Apex Digital 9S-007202 Manuale utente

Categoria
Umidificatori
Tipo
Manuale utente

Questo manuale è adatto anche per

iCH Auto CPAP System
Instruction Manual
Table of Content
User’s manual .................................................................P. 1
Manual de Instrucciones
...............................................
P.25
Instruções de Utilização
................................................
P. 54
Instructions d'utilisation................................................P. 82
Gebrauchsanleitung
......................................................
P. 112
Gebruikershandleiding ..................................................P.141
Manuale di istruzioni ....................................................P.170
Model No.: 9S-007201, 9S-007202
Please read the instruction manual before use.
English
Español
Português
Français
Deutsch
Nederlands
Italiano
iCH Auto/English 1
IMPORTANT SAFEGUARDS
SAVE THESE INSTRUCTIONS
READ ALL INSTRUCTIONS BEFORE USING
WARNING
1. THIS DEVICE IS NOT INTENDED FOR LIFE SUPPORT. It may stop operating due to power
interruption without hazard to patient.
2. If oxygen is used with the device, the oxygen flow must be turned off when the device is not in use.
Explanation of the Warning: When the device is not in operation and the oxygen flow is left on,
oxygen delivered into the tubing may accumulate within the device and create a risk of fire.
3. Oxygen is inflammable. Oxygen should not be used while smoking or in the presence of an open
flame.
4. Always ensure the device is generating airflow before the oxygen supply is turned on. Always turn
off the oxygen supply before stopping the airflow from the device.
5. This device should not be used in the vicinity of a flammable anaesthetic mixture in combination
with oxygen or air and nitrous oxide.
6. The airflow for breathing generated by this device may be as much as C (12.6°F) higher than
the room temperature. This device should not be used if the room temperature is warmer than
35°C (95°F) to prevent the airflow temperature from exceeding 40°C (104°F) and causing irritation
to your airway.
7. If this device overheats, it will stop operating and show message “Error 002” on the display. After
cooling down to the proper temperature, the device can be restarted.
8. This machine should be used only with masks (and connectors) recommended by the
manufacturer, or by your doctor or respiratory therapist. A mask should not be used unless the
CPAP machine is turned on and operating properly. The vent holes for the mask should never be
blocked, to allow for proper exhaling. If the vent hole is blocked, the CPAP machine will stop and
show message Error 002”. After the machine cools down, please re-connect the power cord to
reset the machine.
9. At low CPAP pressure, some exhaled gas may remain in the mask and be breathed in again.
CAUTION
1. Make sure the area around the machine is dry and clean. Dust and foreign particles may affect the
treatment. Keep the air inlet on the back of the machine clear to prevent overheating and damage
of the device. Do not place the machine near a source of hot or cold air. An extremely cold or hot
environment may damage the user's respiratory airway.
2. If there is a possibility of electro-magnetic interference with mobile phones, please increase the
distance between devices or turn off the mobile phone.
3. Do not connect the device to the personal computer for data downloading during the treatment.
This may cause a failure in the CPAP system.
4. To prevent from potential allergic reactions to mask used in the device, This machine shall be
used only with masks (and connectors) recommended by the manufacturer.
5. U.S. Federal law restricts this device to sale by or on the order of a licensed doctor.
iCH Auto/English 2
DANGER -To reduce the risk of electrocution:
1. Always unplug this product immediately after using.
2. Do not use while bathing.
3. Do not place or store product where it can fall or be pulled into a tub or sink.
4. Do not place in or drop into water or other liquid.
5. Do not reach for a product that has fallen into water. Unplug immediately.
6. If product falls into water or other liquids do not reach into the water or other liquids. Unplug
immediately.
WARNING -To reduce the risk of burns, electrocution, fire or injury to persons:
1. This product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used with or near children or invalids.
Choking resulting from a child swallowing a small part that has become detached from the device
or its accessories.
3. Use this product only for its intended use as described in this manual, do not use attachments not
recommended by the manufacturer.
