ResMed Airsense 10 Autoset / Elite Guida utente

Tipo
Guida utente
User guide
English
|
Italiano
|
Español
|
Português
|
Ελληνικά
|
Türkçe
| ةيبرعلا
User guide
English 1
ENGLISH
Welcome
The AirSense10 AutoSet
and AirSense 10 Elite are ResMed's premium auto-adjusting pressure
and Continuous Positive Airway Pressure (CPAP) devices.
WARNING
Read this entire guide before using the device.
Use the device according to the intended use provided in this guide.
The advice provided by your prescribing doctor should be followed ahead of the
information provided in this guide.
Indications for use
AirSense 10 AutoSet
The AirSense 10 AutoSet self-adjusting device is indicated for the treatment of obstructive sleep
apnoea (OSA) in patients weighing more than 30 kg. It is intended for home and hospital use.
The humidifier is intended for single patient use in the home environment and re-use in a
hospital/institutional environment.
AirSense 10 Elite
The AirSense 10 Elite device is indicated for the treatment of obstructive sleep apnoea (OSA) in
patients weighing more than 30 kg. It is intended for home and hospital use.
The humidifier is intended for single patient use in the home environment and re-use in a
hospital/institutional environment.
Clinical benefits
The clinical benefit of CPAP therapy is a reduction in apnoeas, hypopnoeas and sleepiness, as well
as improved quality of life.
The clinical benefit of humidification is the reduction of positive airway pressure related side effects.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the following
pre-existing conditions:
severe bullous lung disease
pneumothorax
pathologically low blood pressure
dehydration
cerebrospinal fluid leak, recent cranial surgery, or trauma.
2
Adverse effects
You should report unusual chest pain, severe headache, or increased breathlessness to your
prescribing physician. An acute upper respiratory tract infection may require temporary
discontinuation of treatment.
The following side effects may arise during the course of therapy with the device:
drying of the nose, mouth, or throat
nosebleed
bloating
ear or sinus discomfort
eye irritation
skin rashes.
At a glance
The AirSense 10 includes the following:
Device
HumidAir
humidifier (if supplied)
Air tubing
Power supply unit
Travel bag
SD card (not available in all devices).
Contact your care provider for a range of accessories available for use with the device including:
Air tubing (heated and non-heated): ClimateLineAir
, SlimLine
, ClimateLineAir Oxy, Standard
Side cover for use without the humidifier
Filter: Hypoallergenic filter, standard filter
Air10
DC/DC converter
SD card reader
Air10 oximeter adapter
Air10 USB adapter
Power Station II
Air10 tubing elbow.
English 3
About your device
1
Air outlet
5
HumidAir humidifier
2
Air filter cover
6
Screen
3
Power inlet
7
Adapter cover
4
Serial number and device number
8
SD card cover
About the control panel
Start/Stop button
Press to start/stop therapy.
Press and hold for three seconds to enter power save
mode.
Dial
Turn to navigate the menu and press to select an option.
Turn to adjust a selected option and press to save your
change.
Home button Press to return to the Home screen.
Different icons may be displayed on the screen at different times including:
Ramp Time
Wireless signal strength (green)
Ramp Time Auto
Wireless transfer not enabled (grey)
Humidity
No wireless connection
Humidifier warming
Airplane Mode
Humidifier cooling
4
Setup
CAUTION
Do not overfill the humidifier as water may enter the device and air tubing.
1. Place the device on a stable level surface.
2. Plug the power connector into the rear of the device. Connect one end of the power cord into
the power supply unit and the other end into the power outlet.
3. Connect the air tubing firmly to the air outlet located on the rear of the device.
4. Open the humidifier and fill it with water up to the maximum water level mark.
Do not fill the humidifier with hot water.
5. Close the humidifier and insert it into the side of the device.
6. Connect the free end of the air tubing firmly onto the assembled mask.
See the mask user guide for detailed information.
Recommended masks are available on www.resmed.com.
