Salter TE-150-EU Manuale del proprietario

Categoria
Termometri digitali per il corpo
Tipo
Manuale del proprietario

Questo manuale è adatto anche per

TE-150-EU
INFRARED EAR THERMOMETER
Instructions and Guarantee
This equipment needs to be installed and put into service in accordance
with the information provided in this manual
2 | GB
GB | 3
INTRODUCTION
This Thermometer is intended to be used as the intermittent measurement of human
body temperature for people of all ages in the home environment.
Utilizing infrared technology, this thermometer takes temperatures in seconds by
measuring heat generated by the ear. This product conforms to the provisions of the EC
directive MDD(93/42/EEC). Its advantages include:
1. Ear measurement
2. Probe cover not required
3. Waterproof probe design
4. One-second reading
5. Auto power o for power saving
6. Low-battery Indicator
7. Warning indication - indications for battery condition and measuring range.
8. Jumbo LCD Display
9. Memory function
IMPORTANT INFORMATION BEFORE USE
When using this product, please be sure to follow all the notes listed below. Any action
against these notices may cause injury or aect the accuracy.
1. Do not disassemble, repair, or remodel the thermometer.
2. Be sure to clean the thermometer lens each time after usage.
3. Avoid direct finger contact with the lens.
4. No modification of this equipment is allowed.
5. It is recommended that user may take 3 temperatures. If they are dierent, use
the highest reading.
6. Do not expose the thermometer to extreme temperature, very high humidity, or
direct sunlight.
7. Avoid extreme shock or dropping the device.
8. Before the measurement, users and thermometer should stay in steady state room
condition for at least 30 minutes.
9. Avoid measuring temperature for 30 minutes after exercise, bathing, or returning
from outdoors.
Probe tip
2 | GB
GB | 3
10. To protect the environment, dispose of empty batteries at appropriate collection
sites according to national or local regulations.
11. It is ill-advised to disassemble the thermometer.
12. Please use the thermometer solely for its intended purpose.
13. Carefully hold the device when in use to avoid dropping the device.
14. Allow one minute between successive measurements as slight variations may
occur if measurements are taken over a short period of time. Use average
temperatures instead.
15. There are no absolute body temperature standards. Keep reliable records of your
personal temperature to serve as a reference for judging a fever.
16. Under any circumstances, the temperature taking result is ONLY for reference.
Before taking any medical action, please consult your doctor.
17. It is recommended calibrate the device every 1 year.
PRODUCT IDENTIFICATION
Memory Recall button
Battery compartment
lid
Probe tip
Power & Measure
button
LCD display
4 | GB
GB | 5
DESCRIPTION OF LCD DISPLAY
TIPS FOR MEASURING HUMAN TEMPERATURE
Bear in mind that the thermometer needs to have been in the room in which the
measurement is taken for at least 30 minutes before use.
NOTE:
Some people produce dierent readings in their left and right ear. In order to record
temperature changes, always measure a person’s temperature in the same ear.
The ear thermometer may be used by children only under adult supervision.
Measurement is usually possible over the age of 6 months. In infants under 6 months,
the ear canal is still very narrow so the temperature of the eardrum often cannot be
recorded and the result displayed is often too low.
The measurement must not be taken in an ear aected by inflammatory diseases
(e.g. discharging pus or secretion), after possible ear injuries (e.g. eardrum damage)
or in the healing period after operative procedures. In all of these cases, please
consult your doctor.
Use of the thermometer on dierent persons can be inappropriate in the event of
certain acute infectious diseases because of the possible spread of germs despite
cleaning and disinfection. If you have any doubts, please consult your doctor.
This thermometer may only be used without a disposable protective cover.
If you have been lying on one ear for some time, the temperature is slightly raised.
Wait 60 seconds or measure in the other ear.
As ear wax can aect the measurement, you should clean the ear before measuring
if necessary.
Temperature in degrees
Celcius scale
Fahrenheit scale
Scan in progressFrowning face
Measurement
≥38.0°C(≥ 100.4°F)
Fever
Smiling face
Measurement
<38.0°C(< 100.4°F)
No Fever
Low battery warning
Memory symbol and sets of records
4 | GB
GB | 5
MEASURING BODY TEMPERATURE IN THE EAR
1. Press the Power & Measure Button for 1 second to switch the thermometer on. All
screen data will display on screen. Then wait for beep sound to let you know the
appliance is ready and followed by last temperature reading.
