GCE PRESSURE WATCH Istruzioni per l'uso

Tipo
Istruzioni per l'uso
INSTRUCTION FOR USE
VOD K POUŽITÍ
ISTRUZIONI PER L'USO
ANVÄNDARANVISNING
GCE HEALTHCARE
CENTRAL GAS SUPPLY SYSTEM EN
PRESSURE WATCH
CS
IT
SYSTÉM CENTRÁLNÍHO ROZVODU PLYNU
SISTEMA CENTRALE DI FORNITURA GAS
SV
CENTRALGASANLÄGGNING
3/80
EN
1. FOREWORD
These instructions for use cover essential information for the entire life cycle of the Pressure
Watch:
Installation
Operation
Alarm conditions, see Alarm Systems
Maintenance
Cleaning
Disposal.
GCE Medical central gas system is medical devices classifi ed as class IIb according to the
Medical Device Directive 93/42/EEC.
Their Compliance with essential requirements of 93/42/EEC Medical Device Directive is based
upon EN ISO 7396-1 standard.
2. INTENDED USE
The Pressure Watch is monitoring and supervising the gas pressure in a clearly stated part of
the Hospital Central Gas Supply System. It is fi tted with quick couplings for emergency supply,
gauges and pressure sensors. The pressure sensors gives a signal for high or low gas pressure
and can be connected to an Alarm system. Together with the Pressure Watch shall always be
used alarm providing all alarms according to standard (like Gas alarm).
This instruction for use details the operational and safety procedures for the Pressure Watch.
The Pressure Watch is available for Oxygen, Nitrous Oxide* and Compressed Air, it can also be
used for Carbon Dioxide*, Nitrogen, Vacuum and gas blends for Medical Applications. These
Pressure Watches are designed to operate at a maximum service pressure of 16 bar.
The system should not be put into service before these procedures are thoroughly read and the
operator is familiar with controls described in this instruction for use.
* The Pressure Watch shall be installed in room with minimal temperature of 10 °C.
Read this instruction before use of the product. Always follow this instruction!
The product shall only be used for the purpose described in this instruction!
The product must be installed by a qualifi ed person and by compliance of all requirements of
EN ISO 7396-1 as amended!
Before use, to guarantee the safety of the patient, check the equipment and the accessories
used with the product, so that data and performance comply with the intended use of the
product!
The product must not, under any circumstances be modifi ed by other than the manufacturer!
The product must not be used for di erent gas than stated on the label.
ENGLISH
INSTRUCTION FOR USE: PRESSURE WATCH
4/80
EN
3. OPERATIONAL, TRANSPORT AND STORAGE SAFETY
REQUIREMENTS
KEEP THE PRODUCT AND ITS ASSOCIATED EQUIPMENT AWAY FROM:
Heat sources (fi re, cigarettes,...)
Flammable materials
Oil or grease (take a great care in the use of hand creams)
Water
Dust
Always maintain oxygen cleanliness standards.
The product and its associated equipment must be prevented from tipping over, turning over
or falling.
Only use the product and its associated equipment in well ventilated area.
National laws, rules and regulations for medical gases, accident prevention and environmental
protection must be observed.
OPERATING CONDITIONS STORAGE AND TRANSPORT
CONDITIONS
MIN MAX MIN MAX
10 °C +40 °C -30 °C +70 °C
20 % 70 % 20 % 70 %
600 mbar 1200 mbar 600 mbar 1200 mbar
It’s very important to assure access to this device only for qualifi ed person – possibility of
misuse.
3.1.OPERATIONAL SAFETY
The panel described in this manual can be emergency supplied by cylinder gases and the
following general practices are our recommendations for the safe, effi cient storage and use of
such compressed gases.
3.2.CYLINDER STORAGE
Storage area shall be designed according to valid standard, here are some recommendations;
Well ventilated with weather protection.
Free from fi re risk. Keep away from sources of heat and ignition. Designated as a ‚no smoking’
area.
Clearly marked as a gas store with appropriate hazard warning signs (e.g. fl ammable, toxic,
radio active etc.).
Kept clear with access restricted to authorised personnel.
Provided with appropriate safety/emergency equipment (e.g. Fire extinguisher, breathing
apparatus etc.)
5/80
EN
3.3.COMPRESSED GAS CYLINDERS IN STORAGE SHOULD BE:
Standing upright where designed for this.
