Operation
OPERATING HAY TESTER (MOISTURE TEST)
1. Insert probe tip (A) into hay so both
brass pieces are in contact with
hay. See Hay Testing Guidelines
for probe information and hay
preparation.
2. Press and release the “Moisture
%” button (B).The tester will
then display a moisture readings
from 14.5% - 45%, with the most
accurate readings between 15%
and 30%.
IMPORTANT: Take readings in at least five
(5) places in the windrow or bale and use
the HIGHEST reading as a guideline. DO
NOT AVERAGE.
High Moisture Readings (above 30%):
Readings over 30% should only be used as
a qualitative indication of very high moisture.
According to most local Agricultural extension agents, hay at
more than 20% moisture should not be baled or stored without
preservative. Hay at more than 25% moisture should not be
baled or stored.
IMPORTANT: If one reading is over the recommended moisture
content, do not bale.
Low Moisture Readings (under 14.5%): Hay tester will not test below
14.5%. If a hay bale being tested is 12% moisture, the tester will still
read at <14.5% (+/–0.3%). If you suspect that the hay being tested is
very low moisture, take a reading of the air before probing the bale.
If the display does not change from the air reading (<14.5%), then the
bale moisture is below the limits of the tester.
10-4
NOTE:
NOTE: When you turn the unit on, it will display the
software version h1.4 or d1.
NOTE - Unit stays on and will shut off after 3 minutes
of no use.
If both temperature and moisture % switches
are pressed simultaneously, the tester will not
be harmed but a meaningless number
will be displayed.
NOTE: