Thermo Fisher Scientific Surveyor Plus Manuale del proprietario

Tipo
Manuale del proprietario
Surveyor Plus
Surveyor LC Pump
Upgrade Guide
60053-97119 Revision B January 2009
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
Xcalibur is a registered trademark and ChromQuest and Surveyor are trademarks of Thermo Fisher Scientific
Inc. in the United States.
The following are registered trademarks in the United States and other countries: Windows is a registered
trademark of Microsoft Corporation.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A released in March 2006. Revision B released in January 2009
Software version: ChromQuest 4.0 SP2, ChromQuest 4.1 SP1, or Xcalibur 1.4 SP1
For Research Use Only. Not regulated for medical or veterinary diagnostic use by U.S. Federal Drug
Administration or other competent authorities.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each
European Union (EU) Member State, and these companies should dispose of or recycle this product.
See www.thermo.com/WEEERoHS for further information on Thermo Fisher Scientific’s compliance
with these Directives and the recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC
erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific hat Vereinbarungen mit Verwertungs-/Entsorgungsfirmen in allen EU-
Mitgliedsstaaten getroffen, damit dieses Produkt durch diese Firmen wiederverwertet oder entsorgt
werden kann. Mehr Information über die Einhaltung dieser Anweisungen durch Thermo Fisher
Scientific, über die Verwerter, und weitere Hinweise, die nützlich sind, um die Produkte zu
identifizieren, die unter diese RoHS Anweisung fallen, finden sie unter www.thermo.com/
WEEERoHS.
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements
Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état
membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage
d'informations sur la conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre
pays et les informations sur les produits Thermo Fisher Scientific qui peuvent aider la détection des
substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.
CAUTION Symbol CAUTION VORSICHT ATTENTION PRECAUCION AVVERTENZA
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Elektroschock: In diesem Gerät werden
Hochspannungen verwendet, die
Verletzungen verursachen können. Vor
Wartungsarbeiten muß das Gerät
abgeschaltet und vom Netz getrennt
werden. Betreiben Sie Wartungsarbeiten
nicht mit abgenommenem Deckel. Nehmen
Sie die Schutzabdeckung von Leiterplatten
nicht ab.
Choc électrique: L’instrument utilise des
tensions capables d’infliger des blessures
corprelles. L’instrument doit être arrêté et
débranché de la source de courant avant
tout intervention. Ne pas utiliser
l’instrument sans son couvercle. Ne pas
elensver les étuis protecteurs des cartes de
circuits imprimés.
Descarga eléctrica: Este instrumento
utiliza altas tensiones, capaces de
producir lesiones personales. Antes de
dar servicio de mantenimiento al
instrumento, éste debera apagarse y
desconectarse de la línea de alimentacion
eléctrica. No opere el instrumento sin sus
cubiertas exteriores quitadas. No remueva
las cubiertas protectoras de las tarjetas
de circuito impreso.
Shock da folgorazione. L’apparecchio è
alimentato da corrente ad alta tensione
che puo provocare lesioni fisiche. Prima di
effettuare qualsiasi intervento di
manutenzione occorre spegnere ed isolare
l’apparecchio dalla linea elettrica. Non
attivare lo strumento senza lo schermo
superiore. Non togliere i coperchi a
protezione dalle schede di circuito
stampato (PCB).
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Chemikalien: Dieses Gerät kann
gefährliche Chemikalien enthalten. Tragen
Sie Schutzhandschuhe beim Umgang mit
toxischen, karzinogenen, mutagenen oder
ätzenden/reizenden Chemikalien.
Entsorgen Sie verbrauchtes Öl
entsprechend den Vorschriften in den
vorgeschriebenen Behältern.
Chimique: Des produits chemiques
dangereux peuven se trouver dans
l’instrument. Proted dos gants pour
manipuler tous produits chemiques
toxiques, cancérigènes, mutagènes, ou
corrosifs/irritants. Utiliser des récipients
et des procédures homologuées pour se
débarrasser des déchets d’huile.
Química: El instrumento puede contener
productos quimicos peligrosos. Utilice
guantes al manejar productos quimicos
tóxicos, carcinogenos, mutagenos o
corrosivos/irritantes. Utilice recipientes y
procedimientos aprobados para
deshacerse del aceite usado.
