Invacare Kirilos, Atlante Manuale utente

  • Ciao! Sono un chatbot AI specificamente addestrato per aiutarti con il Invacare Kirilos Manuale utente. Ho già esaminato il documento e sono pronto a rispondere alle tue domande in modo chiaro e semplice.
InvacarKirilos/Atlante
Kirilos,AtlanteElectric,AtlanteHydraulic
enMobilePatientLift
UserManual...............................3
frLève-personnemobile
Manueld'utilisation.........................52
itSollevatoremobile
Manualed’uso..............................108
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2017InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibited
withoutpriorwrittenpermissionfromInvacare.Trademarksareidentiedbyand®.Alltrademarks
areownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
Tousdroitsréservés.Larepublication,laduplicationoulamodicationdetoutoupartieduprésent
documentestinterditesansl'accordécritpréalabled'Invacare.Lesmarquescommercialessont
identiéesparet®.Touteslesmarquescommercialessontdétenuesparoucédéessouslicence
àInvacareCorporationousesliales,saufstipulationcontraire.
Tuttiidirittiriservati.Èproibitalariproduzione,laduplicazioneolamodicaparzialeocompleta,
salvopreviaautorizzazionescrittadapartediInvacare.Imarchisonocontrassegnatidae®.Tutti
imarchisonodiproprieolicenzadiInvacareCorporationodisueafliate,salvoindicazione
contraria.
Contents
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
1General.........................................4
1.1Introduction..................................4
1.2Symbolsinthismanual..........................4
1.3Intendeduse..................................4
1.4Scopeofdelivery..............................5
1.5ServiceLife...................................6
1.6WarrantyInformation...........................6
1.7Limitationofliability............................6
2Safety..........................................7
2.1Safetyinformation..............................7
2.2OperatingInformation...........................7
2.2.1General...................................7
2.2.2PinchPointsandPositioning....................8
2.3RadioFrequencyInterference.....................8
2.4Productlabeling...............................8
3Components......................................10
3.1Mainpartsofthelift............................10
3.2Accessories...................................11
4Setup...........................................12
4.1SafeAssembly.................................12
4.2AssemblingtheLift.............................12
4.3DisassemblingtheLift...........................13
5Usage..........................................14
5.1Introduction..................................14
5.2Locking/UnlockingtheRearCasters.................14
5.3Raising/LoweringtheLift.........................14
5.3.1Raising/LoweringanElectricLift.................14
5.3.2Raising/LoweringaHydraulicLift................15
5.4Closing/OpeningLegs...........................16
5.4.1Closing/Openinglegsmanually..................16
5.5PerforminganEmergencyStop....................16
5.6ActivatinganEmergencyRelease...................16
5.7Chargingthebattery............................17
6LiftingthePatient.................................19
6.1SafeLifting...................................19
6.2PreparingtoLift...............................21
6.3AttachingtheSlingtotheLift.....................23
6.4LiftingandTransferringthePatientFromaBed........24
6.4.1FloorTransfers(LiftingfromtheFloor)............28
6.4.2CommodeTransferGuidelines..................30
6.4.3WheelchairTransfer..........................32
7Maintenance.....................................34
7.1MaintenanceandSafetyInspection.................34
7.1.1Safetyinspectionchecklist.....................36
7.2LubricatingtheLift.............................37
7.3CleaningtheSlingandLift........................37
7.4Checkingandtighteningbolts.....................38
7.5MaintainingtheHydraulicPump...................38
8AfterUse........................................39
8.1TransportationandStorage.......................39
8.2Disposal.....................................39
8.3Reuse.......................................39
9Troubleshooting...................................40
9.1Identifyingandrepairingfaults....................40
10Technicaldata...................................42
10.1Dimensionsandweight.........................42
10.2Electricalsystem..............................43
10.3Hydraulicsystem..............................44
10.4Environmentalconditions........................44
10.5Materials....................................44
10.6Electromagneticcompliance(EMC)information.......44
10.7Electromagneticcompliance(EMC).................46
InvacarKirilos/Atlante
1General
1.1Introduction
ThankyouforchoosinganInvacareproduct.
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.Inordertoensuresafetywhen
usingtheproduct,readtheusermanualcarefullyandfollow
thesafetyinstructions.
Pleasenotethattheremaybesectionsinthisusermanual,
whicharenotrelevanttoyourproduct,sincethismanual
appliestoallavailablemodels(onthedateofprinting).If
nototherwisestated,eachsectioninthismanualrefersto
allmodelsoftheproduct.