Properly secured to prevent overturning.
Fitted with valve protection devices (e.g. Caps, guards etc.) where supplied.
Segregated in the storage area according to the various categories (e.g. Flammable, oxidant
etc.).
Segregated in the storage area according to content and clearly designated full or empty.
Managed to ensure that the oldest stock is used fi rst.
Checked periodically for general condition.
3.4.PERSONNEL WHO ENTER THE STORAGE AREA SHOULD BE:
Responsible and competent to maintain the gas store and its contents as above.
Familiar with and able to identify the contents of the gas containers and their potential hazards.
3.5.EMERGENCY PROCEDURES:
An emergency results from;
A fi re near or in the cylinder manifold room.
An unintended leak of gas from the cylinders or pipeline.
Some other potential or actual incident which will aff ect the integrity of the manifold installation.
PROCEDURE TO FOLLOW:
Raise the alarm and notify a supervisor and/or call the fi re brigade.
Evacuate all personnel from the immediate danger area.
IN CASE OF FIRE, REMOVE ALL CYLINDERS:
Close the cylinder valve(s) on all cylinder(s)/ cylinder bundles.
Close the bank isolating valve(s).
Disconnect the high pressure hoses.
Move the cylinders to a safe area using a suitable cylinder trolley.
IN THE CASE OF AN UNCONTROLLABLE ESCAPE OF GAS:
Open the doors to the manifold room to ventilate the area.
Do not introduce sources of ignition into the area.
It is important that ready access to these instructions is given at all times and that the product
is not put into operation until the operator is fully familiar with its functions, controls and
safety precautions.
It is the duty of all employers to provide such information, training and supervision according
to national standards as is necessary to ensure so far as is reasonably practical the health and
safety at work of its employees.
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider must ensure that all
personnel handling the product are provided with the operating instructions & performance
data.
Do not use the product without properly familiarization of the product and its safe operation
as defi ned in this Instruction for use. Ensure user is aware of particular information and
knowledge required for the gas in use.
6/80
EN
5. PRODUCT DESCRIPTION
The Pressure Watch is monitoring and supervis ing the gas pressure in a clearly stated part
of the Hospital Central Gas Supply System. It is fi tted with shut off valves, quick couplings for
emer gency supply, gauges and pressure sensors. The pressure sensors gives a signal for high
or low gas pressure and can be connected to an Alarm system. Together with the Pressure
Watch shall always be used alarm providing all alarms accord ing to standard (like Gas alarm).
The Pressure Watch can also be delivered with Emergency Source (ES) and can be installed
recessed or exposed.
5.1.ORDER CODE
Order code contents name and description of the product, see below.
NAME OF THE PRODUCT
PW ; 6G ; NS ; SP ; SW GW ; W L D S E
1: TYPE OF AVSU
PW = Pressure watch PS = Pressure watch system (Pressure watch is produced with alarm)
2: NUMBER OF GASES
1G = 1 3G = 3 5G = 5
2G = 2 4G = 4 6G = 6
3: GAS STANDARD
SS = SS NS = NIST UN = UNI
SD = SSD NF = AFNOR CZ = CZ
DN = DIN BS = BSI
4: ALARM
―― = NO 03 = Touch TMA TFT03 BB - Prepared for alarm
(big box, no alarm)
97 = Standard TMA 9705 SP - SSGM Premium
5: SENSORS
―― = NO SW = Pressure Switches NO/NC TR = Pressure Transmitter 4-20mA
6: FRONT DOOR MATERIAL
GW = Glass (white) GI - Glass custom image NO = (No doors required)
GC = Glass custom colour ST = Steel
7: VISIBLE VALVE HANDLES
W = YES ― = NO
8: LOCKING MECHANISM
― = No lock S = Sealing (plombation) L = Push in lock
9: MICRO SWITCH ON DOOR
― = NO D = YES
123456 7 8 9 10 11
7/80
EN
10: OUTLET PIPE
S = Standard 300 mm A = No pipe (Soldering adapter)
11: PACKAGE
E = Everything mounted D = Door packed separatly
PRODUCT DESCRIPTION
1st branch 2st branch 3st branch 4st branch 5st branch 6st branch
OET15ESN ; NET15SSN ; AET15SSN ; SND15SSN ; CND15SSN ; VND15SSN ; EN ; G
1: GAS TYPE