4. Never operate this product if it has a damaged cord or plug, if it is not working properly, if it has
been dropped or damaged, or dropped into water. Return the product to a service centre for
examination and repair.
5. Keep the cord away from heated surfaces.
6. Never block the air openings of this product or place it on a soft surface, such as a bed or couch,
where the air openings may be blocked. Keep the air opening free of lint, hair, and other similar
particles.
7. Never drop or insert any object into any opening or hose.
8. Follow the national requirementfor disposing of the unit.
9. Do not operate the device before the mask and water chamber have been installed.
10. Disconnect the water chamber from the device when not in use, water entering the device may
result in electric shock hazard or damage.
11. Do not use if water comes in contact with the device or enters the tubing.
12. This device is not for use with patients whose supraglottic airways have been bypassed.
13. Do not cover the device with blankets or clothes .
14. Do not remove water chamber while humidifier is heating. Do not touch the heater plate for 30
minutes after the device is disconnected from the mains.
15. No modification of this equipment is allowed.
16. Do not modify this equipment without authorization of the manufacturer.
17. If this equipment is modified, appropriate inspection and testing must be conducted to ensure
continued safe use of the equipment.
18. Do not leave long lengths of air tubing around the top of your bed. It could twist around your head
or neck while you are sleeping and lead to strangulation.
iCH Auto/English 3
1. Introduction
This manual should be used for initial set up of the system and saved for reference purpose.
1.1 General Information
Obstructive Sleep Apnea (OSA) is a condition that an intermitted and repetitive obstruction of the
upper respiratory tract causes a complete (apnea) or partial (hypopnea) block of breathing airflow
during sleep. The syndrome varies depending on the degree of relaxation of the tongue and soft
palate muscle.
The most common treatment for OSA is Continuous Positive Airway Pressure (CPAP). CPAP
devices can deliver a constant air pressure into your upper airway via a nasal mask. This
constant air pressure can keep your airway open during sleep, therefore prevents the OSA.
This device is a micro-processor controlled continuous positive airway pressure device. It
features the illuminated, menu-driven LCD display, universal power supply and ramp time
adjustment. The ramp time adjustment and ultra quiet operation ensure you to fall asleep
comfortably while air pressure slowly builds up to treatment level. The user compliance meter
records the total system’s operating time for physician’s reference.
The system has been tested and successfully approved to the following standards:
IEC/EN 60601-1
IEC/EN 60601-1-2
IEC/EN 61000-3-2 Class A
IEC/EN 61000-3-3
CISPR 11 Group 1, Class B
For US and CANADA only
The description below is for Canada only
Le produit à é testé avec des équipements médicaux et respecte les normes ANSI/AAMI
ES60601-1 (2005)+AMD 1 (2012) and CAN / CSA C22.2 No. 60601-1 (2014).. prévenant les choc électrique, le
feu et les risques de blessures physiques.
E348970
53DG
Medical Equipment- CPAP with integrated heated humidifier
AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI
ES60601-1 (2005) , ANSI/AAMI ES60601-1 (2005) + AMD 1
(2012), CAN/CSA-C22.2 No. 60601-1 (2008) and
CAN/CSA-C22.2 No. 60601-1 (2014)
iCH Auto/English 4
EMC Warning Statement
This equipment has been tested and found to comply with the limits for medical devices to the
IEC/EN 60601-1-2. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment to an outlet on a circuit different from that to which the other device(s)
are
connected.
- Consult the manufacturer or field service technician for help.
1.2 Intended Use
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of
adult Obstructive Sleep Apnea (OSA). The integrated heated humidifier is designed to increase
the humidity of the air from the CPAP thereby relieving the symptoms of a dry nose and throat that
some people may experience.
Cautions: Some patients might have pre-existing contraindications for CPAP therapy,
or might experience some potential side effects of using a CPAP device.
Please consult your doctor if you have any questions concerning your
therapy.
NOTE: Equipment not suitable for use in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide.