Prodotti chimici. Possibile presenza di
sostanze chimiche pericolose
nell’apparecchio. Indossare dei guanti per
maneggiare prodotti chimici tossici,
cancerogeni, mutageni, o
corrosivi/irritanti. Utilizzare contenitori
aprovo e seguire la procedura indicata per
lo smaltimento dei residui di olio.
Heat: Before servicing the instrument,
allow any heated components to cool.
Hitze: Warten Sie erhitzte Komponenten
erst nachdem diese sich abgekühlt haben.
Haute Temperature: Permettre aux
composants chauffés de refroidir avant
tout intervention.
Altas temperaturas: Permita que lop
componentes se enfríen, ante de efectuar
servicio de mantenimiento.
Calore. Attendere che i componenti
riscaldati si raffreddino prima di
effetturare l’intervento di manutenzione.
Fire: Use care when operating the system
in the presence of flammable gases.
Feuer: Beachten Sie die einschlägigen
VorsichtsmaBnahmen, wenn Sie das
System in Gegenwart von entzündbaren
Gasen betreiben.
Incendie: Agir avec précaution lors de
l’utilisation du système en présence de
gaz inflammables.
Fuego: Tenga cuidado al operar el
sistema en presencia de gases
inflamables.
Incendio. Adottare le dovute precauzioni
quando si usa il sistema in presenza di gas
infiammabili.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
Verletzungsgefahr der Augen:
Verspritzte Chemikalien oder kleine
Partikel können Augenverletzungen
verursachen. Tragen Sie beim Umgang mit
Chemikalien oder bei der Wartung des
Gerätes eine Schutzbrille.
Danger pour les yeux: Dex projections
chimiques, liquides, ou solides peuvent
être dangereuses pour les yeux. Porter des
lunettes de protection lors de toute
manipulationde produit chimique ou pour
toute intervention sur l’instrument.
Peligro par los ojos: Las salicaduras de
productos químicos o particulas que
salten bruscamente pueden causar
lesiones en los ojos. Utilice anteojos
protectores al mnipular productos
químicos o al darle servicio de
mantenimiento al instrumento.
Pericolo per la vista. Gli schizzi di
prodotti chimici o delle particelle presenti
nell’aria potrebbero causare danni alla
vista. Indossare occhiali protettivi quando
si maneggiano prodotti chimici o si
effettuano interventi di manutenzione
sull’apparecchio.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
Allgemeine Gefahr: Es besteht eine
weitere Gefahr, die nicht in den
vorstehenden Kategorien beschrieben ist.
Dieses Symbol wird im Handbuch
auBerdem dazu verwendet, um den
Benutzer auf Anweisungen hinzuweisen.
Danger général: Indique la présence
d;un risque n’appartenant pas aux
catégories citées plus haut. Ce symbole
figure également sur l’instrument pour
renvoyer l’utilisateur aux instructions du
présent manuel.
Peligro general: Significa que existe un
peligro no incluido en las categorias
anteriores. Este simbolo también se utiliza
en el instrumento par referir al usuario a
las instrucciones contenidas en este
manual.
Pericolo generico. Pericolo non
compreso tra le precedenti categorie.
Questo simbolo è utilizzato inoltre
sull’apparecchio per segnalare all’utente
di consultare le istruzioni descritte nel
presente manuale.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
Wenn Sie sich über die Sicherheit eines
Verfahrens im unklaren sind, setzen Sie
sich, bevor Sie fortfahren, mit Ihrer
lokalen technischen
Unterstützungsorganisation für Thermo
Fisher Scientific San Jose Produkte in
Verbindung.
Si la sûreté d’un procédure est incertaine,
avant de continuer, contacter le plus
proche Service Clientèle pour les produits
de Thermo Fisher Scientific San Jose.
Cuando la certidumbre acerca de un
procedimiento sea dudosa, antes de
proseguir, pongase en contacto con la
Oficina de Asistencia Tecnica local para
los productos de Thermo Fisher Scientific
San Jose.
Quando e in dubbio la misura di sicurezza
per una procedura, prima di continuare, si
prega di mettersi in contatto con il
Servizio di Assistenza Tecnica locale per i
prodotti di Thermo Fisher Scientific San
Jose.