O = Oxygen C = Carbon dioxide V = Vacuum
N = Nitrous Oxide I = Nitrogen M = Oxygen/Nitrous Oxide
A = Medical Air R = Argon
S = Surgical Air G = AGSS
2: ES PIPING
N = NO E = YES
3: GAUGE
N = No gauge O = Outlet pressure T = Double inlet gauges
(emergency piping)
TRIPPLE
I = Inlet pressure D = Inlet + Outlet pressure
DOUBLE
4: PIPE LINE SIZE
15 = 15 mm 22 = 22 mm 28 = 28 mm
5: VALVE POSTION SWITCH
N = NO D = Two fully opened
+ fully closed in
I = ES functionality
(IF ES is selected)
S = YES fully open E = ES functionality
(IF ES is selected)
6: PIPE LINE SIZE
N = NO S = Single QC D = Dual QC
7: PIPE LINE SIZE
N = NO F = YES
8: LANGUAGE IFU, ALARM, LABELS
EN = English CS = Czech RO = Romanian
SV = Swedish SK = Slovak IT = Italian
NO = Norwegian DE = German FR = French
1 23 4 56 7 8 9
8/80
EN
DA = Danish NL = Netherlands ES = Spanish
FI = Finnish HU = Hungarian PT = Portuguese
9: CUSTOMER LOGO/BRAND
G = GCE L = Linde
5.2.THE PRESSURE WATCH ON THE PICTURE IS A 2GAS VERSION
A – Inlet tube connection
Tube for gas supply inlet.
C – Shut o valve
In order to be able to open or close the gas supply, the blue handle is turned a quarter of a
revolution. Each valve shall be marked with gas type so that the right gas is shut off at leakage
or service. The marking on the plate shows if the valve is open or closed, i.e. if the handle is in
line with the marking the valve is open and the gas fl ows freely.
D – Outlet pressure gauge
On the gauge is possible to see the outlet pressure.
F – Pressure sensor
The switch gives a signal for high gas pressure of the gas supply.
G – Quick coupler ( gas specifi c)
Via the quick coupler can be connected the reserve gas cylinder when the main gas supply
is disconnect.
H – Outlet tube connection
Tube for gas supply outlet.
Note: The pressure watch is also available with switch for position of valve.
HA
C
G
F
D
9/80
EN
5.3.THE PRESSURE WATCH WITH EMERGENCY SOURCE ON THE
PICTURE IS A 2GAS VERSION
The Pressure Watch with Emergency Source consists of all devices as Pressure Watch only
Emergency Source ES is fi tted to provide the gas delivery when the main gas supply is out of
order.
B – Inlet pressure gauge for fi rst source
On the gauge is possible to see the inlet pressure of the gas supply.
J – Inlet tube connection for reserve source
Tube for emergency gas supply inlet.
K – Inlet pressure gauge for reserve source
On the gauge is possible to see the inlet pressure of the emergency gas supply.
L – Shut o valve for reserve source
In order to be able to open or close the emergency gas supply, the blue handle is turned a
quarter of a revolution. Each valve shall be marked with gas type so that the right gas is shut off
at leakage or service. The marking on the plate shows if the valve is open or closed, i.e. if the
handle is in horizontal position like on the picture the valve is closed and the gas does not fl ow.
D - Outlet pressure gauge
On the gauge is possible to see the outlet pressure of the gas supply.
5.3.1.PHYSICAL ISOLATION
Each Pressure Watch or Pressure Watch ES (except for vacuum system) is equipped with a
physical isolation. These device shall be fi tted to each pipe line when modifi cations are needed
to existing system. It shall be fi tted manually to pipe line by trained person.
Place for physical isolation is marked with this label –
J
K
B
L
D
10/80
EN
6. INSTALLATION
It is necessary that an installer of the product is trained and has an appropriate licence
according to national standards. It is outside the scope of these operating instructions to give
detailed guidance on installations.
The installation must conform to the standard MEDICAL GAS PIPELINE SYSTEMS EN ISO
7396-1 as amended. All tests according this standard have to be fulfi lled.
When working with medical gases it is essential that no oil or grease come into contact with
the gas. This means that hands, tools and work clothes be free from oil or grease before
any work is undertaken. Components used for service must be gas compatible. For oxygen
service must also be degreased and although they will reach you in this condition care must
be taken during storage and handling.