1.3 Contraindications
Studies have shown that the following pre-existing conditions may contraindicate the use of CPAP
therapy for some patients. Should you have any of these conditions, your physician will
determine if CPAP therapy is appropriate for you.
Bullous lung disease
Pneumothorax
Pathologically low blood pressure due to or associated with intravascular volume depletion
Severe cardiac arrhythmias or coronary artery disease
Stroke
Seizures
Pneumocephalus has been reported in a patient using nasal Continuous Airway pressure
iCH Auto/English 5
iCH Auto/English 6
2. Product Description
2.1 Unpacking the Contents
Components include:
(1) Main CPAP device with integrated heated humidifier
(2) Power supply adapter and AC power cord
(3) User manual
(4) Flexible air tubing, 1.8 m length
(5) Carrying bag
(6) SD card and USB cable
NOTE: Please use a SD card (smaller than 4GB) which complies with SDHC standard.
Before using it, please format it to FAT16 to ensure correct data collection.
CAUTION: Patient should not connect the device to a personal computer for data
downloading. This may cause the CPAP system failure.
To secure contents inside, the device and accessories are bundled in a paper packaged box.
Unpack this box by removing the device and accessories and check for any damage, which may
have occurred during shipping. If there are damages, please contact your dealer immediately.
2.2 System Overview
3. Installation
1. LCD Display
2. Up & Down Button
3. Start/Standby Button
4. Menu Button
5. SD Card Slot/mini-USB Connector
6. Air Filter
7. Heater Platform
8. Heater Indicator
9. Heater Control knob
10. Air Outlet
11. Power Socket
12. Air Outlet of the Water Chamber
13. Protruding tab and groove
14. Fill Line
15. Water Chamber
16. Buckle
5
6
7
9
10
11
1
2
3
4
12
15
14
16
13
iCH Auto/English 7
3.1 Disassembling the Water Chamber
1. Turn the device off and allow the heater and water to cool. If necessary, disconnect the
flexible tubing from the water chamber.
2. Place thumb on the top of the water chamber and tip it upward as shown in Figure 3-1.
Remove the water chamber from the system.
3. Unlock the water chamber by pulling the tab as shown in Figure 3-2. Remove the chamber lid
from the base.
3.2 Filling the Water Chamber
1. Fill the chamber to the fill line with distilled
water (approx. 300 ml), as shown in Figure 3-3.
2. Follow the steps to assemble the water chamber:
2.1 Insert the protruding tab into the groove
2.2 Put the top and bottom of the water chamber together
3. If the silicon gasket has been removed, replace it
securely on the top edge of the water chamber base.
Otherwise, the device will not deliver air properly.
4. Follow the steps to install the chamber correctly:
4.1 Position the water chamber on the heater plate
4.2 Tilt the water chamber, align with the air outlet (Figure
3-4, 3-5)
4.3 Push downwards until you hear a click (Figure 3-6)
4.4 Lock on the buckle to secure the chamber (Figure 3-7)
Figure 3-2
Figure 3-1
Figure 3-3
Figure 3-6
Figure 3-5
Figure 3-7
Figure 3-4
iCH Auto/English 8
NOTE: Make sure the silicon connector on the inlet connector fits securely over the
CPAP device air outlet.
CAUTION: Do not overfill the water chamber. Damage to the device may occur.
CAUTION: Use only room temperature water; do not fill the chamber with hot or chilled water.
CAUTION: Do not turn the heater on without the water chamber installed
CAUTION: When installing the water chamber, do not allow water to spill into the device.
CAUTION: Do not move the device when the chamber has water in it.
3.3 Setting Up
1. Place the device on a flat surface and at a lower level than your sleeping position.
2. Connect one end of the air tubing firmly to the air outlet on the back of the device, as shown
in Figure 3-8.
3. Connect the other end of the air tubing to the mask system. Put on the mask and headgear
according to the mask instruction manual.
4. Plug the socket end of the AC power cord into the power supply adapter. Plug the pronged
end of the AC power cord into a main electrical outlet.