CAUTION Symbol CAUTION
Electric Shock: This instrument uses
high voltages that can cause personal
injury. Before servicing, shut down the
instrument and disconnect the instrument
from line power. Keep the top cover on
while operating the instrument. Do not
remove protective covers from PCBs.
Chemical: This instrument might contain
hazardous chemicals. Wear gloves when
handling toxic, carcinogenic, mutagenic,
or corrosive or irritant chemicals. Use
approved containers and proper
procedures to dispose waste oil.
Heat: Before servicing the instrument,
allow any heated components to cool.
Fire: Use care when operating the system
in the presence of flammable gases.
Eye Hazard: Eye damage could occur
from splattered chemicals or flying
particles. Wear safety glasses when
handling chemicals or servicing the
instrument.
General Hazard: A hazard is present that
is not included in the above categories.
Also, this symbol appears on the
instrument to refer the user to instructions
in this manual.
When the safety of a procedure is
questionable, contact your local Technical
Support organization for Thermo Fisher
Scientific San Jose Products.
Thermo Scientific Surveyor LC Pump Upgrade Guide vii
C
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Chapter 1 Upgrade Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Chapter 2 Installing the Interconnect Printed Circuit Board. . . . . . . . . . . . . . . . . . . . . . . . . . .3
Chapter 3 Downloading Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Step 1: Preparing the LC Pump for a Firmware Download . . . . . . . . . . . . . . . . . 8
Step 2: Downloading the Instrument Control Firmware . . . . . . . . . . . . . . . . . . . 9
Step 3: Downloading the Communication Firmware . . . . . . . . . . . . . . . . . . . . 14
Step 4: Returning the Pump to Normal Operation . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 4 Upgrading the Chromatography Data System . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Chapter 5 Adding the Surveyor LC Pump Plus Labels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Chapter 6 Configuring the Surveyor Plus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Configuring the Instrument in ChromQuest. . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Configuring the Surveyor LC Pump Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Configuring the Signal Polarities for the Surveyor AS . . . . . . . . . . . . . . . . . . 30
Configuring the Instrument in Xcalibur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Configuring the LC Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Configuring the Signal Polarities for the Surveyor AS . . . . . . . . . . . . . . . . . . 32
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Contents
Thermo Scientific Surveyor LC Pump Upgrade Guide ix
P
Preface
This installation guide describes how to upgrade the Surveyor™ LC Pump to a Surveyor LC
Pump Plus.
To increase the retention time reproducibility of the LC system, the upgraded LC pump issues
two output signals to the autosampler at the start of a run. The LC pump issues the first
output signal when it senses a stable backpressure. It issues the second output signal when its
cams return to the start position. The autosampler waits for these two signals before it
performs an injection.
Related Documentation
In addition to this guide, Thermo Fisher Scientific provides the following documents for the
Surveyor LC Pump Plus:
Surveyor LC Pump Plus Hardware Manual
Surveyor Plus Getting Started with ChromQuest Guide
Surveyor Plus Getting Started with Xcalibur Guide
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION
notice is accompanied by an appropriate CAUTION symbol.
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Preface
xSurveyor LC Pump Upgrade Guide Thermo Scientific
Contacting Us
There are several ways to contact Thermo Fisher Scientific for the information you need.
YTo contact Technical Support
Find software updates and utilities to download at mssupport.thermo.com.
YTo contact Customer Service for ordering information
YTo copy manuals from the Internet
Go to mssupport.thermo.com and click Customer Manuals in the left margin of the
window.
YTo suggest changes to documentation or to Help
Complete a brief survey about this document by clicking the link below. Thank you
in advance for your help.
Send an e-mail message to the Technical Publications Editor at
techpubs-lcms@thermofisher.com.
Note Highlights information of general interest.
Tip Highlights helpful information that can make a task easier.
Phone 800-532-4752
Fax 561-688-8736
E-mail us.techsupport.analyze@thermofisher.com
Knowledge base www.thermokb.com
Phone 800-532-4752
Fax 561-688-8731
E-mail us.customer-support.analyze@thermofisher.com
Web site www.thermo.com/ms
Thermo Scientific Surveyor LC Pump Upgrade Guide 1
1
Upgrade Kit Contents
Table 1 lists the items supplied in the Surveyor LC Pump upgrade kit (P/N 60053-62045).