6.1.WALL FIXING
Please consult the positioning of the Pressure Watch with gas supplier and designer. Mount the
Pressure Watch at the height to suit either cylinders or cylinder bundles.
Be aware that only upper screw hold the entire weight of the product. Size of used screws
shall conform to total load.
6.1.1.PRESSURE WATCH COMPONENTS
1. Box
2. Sliding door
3. Alarm cover
4. Stopper of sliding door
The middle of the Pressure Watch is best placed 1.5 m over the fl oor. It must be possible to
read the gauges, and to have access to the valves. Attention should be paid to the passing bed
transports so these will not cause harm to the Pressure Watch.
Note: It is possible slide up the door and dismount it so with activate plastic lever inside both
slides at one time.
4
1
2
3
11/80
EN
There is need space above of Pressure Watch for full open slide door.
Put door on lock carefully otherwise then lock can break and injure. Do not use with damaged
lock. Visually check the correct function.
6.1.2.RECESSED MOUNTING:
Make a hole in the wall. Put cross bars or other means of attachment, which will be used for the
mounting of the Pressure Watch, at determined height. You can use hole on back or left and
right side of box. Numbers of holes depends on variant.
Unused hole plug with white sticker.
6.1.3.EXPOSED MOUNTING:
Make a hole in the wall. Put cross bars or other means of attachment, which will be used for
the mounting of the Pressure Watch, at determined height. You can use hole on back of box.
Numbers of holes depends on variant.
Unused hole plug with white sticker.
6.1.4.PIPE CONNECTION
The sequence of the gases is given by the gas piping in the medical facilities and the relevant
local standards and it is defi ned by pressure watch confi guration.
1. The valve need not to be demounted at soldering.
2. Cut the tubes in a perpendicular fashion to suitable length. Grade the tubes, and make sure
there are absolutely no traces of copper splint. Soldering should be done with nitrogen as
the protecting gas inside the tube line according to the EN ISO 7396-1
3. Connect the copper tubes and solder them together.
4. Chill the tubes at soldering.
6.1.5.FINAL MOUNTING
At recessed mounting, the wall is fi nished and the empty space between the box (2) and the
wall is tightened. Fasten the wire into the sealing hole (5).
6.2.FITTING INLET AND OUTLET CONNECTION
Solder pipes according 6.1.4.
6.3.FITTING TO SOURCE OF ELECTRICITY
Any work with electric wiring must be carried out by a trained person with an appropriate
licence according to national standards.
Act upon Instruction For Use for Gas alarm.
The product must be grounded through the anchor bolt at the protective earth label .
12/80
EN
7. OPERATION
7.1.START UP OF THE PRESSURE WATCH
When the Pressure Watch is completely installed.
Make sure that the incoming gas is right for their respective connections.
Open the input valves slowly.
Check the outlet gauges D, if the gas pressure is correct for each gas.
Check all connections for leakage using a 0.5% teepol / water solution.
7.2.START UP OF THE PRESSURE WATCH WITH EMERGENCY SOURCE
When the Pressure Watch with Emergency Source is completely installed.
Make sure that the incoming gas is right for their respective connections.
Ensure the valves B on the main gas line are closed
Before opening the valves L on the Emergency Source gas line check the gauges D on the
Emergency Source gas line, if the gas pressure is correct for each gas.
Open the valves L on the Emergency Source slowly.
Check the gauges D, if the gas pressure is correct for each gas.
Check all connections for leakage using a 0.5% teepol / water solution.
Close the valves L on the Emergency Source.
Before opening the valves C on the main gas line check the gauges B on the main gas line, if
the gas pressure is correct for each gas.
Open the valves C on the main gas line slowly.
Check the gauges D, if the gas pressure is correct for each gas.
Check all connections for leakage using a 0.5% teepol / water solution.
For operating the alarm system see Instruction For Use for the alarm.
HA
C
G
F
D
13/80
EN
7.3.CONTROL THE FUNCTION OF THE PRESSURE SENSORS
1. HIGH PRESSURE OXYGEN.
Close valve C for Oxygen, connect a Oxygen gas cylinder to the female QC for emergency
supply. Increase the pressure to present high pressure level, the alarm gives an signal of high
pressure.
2. LOW PRESSURE OXYGEN.
Remove the gas cylinder and use a male probe (quick connection) to decrease the pressure,
to the preset low pressure level and wait for the alarm.
3. OPEN THE VALVE AGAIN SLOWLY.
Check the alarm levels for all gases in the same way.