2. Make sure that the sensor tip and also the ear canal
are clean. As the ear canal is slightly curved, you
have to pull the ear slightly up and backwards before
inserting the sensor tip. This is important so that the
sensor tip can be pointed directly at the eardrum.
Under 1 year
Have child lay flat with the head sideways, so that the ear is facing
upwards. Gently pull the ear straight back.
3. Insert the probe into the ear canal (the probe must insert well to ensure an
accurate temperature measurement), then press the Power & Measure Button for
about 1 second, release the button and you will hear 1 short beep sound means
complete the measurement.
4. You may take out the device to read the measurement result.
5. The backlight display will remain light for approximately 5 seconds each time the
Power & Measure Button is pressed, after each reading is completed, and each
time the Memory button is pressed to recall a previous reading.
6. The thermometer will automatically shut o after one minute without use.
1 year +
Stand behind and slightly to the side of the child/adult. Gently pull the
ear up and back to straighten the ear canal.
6 | GB
GB | 7
Note:
7. To ensure accuracy, please wait at least 1 minute between successive readings.
8. Clean the thermometer lens again after use. After cleaning, wait at least
10 minutes between readings, to ensure accuracy.
CHANGING THE MEASUREMENT UNIT (ºC/ºF)
Changing °C/ °F, please make sure the device is in power on condition. Press the Power &
Measure and Memory Recall buttons together to toggle between °C and °F.
MEMORY FUNCTION
You can recall up to 10 stored measurements in memory to share with your doctor or
trained healthcare professional.
1. Press the Memory Recall button. The first
reading displayed is the latest measurement
stored in memory.
2. Continue to press the Memory Recall button
button to view the next previously stored
measurement.
3. Any new measurement will be recorded and the
oldest memory deleted without you having to do
anything.
The readings
If the temperature measurement is below 38°C
(100.4°C), a smiling face will be appear next
to the reading.
If the reading is 38°C (100.4°C) or above, a
frowning face will be displayed.
6 | GB
GB | 7
CLEANING AND DISINFECTING
For home use device disinfection, 70% Ethanol or Isopropyl
alcohol (available in the pharmacy) can be used.
Clean the measuring sensor after each use.
Use a clean cloth or cotton bud that can be moistened with
70 % alcohol.
Lens/ Measurement Sensor
Gently clean with an alcohol swab.
Do not use water to wash the thermometer lens directly.
Thermometer
Clean with a soft, dry cloth.
Do not use water to rinse the device.
APPLIED STANDARDS
This product conforms to the provisions of the EC directive MDD(93/ 42/ EEC).
The following standards apply to design and/or manufacture of the products:
ISO 80601-2-56
Medical electrical equipment -- Part 2-56: Particular requirements for basic safety and
essential performance of clinical thermometers for body temperature measurement
IEC/EN 60601-1
Medical electrical equipment- Part 1: General requirement for safety
IEC/EN 60601-1-2
Medical electrical equipment- Part 2: Collateral standard: Electromagnetic compatibility
- Requirements and tests
IEC/EN 60601-1-11
Medical electrical equipment - Part 1-11: General requirements for basic safety and
essential performance - Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment
8 | GB
GB | 9
BATTERY INSTALLATION
Low battery warning
When the batteries get weak, the battery warning symbol
appears . It is still possible to measure temperature.
The batteries must be replaced. When the battery symbol
flashes and Lo appears in the display, the batteries must be replaced. If the
batteries are too flat, then the thermometer will switch o automatically.
NOTE:
It is recommended to remove the batteries if the unit will not be used for an extended
period of time.
Please store batteries away from small children and heat.
Do not use rechargeable batteries.
Dispose of used batteries in accordance with the applicable legal regulations.
Never dispose of batteries in the normal household waste.
Replacing the Battery
1. Use a Phillips head screwdriver to loosen battery cover screw. Remove the battery
cover.