5. Plug the power supply cord’s adapter connector into the power inlet on the right side of the
device, as shown in Figure 3-9.
6. Once the power supply cord’s adapter connector is plugged into the power inlet, the CPAP
system is in ready to operate position (“STANDBY” sign appears in LCD display)
NOTE: The AC power cord also serves to disconnect the device.
NOTE: Do not position the equipment so that it is difficult to operate the
disconnecting device.
Figure 3-8
Figure 3-9
iCH Auto/English 9
4. Operation
NOTE: Always read the operating instructions before use.
4.1 Control Panel Description
Buttons on control panel and main use of the buttons:
START/STANDBY
To start the treatment, simply press the "START/STANDBY" button. To stop the
treatment, press the "START/STANDBY" button again. The display will switch
between [STANDBY] and Therapy Pressure [ XX.X cmH
2
O] or [ XX.X hPa] in
cmH
2
O or hPa unit.
MENU
Press the "MENU" button to enter the setting mode when device is in standby mode.
The adjustment setting includes ramp time selection, ramp starting pressure, therapy
pressure adjustment, alarm ON/OFF, clock alarm setting, compliance meter, and total
operating meter. When each setting's value has been changed, press "MENU" for
confirmation and press "MENU" again for next setting selection. Please refer to 4.3
Function Description section for detailed information.
UP
Press the "UP" button to increase the selected value.
DOWN
Press the "DOWN" button to decrease the selected value.
iCH Auto/English 10
4.2 Function Description
Getting Started First time use.
For first-time use, follow the steps below:
1. Make sure that your doctor or care provider has initialized and configured the appropriate
treatment values into the unit.
2. Make sure that your package contains a face or nasal mask with headgear. If not, your
doctor or care provider can recommend the type of mask and headgear you must obtain
3. Attach tubing and mask (see Setup section).
4. Attach power cord to the unit and an electrical outlet (see Setup section). Unit will power-on
automatically.
5. Use the “MENU button to set the Current Time, Ramp Up Time, and Ramp Starting
Pressure, (see “MENU” Button Section for description of these options).
Note that these values may have already been set by your doctor or care provider.
6. Press the “START/STANDBY” button Airflow to the mask will begin.
7. You can now place the Headgear and Mask over your head and face and begin treatment.
4.3 Menu Button - Option Description
(1) Pressure Unit
The first selection of pressing "MENU" button is to set Pressure Unit [cmH
2
O / hPa] menu,
press "UP" or "DOWN" button to set the preferred pressure unit and press "MENU" for
confirmation. There are two pressure unit, cmH
2
O and hPa.
(2) Ramp Time
Ramp time function allows user to fall into sleep with a lower, comfortable pressure and helps
users gradually get accustomed to increasing treatment pressure. The first selection of
pressing "MENU" is [Ramp XX MIN]. When the "MENU" setting is in [Ramp XX MIN]
mode, press "UP" or "DOWN" button to set the preferred ramp time and press "MENU" for
confirmation. There are 10 adjustable levels, 0, 5, 10, 15, 20, 25, 30, 35, 40 and 45 minutes.
(3) Ramp Starting Pressure (only CPAP mode)
Press "MENU" button to select [Ramp P XX.X] menu, press "UP" or "DOWN" button to
set the preferred ramp starting pressure and press "MENU" for confirmation. The ramp
starting pressure can be changed from 4 cmH
2
O / hPa to “Therapy Pressure 1” cmH
2
O /
hPa. For example, if your therapy pressure is 10 cmH
2
O / hPa, the maximum ramp starting
pressure you can select is 9 cmH
2
O / hPa.
(4) Therapy Pressure (only CPAP mode)
Press "MENU" button to select the [P XX.XcmH
2
O] or [P XX.XhPa] menu. You can
view the current pressure setting displayed in cmH
2
O or hPa unit. Therapy pressure is
adjustable only by the provider, a respiratory therapist or doctor.
NOTE: The therapy pressure is to only be prescribed by a physician.