English 5
Starting therapy
1. Fit your mask.
2. Press Start/Stop or breathe normally if SmartStart
is enabled.
You will know that therapy is on when the Sleep Report screen is displayed.
The current treatment pressure is shown in green.
During ramp time the pressure is gradually increasing and you will see a
spinning circle. Once the prescribed treatment pressure is reached, the entire
circle will be green.
The screen will go black automatically after a short period of time. You can press Home or the dial to
turn it back on. If power is interrupted during therapy, the device will automatically restart therapy
when power is restored.
The AirSense 10 device has a light sensor that adjusts the screen brightness based on the light in
the room.
Stopping therapy
1. Remove your mask.
2. Press Start/Stop or if SmartStart is enabled, therapy will stop automatically after a few seconds.
The Sleep Report now gives you a summary of your therapy session.
Usage hoursIndicates the number of hours of therapy you received last
session.
Mask SealIndicates how well your mask sealed:
Good mask seal.
Needs adjusting, see Mask Fit.
HumidifierIndicates if your humidifier is working properly:
Humidifier working.
Humidifier might be faulty, contact your care provider.
If set by your care provider, you will also see:
Events per hourIndicates the number of apnoeas and hypopnoeas experienced per hour.
More InfoTurn the dial to scroll down to view more detailed usage data.
6
Power save mode
Your AirSense 10 device records your therapy data. In order to allow it to transmit the data to your
care provider, you should not unplug the device. However, you can put it into power save mode to
save electricity.
To enter power save mode:
Press and hold Start/Stop for three seconds.
The screen goes black.
To exit power save mode:
Press Start/Stop once.
The Home screen is displayed.
My Options
Your AirSense 10 device has been set up for your needs by your care provider, but you may find you
want to make small adjustments to make your therapy more comfortable.
Highlight My Options and press the dial to see your
current settings. From here, you can personalise your
options.
Ramp Time
Designed to make the beginning of therapy more comfortable, Ramp Time is the period during
which the pressure increases from a low start pressure to the prescribed treatment pressure.
You can set your Ramp Time to Off, 5 to 45 minutes or Auto. When Ramp Time is set to Auto, the
device will detect when you have fallen asleep and then automatically rise to the prescribed
treatment pressure.
To adjust Ramp Time:
1. In My Options, turn the dial to highlight Ramp
Time and then press the dial.
2. Turn the dial to adjust the ramp time to your
preferred setting and press the dial to save the
change.
English 7
Humidity Level
The humidifier moistens the air and is designed to make therapy more comfortable. If you are
getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask, turn
down the humidity.
You can set the Humidity Level to Off or between 1 and 8, where 1 is the lowest humidity setting
and 8 is the highest humidity setting.
To adjust the Humidity Level:
1. In My Options, turn the dial to highlight Humidity
Level and then press the dial.
2. Turn the dial to adjust the humidity level and press
the dial to save the change.
If you continue to get a dry nose or mouth, or moisture in your mask, consider using ClimateLineAir
heated air tubing. ClimateLineAir together with Climate Control delivers more comfortable therapy.
Mask Fit
Mask Fit is designed to help you assess and identify possible air leaks around your mask.
To check Mask Fit:
1. Fit the mask as described in the mask user guide.
2. In My Options, turn the dial to highlight Run Mask Fit and then press the
dial.
The device starts blowing air.
3. Adjust the mask, mask cushion and headgear until you get a Good result.
To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a good mask seal, talk to your
care provider.
More options
There are some more options on your device which you can personalise.
When Pressure Relief is enabled, you may find it easier to breathe out. This
can help you get used to therapy.
When SmartStart is enabled, therapy starts automatically when you breathe
into your mask. When you remove your mask, it stops automatically after a
few seconds.
*When enabled by your care provider.
8
Caring for your device
It is important that you regularly clean your AirSense 10 device to make sure you receive optimal
therapy. The following sections will help you with disassembling, cleaning, checking and
reassembling your device.