2. Remove used battery.
3. Replace with a lithium 3V CR2032 battery in designated area. The + sign on the
battery should be facing upwards.
4. Place the battery cover on the thermometer and tighten the screw to secure it
in place.
8 | GB
GB | 9
CLINICAL ACCURACY VALIDATION METHOD
This device is an adjusted mode clinical thermometer
the validated information for clinical accuracy in each adjusted mode are:
GroupA1: ∆cb=-0.01°C, L
A
=0.18, δr = ±0.08°C
GroupA2: ∆cb= 0.06°C, L
A
=0.22, δr = ±0.08°C
GroupB : ∆cb=-0.01°C, L
A
=0.20, δr = ±0.07°C
GroupC : ∆cb=-0.01°C, L
A
=0.18, δr = ±0.07°C
∆cb: CLINICAL BIAS
L
A
: LlMITS OF AGREEMENT
δr: CLINICAL REPEATABILITY
ERROR CODES
When a malfunction or incorrect temperature measurement occurs, an error message will
appear as described below.
LCD Display Cause Solution
The temperature measured
is lower than 34°C (93.2°F)
Operate the thermometer only
between the specified temperature
ranges. If necessary, clean the
sensor tip. In the event of a
repeated error message, contact
your retailer or Customer Services.
The temperature measured
is higher than 43°C
(109.4°F)
The operating temperature
is not in the range
15°C~35°C (59°F~95°F)
Operate the thermometer only
between the specified temperature
ranges.
10 | GB
GB | 11
TECHNICAL SPECIFICATION
Measuring range:
Human Body: 34°C ~43°C (93.2°F ~ 109.4°F)
Laboratory Accuracy:
34°C ~43°C±0.2°C (93.2°F ~ 109.4°F ±0.4°F)
Measuring site: Ear
Reference body site: Oral
Calibration Accuracy:
±0.2°C(±0.4°F): from 35°C ~42°C (95°F ~ 107.6°F)
±0.3°C(±0.5°F): out of range
Display resolution: 0.1°C/°F
Operating environment:
15°C~35°C (59°F ~ 95°F) with relative humidity up to 95% (non condensing)
Storage/transportation environment:
-25 to 55 °C(-13 to 131 °F) with relative humidity up to 95% (non condensing)
Power supply: A lithium 3V CR2032 battery
Memory: 10 sets
Weight: approx. 49g (with battery)
Dimensions: approx. 112.8mm×34.4mm×44.5mm (L×W×H)
Shelf life: 3 years
EMC TABLES
Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information.
1. Electromagnetic/Radio Frequency Interference (EMC/RFI): Readings may be
aected if the unit is operated within a radio frequency electromagnetic
field strength of approximately 3 volts per metre, but the performance of the
instrument will not be permanently aected. Care should be taken to keep the
thermometer at least 15cm/6 inches away from R/C transmitters to avoid radio
frequency interference.
2. Avoid keeping the thermometer too close to objects that continuously generate
high heat (like a hot plate) for long periods of time, which can cause overheating
of the thermometer.
10 | GB
GB | 11
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic emissions
The Ear Thermometer is intended for use in the electromagnetic environment
specified below. The customer or the user of the Ear Thermometer should assure that
it is used in such an environment.
Emissions test Compliance
Electromagnetic environment
guidance
RF emissions CISPR 11 CISPR 11
The Ear Thermometer uses RF energy
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B
The Ear Thermometer is suitable
for use in all establishments other
that domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
Flicker emissions
IEC 61000-3-3
Not applicable
12 | GB
GB | 13
Table 2 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The Ear Thermometer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ear Thermometer should assure that it is used in such an
environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV
contact
±8kV air
Floor should be wood, concrete or
ceramic tile. If floors are covered
with shythetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV air for input/
output lines
Not
applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC
61000-4-5
±1kV dierential
mode
±2kV common mode
Not
applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip
in UT for 0.5 cycle
40% UT (60% dip in
UT for 5 cycles
70% UT (30% dip in
UT for 25 cycles
5% UT (>95% dip in
UT for 5 sec
Not
applicable
Mains power quality should be that
of a typical commercial or hospital
environment. If the use of the Ear
Thermometer requires continued
operation during power mains
interruptions, it is recommended that
the Ear Thermometer be powered
from an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
12 | GB
GB | 13
Table 3 For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration-electromagnetic immunity
The Ear Thermometer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ear Thermometer should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz to
80 MHz
3V/m
80MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Ear Thermometer, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance.
