Steris Anesthesia / Respiratory Rack Istruzioni per l'uso

Tipo
Istruzioni per l'uso
OPERATOR MANUAL
Anesthesia/Respiratory Rack – Long
(2022-04-25) 10112981
i
Introduction Operator Manual 10112981
This manual contains important information on the proper use and
routine maintenance of the Anesthesia/Respiratory Rack – Long. All
personnel involved in the use and routine maintenance of this
equipment must carefully review and comply with the Safety
Precautions and instructions contained in this manual. These
instructions are important to protect the health and safety of
personnel using this Anesthesia/Respiratory Rack – Long to process
items in the AMSCO® 7053L Single-Chamber Washer/Disinfector,
and should be retained in a conveniently accessible area for quick
reference.
Advisory IMPORTANT: A listing of the Safety Precautions to be observed
when using and servicing this equipment can be found in SECTION 1
of this manual. Do not use the equipment until you have become
familiar with this information.
STERIS does not intend, recommend or represent in any way that
this Anesthesia/Respiratory Rack Long be used in the AMSCO 7053L
Single-Chamber Washer/Disinfector for the terminal disinfection or
sterilization of any regulated medical device. The Anesthesia/
Respiratory Rack – Long is intended only to perform an initial step
in the processing of soiled, reusable medical devices in the
washer/disinfector. If medical devices will be contacting blood or
compromised tissues, such devices must be terminally processed
in accordance with Good Hospital Practices before each use in
human patients.
STERIS does not recommend processing flexible endoscopes, single-
use instruments, rigid telescopes (optical lens), fiber-optic cables,
multi-part instruments that cannot be disassembled, heat-sensitive
devices and coagulating electrodes using the Anesthesia/Respiratory
Rack – Long in the washer/disinfector. Processing these types of
instruments may result in instrument or equipment damage as well as
improperly cleaned items. All items processed using the Anesthesia/
Respiratory Rack – Long in the washer/disinfector must withstand
temperatures of up to 116°C (240°F).
This Anesthesia/Respiratory Rack – Long is specifically designed to
only hold items as outlined in this manual. If there is any doubt about
a specific material or product, contact the instrument manufacturer
for recommended washing techniques.
Any alteration of this equipment not authorized or performed by
STERIS will void the warranty. Alteration of this equipment which
could adversely affect its operation and efficacy may violate national,
state and local laws or regulations.
Serious incidents that have occurred in relation to this medical
device should be reported to the manufacturer and competent
authority in the country where the incident occurred.
©2022, STERIS Corporation. All rights reserved. Printed in Canada.
A WORD FROM STERIS
ii
10112981 Operator Manual Introduction
Indications For Use The Anesthesia/Respiratory Rack – Long is designed to be used
in the AMSCO 7053L Single-Chamber Washer/Disinfector for the
efficient cleaning, rinsing, drying and intermediate level disinfection
of different types of anesthesia tubing, bags, masks and respiratory
goods.
IMPORTANT: The Anesthesia/Respiratory Rack – Long is designed
to be used only with the Anesthesia/Respiratory Goods cycle.
Service Information A thorough preventive maintenance program is essential to help
ensure safe and proper washer/disinfector and rack operation. This
manual contains a routine maintenance procedure which should be
followed for satisfactory equipment performance.
Customers are encouraged to contact STERIS concerning extended
service maintenance agreements to give your washer/disinfector
and rack planned maintenance, which will help ensure equipment
performance according to factory specifications. STERIS maintains a
global staff of well-equipped, qualified service technicians to provide
this service, as well as professional repair services. Please contact
STERIS for details.
Technical
Specifications
The following specifications apply to the Anesthesia/Respiratory
Rack – Long:
Overall Size: 648 x 825 x 573 mm (25.5 x 32.5 x 22.5")
Weight: 15 kg (33.5 lb)
Material: Stainless steel, Plastic, Silicone, Polypropylene
The Anesthesia/Respiratory Rack – Long includes (refer to Figure 2-1):
Large hollow jets (x 12);
Medium hollow jets (x 12);
Small hollow jets (x 12);
Hooks (x 8);
Spring stoppers (large hollow jets) (x 12);
Helicline for anesthesia tubing (x 1);
Adapters for anesthesia tubing (secured in bottom of rack) (x 8);
Additional adapters for anesthesia tubing (x 8).
Certification This AMSCO 7053L Single-Chamber Washer/Disinfector, compatible
with the Anesthesia/Respiratory Rack - Long, meet the applicable
requirements of certain standards. Refer to washer/disinfector
Operator Manual (10085881) for the current list of standards.
iii
10112981 Operator Manual Introduction
Manufactured by:
STERIS Canada ULC
490 boulevard Armand-Paris
Québec (Québec)
Canada G1C 8A3
The base language of this document is
ENGLISH. Any translation must be made
from the base language document.