WARNING
Regularly clean your tubing assembly, humidifier and mask to receive optimal therapy and to
prevent the growth of germs that can adversely affect your health.
Disassembling
1. Hold the humidifier at the top and bottom, press it gently and pull it away from the device.
2. Open the humidifier and discard any remaining water.
3. Hold the cuff of the air tubing and gently pull it away from the device.
4. Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Cleaning
You should clean the device weekly as described. Refer to the mask user guide for detailed
instructions on cleaning your mask.
1. Wash the humidifier and air tubing in warm water using mild detergent.
2. Rinse the humidifier and air tubing thoroughly and allow to dry out of direct sunlight and/or heat.
3. Wipe the exterior of the device with a dry cloth.
English 9
Notes:
The humidifier may be washed in a dishwasher on the delicate or glassware cycle (top shelf only).
It should not be washed at temperatures higher than 65ºC.
Do not wash the air tubing in a dishwasher or washing machine.
Empty the humidifier daily and wipe it thoroughly with a clean, disposable cloth. Allow to dry out
of direct sunlight and/or heat.
Checking
You should regularly check the humidifier, air tubing and the air filter for any damage.
1. Check the humidifier:
Replace it if it is leaking or has become cracked, cloudy or pitted.
Replace it if the seal is cracked or torn.
Remove any white powder deposits using a solution of one part household vinegar to
10 parts water.
2. Check the air tubing and replace it if there are any holes, tears or cracks.
3. Check the air filter and replace it at least every six months. Replace more often if there are any
holes or blockages by dirt or dust.
To replace the air filter:
1. Open the air filter cover and remove the old air filter.
The air filter is not washable or reusable.
2. Place a new air filter onto the air filter cover and then close it.
Make sure the air filter is fitted at all times to prevent water and dust from entering the device.
Reassembling
When the humidifier and air tubing are dry, you can reassemble the parts.
1. Connect the air tubing firmly to the air outlet located on the rear of the device.
2. Open the humidifier and fill it with room temperature water up to the maximum water level
mark.
3. Close the humidifier and insert it into the side of the device.
4. Connect the free end of the air tubing firmly onto the assembled mask.
10
Therapy data
Your AirSense 10 device records your therapy data for you and your care provider so they can view
and make changes to your therapy if required. The data is recorded and then transferred to your care
provider wirelessly, if a wireless network is available, or via an SD card.
Data transmission
Your AirSense 10 device has the capability of wireless communication so that your therapy data can
be transmitted to your care provider to improve the quality of your treatment. This is an optional
feature that will only be available if you choose to benefit from it and if a wireless network is
available. It also allows your care provider to update your therapy settings in a more timely manner
or upgrade your device software to ensure you receive the best therapy possible.
The data is usually transmitted after therapy has stopped. In order to make sure that your data is
transferred, leave your device connected to the mains power at all times and make sure that it is not
in Airplane Mode.
Notes:
Therapy data might not be transmitted if you use it outside of the country or region of purchase.
Wireless communication depends on network availability.
Devices with wireless communication might not be available in all regions.
SD card
An alternative way for your therapy data to be transferred to your care provider is via the SD card.
Your care provider may ask you to send the SD card by mail or to bring it in. When instructed by
your care provider, remove the SD card.
Do not remove the SD card from the device when the SD light is flashing, because data is being
written to the card.
To remove the SD card:
1. Open the SD card cover.
2. Push in the SD card to release it. Remove the SD card from the device.
Place the SD card in the protective folder and send it back to your care provider.
For more information on the SD card refer to the SD card protective folder provided with your
device.
Note: The SD card should not be used for any other purpose.
English 11
Travelling
You can take your AirSense 10 device with you wherever you go. Just keep the following in mind:
Use the travel bag provided to prevent damage to the device.
Empty the humidifier and pack it separately in the travel bag.
Make sure you have the appropriate power cord for the region you are travelling to. For
information on purchasing, contact your care provider.