d=1.2
d=1.2 80 MHz to 800 MHZ
d=2.3 800 MHz to 2,5 Ghz
where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is
the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.” Interference may occur
in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Ear Thermometer is used exceeds
the applicable RF compliance level above, the Ear Thermometer should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Ear
Thermometer. Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
14 | GB
GB | 15
Table 4 For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment
and the Ear Thermometer
The Ear Thermometer is intended for use in an electromagnetic environment in which radiated
RF distances are controlled. The customer or the user of the Ear Thermometer can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Ear Thermometer as recommended below
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHZ
d=1.2
80 MHz to 800 MHZ
d=1.2
800 MHz to 2.5 GHz
d=2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
2 YEAR GUARANTEE
FKA Brands Ltd guarantees this product from defect in material and workmanship for a period of
2 years from the date of purchase, except as noted below. This FKA Brands Ltd product guarantee
does not cover damage caused by misuse or abuse; accident; the attachment of any unauthorised
accessory; alteration to the product; or any other conditions whatsoever that are beyond the control
of FKA Brands Ltd. This guarantee is eective only if the product is purchased and operated in the
UK / EU. A product that requires modification or adaptation to enable it to operate in any country
other than the country for which it was designed, manufactured, approved and / or authorised,
or repair of products damaged by these modifications is not covered under this guarantee. FKA
Brands Ltd shall not be responsible for any type of incidental, consequential or special damages.
To obtain guarantee service on your product, return the product post-paid to your local service
centre along with your dated sales receipt (as proof of purchase).
Upon receipt, FKA Brands Ltd will repair or replace, as appropriate, your product and return it to
you, post-paid. Guarantee is solely through Salter Service Centre. Service of this product by anyone
other than the Salter Service Centre voids the guarantee.
This guarantee does not aect your statutory rights.
For your local Salter Service Centre, go to www.salterhousewares.co.uk/servicecentres
14 | GB
GB | 15
EXPLANATION OF SYMBOLS
The CE marking with the Registration Number of the Notified Body. This denotes the
compliance of European Medical Device Directive 93/42/EEC
Consult the instruction for use
Disposal information: Should you wish to dispose of the article, do so in accordance
with current regulations. Details are available from your local authority
Type of protection of applied part against electric shock, body floating
This product meets the basic safety and essential performance requirements
indicated in the IP22 conditioning test (protection against solid foreign objects of
12.5mm Ø and greater and against vertically falling water drops when enclosure
tilted up to 15°)
Temperature limits
European Authorized Representative
Manufacturer’s name and address
SN YYMWWWXXXXX
SN: Product Serial Number
YY: year, MM: month, WWW: working sheet, XXXXX: serial no.
The empty, completely flat batteries must be disposed of through specially
designated collection boxes, recycling points or electronics retailers. You are legally
required to dispose of the batteries.
RoHS
This product fulfilling the requirements of the RoHS Directive 2011/65/EU.
REACH
This product fulfilling the requirements of the REACH Directive EC 1907/2006 and its
amendments, do not contain Substances of Very High Concern in concentration above
the limit of 0.1 %. No substance(s) is/are present in the parts of the product above
the concentration of 0.1 % weight by weight.
16 | F
F | 17
PRÉSENTATION
Ce thermomètre est destiné à eectuer des relevés intermittents de la température
corporelle humaine. Il est adapté pour les personnes de tous âges et à un usage
domestique.
Ce thermomètre fonctionne avec la technologie infrarouge et relève la température
en quelques secondes en mesurant la chaleur dégagée par l’oreille. Ce produit est
conforme aux dispositions contenues dans la directive de l’UE relative aux dispositifs
médicaux 93/42/CEE. Les avantages :
1. Mesure auriculaire
2. Aucun couvre sonde requis
3. Conception de sonde étanche
4. Résultat du relevé de sonde en une seconde
5. Mise en marche/arrêt automatique pour une économie d’énergie
6. Voyant indicateur de pile faible
7. Avertissement - mises en garde liées à l’état de la pile et à la plage de mesure.
8. Grand écran d’achage LCD
9. Fonction mémoire
INFORMATIONS IMPORTANTES AVANT UTILISATION
Lors de l’utilisation de ce produit, s’assurer de suivre toutes les indications fournies
ci-dessous. Tout non-respect de ces avis peut entraîner des blessures ou aecter la
précision de l’appareil.