7.4.PHYSICAL ISOLATION
Each Pressure Watch or Pressure Watch ES (except for vacuum system) is equipped with a
physical isolation. These device shall be fi tted to each pipe line when modifi cations are needed
to existing system. It shall be fi tted manually to pipe line by trained person.
7.4.1.FITTING PROCEDURE OF PHYSICAL ISOLATION
1. Disassemble the front plate and the inner cover
2. Close the shut off valve on adequate pipe line and ensure the gas is no more fl owing through
3. Disassemble both switches and unscrew the cap nut from the valve
4. Put the physical isolation between the valve and the connection
J
K
B
L
14/80
EN
Physical isolation – Copper blanking plate
5. Screw back and tighten the cap nut on the valve
6. Slowly open the shut off valve and check if the gas is fl owing through
7. If yes, please check the assembly of the physical isolation again
8. If the gas is no more fl owing through the physical isolation is fi tted correctly
9. Close the shut off valve
When modifi cations are ready then the physical isolation has to be disassembled from the
pipe line!
7.4.2.DISASSEMBLY OF THE PHYSICAL ISOLATION
1. Ensure the shut off valve is closed
2. Unscrew the cap nut from the valve
3. Take out the physical isolation
4. Screw back and tighten the cap nut on the valve
5. Screw back and tighten the switches – low pressure switch to LP side and high
pressure on the HP side (see the LP and HP marking on the brass block)
6. Slowly open the shut off valve and check if the pressure gauge indicates any pressure
7. If yes, check the external leakage of the connections
7.5.ALTERNATIVE OF PRESSURE SENSORS
ALTERNATIVE 1. PRESSURE SWITCHES:
To set the pressure switches ”PS” (repeat for every PS):
1. Connect an Ohm-meter to the two pins of PS.
2. Rise and sink the pressure by using the quick coupling (and if necessary a spare cylinder).
3. Note at what pressure the PS change from closed to open (seen on the Ohm-meter,
closed=short circuit, open=no contact between the pins). Closed PS = No alarm. Open PS=
Alarm
4. If needed; adjust the PS by operating the screw positioned between the two pins. Turning
the screw clockwise sets the alarm point corresponding to a higher pressure.
5. Reiterate point 2 to 4 until you are pleased with the settings.
6. Connect the PS to Gas alarm according to enclosed Gas alarm instruction.
ALTERNATIVE 2. PRESSURE TRANSMITTER 420 mA:
4 mA is equivalent to 0 bar and 20 mA to the maximum pressure of the pressure transmitter.
The pressure transmitter is linear and you connect it directly to central computer system or local
alarm system. See also IFU for the master system.
15/80
EN
8. EMERGENCY GAS SUPPLY
8.1.EMERGENCY GAS SUPPLY
On signal of alarm condition, the display is indicating low or high pressure for one or more
gases. The alarm condition should be recorded and the necessary action taken.
1. Pressure Watch not equipped with central Emergency Supply, follow instructions below.
1. GAS TYPE
Mute the acoustic signal by pressing the button.
Display of the alarm gives visual indication which
gas (gases) has incorrect pressure.
2. CLOSE THE VALVE
Open the hatch and close the inlet shut off valve
for the gas which causes the alarm.
3. CONNECT THE RESERVE GAS CYLINDER
Connect the suitable hose from the cylinder regulator
to the quick connector in the Pressure Watch.
4. OPEN THE CYLINDER VALVE CAREFULLY
Put both hands on top of the cylinder valve and
open the emergency gas slowly.
5. CHECK THE ALARM
Check on the display that the alarm returns to
normal mode. If the alarm does not return to normal
mode, see point 1.
6. INFORM THE RESPONSIBLE PERSONNEL
Telephone number to responsible technician. The
responsible personnel checks the equipment, replaces
the gas bottles if necessary and resets the equipment
to normal mode.
ALARM
ALARM
16/80
EN
2. Pressure Watch equipped with central Emergency Supply, follow instructions below.
1. GAS TYPE
Mute the acoustic signal by pressing the button.
Display of the alarm gives visual indication which
gas (gases) has incorrect pressure.
Open the hatch and close the inlet shut off valve
for the gas which causes the alarm.
3. OPEN THE VALVE
2. CLOSE THE VALVE
Open the shut off valve on the Emergency Supply for
the gas which causes the alarm.