(5) Maximum Pressure (only APAP mode)
Press "MENU" button to select [Max. XX.XcmH
2
O] or [Max. XX.XhPa] menu, you can
view the current pressure setting displayed in cmH
2
O or hPa unit. Maximum pressure is
adjustable only by the provider, a respiratory therapist or doctor.
NOTE: The maximum pressure is only to be prescribed by a doctor.
(6) Minimum Pressure (only APAP mode)
Press "MENU" button to select [Min. XX.XcmH
2
O] or [Min. XX.XhPa] menu, you can
view the current pressure setting displayed in cmH
2
O or hPa unit. Minimum pressure is
adjustable only by the provider, a respiratory therapist or doctor.
iCH Auto/English 11
NOTE: The minimum pressure is only to be prescribed by a doctor.
(7) Initial Pressure (only APAP mode)
Press "MENU" button to select [Init. XX.XcmH
2
O] or [Init. XX.XhPa] menu. Here you
can view the current pressure setting displayed in cmH
2
O or hPa unit. Initial pressure is
adjustable only by the provider, a respiratory therapist or doctor.
NOTE: The initial pressure is only to be prescribed by a doctor.
(8) PVA (Pressure Variation Algorithm)
Press "MENU" button to select [c on/off] menu, you can view if the PVA function is
enabled by the provider, a respiratory therapist or physician. The PVA function delivers
pressure relief upon exhalation to improve comfort. If the PVA function is enabled, press
"UP" or "DOWN" button to adjust the comfort level. There are 3 adjustable levels of
pressure relief that you feel while exhalation. Level 1 provides small pressure variation; and
level 3 provides higher pressure variation.
(9) Low Pressure Auto Off (only CPAP mode)
Press "MENU" button to select [LOW P on/off] menu, you can view if the Low Pressure Auto
Off setting is enabled and the device will be automatically turned off while detecting a large
leak for 3 minutes. This setting is adjustable only by the provider, a respiratory therapist or
physician.
(10) Compliance Meter
Press "MENU" button to select the [CM XXXX.X hr] menu. The compliance meter records
the total therapy hours for the device. The compliance meter should be re-set only by the
provider, a respiratory therapist or by a doctor.
(11) Alarm
Press "MENU" button to select the [Alarm on/off] menu, and press "UP" or "DOWN" button
to set the alarm on or off. When the alarm is turned on, the audible alarm will activate with
warning messages shown on the LCD display. Set alarm off to mute the audible alarm.
(12) Clock Alarm
Press "MENU" button to select the [Clock Alarm on/off] menu and press "UP" or "DOWN"
button to set the clock alarm on or off. When the clock alarm is set on, the display will show
the time on the left side. Press "UP" or "DOWN" button to set the time to wake you up. Once
the clock alarm is activated, press the start/standby button to mute the audible alarm.
(13) Clock
Press "MENU" button to select the [Clock XX:XX] menu, press "UP" or "DOWN" button to
set the current time.
NOTE: If you set a new time that occurs in the past then the "Invalid Data" which does
not exist for this new period would be erased.
(14) Turning off the Device
Remove the power cord from the electrical outlet, and disconnect power cord from the power
socket on the right side of the device.
NOTE: Once the setting is confirmed, press the “MENU” button. Otherwise, the device
will automatically go back to standby without saving the modification if no action
is taken in 20 seconds.
iCH Auto/English 12
(15) Compliance Information
Hold "UP" and "DOWN" button to read compliance information while in the standby screen.
Press “START/STANDBY” button to go back to standby screen
i. Period
When entering the compliance information mode, the LCD screen should start
blinking to allow you to select the period. Press "UP" or "DOWN" button to select the
period from 1 day to 90 days. After selecting the period, press "MENU" to confirm.
Press “DOWN” button to read more compliance information.
ii. Usage Hours / Compliance Hours
The Usage Hours [XXX hrs] records the therapy hours under the set time frame.