1.2Symbolsinthismanual
Symbolsandsignalwordsareusedinthismanualandapply
tohazardsorunsafepracticeswhichcouldresultinpersonal
injuryorpropertydamage.Seetheinformationbelowfor
denitionsofthesignalwords.
WARNING
Indicatesahazardoussituationthatcouldresult
inseriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresult
inminororslightinjuryifitisnotavoided.
IMPORTANT
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
TipsandRecommendations
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.Thelaunchdate
ofthisproductisstatedintheCEdeclaration
ofconformity.
Manufactureroftheproduct.
1.3Intendeduse
Mobilepatientliftsareindoortransferdevices,designed
tobeusedinmostofthecommonliftingsituationsin
hospitals,nursingfacilitiesanddomesticareas,forexample:
Betweenthebedandawheelchair
Toandfromthetoilet
Loweringandraisingpatientsto/fromtheoor
Themobilepatientliftcanbeusedtotransferandposition
completelyorpartiallyimmobilepatients,whocannotbe
transferredwithothertypesofliftsortransferaids.All
positionchangesarepossiblewithoutassistanceofthe
patient.Themobilepatientliftisonlyintendedtolift
patientsuptothemaximumweightlimitstatedintechnical
data.Therearenoknowncontraindicationsforthisproduct.
Selectingtheappropriateslingsandaccessoriesforeach
individualisimportanttoassuresafetywhenusingapatient
lift.SeeInvacare’sslingandaccessoryusermanualsfor
furtherinformationonthosedevices.
4
1551341-E
General
Invacarerecommendsthatthepatientbetransferredtoa
showerchairorothermeansforbathing.
Themobilepatientliftcanbeturned(rotated)inplacefor
transferswithlimitedoorspace.
1.4Scopeofdelivery
Theitemslistedinthetableareincludedinyourpackage.
A
Mastassemblyincludingboom,hanger
barandactuatororhydraulicpump(1
piece)
B
Battery(1piece,KirilosandAtlante
Electriconly)
C
Handcontrol(1piece,Kirilosand
AtlanteElectriconly)
D
Legspreaderhandle(1piece)
E
Basewithlegs(1piece)
F-notshown
Cableforcharging(1piece,Kirilosand
AtlanteElectriconly)
G-notshown
Sling(1piece)
H-notshown
UserManual(1piece)
1551341-E5
InvacarKirilos/Atlante
1.5ServiceLife
WARNING!
RiskofInjuryorDamage
Theproducthasbeentestedfortheservicelife
statedinthismanual.Useoftheproductbeyond
thistimeperiodmaycauseproductdamageand
injury.
Onlyusetheproductfortheservicelifestated
inthismanual.DONOTexceedtheservicelife
oftheproduct.
Removetheproductfromservicewhenthe
servicelifehasbeenmet.DONOTcontinueto
usetheproduct.
Performallmaintenanceaccordingtothe
recommendedscheduleinthismanual.
Theproducthasanexpectedlifetimeof8yearswhenused
inaccordancewithsafetyinstructions,maintenanceintervals
andcorrectuse,statedinthismanual.Theeffectiveproduct
lifetimecanvaryaccordingtofrequencyandintensityofuse.
ActuatorServiceLife
Numberofliftsperday
Servicelifeofactuator
(inyears)
1–210
39
46
55
64
Numberofliftsperday
Servicelifeofactuator
(inyears)
7–93
10–132
14–271
1.6WarrantyInformation
Termsandconditionsofthewarrantyarepartofthegeneral
termsandconditionsparticulartotheindividualcountriesin
whichthisproductissold.
ContactinformationforyourlocalInvacareofceislocated
insidethebackcoverofthismanual.
1.7Limitationofliability
Invacareacceptsnoliabilityfordamagearisingfrom:
Non-compliancewiththeusermanual
Incorrectuse
Naturalwearandtear
Incorrectassemblyorset-upbythepurchaserora
thirdparty
Technicalmodications
Unauthorizedmodicationsand/oruseofunsuitable
spareparts
61551341-E
Safety
2Safety
2.1Safetyinformation
WARNING!
Donotusethisproductoranyavailable
optionalequipmentwithoutrstcompletely
readingandunderstandingtheseinstructions
andanyadditionalinstructionalmaterialsuch
asusermanuals,servicemanualsorinstruction
sheetssuppliedwiththisproductoroptional
equipment.Ifyouareunabletounderstand
thewarnings,cautionsorinstructions,contact
ahealthcareprofessional,dealerortechnical
personnelbeforeattemptingtousethis
equipment-otherwise,injuryordamagemay
occur.