5. INFORM THE RESPONSIBLE PERSONNEL
Telephone number to responsible technician. The
responsible personnel checks the equipment,
replaces the gas bottles if necessary and resets the
equipment to normal mode.
NORMAL
SUPPLY
EMERGENCY
SUPPLY
ALARM
EMERGENCY
SUPPLY
NORMAL
SUPPLY
17/80
EN
9. ALARM CONDITIONS
GENERAL FUNCTION OF ALARM SYSTEMS
Gas alarm monitors the alarm conditions from the Pressure Watch. The display shows high or
low pressure for 1-6 gases.
See instruction for use for Gas alarm.
9.1.ALARM PROCEDURE
On signal of alarm condition, the appropriate visual indicator is appeared on the gas alarm.
The alarm condition should be recorded, and the necessary action taken, as detailed in the
sections below.
9.2.STATUS INDICATED BY ALARM
9.2.1.ABNORMAL POSITION OF VALVES
Abnormal positions of valves can be detected if there are position switches on valves. In this
case should be the necessary action taken for safety operation and continuous gas supply.
9.2.2.HIGH SUPPLY PRESSURE
Check that the outlet pressure is as preferred. If not, fi x the problem with gas source or use
diff erent source.
9.2.3.LOW SUPPLY PRESSURE
Check that the outlet pressure is as preferred. If not, fi x the problem with gas source or use
diff erent source.
9.2.4.SLIDE DOOR ARE OPEN
Open position of door can be detected if there are position switch on door. In this case should
be the necessary action taken for safety operation and continuous gas supply.
10. ROUTINE MAINTENANCE INSTRUCTION
Service shall be carried out only by authorised person according to national standards.
Use only original components. Contact GCE for further information about repair procedures.
THE FOLLOWING ROUTINE CHECKS SHOULD BE MADE ON A WEEKLY BASIS:
Check all valves for correct operation and leak test using a 0.5% teepol in water solution.
Check pipe work for leaks using a 0.5% teepol in water solution.
Check that all notices are in place, can be easily read and are not obstructed from view.
Inspect the vicinity of cylinders for anything introduced since last inspection, which could
aff ect the continuing safe operation of the manifold.
GENERAL MAINTENANCE SHOULD BE UNDERTAKEN IN ACCORDANCE WITH A PLANNED
MAINTENANCE SCHEDULE, WHICH SHOULD INCLUDE THE FOLLOWING ITEMS:
Leak test all joints using a 0.5% teepol in water solution.
Examine mounting frames and cylinder securing chains for damage deterioration and security
of attachment. programme.
Valves to be used in normal or emergency operation are accessible and easy to operate.
Valves at service point’s outlets and purge valves should be checked for gas tightness.
No item of equipment is overdue for any periodic inspection/test which may be applicable.
The setting and operation of regulators is satisfactory.
The necessary safety devices and instruments are fi tted, are of the correct type and do not
show signs of deterioration or unauthorised interference.
The equipment downstream of the outlet point is suitable for further service.
Check connecting nuts and seating faces are undamaged.
Check o-rings are undamaged.
Check gauges operate smoothly and zero correctly.
Check isolating valves for closure tightness and gland leakage.
The gas manifold room must always remain thoroughly clean. Materials other than those
required for operating the manifold must not be stored in the room.
18/80
EN
Check that all ventilation grills are free from obstruction and that the extraction system (when
installed) is functioning correctly.
Ensure that all restrictions on smoking and open fi re are strictly enforced.
Replace all worn or damaged parts with original spare parts. Remember that gaskets, o-rings
and sealing surfaces must be clean and free from damage to work eff ectively.
SAFETY NOTE  AS PART OF THE PLANNED MAINTENANCE SPECIAL CHECKS SHOULD BE
MADE TO ENSURE:
All changes (including removal and addition of parts) and extensions conform to the right code
of practice.
Changes in the vicinity of the installation do not aff ect the safety of the operation e.g. proximity
of electricity to fuel gases, oils and combustibles near to oxygen, illicit use of a pipeline as an
electrical conductor or as a support for other items.
11. SCHEME OF EXAMINATION
GCE recommendations for a Written Scheme of Examination:
EQUIPMENT EXAMINATION
INTERVAL
GUIDELINES OF WRITTEN SCHEME
1. HOSES 5 years a. Visual, external examination.
b. Replace with new hose.
c. Record replacement.