European Community Representative:
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865
Ireland
Sales and Service -- Americas, Oceania and Asia:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
800-548-4873
Website:
www.steris.com
Sales and Service -- Europe, Middle-East and Africa:
STERIS
116 Avenue de Magudas, 33185
Le Haillan
FRANCE
Support Service, by email:
Support_EMEA_IPT_SE@steris.com
All other requests, by phone: +33 5 56 93 94 94
iv
10112981 Operator Manual Introduction
Healthcare
Capital Equipment
v
10112981 Operator Manual Table of Contents
TABLE OF CONTENTS
Section
Number Description Page
A WORD FROM STERIS........................................................................................... I
Advisory ........................................................................................................................................................... i
Indications For Use ..........................................................................................................................................ii
Service Information ..........................................................................................................................................ii
Technical Specifications ..................................................................................................................................ii
Certification ......................................................................................................................................................ii
1 SAFETY PRECAUTIONS .....................................................................................1-1
2 LOADING INSTRUCTIONS ..................................................................................2-1
2.1 Anesthesia/Respiratory Rack – Long – Components ............................................................................2-1
2.2 Preparing Instruments ...........................................................................................................................2-2
2.3 Loading Rack ........................................................................................................................................2-2
2.3.1 General ...........................................................................................................................................2-2
2.3.2 DIN Tray .........................................................................................................................................2-3
2.3.3 Helicline, Hollow Jets and Hooks ...................................................................................................2-3
3 OPERATING INSTRUCTIONS .............................................................................3-1
4 UNLOADING INSTRUCTIONS .............................................................................4-1
4.1 Unloading Anesthesia/Respiratory Rack – Long....................................................................................4-1
4.2 Visual Inspection ...................................................................................................................................4-2
5 ROUTINE MAINTENANCE AND REPLACEMENT PARTS ....................................5-1
5.1 Routine Maintenance .............................................................................................................................5-1
5.2 Replacement Parts ................................................................................................................................5-2
vi
10112981 Operator Manual Table of Contents
Number Description Page
2 LOADING INSTRUCTIONS ..................................................................................2-1
2-1 Anesthesia/Respiratory Rack Long – Components ...................................................................2-1
2-2 Example of Anesthesia Items and Respiratory Goods ...................................................................2-2
2-3 Example of DIN Tray Loaded in Rack ............................................................................................2-3
2-4 Example of Anesthesia Items and/or Respiratory Goods on Large Hollow Jets ............................2-3
2-5 Example of Anesthesia Items and/or Respiratory Goods on Medium Hollow Jets ........................2-4
2-6 Example of Anesthesia Items and/or Respiratory Goods on Small Hollow Jets ............................2-4
2-7 Example of Anesthesia Items and/or Respiratory Goods on Hooks ..............................................2-5
2-8 Adapters in Bottom of Rack ...........................................................................................................2-5
2-9 Example of Anesthesia Tubing With Adapters on Helicline ...........................................................2-5
2-10 Example of Loaded Anesthesia/Respiratory Rack – Long..............................................................2-6
LIST OF TABLES
Table
Number Description Page
5 ROUTINE MAINTENANCE AND REPLACEMENT PARTS ....................................5-1
5.1 Replacement Parts ................................................................................................................................5-2
LIST OF ILLUSTRATIONS
Figure
1-1
Safety Precautions Operator Manual 10112981
1
The following Safety Precautions must be observed when operating or servicing the Anesthesia/Respiratory
Rack – Long. WARNING indicates the potential for personal injury and CAUTION indicates the potential for
damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is
important to review ALL Safety Precautions before operating or servicing the equipment.
Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit
and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these Safety Precautions.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING – BURN HAZARD:
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Always wear Personal Protective Equipment (PPE) appropriate to the task to perform.
Always load or unload Anesthesia/Respiratory Rack – Long on appropriate loading cart or surface.
Incorrect rack loading or overloading could lead to injury and/or damage to the equipment or
instruments.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this
equipment. Contact STERIS to schedule preventive maintenance.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Repairs and adjustments performed by unqualified personnel or installation of unauthorized
parts could cause personal injury, result in improper equipment performance, void the warranty or result
in costly damage. Contact STERIS regarding service options.
Anesthesia/Respiratory Rack – Long and items are extremely hot when cycle is completed. Allow rack
and items to cool to room temperature and always wear appropriate Personal Protective Equipment
(PPE) before reaching into chamber.