Press “MENU” button to reselect the time frame, or press “DOWN” button to read
more compliance information.
iii. Average Compliance Hours per day
The Average Compliance Hours per day [X.X hrs/d] records the average
compliance hours per day over the different time frame. Press "UP" or "DOWN"
button to read last or next compliance information.
iv. Days4 Hours
The Days4 Hours [
4 X.X d] records the cumulative number of days that
exceeded 4 hours over the different time frame. Press "UP" or "DOWN" button to
read last or next compliance information.
v. AHI
The [AHI XX.X] records the average Apnea Hypopnea Index (AHI) over the
different time frame. Press "UP" or "DOWN" button to read last or next compliance
information.
vi. Average Pressure
The [P XX.XcmH
2
O] or [P XX.XhPa] records the average pressure over the
different time frame. Press "UP" or "DOWN" button to read last or next compliance
information.
vii. Average Leak Rate
The [Leak XX.X lpm] records the average leak over the different time frame.
Press "UP" button to read last or next compliance information or press "MENU" to
come back to the main menu.
(16) Check Value
Press "MENU" button to select [CHXXX] menu, which displays a check value. Your care
provider may ask for this information
iCH Auto/English 13
4.4 Flowchart of Menu settings
Enter the user's menu mode by pressing the “MENU button.
In each setting, when the preferred value has been selected, press "MENU" for confirmation and
press "MENU" again to enter next selection.
NOTE: For doctors, please refer to a separate “Doctor’s Additional Instructions” manual.
APAP
Mode
CPAP
Mode
Clock
< hh:mm >
Clock Alarm <On>
< hh:mm >
CM
(Compliance Meter)
< xxxx.x > hr
Alarm
< on/off >
APAP
Max. P
(Maximum Pressure)
< 5.0 ~ 20.0 > cmH
2
O / hPa
Min. P
(Minimum Pressure)
< 4.0 ~ 19.0 > cmH
2
O / hPa
Init. P
(Initial Pressure)
< 4.0 ~ 19.0 > cmH
2
O / hPa
Pressure Unit
< cmH
2
O / hPa >
C
(PVA Comfort Level)
< on/off >
CPAP
Ramp
(Ramp Time)
< 0 ~ 45 > min
P
(Therapy Pressure)
< 4.0 ~ 20.0 > cmH
2
O / hPa
Ramp P
(Ramp Starting Pressure)
<4.0~19.0>cmH
2
O / hPa
Pressure Unit
< cmH
2
O / hPa >
Low P
(Low Pressure Shut Off)
< on/off >
C
(PVA Comfort Level)
< on/off >
PVA <On>
< 0, 1, 2, 3 >
PVA <On>
< 0, 1, 2, 3 >
Clock Alarm
< on/off >
Ramp
(Ramp Time)
< 0 ~ 45 > min
iCH Auto/English 14
4.5 Flowchart of Compliance Information reading
4.6 Using the heated humidifier
1. Turn the control knob on the back of the device to turn on the heater and the heater on
indicator
above the knob will light up (orange light), as shown in Figure 4-1
2. Adjust the control knob to increase the humidity in the range of 1 to 6 (see diagram). 1 is
minimum humidity, 6 is maximum humidity. The optimum setting is dependent on the ambient
temperature and humidity of your room.
NOTE: When the ‘Low P’ sign is displayed on the screen while large air leakage is
Figure 4-1
Period
(Time Frame)
< 1 d /7d / 30 d / 90 d>
Usage Hours
< xxx > hrs
4
(Days4hrs)
< xx > d
Average Compliance
Hours per day
< x.x > hrs /d
AHI
< xx.x >
P
(Average Pressure)
< xx.x > cmH
2
O/hPa
Leak
(Average Leak Rate)
< xx.x > lpm
CH
(Check Value)
< xxx >
iCH Auto/English 15
detected, heating will automatically stop with the heater indicator lights off. (e.g.
removing the mask will stop the machine from operating.) If the air leakage
problem is eliminated, heater will resume again.