2. PIPE WORK 1 year a. Visual, external examination.
b. Pressure test.
c. Record all details of examination.
3. REGULATORS FOR
EMERGENCY SUPPLY 5 years a. Replace with new or manufacturer’s refurbished
unit.
b. Record all details.
4. VALVES 5 years a. Replace seals/glands/o-rings.
b. Functional test.
c. Record all details of examination.
12. ACCESSORIES
Standard accessories – Gas alarm.
GCE recommends the use of GCE Shut Off Valves, GCE Collecting Branches, GCE Pressure
hoses.
Contact supplier of the system or GCE dealer for spare part lists.
Use only genuine GCE spare parts!
13. CLEANING OF EXTERNAL PARTS
Remove dirt with a soft cloth damped in oil free oxygen compatible soap water and rinse with
clean water.
If other cleaning solutions are used, check that they are not abrasive and that they are
compatible with: Cu, Fe, Al alloy, brass, plastic materials of components, labels and gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not clean electrical connections when the power is on. Make sure the connections are dry
before turning power back on.
19/80
EN
14. PRODUCT LIFE TIME AND MAINTENANCE
14.1.PRODUCT LIFE TIME AND WASTE MANAGEMENT
Maximum life time of the product is 10 years from the manufacturing date at compliance of
Examination schedule – see chapter 11. At the end of the product’s life time (10 years maximum),
the product must be withdrawn from service. The owner of the device shall prevent the reuse of
the product and handle the product in compliance with “Directive of European Parliament and
Council 2008/98/EC on waste“.
In accordance to Article 33 of REACH GCE, s.r.o. as responsible manufacturer shall inform all
customers if materials containing 0.1% or more of substances included in the list of Substance
of Very High Concern (SVHC).
The most commonly used brass alloys used for bodies and other brass components contain
2-3% of lead (Pb), EC no. 231-100-4, CAS no. 7439-92-1. The lead will not be released to the gas
or surrounding environment during normal use. After end of life the product shall be scrapped
by an authorized metal recycler to ensure effi cient material handling with minimal impact to
environment and health.
To date we have no information that indicates that other materials containing SVHC of
concentrations exceeding 0.1% are included in any GCE product.
14.2.MANUFACTURING DATE
The manufacturing date is included in the product serial number which is marked on the label.
Serial number composition is the following: YYMMXXXXX
Y: 2 last numbers of the year of manufacture
M: month of manufacture
X: production number
The perforation in the device label indicates the year when the equipment has to be disposed
o ff .
GCE recommends to use original packaging when is withdraw from operation.
14.3.MAINTENACE
Repair activities of GCE product must be carried out by GCE or authorized repair centres.
Contact GCE for further information about service available in your area.
All labels on the equipment must be kept in good, legible condition by the owner and the user
during the entire product life time.
Use only genuine GCE components.
15. TRANSPORT AND STORAGE
Before fi rst use, the product shall be kept in its original packaging.
The product owner or user must ensure that during the product life cycle transport and storage
environmental conditions are in accordance with the above “Safety instruction” section
requirements, in order to maintain product integrity and cleanliness.
20/80
EN
16. GLOSSARY
Consult Instruction for use Take back equipment for
recycling. Do not dispose
equipment into unsorted
municipal waste
Caution!
Keep away from heat and
ammable materials Suitable for Hospital care use
Keep away from oil and
grease! Weight of product
Humidity limit SN Serial number
Temperature limit REF Catalogue number
Keep dry! LOT Batch code
Date of manufacture Manufacturer
Use by date Fragile, handle with care
Inlet parameter Outlet parameter
17. WARRANTY
The Standard Warranty period is two years from date of receipt by the GCE Customer (or if this
is not known 2 years from time of the product manufacture shown on the product).
The standard warranty is only valid for products handled according to Instruction for use
(IFU) and general industry good practice and standards.
APPENDIX:
Nr 1 – Technical specifi cations and performance data
Nr 2 – Testing report for relevant Serial Number of the Pressure Watch
MANUFACTURER:
GCE, s.r.o. Tel : +420 569 661 111
Zizkova 381 Fax : +420 569 661 602
583 01 Chotebor http://www.gcegroup.com
Czech Republic © GCE, s.r.o.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80

GCE PRESSURE WATCH Istruzioni per l'uso

Tipo
Istruzioni per l'uso