Theinformationcontainedinthisdocumentis
subjecttochangewithoutnotice.
Checkallpartsforshippingdamagebeforeusing.Incaseof
damage,Donotusetheequipment.Contactthedealeror
Invacarerepresentativeforfurtherinstructions.
2.2OperatingInformation
Thissectionofthemanualcontainsgeneralsafety
informationaboutyourproduct.Forspecicsafety
information,refertotheappropriatesectionofthemanual
andprocedureswithinthatsection.Forinstance,forsafety
informationrelatedtoassemblingthelift,refertotheSetup
sectionofthemanual.
2.2.1General
WARNING!
RiskofFalling
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction
withInvacarepatientlifts.
WARNING!
RiskofFalling
Donotattemptanytransferwithoutapproval
ofthepatientsphysician,nurseormedical
assistant.Thoroughlyreadtheinstructionsin
thisUserManual,observeatrainedteamof
expertsperformtheliftingproceduresandthen
performtheentireliftprocedureseveraltimes
withpropersupervisionandacapableindividual
actingasapatient.
Usecommonsenseinalllifts.Specialcare
mustbetakenwithpeoplewithdisabilities
whocannotcooperatewhilebeinglifted.
Usesteeringhandleonthemastatalltimesto
pushorpullthepatientlift.
Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to
ensurethatitisproperlyattachedbeforethe
patientisremovedfromastationaryobject
(bed,chairorcommode).
1551341-E
7
InvacarKirilos/Atlante
WARNING!
Thepatientliftcanbeusedindoorsonalevel
surface.Ifthepatientliftisusedintheareaof
ashowerorbath,ensurethatthepatientliftis
wipedcleanofanymoistureafteruse.
Periodicallyinspectallcomponentsofthe
patientliftforsignsofcorrosion.Replaceall
partsthatarecorrodedordamaged.
2.2.2PinchPointsandPositioning
WARNING!
RiskofInjury
Pinchpointsarepresentinseverallocationson
theliftandngerscouldbepinched.
Thehangerbarcanmovesuddenlyandcause
injury.
Alwayskeephandsandngersclearofmoving
parts.
Whenpositioninglift,beawareoftheposition
ofthehangerbarandthepatient.
2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
MostelectronicequipmentisinuencedbyRadio
FrequencyInterference(RFI).
CAUTIONshouldbeexercisedwithregardtothe
useofportablecommunicationequipmentinthe
areaaroundsuchequipment,otherwiseinjury
ordamagemayoccur.
IfRFIcauseserraticbehavior:
PUSHtheRedPowerSwitchOFFIMMEDIATELY .
DONOTturnthePowerSwitchONwhile
transmissionisinprogress.
2.4Productlabeling
Labellocation
Theproductlabelisplacedonthemastofthepatient
liftandcontainsthemainproductinformation,including
technicaldata.
Product name
YYYY-MM
= xxx kg
Invacare Portugal, Lda
Rua Est rada Ve lha 949
4465-784 Leça do Balio, Portug al
XXXXXX XXXXX_XX
XXXXXXXXXXXXX
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
Xxxxxxx xxxxx xxx
ISO 10535
1590833-A
81551341-E
Safety
Symbolsontheproductlabel
Manufactureraddress
Dateofmanufacture
TypeReferenceNumber
SerialNumber
SafeWorkingLoad
ClassIIequipment
(KirilosandAtlanteElectriconly)
TypeBappliedpart
(KirilosandAtlanteElectriconly)
ThisproductcomplieswithDirective
93/42/EECconcerningmedicaldevices.The
launchdateofthisproductisstatedinthe
ECdeclarationofconformity.
WEEEconform.
(KirilosandAtlanteElectriconly)
ReadManual
(AtlanteHydrauliconly)
Otherlabelsandsymbols
ReadManual
(KirilosandAtlanteElectriconly)
1551341-E9
InvacarKirilos/Atlante
3Components
3.1Mainpartsofthelift
A
Base
B
Leg
C
Caster
D
Rearcasterwithbrake
E
Boom
F
Mast
G
Hangerbar
H
Hookforsling
I
Actuator(KirilosandAtlanteElectriconly)
J
Controlunitwithbattery(KirilosandAtlanteElectric
only)
K
EmergencyStop(KirilosandAtlanteElectriconly)
L
Steeringhandle
M
Handcontrol(KirilosandAtlanteElectriconly)
N
Legspreaderhandle
OHydraulicpumpreplacesactuatorIforAtlante
Hydraulic
P
Manualemergencyloweringdevice(Kirilosand
AtlanteElectriconly)
101551341-E
Components
3.2Accessories
CAUTION!