In addition to these items, you must also have the items listed in Tab l e 2 on page 7 to upgrade
the communication firmware. The items listed in Table 2 are included in the Surveyor LC
Pump accessory kit.
To upgrade the Surveyor LC Pump to a Surveyor LC Pump Plus, do the following:
1. Install the new interconnect PCB by following the instructions in Chapter 2, “Installing
the Interconnect Printed Circuit Board.
2. Download the new CPU and communication firmware by following the instructions in
Chapter 3, “Downloading Firmware.
3. Install the service pack for your chromatography data system by following the
instructions in Chapter 4, “Upgrading the Chromatography Data System.
4. Attach the back panel labels for the Surveyor LC Pump Plus by following the instructions
in Chapter 5, “Adding the Surveyor LC Pump Plus Labels.
5. Configure the autosampler to recognize the output signals from the upgraded LC pump
by following the instructions in Chapter 6, “Configuring the Surveyor Plus System.
Table 1. Parts list for the Surveyor LC Pump Upgrade kit
Description Part Number
LC Pump Plus interconnect PCB assembly 00950-01-00022
ChromQuest 4.0 SP1 software CD-ROM CHROM-98002
ChromQuest 4.1 SP1 software CD-ROM CHROM-98003
Label set, Surveyor LC Pump Plus 60053-50105
Label, Surveyor LC Pump Plus upgrade 60053-50110
Xcalibur Surveyor LC Pump 1.4 SP1 software XCALI-64069
Manual CD-ROM 60053-64000
Thermo Scientific Surveyor LC Pump Upgrade Guide 3
2
Installing the Interconnect Printed Circuit Board
The new interconnect PCB contains the circuitry that enables the Pump RDY output signal
and the INJ HOLD output signal, generated from pins 5 and 6, respectively, on the back
panel of the LC pump. See Figure 24 on page 23.
YTo install the new PCB
1. Turn off the power to the Surveyor LC Pump by pressing the power button below the left
door of the pump, and then unplug the pump from line power.
2. Gain access to the interconnect PCB as follows:
a. Detach the LED cable from the inner right-door of the pump.
Figure 1. Detaching the LED cable
b. Pull the drainage funnel forward off the housing of the pump. See Figure 2.
c. Use a #2 Phillips screwdriver to remove the four screws that secure the chassis of the
pump to its housing. See Figure 2.
CAUTION Because the Surveyor LC Pump contains a high voltage supply, you must
turn off the power to the pump and unplug it from line power before you pull the
chassis out of the housing.
LED cable
2 Installing the Interconnect Printed Circuit Board
4Surveyor LC Pump Upgrade Guide Thermo Scientific
Figure 2. View of the LC pump, showing the chassis pulled out of the housing
d. Pull the chassis forward out of the housing. See Figure 3.
Figure 3. View of the LC pump without its housing
3. Remove the interconnect PCB as follows:
a. Detach the eight cables connected to the interconnect PCB. See Figure 4.
Screws
Drainage
funnel
Screws
Interconnect
PCB
CAUTION Because it is easy to drop the small screws that secure the interconnect
PCB to the chassis, take care as you remove them.
2 Installing the Interconnect Printed Circuit Board
Thermo Scientific Surveyor LC Pump Upgrade Guide 5
b. Using a #1 Phillips screwdriver, carefully remove the four screws that secure the
interconnect PCB to the chassis. See Figure 4.
Figure 4. View of the interconnect PCB, showing the cable connections
4. Install the new interconnect PCB as follows:
a. Use the screws you removed in step 3b to secure the new interconnect PCB to the
chassis of the pump.
b. Reattach the eight cables shown in Figure 4.
5. Push the chassis back into the housing. Then secure it to the housing with the four screws
you removed in step c on page 3.
6. Reconnect the LED cable to the inner right-door of the pump. See Figure 1 on page 3.
7. Reinstall the drainage funnel. See Figure 2 on page 4.
8. Reconnect the LC pump to line power, but do not turn the pump on.
Screws (4x)
Thermo Scientific Surveyor LC Pump Upgrade Guide 7
3
Downloading Firmware
Table 2 lists the hardware that you must have to download firmware to the CPU board.