1. Ne pas démonter, réparer ou altérer le thermomètre.
2. Il convient de nettoyer la lentille du thermomètre après chaque utilisation.
3. Éviter tout contact direct des doigts sur la lentille.
4. Aucune modification de cet équipement n’est autorisée.
5. Nous conseillons d’eectuer 3 relevés de température. En cas de divergence entre
ces relevés, retenez la mesure la plus élevée.
6. Ne pas exposer le thermomètre à des températures extrêmes, à une forte humidité
ou à la lumière directe du soleil.
7. Éviter tout choc violent ou de faire tomber le thermomètre.
8. Avant d’eectuer un relevé, l’utilisateur et le thermomètre doivent rester dans une
pièce à une température ambiante constante pendant au moins 30 minutes.
9. Éviter de relever la température dans les 30 minutes qui suivent toute activité
physique, après avoir pris un bain ou être sorti(e).
Pointe de la sonde
16 | F
F | 17
10. Dans un souci de protection de l’environnement, jeter les piles usagées dans
un site une collecte de déchets appropriée conformément aux règlementations
locales ou nationales.
11. Il est déconseillé de démonter le thermomètre.
12. Veuillez utiliser le thermomètre uniquement aux fins pour lesquelles il a été
conçu.
13. Tenir le thermomètre avec précaution pendant son utilisation pour éviter de le
faire tomber.
14. Laisser une minute entre chaque relevé consécutif car de légères variations
peuvent se produire si les mesures sont réalisées sur une courte période. Utiliser
plutôt une température moyenne.
15. Il n’y a pas de normes de température corporelle absolues. Conserver des relevés
fiables de votre température corporelle à titre de référence pour déterminer si
vous avez de la fièvre.
16. Le relevé de température est fourni UNIQUEMENT à titre indicatif. Avant de prendre
toute mesure médicale, veuillez consulter votre médecin.
17. Il est recommandé de calibrer le thermomètre tous les ans.
IDENTIFICATION DU PRODUIT
Bouton de mémoire
Couvercle du
compartiment à pile
Pointe de la sonde
Bouton
d’alimentation et
de relevé
Écran LCD
18 | F
F | 19
DESCRIPTION DE L’ÉCRAN D’AFFICHAGE
CONSEILS POUR RELEVER LA TEMPÉRATURE CORPORELLE HUMAINE
N’oubliez pas que le thermomètre doit être resté dans la pièce dans laquelle le relevé
sera eectué pendant au moins 30 minutes avant son utilisation.
REMARQUE :
Pour certaines personnes, le relevé sera diérent selon qu’il est eectué dans
l’oreille gauche ou droite. Pour relever les changements de température, toujours
prendre la température dans la même oreille.
Le thermomètre auriculaire peut être utilisé par des enfants uniquement sous la
surveillance d’un adulte. Il est généralement possible d’eectuer des relevés dès
l’âge de 6 mois. Chez les nourrissons de moins de 6 mois, le canal auditif étant
encore très étroit, il est souvent impossible de relever la température du tympan et la
mesure achée est souvent trop faible.
Ne pas eectuer de relevé dans une oreille comportant une atteinte inflammatoire
(par ex. en cas d’écoulement de l’oreille), après une blessure (par ex. tympan
endommagé) ou pendant la période de guérison après une procédure opératoire. Dans
chacun de ces cas, veuillez consulter votre médecin.
Même si le thermomètre est nettoyé et désinfecté, il est déconseillé de l’utiliser pour
plusieurs personnes, en cas de certaines maladies infectieuses graves en raison
du risque de propagation des bactéries. En cas de doute, veuillez consulter votre
médecin.