Compatibilityofslingsandhangerbars/strap
hooks
Invacare®usesa"LoopandCoatHanger
Bar/StrapHookSystem"asdomanyother
manufacturers.Thereforeothersuitablepatient
transfersystems(slings),manufacturedbyother
companies,canbeusedontheInvacarepatient
liftrangeaswell.
Howeverwedorecommend:
Ariskassessmentisalwaystobecarriedoutby
aprofessionalpriortoissuingliftingequipment.
ItisimportantthattheTask,Individual,Load,
EnvironmentandEquipmentareconsideredin
theriskassessment.
Alwayschoosetheslingdesignandsize
accordingtothepatient’sweight,sizeand
physicalabilitywhilstconsideringthetypeof
transfertobecarriedout.
Doonlyuseslingsthataresuitablefora"Loop
andCoatHangerBar/StrapHookSystem".
Donotuseslingsfor"KeyholeHangerBar"or
for"TiltingFrameHangerBar"designs.
AvailableAccessories
Slingmodelsfor"LoopandCoatHangerBarSystem":
Fullbodysupportslingswithoutheadsupport
Fullbodysupportslings–withheadsupport
Slingsfordress/toiletingwithorwithoutheadsupport
Slingsforamputee
1551341-E
11
InvacarKirilos/Atlante
4Setup
4.1SafeAssembly
WARNING!
RiskofInjury
Improperassemblymaycauseinjuryordamage.
AssemblyMUSTbeperformedonlybyqualied
personnel.
UseonlyInvacarepartsintheassemblyof
thispatientlift.Thebaselegs,themast,
boom,pumpassemblyandthehangerbar
aremanufacturedtospecicationsthatassure
correctalignmentofallpartsforsafefunctional
operation.
DONOTovertightenthemountinghardware.
Thiswilldamagethemountingbracket.
Ifthereareanyissuesorquestionsduringassembly,
contactalocalInvacarerepresentative.Refertothe
contactinformationonthelastpageinthismanual.
4.2AssemblingtheLift
WARNING!
Checkallpartsforvisibledefectsordamage
beforeassembly.Incaseofanydamage,do
notusetheproductandcontactInvacare.
Makesuretheemergencystopisactivated
beforeassemblyordisassembly.
Takecarewhenliftingcomponentsduring
assembly.Somepartsareheavy.Always
remembertoadoptthecorrectliftingposition.
Assemblingthemasttothebase
12
1551341-E
Setup
1.PutthebaseAonalevelsurfaceandlocktherear
castorsB.
2.LoosenthemastknobC,butleaveitscrewedintothe
baseA.
3.InsertthetubeofthemastDintotheholeofthebase
A.
4.WhilesupportingthemastassemblyD,tightenthe
mastknobC.
5.PlacethelegspreaderhandleEontoitssupportF
andpushthespringpin,tolockthehandleEonits
supportF.
Installingtheactuatororhydraulicpumptotheboom
2x17mmwrench
1.LoosenthenutAandremoveboltBfromtheboom
mountingbracketC.
2.PlacetheactuatorDorhydraulicpumpEintheboom
mountingbracketCandaligntheholes.
3.ReinstalltheboltBandtightenthenutA.
Wiring
(KirilosandAtlanteElectriconly)
1.PlugthehandcontrolcableonsocketAandmotor
cableonsocketBcontrolunit.
2.PlacethebatteryConthecontrolunit.Pushthetop
ofthebatteryagainstthemountingbracketuntilthere
isanaudibleclick.
4.3DisassemblingtheLift
1.Putthebaseonalevelsurfacesothatallcastorsmake
contactwiththeoor.
2.Locktherearcastors.
3.Reversetherelevantstepsoftheassemblingprocedure
foryourmodel.
1551341-E13
InvacarKirilos/Atlante
5Usage
5.1Introduction
Theoperationofthepatientliftisaneasyandsafe
procedure.
Beforeusingtheliftwithapatient,refertothe
followingproceduresforsafetyinformationand
instruction:
OperatingInformationintheSafetysection
LiftingandTransferringthePatientintheLifting
thePatientsection
5.2Locking/UnlockingtheRearCasters
A
B
1.ToLocktheCasterA:PresstabBdown.