When cycle is completed, wait until water flow stops before opening door and unloading washer/
disinfector. Hot water or steam may be sprayed through door opening if door is opened too soon. Wear
appropriate Personal Protective Equipment (PPE) whenever reaching into chamber.
To avoid equipment damage and maximize processing efficiency, always load soiled items as instructed
in this Anesthesia/Respiratory Rack – Long Operator Manual.
To prevent voiding the warranty or damaging the equipment, use only authorized STERIS replacement
parts.
Use nonabrasive cleaners when cleaning Anesthesia/Respiratory Rack – Long. Follow directions on
containers and rub in a back-and forth motion (in same direction as surface grain). Abrasive cleaners
will damage stainless steel.
SAFETY PRECAUTIONS
1-2
10112981 Operator Manual Safety Precautions
Healthcare
Capital Equipment
2
2-1
Loading Instructions Operator Manual 10112981
2. 1 Anesthesia/
Respiratory Rack –
Long – Components
See Figure 2-1.
Before using the Anesthesia/Respiratory Rack – Long, it is important
to become familiar with the location of the rack components.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
Bar code tag location.
Use bar code Anesthesia/
Respiratory Goods (05)
Figure 2-1. Anesthesia/Respiratory Rack – Long – Components
Large
Hollow Jets
(x 12)
Small
Hollow Jets
(x 12)
Medium
Hollow Jets
(x 12)
Adapters
(x 8)
Spring
Stoppers
(x 12)
Helicline
Hooks
(x 8)
Additional
Adapters
(x 8)
LOADING INSTRUCTIONS
2-2
10112981 Operator Manual Loading Instructions
2. 2 Preparing
Instruments See Figure 2-2.
Disassemble all anesthesia tubing, bags, masks and respiratory
goods prior to loading them in the Anesthesia/Respiratory Rack –
Long.
Follow device manufacturer’s instructions for use for manually
cleaning items before placing them in the washer.
2. 3 Loading Rack
2.3.1 General Anesthesia items and respiratory goods are loaded on the
Anesthesia/Respiratory Rack – Long components.
A DIN tray can also be used to load smaller anesthesia items and
respiratory goods.
Figure 2-2. Example of Anesthesia Items and
Respiratory Goods
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload Anesthesia/Respiratory
Rack – Long on appropriate
loading cart or surface.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: To avoid equipment
damage and maximize
processing efficiency, always
load soiled items as instructed
in this Anesthesia/Respiratory
Rack – Long Operator Manual.
2-3
Loading Instructions Operator Manual 10112981
2.3.2 DIN Tray See Figure 2-3.
To load the Anesthesia/Respiratory Rack – Long using a DIN tray,
proceed as follows:
1. Pull out DIN tray from rack.
2. Place small anesthesia items and respiratory goods in DIN tray.
IMPORTANT: Do not place any tubing in DIN Tray.
IMPORTANT: Trays should not be overloaded with instruments
to ensure cleaning efficacy.
3. Place DIN tray back in position in rack.
2.3.3 Helicline, Hollow Jets
and Hooks See Figure 2-1 and Figures 2-4 to 2-10.
To load tubing as well as anesthesia items and respiratory goods
using helicline, hollow jets and hooks, proceed as follows:
1. Place desired items on large hollow jets (see Figure 2-4).
2. Spring stoppers can be used to adjust height of items to be
processed (see Figure 2-1).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
Figure 2-3. Example of DIN Tray Loaded in Rack
Connect
Small Tube
to Hollow Jet
using
Irrigation
Hose for
Anesthesia
Washing
Rack - 7053L
accessory
(FD509)
Figure 2-4. Example of Anesthesia Items and/
or Respiratory Goods on Large Hollow Jets
2-4
10112981 Operator Manual Loading Instructions
NOTE: Some items are equipped with small tubes. These small
tubes can be connected to hollow jets to ensure complete
coverage of tubes during the cycle. Ensure other extremity of
each tube is free for evacuation (see Figure 2-4).
3. Place desired items on medium hollow jets (see Figure 2-5).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
4. Place desired items on small hollow jets (see Figure 2-6).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
Figure 2-5. Example of Anesthesia Items and/
or Respiratory Goods on Medium Hollow Jets
Figure 2-6. Example of Anesthesia Items and/
or Respiratory Goods on Small Hollow Jets
2-5
Loading Instructions Operator Manual 10112981
5. Place desired items on hooks (see Figure 2-7).
6. To place desired anesthesia tubing on helicline, proceed as
follows:
a. To obtain maximum coverage during cycle, connect one
extremity of anesthesia tubing on adapter located in bottom
of rack (see Figure 2-8).
b. Place tubing on helicline.
c. To obtain maximum coverage during cycle, insert one of
additional adapters in other extremity of anesthesia tubing
(see Figures 2-8 and 2-9).