Ce thermomètre doit être utilisé uniquement sans couvre sonde jetable. Si vous êtes
resté(e) couché(e) sur une oreille pendant une période prolongée, la température
Température en degrés
Échelle Celsius
Échelle Fahrenheit
Relevé en coursVisage mécontent
Relevé
≥38.0°C(≥ 100.4°F)
Fièvre
Visage souriant
Relevé
<38.0°C(< 100.4°F)
Pas de fièvre
Voyant indicateur de
pile faible
Symbole de mémoire et jeux d’enregistrements
18 | F
F | 19
sera légèrement augmentée.
Attendre 60 secondes ou eectuer le relevé dans l’autre oreille.
La présence de cérumen pouvant aecter le relevé, nettoyer l’oreille avant de prendre
la mesure si cela est nécessaire.
MODE D’EMPLOI DU THERMOMÈTRE
1. Appuyer sur le bouton d’alimentation et de relevé pendant 1 seconde pour mettre en
marche le thermomètre. Les données s’achent sur l’écran. Attendre le bip sonore
indiquant que le thermomètre est prêt à l’emploi, le relevé précédent s’ache sur
l’écran.
2. S’assurer que la pointe de la sonde et le canal auditif
sont propres. Tirer légèrement l’oreille vers le haut et
vers l’arrière pour insérer la pointe de la sonde dans
le canal auditif qui est légèrement incurvé. Ceci est
important pour s’assurer que la pointe de la sonde soit
orientée directement sur le tympan.
Enfants de moins de 1 an
L’enfant doit être couché la tête tournée d’un côté pour que l’oreille
dans laquelle le relevé sera eectué soit orientée vers le haut. Tirer
doucement sur l’oreille vers l’arrière.
3. Insérer la sonde dans le canal auditif (la sonde doit être susamment insérée
pour assurer un relevé précis de la température), puis appuyer sur le bouton
d’alimentation et de relevé pendant environ 1 seconde, relâcher le bouton, un bip
sonore bref retentit indiquant que le relevé est terminé.
4. Sortir le thermomètre de l’oreille pour lire les résultats du relevé.
5. L’achage rétroéclairé reste à l’écran pendant environ 5 secondes à chaque fois
que le bouton d’alimentation et de relevé est actionné, une fois chaque relevé
terminé, et à chaque fois que le bouton mémoire est actionné pour acher le
Enfants à partir de 1 an
Se tenir derrière et légèrement de côté par rapport à l’enfant/la
personne. Tirer doucement sur l’oreille vers le haut puis vers l’arrière
pour redresser le canal auditif.
20 | F
F | 21
résultat précédent.
6. Le thermomètre s’éteindra automatiquement après une minute d’inactivité.
Remarque :
7. Afin d’assurer une mesure précise, il convient d’attendre au moins 1 minute entre
chaque relevé consécutif.
8. Nettoyer de nouveau la lentille du thermomètre après utilisation. Après le
nettoyage, attendre au moins 10 minutes entre les relevés afin d’assurer la
précision des mesures.
CHANGER L’UNITÉ DE MESURE (°C/°F)
Pour basculer entre °C/°F, s’assurer que l’appareil est en marche. Appuyer simultanément
sur les boutons d’alimentation et de relevé et de mémoire pour basculer entre °C et °F.
FONCTION DE MÉMOIRE
Vous pouvez rappeler jusqu’à 10 mesures en mémoire
à partager avec votre médecin ou un professionnel de
santé compétent.
1. Appuyer sur le bouton mémoire. Le premier relevé
aché correspond à la dernière mesure mémorisée.
2. Maintenir le bouton mémoire enfoncé pour acher
l’avant-dernière mesure mémorisée.
3. Toute nouvelle mesure sera mémorisée et la mémoire la
plus ancienne sera supprimée automatiquement.
Les relevés
Si la mesure de température est inférieure à 38°C
(100.4°F), un visage souriant s’ache à côté
de la mesure.
Si la mesure de température est égale ou
supérieure à 38°C (100.4°F), un visage mécontent
s’ache.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72

Salter TE-150-EU Manuale del proprietario

Categoria
Termometri digitali per il corpo
Tipo
Manuale del proprietario
Questo manuale è adatto anche per