2.ToUnlocktheCasterA:PresstabBup.
5.3Raising/LoweringtheLift
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
Invacaredoesrecommendthattherearcasters
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedor
anystationaryobject.
5.3.1Raising/LoweringanElectricLift
Thehandcontrolisusedtoraiseorlowerthelift.
1.ToraisetheliftPress
andholdtheUPAbutton
toraisetheboomandthe
patient.
2.TolowertheliftPress
andholdtheDOWNB
buttontolowertheboom
andthepatient.
Releasethebuttontostopraisingorloweringthelift.
14
1551341-E
Usage
5.3.2Raising/LoweringaHydraulicLift
Thepumphandleandcontrolvalveareusedtoraiseor
lowerthehydrauliclift.
RefertothesafetyinformationintheIntroductionof
theUsagesectionbeforeperformingthisprocedure.
Therearetwocontrolsonthepumpassembly:
ThecontrolvalveA
ThepumphandleB
Toraisethelift:
1.TurnthecontrolvalveAclockwisetotheCLOSED
position.
2.MovethepumphandleBupanddowntoelevatethe
boomC.
Tolowerthelift:
1.TurnthecontrolvalveAcounterclockwisetotheOPEN
position.
2.TheweightofthepatientintheslingDwillcausethe
boomtodescend.
3.Ifthereisnopatientinthesling,pulldownonthe
boomC.
Therateofdescentoftheboomiscontrolledbythe
amountthatthecontrolvalveisopen.
Asafetygateispartofthehydraulicsystem
thatcontrolsthemaximumdescentoftheboom
regardlessofhowfarthecontrolvalveisopened.
1551341-E15
InvacarKirilos/Atlante
5.4Closing/OpeningLegs
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegs
oftheliftonlyaslongasittakestoposition
theliftoverthepatientandliftthepatientoff
thesurfaceofthebed.Whenthelegsofthe
liftarenolongerunderthebed,returnthe
legsofthelifttothemaximumopenposition.
5.4.1Closing/Openinglegsmanually
A
1.Toopenthelegs,pullthelegspreaderhandleAoffand
thentotheleftuntilitlocksinthenotchofthebracket.
2.Toclosethelegs,pushthelegspreaderhandleAoff
andthentotherightuntilitlocksinthenotchofthe
bracket.
TheshifterhandleMUSTlockintoitsmountingslot
tolockthelegsinthefullclosedposition.
WARNING!
RiskofInjury
Theliftcouldtipandendangerthepatientand
assistants.
IfthelegspreaderhandleisNOTpositioned
completelyintoitsmountingslot,DONOTuse
thepatientliftuntilthelegspreaderhandle
isproperlyseatedandthelegsofthepatient
liftarelockedinplace.
5.5PerforminganEmergencyStop
(KirilosandAtlanteElectriconly)
1.PresstheredemergencybuttonAonthecontrolunit
tostoptheboomandpatientfromraisingorlowering.
2.Toreset,rotatetheemergencybuttonclockwise.
5.6ActivatinganEmergencyRelease
Electricalemergencyreleasewithcontrolunit
(KirilosandAtlanteElectriconly)
161551341-E
Usage
1.InsertapenintotheholelabeledEmergencyDownA
onthecontrolunitB
Secondary(manual)emergencyreleasewithactuator
(KirilosandAtlanteElectriconly)
Incaseofpartialortotalpowerfailure,orifthebattery
runsdownwhileusingthelift,itisequippedwithamanual
emergencyreleasesystemlocatedatthebottomofthe
actuator.
Itisrecommendedthattheprimaryemergency
releasebeused.Thesecondary(manual)emergency
releaseisonlyaback-uptotheprimaryemergency
release.
1.RotatetheemergencyreleaseknobAcounterclockwise.
Theboomwilllowerslowlywithoutadditional
downwardforceontheboom.
Emergencyreleasewithhydraulicpump
(AtlanteHydrauliconly)
1.TurnthecontrolvalveAcounterclockwisetotheOPEN
position.
5.7Chargingthebattery
(KirilosandAtlanteElectriconly)
1551341-E
17
InvacarKirilos/Atlante
IMPORTANT!
Makesuretheemergencystopisnotactivated
whilechargingthebattery.
Donotusetheliftwhilechargingthebattery.
Makesurethatchargingtakesplaceinaroom
withgoodairventilation.
Donotuseormovetheliftwithoutunplugging
fromthewalloutletaftercharging.