Figure 2-7. Example of Anesthesia Items
and/or Respiratory Goods on Hooks
Figure 2-8. Adapters in Bottom of Rack
Figure 2-9. Example of Anesthesia Tubing
with Adapters on Helicline
2-6
10112981 Operator Manual Loading Instructions
7. Anesthesia/Respiratory Rack – Long is loaded and anesthesia
items and respiratory goods are now ready to be processed (see
Figure 2-10).
Refer to AMSCO ® 7053L Single-Chamber Washer/Disinfector
Operator Manual (10085881) for information about processing
Anesthesia/Respiratory Goods cycle.
3
3-1
Operating Instructions Operator Manual 10112981
OPERATING INSTRUCTIONS
For information about the
Anesthesia/Respiratory
Goods Cycle, refer to the
AMSCO®7053L Single-
Chamber Washer/Disinfector
Operator Manual
(10085881)
3-2
10112981 Operator Manual Operating Instructions
Healthcare
Capital Equipment
4
4-1
Unloading Instructions Operator Manual 10112981
4.1 Unloading
Anesthesia/Respiratory
Rack – Long
IMPORTANT: When Anesthesia/Respiratory Goods cycle is
completed, allow load to cool before removing Anesthesia/
Respiratory Rack – Long from washer/disinfector chamber.
To unload Anesthesia/Respiratory Rack – Long, proceed as follows:
1. Remove Anesthesia/Respiratory Rack – Long from washer/
disinfector chamber and place rack on loading cart.
2. If desired, step back to pull loading cart away from washer/
disinfector.
3. Items need to be inspected before removing items from
Anesthesia/Respiratory Rack – Long (see SECTION 4.2, VISUAL
INSPECTION, for more details).
IMPORTANT: STERIS recommends unloading Anesthesia/
Respiratory Rack – Long on an appropriate surface or work table.
4. When inspection is completed and all anesthesia items and
respiratory goods are properly cleaned, carefully empty rack.
5. Slide out DIN tray from bottom of rack.
6. Carefully empty tray.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload Anesthesia/Respiratory
Rack – Long on appropriate
loading cart or surface.
WARNING – BURN HAZARD:
Anesthesia/Respiratory Rack
– Long and items are
extremely hot when cycle is
completed. Allow rack and
items to cool to room
temperature and always
wear appropriate Personal
Protective Equipment (PPE)
before reaching into
chamber.
When cycle is completed,
wait until water flow stops
before opening door and
unloading washer/disinfector.
Hot water or steam may be
sprayed through door
opening if door is opened too
soon. Wear appropriate
Personal Protective Equipment
(PPE) whenever reaching
into chamber.
UNLOADING INSTRUCTIONS
4-2
10112981 Operator Manual Unloading Instructions
4.2 Visual Inspection IMPORTANT: When Anesthesia/Respiratory Goods cycle is
completed, allow load to cool before removing Anesthesia/
Respiratory Rack – Long from washer/disinfector chamber.
IMPORTANT: Good Hospital Practices dictate all instruments be
inspected for visible debris after processing in washer. Any
instrument with visible debris must be reprocessed until clean and
free of visible debris prior to terminal processing. Failure to reprocess
until all visible debris have been removed may impede terminal
processing.
1. Remove Anesthesia/Respiratory Rack – Long from washer/
disinfector chamber and place rack on loading cart.
2. Once items are cool, check each item to detect residues or
debris.
3. Any item that has not been properly cleaned must be
reprocessed in washer/disinfector until clean.
4. See SECTION 4.1, UNLOADING ANESTHESIA/RESPIRATORY RACK
LONG, to properly unload Anesthesia/Respiratory Rack – Long.
IMPORTANT: STERIS recommends unloading Anesthesia/
Respiratory Rack – Long on an appropriate surface or work table.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload Anesthesia/Respiratory
Rack – Long on appropriate
loading cart or surface.
WARNING – BURN HAZARD:
Anesthesia/Respiratory Rack
– Long and items are
extremely hot when cycle is
completed. Allow rack and
items to cool to room
temperature and always
wear appropriate Personal
Protective Equipment (PPE)
before reaching into
chamber.
When cycle is completed,
wait until water flow stops
before opening door and
unloading washer/disinfector.
Hot water or steam may be
sprayed through door
opening if door is opened
too soon. Wear appropriate
Personal Protective Equipment
(PPE) whenever reaching
into chamber.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120