If you are using an external battery, you should turn off the humidifier in order to maximise the
life of your battery. Do this by turning the Humidity Level to Off.
Travelling by plane
Your AirSense 10 device may be taken on board as carry-on luggage. Medical devices do not count
toward your carry-on luggage limit.
You can use your AirSense 10 device on a plane as it meets the Federal Aviation Administration
(FAA) requirements. Air travel compliance letters can be downloaded and printed from
www.resmed.com.
When using the device on a plane:
Make sure the humidifier is completely empty and inserted into your device. The device will not
work without the humidifier inserted.
Turn on Airplane Mode.
To turn on Airplane Mode:
1. In My Options, turn the dial to highlight Airplane
Mode and then press the dial.
2. Turn the dial to select On and then press the dial to
save the change.
The Airplane Mode icon
is displayed at the top
right of the screen.
CAUTION
Do not use the device with water in the humidifier on a plane due to the risk of inhalation of
water during turbulence.
12
Troubleshooting
If you have any problems, have a look at the following troubleshooting topics. If you are not able to
fix the problem, contact your care provider or ResMed. Do not try to open the device.
General troubleshooting
Problem/possible cause
Solution
Air is leaking from around my mask
Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
I am getting a dry or blocked nose
Humidity level may be set too low. Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the
ClimateLineAir user guide.
I am getting droplets of water on my nose, in the mask and air tubing
Humidity level may be set too high. Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the
ClimateLineAir user guide.
My mouth is very dry and uncomfortable
Air may be escaping through your mouth. Increase the Humidity Level.
You may need a chin strap to keep your mouth closed or a
full face mask.
Air pressure in my mask seems too high (it feels like I am getting too much air)
Ramp may be turned off. Use the Ramp Time option.
Air pressure in my mask seems too low (it feels like I am not getting enough air)
Ramp may be in progress. Wait for air pressure to build up or turn Ramp Time off.
My screen is black
Backlight on the screen may have turned off. It turns off
automatically after a short period of time.
Press Home or the dial to turn it back on.
Power may not be connected. Connect the power supply and make sure the plug is fully
inserted.
I have stopped therapy, but the device is still blowing air
Device is cooling down. Device blows a small amount of air in order to avoid
condensation in the air tubing. It will stop automatically after
30 minutes.
English 13
Problem/possible cause
Solution
My humidifier is leaking
Humidifier may not be assembled correctly. Check for damage and reassemble the humidifier correctly.
Humidifier may be damaged or cracked. Contact your care provider for a replacement.
My therapy data has not been sent to my care provider
Wireless coverage may be poor. Make sure that the device is placed where there is coverage
(ie, on your bedside table, not in a drawer or on the floor).
The Wireless signal strength icon
indicates good
coverage when all bars are displayed, and poor coverage
when fewer bars are displayed.
The No wireless connection icon
is displayed on the
top right of the screen. no wireless network available.
Make sure that the device is placed where there is coverage
(ie, on your bedside table, not in a drawer or on the floor).
If instructed to do so, send the SD card to your care provider.
The SD card also contains your therapy data.
Device may be in Airplane Mode. Turn off Airplane Mode, see Travelling by plane.
Data transfer is not enabled for your device. Talk to your care provider about your settings.
My screen and buttons are flashing
Software upgrade is in progress. Software upgrade takes approximately 10 minutes to
complete.
Device messages
Device message/possible cause
Solution
High leak detected, check your water tub, tub seal or side cover
Humidifier may not be inserted properly. Make sure the humidifier is correctly inserted.
Humidifier seal may not be inserted properly. Open the humidifier and make sure that the seal is correctly
inserted.
High leak detected, connect your tubing
Air tubing may not be connected properly. Make sure the air tubing is firmly connected at both ends.
Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Tubing blocked, check your tubing
Air tubing may be blocked. Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly. Remove and reinsert the SD card.
14
Device message/possible cause
Solution
Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.
Move the switch on the SD Card from the lock position
to the unlock position
and then re-insert it.
