable connection on the power supply transformer as well.
3) Ensure that the power verification lights are on.
- There will be a green light on the power supply box.
- There will be a green light on the eQuinox concentrator control panel in the shape of a power cord.
4) If power verification LEDs are not lit, remove all connections of the power supply for 20 seconds and
reconnect.
5) If the condition persists, contact your healthcare provider.
The eQuinox is not
properly communi-
cating the battery
pack to charge it.
1) Remove battery pack and re-install to ensure that it is secure in the concentrator.
2) Inspect the eQuinox for a solid yellow light when the battery is installed or a flashing battery icon
when attempting to charge. If either of these symptoms is present, continue to step 3
3) If possible, install a separate battery pack in the eQuinox. If the replacement battery works properly,
then the original battery needs to be replaced.
4) If this problem persists contact healthcare provider.
The eQuinox beeps 3
times when it is first
powered on.
The internal battery
that powers the
alarms is low
1) Contact your healthcare provider.
eQuinox (Model 4000) User Manual — 19
Air Inlet Filter
Ambient air is drawn into the device through the air inlet located
on the top rear of the device. Cleaning the air inlet lter is the
most important maintenance activity you will perform as a user.
Cleaning the air inlet lter will keep your eQuinox performing
properly. Check this air inlet lter daily and clean it at least once
a week using the following procedure:
1. Remove the air inlet filter
2. Wash the filter in warm tap water using a mild soap
detergent solution
3. Rinse the filter thoroughly with clear tap water and squeeze
out the excess water
4. Allow the filter to air dry
5. Reinsert the filter in the cabinet
Cleaning tip: Keep a clean second lter as a replacement for use
when one lter is drying.
Routine Maintenance
Only a service technician trained and qualied by CAIRE should
work on the internal parts of the eQuinox.
Preventative Maintenance (PM) should be done every year.
Talk to your provider to arrange a PM schedule. Your home
care provider or qualied technician will inspect the following
parts only when needed: alarms, internal 9-volt battery, battery,
internal lters and other internal parts.
User Care and Cleaning of the Device
Cannula Replacement
Replace your supply tubing and cannula as often as your
home care provider recommends. Your physician or home
care provider will provide you with cleaning, disinfection and
replacement information.
The air inlet lter should be replaced annually. If the eQuinox is
used in a dusty environment, the lter may need to be replaced
more frequently. You should not operate the eQuinox without an
air inlet lter installed for more than 5 minutes.
Cleaning, Care & Routine Maintenance
CAUTION: Do not operate the Oxygen Concentrator
without the air intake filter in place. If a second filter is
provided, insert the “replacement” filter before you clean
the dirty filter. Clean the dirty filter in a warm soap and water
solution then dry thoroughly prior to use.
Note: The Manufacturer does not recommend the sterilization of this
equipment.
Note: Do not operate the Oxygen Concentrator without the air intake
lter in place. If a second lter is provided, insert the “replacement” lter
before you clean the dirty filter. Clean the dirty filter in a warm soap and
water solution then dry thoroughly prior to use.
Note: Always follow the cannula manufacturer’s instructions for proper
use. Replace the disposable cannula as recommended by the cannula
manufacturer or your Equipment Provider. Additional supplies are avail-
able from your Equipment Provider.
Note: If three beeps sound after the power ON self-test, this indicates the
9v battery is low in power. Contact your home care provider if you hear
three beeps.
Air inlet
Filter
20 — 4805-C1 C (English)
Optional Humidifier
If used, clean your humidier every day to decrease the
possibility of contamination. Follow the cleaning directions from
the home care provider or the manufacturer of the humidier.
Replace the humidier every month, or as directed by the
manufacturer or your home care provider. (Do not overll the
humidier.)
WARNING: DO NOT USE LIQUID DIRECTLY ON THE UNIT.
A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES
BUT IS NOT LIMITED TO THE FOLLOWING: ALCOHOL AND
ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLORINE-BASED
PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED PRODUCTS
(PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO CLEAN
THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS THEY CAN
DAMAGE THE UNIT’S PLASTIC.
CAUTION: DO NOT operate the eQuinox and ambulate
while a humidifier is attached. Remove the humidifier bot-
tle before walking. DO NOT lay the eQuinox down while
attached to a humidifier bottle.
eQuinox Cart
Before cleaning the eQuinox Cart, remove it from the eQuinox.
Use a damp (not wet) cloth or sponge and mild cleaning solution
to clean the cart.
Battery Pack (Battery)
The Battery in the eQuinox needs special care to make sure
it has a longer life and the highest level of performance. The
CAIRE Battery is the only Battery recommended for use with
the eQuinox.
Use a damp (not wet) cloth or sponge to clean the Battery. First
spray the cloth or sponge with a mild detergent, then clean the
Battery case and the latch.
CAIRE recommends replacing the battery pack after 2 years.
Battery Disposal
Your Battery is rechargeable and can be recycled. Always return
it to your home care provider for proper disposal. You can also
contact your local city or town ofces for directions on proper
disposal of the Battery.
eQuinox Oxygen System Disposal
Local environmental laws may not allow disposal of electrical
and/or electronic equipment such as the eQuinox or AC Power
Supply. Contact the local city or town ofces for instructions on
the correct way to dispose of electrical or electronic equipment.
You may also contact CAIRE Inc. for disposal information.
AC Power Supply, 12VDC Cable, Power Cord and Outer Case:
Turn OFF the eQuinox and disconnect from AC or DC power
before you clean or disinfect. DO NOT spray right onto the outer
case. Use a damp (not wet) cloth or sponge. Spray the cloth or
sponge with a mild cleaning solution to clean the cabinet and
power supply. To disinfect the eQuinox, use Lysol® Brand II
disinfectant (or equivalent). Follow the directions of the cleaner
manufacturer.
WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE
UNIT AND DISCONNECT THE POWER CORD FROM THE
ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO PRE-
VENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY
YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN
SHOULD REMOVE THE COVERS OR SERVICE THE UNIT. DO NOT USE
LIQUID DIRECTLY ON THE UNIT. A LIST OF UNDESIRABLE CHEMICAL
AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING: AL-
COHOL AND ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLO-
RINE-BASED PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED
PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED
TO CLEAN THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS
THEY CAN DAMAGE THE UNIT’S PLASTIC.
WARNING: ELECTRICAL SHOCK HAZARD. DISCONNECT THE POWER
CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT
TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD.
ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECH-
NICIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.
WARNING: CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CON-
CENTRATOR AND THE BATTERY FROM GETTING WET OR ALLOWING
FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE A MALFUNCTION OR
SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL
SHOCK OR BURNS.
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER
CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A
DAMP (NOT WET) CLOTH OR SPONGE, AND THEN WIPE ALL SURFAC-
ES DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE. PAY
SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA
CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST, WATER,
AND PARTICLES.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184

Equinox Systems 4000 Manuale utente

Tipo
Manuale utente