Donotattempttousetheliftifthebattery
housingisdamaged.
Replaceadamagedbatteryhousingbefore
furtheruse.
Itisrecommendedtochargethebatterydailytoensure
optimaluseoftheliftandprolongthelifeofthebattery.
Furthermore,itisrecommendedtochargethebattery
beforerstuse.
Controlunit
Thecontrolunitisequippedwithasoundsignal.Abeep
indicatesthatthebatteryhaslowcapacity,butloweringthe
patientisstillpossible.Itisrecommendedtochargethe
batteriesassoonasthesoundsignalisheard.
1.PlugthecableforchargingCintoapoweroutlet.
Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhen
thebatteriesarefullycharged.
TherightyellowchargediodeAwillcontinuous
lightduringchargingandswitchoffwhenfully
charged.
TheleftgreendiodeBwilllightcontinuously
whilethecontrolunitisconnectedtothemains,
andlightsupwhenanybuttononthehand
controlispressed.
2.Disconnectthepowercordfromthepoweroutletafter
thebatteryhasbeenfullycharged.
181551341-E
LiftingthePatient
6LiftingthePatient
6.1SafeLifting
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
Theliftcouldtipandendangerthepatientand
assistants.
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
WheelchairandbedwheellocksMUSTbein
alockedpositionbeforeloweringthepatient
ontoorliftingthepatientoffofthewheelchair
orbedtopreventthewheelchairorbedfrom
movingduringtransfer.
Beforetransferring,checkthatthewheelchair
(bed,commodeorothersurface)weight
capacitycanwithstandthepatient’sweight.
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
Theliftcouldtipandendangerthepatientand
assistants.
TheInvacaremobilepatientliftisNOTatransport
device.Itisintendedtotransferanindividual
fromonerestingsurfacetoanother(suchasa
bedtoawheelchair).
Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety.
Ifitisnecessarytoclosethelegsoftheliftto
maneuvertheliftunderabed,closethelegs
oftheliftonlyaslongasittakestoposition
theliftoverthepatientandliftthepatientoff
thesurfaceofthebed.Whenthelegsofthe
liftarenolongerunderthebed,returnthe
legsofthelifttothemaximumopenposition.
Invacarerecommendslockingtherearcasters
oftheliftONLYwhenpositioningorremoving
theslingfromaroundthepatient.
Invacarerecommendsthattherearcasters
beleftunlockedduringliftingproceduresto
allowthepatientlifttostabilizeitselfwhenthe
patientisinitiallyliftedfromachair,bedor
anystationaryobject.
1551341-E19
InvacarKirilos/Atlante
WARNING!
RiskofDeath,InjuryorDamage
Improperuseofthisproductmaycausedeath,
injuryordamage.
Usethesteeringhandleonthemastatall
timestopushorpulltheproduct.
Avoidusingtheproductonanincline.Invacare
recommendsthattheproductonlybeusedon
aatsurface.
Duringtransfer,withthepatientsuspended
inaslingattachedtothelift,DONOTrolllift
overunevensurfacesthatcouldcausethelift
totipover.
WARNING!
RiskofInjuryorDamage
Damagetopartsofthelift(pendant,casters,
etc.)causedbyimpactwiththeoor,wallsor
otherstationaryobjectsmaycausedamageto
theproductandleadtoinjury.
DONOTallowpartsofthelifttoimpactthe
oor,wallsorotherstationaryobjects.
ALWAYSstorethependantproperlywhennot
inuse.
WARNING!
RiskofDeath
Thependantcordcancauseinjuryifimproperly
positionedandsecured.
ALWAYSbeawareofthelocationofthe
pendantcordinrelationtothepatientand
caregivers.
DONOTallowthependantcordtobecome
entangledaroundthepatientandcaregivers.
Thependantmustbesecuredproperly.ALWAYS
storethependantproperlywhennotinuse.
WARNING!
RiskofEntrapmentorStrangulation
Itemsinthepatientssurroundingscancause
entrapmentstrangulationduringlifting.Toavoid
entrapmentorstrangulation:
Beforelifting,checkthatthepatientis
completelyfreeofhis/hersurroundings.
WARNING!
RiskofEntrapment
Thereisariskofentrapmentbetweenthehanger
barhooksandthesling.
Usecautionwhenlifting.
NEVERputhandsorngersonornearthe
hookswhenlifting.
Ensurethepatientshandsandngersareaway
fromthehooksbeforelifting.
201551341-E
1/164