System fault, refer to user guide, Error 004
Device may have been left in a hot environment. Allow to cool before re-use. Disconnect the power supply
and then reconnect it to restart the device.
Air filter may be blocked. Check the air filter and replace it if there are any blockages.
Disconnect the power supply and then reconnect it to restart
the device.
Air tubing may be blocked. Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
There may be water in the air tubing. Empty the water from the air tubing. Disconnect the power
supply and then reconnect it to restart the device.
All other error messages, for example, System fault, refer to user guide, Error 0XX
An unrecoverable error has occurred on the device. Contact your care provider. Do not open the device.
English 15
Reassembling parts
Some parts of your device are designed to easily come off in order to avoid damage to the parts or
the device. You can easily reassemble them as described below.
To insert the humidifier seal:
1. Place the seal into the lid.
2. Press down along all edges of the seal until it is firmly in place.
To reassemble the humidifier lid:
1. Insert one side of the lid into the pivot hole of the base.
2. Slide the other side down the ridge until it clicks into place.
WARNING
Make sure that you arrange the air tubing so that it will not twist around the head or neck.
Regularly inspect power cords, cables, and power supply for damage or signs of wear.
Discontinue use and replace if damaged.
Keep the power cord away from hot surfaces.
If you notice any unexplained changes in the performance of the device, if it is making
unusual sounds, if the device or the power supply are dropped or mishandled, or if the
enclosure is broken, discontinue use and contact your care provider or your ResMed
Service Centre.
Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorised ResMed service agent.
Beware of electrocution. Do not immerse the device, power supply or power cord in water.
If liquids are spilled into or onto the device, unplug the device and let the parts dry. Always
unplug the device before cleaning and make sure that all parts are dry before plugging it
back in.
16
Supplemental oxygen must not be used while smoking or in the presence of an open
flame.
Always make sure that the device is turned on and airflow generated before the oxygen
supply is turned on. Always turn the oxygen supply off before the device is turned off, so
that unused oxygen does not accumulate within the device enclosure and create a risk of
fire.
Do not perform any maintenance tasks while the device is in operation.
The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation in the
configuration in which it will be used.
The use of accessories other than those specified for the device is not recommended. They
may result in increased emissions or decreased immunity of the device.
Regularly check the antibacterial filter for signs of moisture or other contaminants,
particularly during nebulization or humidification. Failure to do so could result in increased
breathing system resistance.
CAUTION
Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the
effectiveness of the treatment and/or damage the device.
Use only vented masks recommended by ResMed or by the prescribing doctor with this
device. Fitting the mask without the device blowing air can result in rebreathing of exhaled
air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow
of the fresh air into the mask.
Be careful not to place the device where it can be bumped or where someone is likely to
trip over the power cord.
Blocking the air tubing and/or air inlet of the device while in operation could lead to
overheating of the device.
Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding)
that could block the air inlet or cover the power supply unit.
Do not place the device on its side as water might get into the device.
Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is
correctly set up.
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or
antibacterial soaps or scented oils to clean the device, the humidifier or air tubing. These
solutions may cause damage or affect the humidifier performance and reduce the life of
the products. Exposure to smoke, including cigarette, cigar or pipe smoke, as well as ozone
or other gases, may damage the device. Damage caused by any of the foregoing, will not
be covered by ResMed's limited warranty.
If you use the humidifier, always place the device on a level surface lower than your head
to prevent the mask and air tubing from filling with water.
Leave the humidifier to cool for ten minutes before handling to allow the water to cool and
to make sure that the humidifier is not too hot to touch.
Make sure that the humidifier is empty before transporting the device.
Note: For any serious incidents that occur in relation to this device, these should be reported to
ResMed and the competent authority in your country.
English 17
Technical specifications
Units are expressed in cm H
2
O and hPa. 1 cm H
2
O is equal to 0.98 hPa.