Danger: In order to avoid the destruction or damage to the device or even the
hazard of electric shock, when the water chamber is attached, DO NOT fill
water from the outlet port of the water chamber. Users must fill water no
more than the maximum mark after the water chamber is removed from the
device.
4.7 Using the SD card
If doctors need to review the usage data, they may ask you to use the SD card to copy data from
the device, and to return the card to them. Data that is copied to the SD Card is still stored and
available on the device.
1. Insert the SD Card when the device is in standby mode.
2. Data copying starts automatically when the SD Card is inserted into the slot.
3. The [COPY] & [Card] message is displayed on the LCD while data is being copied.
4. The [End] & [OK] message is displayed on the LCD when copying has finished.
5. Remove the SD Card and mail it to the clinician.
6. The SD Card should be stored in the plastic card case when not in use. The SD card
does not need to be uninstalled for the device to work properly.
CAUTION: Do not remove the SD card until [End] & [OK] message is displayed on the LCD,
or data copied to the card may be corrupted or missing.
5. Cleaning & Maintenance
5.1 Device
The device should be checked and dusted regularly (at least every 30 days). Wipe with a damp
cloth and a mild detergent and keep it free from dust. If other detergent is used, choose one that
will have no chemical effects on the surface of the plastic case. All parts should be air-dried
thoroughly before use. Inspect the device and all circuit parts for any damage after cleaning and
replace if necessary.
WARNING: Don’t try to open this device. Repairs and internal servicing should only performed by
an authorized service agent. Do not drop any object into the air tube or air outlet.
5.2 Tubing and Mask
The tubing and mask should be checked and cleaned regularly. Please refer to the cleaning
instruction s that come with the accessories.
1. Disconnect the air tubing from the air outlet of the device.
iCH Auto/English 16
2. Remove the air tubing and headgear straps from the mask.
3. Wash the mask system according to the instructions supplied with it.
4. Wash the air tubing in warm water using mild detergent. Rinse thoroughly, hang and allow to
dry.
5. Before next use, assemble the mask and headgear according to the mask user instructions.
6. All items of the mask and air tubing are subject to normal wear and tear and may eventually
be
replaced. Replace the mask and the air tubing if they are damaged.
CAUTION: Do not use bleach, chlorine-, alcohol-, or aromatic-based (including all scented oils),
moisturizing or antibacterial soaps to clean the cushion, mask or air tubing. These solutions
may cause hardening and reduce the life of the product.
CAUTION: Do not wash or dry the mask or air tubing at a temperature above 70°C
(158°F)
WARNING: Do not use any cleaner containing fragrance or conditioners as they will leave a
residue.
WARNING: The mask must not be re-used by another person. This is to avoid the risk of
cross-infection.
5.3 Air Filter
For an optimum operation of the device, the air filter can be cleaned by the user at least once
every two weeks or more often if this device is operated in a dusty environment. It is
recommended that the filter be replaced with a new one after 6 months of use; however,
depending on the air quality, the replacement time may vary.
N
OTE: Please follow national requirements to dispose of the unit
properly.
5.4 Water Chamber
1. Turn the heater off and allow the heater and water to cool.
2. Disconnect the flexible tubing from the water chamber. Place thumb on the top of the water
chamber and lift it to separate it from the system. Open the chamber and empty the
remaining water.
CAUTION: Dirty air filter may cause high operating temperatures
that affect device performance. Ensure the air filter is
cleaned and fitted at all times
1. Remove the dirty filter from the enclosure on the rear of the device.
2. Wash the filter in warm water with a mild detergent, and rinse with
water. Allow the filter to air dry completely before reinstalling. Do not
use a filter that is not completely dry. If the filter is torn, replace it.
3. Reinstall the filter.
iCH Auto/English 17
3. Use a mild detergent to wash all the chamber parts. Rinse all the parts with clear water and
allow them to air dry.
4. All items of the chamber are subject to normal wear and tear and may eventually be replaced.
Replace the chamber parts if any damage is present.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212