To download the instrument control firmware and the communication firmware, perform
these steps:
Step 1: Preparing the LC Pump for a Firmware Download
Step 2: Downloading the Instrument Control Firmware
Step 3: Downloading the Communication Firmware
Step 4: Returning the Pump to Normal Operation
Table 2. Hardware required to download firmware to the CPU board
Figure Description Part Number
8-pin, minicombicon connector 00004-02511
RJ45 to DB9 cable adapter A3538-010
Serial communication cable 60053-63003
3 Downloading Firmware
Step 1: Preparing the LC Pump for a Firmware Download
8Surveyor LC Pump Upgrade Guide Thermo Scientific
Step 1: Preparing the LC Pump for a Firmware Download
YTo prepare the LC pump for a firmware download
1. Ensure that the LC Pump is turned off.
2. Use a flathead screwdriver to set the rotary switches located on the back panel of the LC
pump to 00. See Figure 5.
Figure 5. View of rotary switches, showing the Unit ID set to 00
3. Plug an 8-pin, 3.81 mm pitch, minicombicon connector (P/N 00004-02511) into
positions 9 through 16 on the back panel of the LC pump. Then create a jumper across
pins 9 and 15 with a piece of wire, such as a paperclip. See Figure 6.
Figure 6. View of the LC pump back panel, showing a jumper between pins 9 and 15
4. Plug an RJ-45 to DB9 adapter (P/N A3538-010) into a COM port on your computer
and record the port number.
5. Using a serial communication cable (P/N 60053-63003), connect the COM port on the
back panel of the computer to the RS232 connection located on the back panel of the
LC pump. See Figure 7.
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
0
UNIT ID
NC
FW DOWNLOAD
TF1
PRESSURE -
PRESSURE+
+5V @ 15OmA
GND
PUMP ON
GND
NC
NC
PROG. START
+5V @ 150mA
GND
NC
TF2
RS232 ENET
Minicombicon plug
Shorting wire
[connecting pin 9 (GND) to pin 15 (FW Download)]
3 Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
Thermo Scientific Surveyor LC Pump Upgrade Guide 9
Figure 7. Back panel connections between the computer and the Surveyor LC Pump
6. Turn on the power to the LC pump.
7. Verify that the Power LED is green, the Comm LED is flashing amber, and the other two
LEDs are not illuminated. See Figure 8.
Figure 8. Surveyor LC Pump LEDs
Step 2: Downloading the Instrument Control Firmware
The ChromQuest software CD-ROM contains the Update.exe utility that allows you to
download the instrument control firmware.
YTo download the control firmware
1. Use Windows™ Explorer to locate the Update.exe utility on the ChromQuest software
CD-ROM. As Figure 9 shows, the Update.exe utility and the firmware file (*lcb) are
located in the following folder:
CD-ROM drive:\ChromQuest\Firmware\60053-98020 LPump v2.00 \LPump Control
Firmware
UNIT ID
NC
FW DOWNLOAD
NC
PRESSURE -
PRESSURE+
+5V @ 15OmA
GND
PUMP ON
GND
+5V @ 150mA
GND
NC
NC
RS232 ENET
12345678910 11 12 13 14 15 16
NC
Ethernet port
Back panel of Surveyor LC PumpComputer
Serial port with
DB9 to RJ45 adapter
Serial communication cable
Ethernet switch
Ethernet
port
Power Comm Run Degas
3 Downloading Firmware
Step 2: Downloading the Instrument Control Firmware
10 Surveyor LC Pump Upgrade Guide Thermo Scientific
Figure 9. View of the contents of the LPump Control Firmware folder
2. Double-click the Update.exe icon to start the utility.
The Update Windows Application – Update dialog box appears. See Figure 10.
Figure 10. Update Windows Application Update dialog box
3. Set up the appropriate communication settings as follows:
a. Choose Communication > Settings to open the COM Port Setting dialog box. See
Figure 11.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

Thermo Fisher Scientific Surveyor Plus Manuale del proprietario

Tipo
Manuale del proprietario