90W power supply unit
AC input range:
100240V, 5060Hz 1.01.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
DC output:
24V 3.75A
Typical power consumption: 53W (57VA)
Peak power consumption:
104W (108VA)
Environmental conditions
Operating temperature:
+5°C to +35°C
Note:
The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the
device remains safe.
Operating humidity:
10 to 95% relative humidity, non-condensing
Operating altitude:
Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
Storage and transport temperature:
-20°C to +60°C
Storage and transport humidity: 5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to
IEC 60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile
communication devices are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com/downloads/devices
Classification: EN60601-1:2006/A1:2013
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Internally located at device outlet, analog gauge pressure
type, 0 to 40 cm H
2
O (0 to 40 hPa)
Flow sensor:
Internally located at device inlet, digital mass flow
type, -70 to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
2
O (30 hPa) for more than 6 sec or 40 cm H
2
O (40 hPa) for more than 1 sec.
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
25 dBA with uncertainty of 2 dBA
Standard:
25 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
33 dBA with uncertainty of 2 dBA
Standard:
33 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
35 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
18
Physical - device and humidifier
Dimensions (H x W x D):
116 mm x 255 mm x 150 mm
Air outlet (complies with ISO 5356-1:2015):
22 mm
Weight (device and cleanable humidifier):
1248 g
Housing construction:
Flame retardant engineering thermoplastic
Water capacity:
To maximum fill line 380 mL
Cleanable humidifier - material:
Injection moulded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate: 68°C
Cut-out:
74°C
Maximum gas temperature:
41°C
Air filter
Standard:
Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Hypoallergenic:
Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used: 4G, 3G, 2G
It is recommended that the device is a minimum distance of 0.8" (2 cm) from the body during operation. Not applicable to
masks, tubes or accessories. Technology may not be available in all regions.
Declaration of Conformity (DoC to the Radio Equipment Directive)
ResMed declares that the AirSense 10 device (models 370xx or 371xx) is in compliance with the essential requirements
and other relevant provisions of Directive 2014/53/EU (RED). A copy of the Declaration of Conformity (DoC) can be found on
Resmed.com/productsupport
The 2G radio equipment operates with the following frequency bands and maximum radio-frequency power:
GSM 850/900: 35dBm
GSM 1800/1900: 32dBm
The 4G device can be used in all European countries without any restrictions.
All ResMed devices are classified as medical devices under the Medical Device Directive. Any labelling of the product and
printed material, showing
0123, relates to the Council Directive 93/42/EEC including the Medical Device Directive
amendment (2007/47/EC).
Operating pressure range
AutoSet, CPAP:
4 to 20 cm H
2
O (4 to 20 hPa)
Supplemental oxygen
Maximum flow:
4 L/min
English 19
Pneumatic flow path
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
Design life
Device, power supply unit: 5 years
Cleanable humidifier:
2.5 years
Air tubing: 6 months
General
The patient is an intended operator.
Humidifier performance
Mask Pressure
cm H
2
O (hPa)
RH output % at 17°C
ambient
temperature
RH output % at 22°C
ambient
temperature
Nominal system output AH
1
, BTPS
2
Setting 4
Setting 8
Setting 4
Setting 8
4
85
100
6
>10
10 85 100 6 >10
20
85
90
6
>10
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
Length
Inner diameter
ClimateLineAir
Flexible plastic and electrical components
2 m
15 mm
ClimateLineAir Oxy
Flexible plastic and electrical components
1.9 m
19 mm
SlimLine
Flexible plastic
1.8 m
15 mm
Standard
Flexible plastic
2 m
19 mm
Heated air tubing temperature cut-out: 41°C
Notes:
The manufacturer reserves the right to change these specifications without notice.
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
Do not use electrically conductive or antistatic air tubing.
The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Range
Display resolution
Pressure sensor at air outlet:
Mask pressure 420 cm H
2
0 (420 hPa) 0.1 cm H
2
O (0.1 hPa)
Flow derived values:
Leak 0–120 L/min 1 L/min
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166

ResMed Airsense 10 Autoset / Elite Guida utente

Tipo
Guida utente